Maxtec UltraMaxO2 User manual

UltraMaxO2™
OPERATING MANUAL &
INSTRUCTIONS FOR USE
R221P11
R221M11 Rev. C
OXYGEN ANALYZER


866.4.Maxtec www.maxtecinc.com I
INDICATIONS FOR USE:
The UltraMaxO2is an oxygen analyzer for checking the oxygen concentration, flow and outlet
pressure of oxygen concentrators.
WARRANTY:
Under normal operating conditions, Maxtec warrants the UltraMaxO2to be free from defects of
workmanship or materials for a period of Three (3) years from the date of shipment from Maxtec,
provided that the unit is properly operated and maintained in accordance with Maxtec’s operating
instructions. Based on Maxtec product evaluation, Maxtec’s sole obligation under the foregoing
warranty is limited to making replacements, repairs, or issuing credit for equipment found to be
defective. This warranty extends only to the buyer purchasing the equipment directly from Maxtec
or through Maxtec’s designated distributors and agents as new equipment.
Routine maintenance items, such as batteries, are excluded from warranty. Maxtec and any other
subsidiaries shall not be liable to the purchaser or other persons for incidental or consequential
damages or equipment that has been subject to abuse, misuse, mis-application, alteration,
negligence or accident.
These warranties are exclusive and in lieu of all other warranties, expressed or implied, including
warranty of merchantability and fitness for a particular purpose.
CLASSIFICATIONS:
Classification:.............................................. Class II electrical device
Protection against electric shock: ...........................Internally powered equipment
Protection against water: ....................................................IPX1
Mode of operation:....................................................Continuous
Sterilization: ......................................................See section 6.0
Flammable anaesthetic mixture:......Not for use in presence of flammable anaesthetic mixtures
Power specification: .......................................1.8-3.2V 32mW10mA
Caution: Federal law restricts this device to sale by or on the order of a physician or other licensed
healthcare practitioner.
NOTE: The UltraMaxO2is for use only by trained personnel. Before use, all individuals using the
UltraMaxO2should become familiar with the information contained in this Operation Manual.
Adherence to these instructions is necessary for safe, effective product performance. Thoroughly
read all instructions and labeling provided with this device and any other equipment that will
be used.
Maxtec TEL:(800) 748.5355
6526 South Cottonwood Street FAX:(801) 270.5590
USA WEBSITE:www.maxtecinc.com
Authorized Representative:
QNET BV
Hommerterweg 286
6436 AM Amstenrade
The Netherlands EC REP
Product Disposal Instructions:
The batteries and circuit board are not suitable for regular trash disposal. Follow local
guidelines for proper disposal

866.4.Maxtec www.maxtecinc.comII
PRINCIPLE OF OPERATION:
The UltraMaxO2Oxygen Analyzer measures oxygen concentration and flow using
ultrasound technology and measures pressure using a piezoresistive silicon
pressure sensor.
WARNINGS:
Indicates a potentially hazardous situation, if not avoided, could result in death or
serious injury.
• Not for use in an MRI environment.
• Improper use of the UltraMaxO2can cause inaccurate oxygen readings leading to
improper treatment and/or patient harm. Follow the procedures outlined in this
user manual.
• The UltraMaxO2is for checking oxygen concentrators only.
• DO NOT use the UltraMaxO2for continuous oxygen monitoring.
• DO NOT use the UltraMaxO2 to measure the oxygen concentration of a concentrator
when flowing at rates lower than its optimal performance as specified by the
concentrator manufacturer; generally 4 LPM or less on concentrators that have a
maximum flow of 10 LPM, and 1 LPM or less on concentrators that have a maximum
flow of 5 LPM.
• Not for use in anesthesia applications or for measuring oxygen concentration from
any sources other than conventional oxygen concentrators.
• Not for use with inhalation agents. Operating the UltraMaxO2in flammable or
explosive environments may result in fire or explosion.
• Not suitable for use in the presence of flammable anesthetic mixtures.
• Users must become thoroughly familiar with the information contained in this
Operation Manual before use. Strict adherence to the operating instructions is
necessary for safe, effective product performance. This product will perform only as
designed if operated in accordance with the manufacturer’s operating instructions.
• Use only genuine Maxtec accessories. Failure to do so may seriously impair the
performance of the UltraMaxO2. Repair or alteration of the UltraMaxO2by anyone
other than an authorized Maxtec service representative could cause the product to
fail to perform as designed.
• Use of the UltraMaxO2near devices that generate electrical fields may cause
erratic readings.
• If the UltraMaxO2is ever exposed to liquids from spills or immersion, immediately
remove the batteries and let the device dry completely. When dry, replace the
batteries and check for proper operation.
• DO NOT autoclave or expose the UltraMaxO2to high temperatures (>60°C).
• DO NOT use ethylene oxide sterilization. performance in flow and pressure readings.
!

