mectron Piezosurgery User manual

Manuale d’uso e manutenzione
Manual of use and maintenance
Gebrauchs- und Wartungshandbuch
Notice d’utilisation et d’entretien
Manual de uso y mantenimiento
IT
EN
DE
FR
ES

52 Manualofuseandmaintenance
Table of contents
1 INTRODUCTION 54
1.1 Preface 54
1.2 Typetesting 54
1.3 Descriptionofthedevice 54
1.4 IntendedUse 55
1.5 IntendedUse 55
1.5.1 Importantsafetyrecommendations 55
1.6 Symbolsused 57
2 IDENTIFICATION DATA 59
2.1 PiezosurgeryMedicaldeviceidentication 59
2.2 PiezosurgeryMedicalhandpieceidentication 59
3 DELIVERY 60
3.1 Deliveryofthedevice 60
3.2 Standardlistofmaterialssupplied 60
3.3 Accessories 60
4 INSTALLATION 62
4.1 Safetymeasurestoapplyintheinstallationphase 62
4.2 Firstinstallation 63
5 ASSEMBLY, CONFIGURATION AND USE 65
5.1 Safetymeasurestotakebeforeusing 65
5.2 Assemblyandswitchingonthesystem 67
5.2.1 Additionalearthconnection 67
5.2.2 Powersupplycableconnection 67
5.2.3 Footswitch-stirrupopening 68
5.2.4 Connectingthefootswitchconnectortothebodyofthedevice 68
5.2.5 Insertionofthepolesforsupportingthephysiologicalsolution 69
5.2.6 Switchingonthedevice 69
5.2.7 Choiceoflanguage 70
5.2.8 Settinguptheirrigationcircuit 70
5.2.9 PiezosurgeryMedicalHandpiececonnection 72
5.2.10 Operatinginstructions 74
5.3 Congurationofthesystem 76
5.3.1 Controls 76
5.3.2 Settings 78
5.3.3 Safetymeasureswhenusing 82
5.4 Switchingoffthedevice-“SHUTDOWN” 84
6 PERIODIC MAINTENANCE 85
6.1 Idledevice 85
6.2 Powersupplycable 85
7 REPLACING THE FUSES 86
8 DISPOSAL PROCEDURES AND PRECAUTIONS 87
9 INSERTS 88
10 DESCRIPTION OF SHIELDS, ALARMS AND SOLUTIONS 89
11 PROBLEM-SOLVING 92
12 TECHNICAL DATA AND ELECTROMAGNETIC COMPATIBILITY 94
12.1 Technicaldata 94
12.2 ElectromagneticcompatibilityEN60601-1-2 95
13 GUARANTEE 99
14 PACKING 100

Manualofuseandmaintenance
EN
53
1 INTRODUCTION
1.1 Preface
Carefullyreadthismanualbeforeproceedingwithinstallationoperations,use,maintenanceorother
interventionsonthedevice.Athoroughunderstandingofthemanualisneededbeforeusingthe
devicesoastoavoidpotentialinjurytotheuserand/orpatientsand/ordamagetothedeviceitself.
Alwayskeepthismanualathand.
Important:Toavoidinjurytopeopleordamagetoproperty,carefullyreadtheparagraphson“Safety
Measures” contained in this manual. Depending on the degree of gravity, the prescriptions are
classiedwiththefollowingindications:
DANGER(alwaysreferredtoinjuriestopeople)
WARNING(referredtopossibledamagetoproperty)
The purpose of this manual is to inform the operator of the safety measures, the installation
procedures,theoperatinginstructionsandinstructionsforacorrectmaintenanceofthedevice.
DANGER
Thismanualmustbeusedinassociationwiththe“InstructionsforCleaningandSterilizing”manual
providedwiththePiezosurgeryMedicaldevice.
Ifanyanomalyisdetected,pleasecontactMectronAssistanceCentre.
Anyattempttotamperwithormodifythedevicebytheoperatorornonauthorizedpersonnelwill
invalidatetheguaranteeandrelievetheManufacturerofanyresponsibilityincaseofinjurytopeople
ordamagetoproperty.
Alltheinformationandillustrationshavebeenupdatedtothedateofissuereferredtoonthelast
page.
MECTRONhasundertakentocontinuouslyupgradeits products with possiblemodicationsand
accessories.
Inthecaseofdiscrepanciesbetweenthedescriptionscontainedinthismanualandthecomponents
ofthedevice,contactyourRetailerorMECTRON’sPost-saleService.
1.2 Type testing
AlldevicesproducedbyMectronS.p.A.arerigorouslycheckedandtypetestedineverycomponent.
Duringthistypetestingthecomponentsundergoaseriesofworkcycles.Thisprocedureguarantees
theproduct’sreliabilityinallitscomponents.
1.3 Description of the device
ThePiezosurgeryMedicaldeviceisapiezoelectricultrasounddeviceforbonecutting:thedevice
usespiezoelectricultrasoundtechnologytogeneratemechanicalmicrovibrationswhichcutthrough
mineralizedstructureswithoutdamagingthesofttissues.