866.4.Maxtec www.maxtecinc.com III
• DO NOT expose the UltraMaxO2to irradiation, vacuum, steam, or harsh chemicals.
• DO NOT expose the UltraMaxO2to pressure greater than 50 psi. Exposure to
pressure above 50 psi could cause leaks in the device which may adversely affect
CAUTIONS:
Indicates a potentially hazardous situation, if not avoided, could result in minor or
moderate injury and property damage.
• Replace the batteries with high quality AA Alkaline or Lithium batteries. DO NOT use
rechargeable batteries.
• When not in use for periods greater than 30 days remove the batteries to protect the
UltraMaxO2from potential battery leakage.
• Avoid dropping the UltraMaxO2to prevent damage which may adversely affect
its performance. If damage to the device is suspected, perform the calibration
verification procedure in Section 2.3 of this operating manual.
• Avoid foreign matter entry into the UltraMaxO2.

866.4.Maxtec www.maxtecinc.comIV
TABLE OF CONTENTS
Classifications ...................................................... I
Indications for Use ................................................... I
Warranty........................................................... I
Principle of Operation .................................................II
Warnings ..........................................................II
Cautions .......................................................... III
1.0 SYSTEM OVERVIEW................................................1
1.1 Description ....................................................1
1.2 Component Identification .........................................1
1.3 Symbol Guide ..................................................2
1.4 Component Description...........................................2
2.0 OPERATING INSTRUCTIONS ..........................................3
2.1 Oxygen, Flow and Pressure Measurement .............................3
2.2 Switching Pressure Units of Measure ................................3
2.3 Calibration Verification Procedure ...................................3
3.0 FACTORS INFLUENCING ACCURATE READINGS .............................4
3.1 Effects of Temperature ...........................................4
3.2 Effects of Humidity ..............................................4
3.3 Effects of Other Gases............................................4
3.4 Effects of Low Flow ..............................................4
4.0 ERROR CODES ....................................................4
5.0 CHANGING THE BATTERIES...........................................5
6.0 CLEANING AND MAINTENANCE ........................................5
6.1 Cleaning ......................................................5
6.2 Maintenance...................................................5
7.0 SPECIFICATIONS ..................................................6
8.0 SPARE PARTS AND ACCESSORIES......................................7
8.1 Included with your Unit...........................................7
8.2 Optional Accessories.............................................7

866.4.Maxtec www.maxtecinc.com 1
1.0 SYSTEM OVERVIEW
1.1 Description & Priciple of Operation
The UltraMaxO2is an oxygen analyzer designed to check the oxygen concentration,
flow and outlet pressure of oxygen concentrators. The UltraMaxO2provides unparalleled
performance and reliability from its advanced design that includes the following
features and operational benefits:
• Accurate oxygen measurements.
• No in-field calibration required.
• Convenient ability to measure pressure in PSI or kPa.
• Durable, compact design.
• Large, easy-to-read, liquid crystal display (LCD).
• Shielded, reinforced sample gas inlet port.
• Long battery life with 2 AA batteries.
• Auto-off after 4 minutes.
• Low battery indication.
• Self-diagnostics.
• Easy to clean.
1.2 Component Identification
LCD Display
Mode Button
ON/OFF Button
Gas Sample Inlet
Gas Sample Outlet
Battery Door
1
2
3
4
5
6
1
2
4
5
6
3