54 Manualofuseandmaintenance
1.4 Intended Use
IlPiezosurgeryMedicalisapiezoelectricdeviceforbonesurgerysuppliedwithspecialinsertswhich
enablethesurgeontoapplyosteotomyandosteoplastictechniquesandtodrillholesinavarietyof
surgicalprocedures:
- Otorhinolaryngology
- Oral-maxillofacialsurgery
- Handandfootsurgery
- reconstructiveplasticsurgery.
Furthermore it is possible to execute osteotomy and osteoplastic techniques and drill holes in
neurosurgeryandsurgeryofthespinalcolumn.
Thedevicecanalsobeusedwiththeaidofvisualendoscopictechniquestocarryouttheabove
listedprocedures.
Do not use the Piezosurgery Medical device for any other purposes other than those above
mentioned.Onthecontrary,injurytothepatientsand/ortheusersortothedeviceitselfmaybe
caused.
1.5 Safety measures
Mectrondeclinesallresponsibilityfordirectorindirectinjurytopersonsordamagetopropertyinthe
followingcases:
1 Thedeviceisnotusedforthepurposeforwhichitwasdesigned.
2 Thedeviceisused not in conformitywiththeinstructions and prescriptionsdescribedinthis
manual.
3 Theelectricsystemtowhichthedeviceisconnecteddoesnotconformtotheexistinglawsand
relativeprescriptions.
4 The assembly, extension, adjustment, modication and repair operations are carried out by
personnelotherthanMectronauthorizedpersonnel.
5. The environmental conditions in which the device is kept and stored do not conform to the
prescriptionsindicatedinparagraph12.1“Technicaldata”.
6. TheinsertsusedarenotoriginalMectron.
DANGER:UsingnonoriginalMectroninsertscausespermanentdamagetothethreading
ofthehandpiecetherebyprejudicingthecorrectperformanceandriskinginjurytothepatient.In
this case the manufacturer’s guarantee and the type-approval of the device will no longer
be valid.
1.5.1 Important safety recommendations
DANGER:
Readtherecommendationsthoroughlyandrespectthesetoavoidcompromisingthesafetyofthe
patientand/oruser.Thenonobservanceoftheserecommendationsmaycauseseriousinjurytothe
patientand/oroperator.
DANGER: Qualied and specialized personnel.
The device must be used exclusively by specialized personnel opportunely trained such as the
surgeon.
Improper use may cause injury to patients and/or users and damage to the components of the
device.
WARNING:
FederalLaws(intheUnitedStatesofAmerica)limittheuseofthisdeviceonlyto,orfor,qualied
doctors.
DANGER:
Eachtimebeforeusinginspecteachcomponenttoidentifyeventualdamage.Ifdamageisdetected
donotuse.