866.4.Maxtec www.maxtecinc.com2
1.3 Symbol Guide
The following symbols and safety labels are found on the UltraMaxO2:
1.4 Component Description
3 1/2 Digit Display -The LCD provides direct readout of oxygen concentration, gas
flow and gas pressure. The LCD also displays error codes as necessary.
ON/OFF Button -Turns the device on or off.
Mode Button - Switches between measuring the concentration of gas produced by an
oxygen concentrator and pure oxygen (for calibration verification).
Low Battery Indicator – Indicates the voltage of the batteries is below normal
operating levels.
LPM – Illuminated next to the flow measurement. (Not shown when in calibration
verification mode).
KPA – Indicates the pressure measurement is in units of kilopascals.
PSI – Indicates the pressure measurement is in units of pounds per square inch.
!
LPM
PSI
kPa
%
Attention, consult accompanying
documents On/off button
Mode button
Low battery
Liter per minute flow
Pounds per square inch
Kilopascals
Percent
Gas sample inlet
Gas sample outlet
Consult accompanying
documents
Direct current
Evaluated by ETL (Intertek Testing
Laboratories)
Do not throw away. Follow local
guidelines for disposal
Manufacturer
Latex free
EC REP Authorized representative in
the European Community
SN Serial number
IPX1 Ingress protection
only Restricted to sale by or on order of a
medical professional
REF Catalog number
BAT
LATEX

866.4.Maxtec www.maxtecinc.com 3
“%” symbol – Illuminated next to the concentration measurement.
Gas Sample Inlet – Used to receive the gas sample.
Gas Sample Outlet – Used as an outlet for the gas sample and as a trigger for pressure
measurement when occluded.
Gas Sample Tubing – Used to connect to gas sample sources.
2.0 OPERATING INSTRUCTIONS
2.1 Oxygen, Flow and Pressure Measurement
To check oxygen concentration, flow and pressure of a gas sample from a concentrator:
1. Connect the gas sample tubing to the gas sample inlet of the UltraMaxO2.
2. Attach the other end of the gas sample tubing to the oxygen concentrator.
3. Initiate the flow of gas to the UltraMaxO2at a rate of 1-10 liters per minute (2 liters
per minute is recommended). Ensure the concentrator’s output is stable per the
concentrator manufacturer’s recommendations.
4. Turn on the UltraMaxO2.
5. Allow the oxygen reading to stabilize for approximately 10 seconds before reading
the oxygen concentration and flow.
6. To check pressure, cover the gas sample outlet with thumb or finger while gas is
flowing.
7. Wait 5 seconds for the display to read pressure.
2.2 Switching Pressure Units of Measure
The UltraMaxO2can measure pressure in PSI or kPa. The UltraMaxO2is factory set to
measure in PSI. To switch to kPa:
1. Using a #1 Phillips screwdriver loosen the battery door screw and remove the
battery door.
2. Toggle the switch inside the battery compartment.
3. Replace the battery door and tighten the battery door screw.
2.3 Calibration Verification Procedure
A calibration verification mode is provided to verify that the UltraMaxO2is functioning
properly. To perform the calibration verification:
1. Turn on the UltraMaxO2.
2. Connect a source of pure oxygen (≥99.95%) to the gas sample inlet.
3. Flow 2-5 LPM of gas into the UltraMaxO2. Ensure that the gas flowing to the
UltraMaxO2is at a stable temperature.
4. Press and hold the mode button. While holding the mode button, the gas
measurement should read between 98.5 and 101.5% oxygen. If the gas
measurement is not within this range, call Customer Service. Calibration
verification mode is indicated by “CAL” and “VER” flashing on screen beneath the
gas measurement .

866.4.Maxtec www.maxtecinc.com4
3.0 FACTORS INFLUENCING ACCURATE READINGS
3.1 Effects of Temperature
The UltraMaxO2compensates for temperature and will perform within specifications
throughout the operating temperature range. However, taking measurements during
rapid changes in gas temperature should be avoided.
3.2 Effects of Humidity
The UltraMaxO2has a humidity sensor to detect and compensate for the humidity of
gas entering the device. However, high levels (condensing) of humidity can affect the
accuracy and reliability of the UltraMaxO2. To prevent possible damage:
• Avoid usage in environments of greater than 95% relative humidity.
• DO NOT use this device in a breathing circuit.
• DO NOT breathe or blow into the UltraMaxO2.
3.3 Effects of Other Gases
The UltraMaxO2is designed to measure two different types of gas mixtures:
• Oxygen, nitrogen and argon from oxygen concentrators.
• Pure oxygen during calibration verification mode.
Any other concentrations or combinations of gases will cause the UltraMaxO2to
measure oxygen concentration incorrectly.
3.4 Effects of Low Flow
Oxygen concentrators function on the principle of removing nitrogen gas from air,
leaving concentrated oxygen and argon at a specific oxygen to argon ratio. This
operating principle may be altered when concentrators are set to flow at the low end of
their operational range. At low flows they may output a low oxygen concentration, e.g.
85% to 91%, for reasons other than high nitrogen, possibly due to an increase in argon
content. The UltraMaxO2 requires that the ratio of oxygen to argon remain constant in
order to guarantee an accuracy of +/-1.5% oxygen.
• DO NOT use the UltraMaxO2 to measure the oxygen concentration of a
concentrator when flowing at rates lower than its optimal performance as specified
by the concentrator manufacturer; generally 4 LPM or less on concentrators that
have a maximum flow of 10 LPM, and 1 LPM or less on concentrators that have a
maximum flow of 5 LPM.
4.0 ERROR CODES
The UltraMaxO2 has self diagnostic features built into the software to detect
faulty readings outside of normal operating ranges. The codes, descriptions and
recommended actions are:
• E01: Oxygen measurement out of range Hi (≥102.0% calculated by algorithm).
Recommended Action: Verify that the UltraMaxO2is being used in the correct
mode (Concentrator or Calibration Verification mode). If error code repeats;
perform a calibration verification per section 2.3 of this manual. If error code
repeats again; contact Customer Service.

866.4.Maxtec www.maxtecinc.com 5
• E02: Oxygen measurement out of range Low (≤-2.0% calculated by algorithm).
Recommended Action: Verify that the UltraMaxO2is being used in the correct
mode (Concentrator or Calibration Verification mode). If error code repeats;
perform a calibration verification per section 2.3 of this manual. If error code
repeats again; contact Customer Service.
• E03: Device memory corrupt or missing.
Recommended Action: Return the UltraMaxO2to the manufacturer for factory
repair.
• E04: Signal reading not stable.
Recommended Action: Return the UltraMaxO2to the manufacturer for factory
repair.
• E05: Pressure measurement out of Range Hi (≥50 PSI).
Recommended Action: Check the pressure on a known gas source pressure. If
error code repeats; contact Customer Service.
• E06: Outside of operating temperature Hi (≥40° C).
Recommended action: The UltraMaxO2is too hot, cool the device closer to room
temperature before use.
• E07: Outside of operating temperature Low (≤15° C).
Recommended action: The UltraMaxO2is too cold, warm the device closer to room
temperature before use.
• E08: Device self check found error.
Recommended Action: Remove and replace the batteries. If error code repeats;
return the UltraMaxO2to the manufacturer for factory repair.
5.0 CHANGING THE BATTERIES
Batteries should be changed when the battery icon illuminates. The battery icon will
remain lit until the batteries are changed. If the battery power level is too low the
UltraMaxO2will not power on until the batteries are changed.
5.1 Battery Replacement Procedure
1. Using a #1 Phillips screwdriver loosen the battery door screw and remove the
battery door.
2. Remove the batteries.
3. Insert new batteries ensuring correct placement. DO NOT use rechargeable
batteries.
4. Replace the battery door and tighten the battery door screw.
5. If the UltraMaxO2does not power on when done verify the batteries are installed
correctly and that the batteries are fresh.
6.0 CLEANING AND MAINTENANCE
• Use caution to prevent any fluid from entering the UltraMaxO2.
• DO NOT soak or immerse the UltraMaxO2 in fluid.
• DO NOT autoclave or expose the UltraMaxO2to ethylene oxide sterilization.
6.1 Cleaning
• Wipe down the exterior surfaces of the UltraMaxO2with a moist cloth and mild
hand or dish soap (pH 6-8).

866.4.Maxtec www.maxtecinc.com6
6.2 Maintenance
Replace the batteries with high quality AA Alkaline or Lithium batteries. DO NOT use
rechargeable batteries.
• When not in use for periods greater than 30 days, remove the batteries to protect
the UltraMaxO2from potential battery leakage.
• Store the UltraMaxO2between -15˚C and 60˚C (5˚F - 140˚F)
7.0 SPECIFICATIONS
OXYGEN
Oxygen Measurement Range:
(from a concentrator)
20.9 - 96%
Oxygen Measurement Accuracy: ±1.5 % of full scale at
constant temperature and optimal flow*
Oxygen Measurement Resolution: 0.1% Oxygen
FLOW
Flow Measurement Range: 0 - 10 LPM
Flow Measurement Accuracy: ±0.2 LPM
Flow Measurement Resolution: 0.1 LPM
PRESSURE PSI kPa
Pressure Measurement Range: 0.5 - 50 3.4 - 344
Pressure Measurement Accuracy: ±0.5% ±0.5%
Pressure Measurement Resolution: 0.1 0.1 up to 199,
1 from 200 to 344
Response Time: ≤17 seconds
Warm-up Time: < 1 second
Operating Temperature: 15˚C - 40˚C (59˚F-104˚F)
-15˚C - 60˚C (5˚F-140˚F)Storage Temperature:
Pressure: 800 - 1000 mBars
Humidity: 0 - 95% (non-condensing)
Power Requirements: 2 AA Alkaline batteries (2 x 1.5 Volts)
Battery Life: ≥ 1,100 hours (16,500 read cycles)
Low Battery Indication: "Low Battery" icon displayed on LCD
Dimensions: 3.16” x 5.10” x 1.04”
[80.3mm x 129.5mm x 26.4mm]
Weight: 0.4 lbs (181 g)
*See section 3.4 (pg. 4) of this operating manual for information on the effects of low flow.

866.4.Maxtec www.maxtecinc.com 7
8.0 SPARE PARTS AND ACCESSORIES
8.1 Included with Your Unit
Part Number: Item:
R211M11 Operating Manual and Instructions for Use*
RP46P11-003 Gas Sample Tubing
8.2 Optional Accessories
Part Number: Item:
R221P15 Soft Cover
* An electronic version of this Operating Manual and Instructions for Use are available on-line at
www.maxtecinc.com.
Table of contents
Other Maxtec Analytical Instrument manuals

Maxtec
Maxtec MaxN2+ User manual

Maxtec
Maxtec MAXO2+ A User manual

Maxtec
Maxtec Handi+ User manual

Maxtec
Maxtec UltraMax O2 User manual

Maxtec
Maxtec Handi+ N2 User manual

Maxtec
Maxtec MaxO2 OM-25A User manual

Maxtec
Maxtec MaxO2 Plus User manual

Maxtec
Maxtec MaxO2 Plus User manual

Maxtec
Maxtec MaxBlend 2+p User manual