Manualofuseandmaintenance
EN
55
DANGER: Use.
Onlyusethedeviceforthepurposeforwhichitwasdesigned(seeparagraph1.4“Useindications”).
Failuretocomplywiththisprescriptionmaycauseseriousinjurytothepatient,theoperatorand
damage/faultstothedevice.
DANGER:
TheremovablepanelplacedonthebackofthedeviceisaccessibleonlytoMectronspecialized
personnel.Usersmustnotattempttoremoveitortamperwithitasthiswouldinvalidatetheguarantee
andcoulddamagethedevice.
DANGER: Contraindications.
Do not use the Piezosurgery Medical on patients with cardiac stimulator (Pace-maker) or other
electronicimplants.Thisprescriptionisintendedalsofortheoperator.Incaseofdoubt,contactthe
produceroftheelectronicimplant.
DANGER: Interference with other devices.
Theelectrosurgicallancetmayinterferewiththecorrectfunctioningofthedevice.
DANGER: Interference with other devices.
Il Piezosurgery Medical could emit frequencies which interfere with other devices in the vicinity.
InstallthePiezosurgeryMedicaldeviceatanappropriatedistancefromlifesupportdevice.Before
operatingalwayscheckthecorrectfunctioningofalldevices.
DANGER:
Donotinstallthedeviceinplaceswherethereisariskofexplosion.Thedeviceisnotdesignedto
operateinenvironmentsinwhichthereareammableatmospheres(anestheticmixtures,oxygen,
etc.).
DANGER: Always check the device before treatment.
- Alwayscheckthatthereisnowaterbeneaththedevice.
- Beforeusingfortreatment,checkthatthedeviceisfunctioningproperlyandthattheaccessories
areefcient.
Ifanomaliesin function aredetected,do not starttreatment.If the anomaliesregard the device,
contactMectron’sauthorizedtechnicalassistance.
DANGER: Cleaning and sterilizing.
Allneworrepairedproductsaredeliveredun-sterilized.Beforeusingfortreatmentalltheseproducts
mustbecleanedandsterilizedscrupulouslyfollowingtheinstructionscontainedinthe“Instructions
for Cleaning and Sterilizing” manual provided with the Piezosurgery Medical device.After each
treatment,andbeforeusingagain,itisnecessarytorepeattheprocedureforallaccessoriesused.
DANGER:
OnlyuseoriginalPiezosurgeryMedicalpowercables,accessoriesandspares.
DANGER:
Donotactivatethedeviceifthehandpieceisfaultyorbroken.Immediatelyreplacethehandpiece.
DANGER:
Iftheperistalticpumpisfaulty,usetheotherchannel.
DANGER:
Theelectricsystemofthepremisesinwhichthedeviceisinstalledandusedmustconformtothe
lawsinforceandtherelativeprescriptionsregardingelectricalsafety.
DANGER: Inserts with water irrigation.
Checktheirrigationreleasebeforeandduringuse.Makesuretheliquidcomesoutoftheinsert’s
point.
Donotusethedeviceifthephysiologicalsolutionisnotreleasedorthepumpisfaulty.

56 Manualofuseandmaintenance
DANGER: Injury to people.
Makesurethecablesdonothinderthefreepassageofpeople.
DANGER: Injury to people.
Whentheperistalticpumpapisopen(Pic.12-Ref.C)donotactivatethePiezosurgeryMedical
footswitch.Themovingpartscouldinjuretheoperator.
WARNING:
Donotusepointedorsharpobjectsonthelcdscreen.ThiscoulddamagetheTouchScreenorthe
screenitself.Alightpressureofthenger,eveninalatexglove,allowstheoperatortochoosethe
desiredsettings.
WARNING: Contraindication.
Do not use to treat metal or ceramic prostheses. The ultrasound vibrations could cause the de-
cementationoftheprostheses.
1.6 Symbols used
WARNING,Seeinstructionsforuse
Type“B”appliedpartinconformitywithtechnicalnormEN60601-1
Serialnumber
Lotnumber
Productcode
Manufacturer
0476
Device manufactured in conformity with directive 93/42/EEC including
technicalnormsEN60601-1andEN60601-1-2.Cermetnotiedbody.
METTrademark
UL 60601-1 CSA C22.2
No. 601-1 - E113015 ConformitytonormsUL-CSA

Manualofuseandmaintenance
EN
57
Thedeviceanditsaccessoriesmustnotbedisposedofortreatedassolid
urbanwaste
Footswitchconnector
Maybesterilizedatamaximumtemperatureof135°C
AlternateCurrent
Donottouchthemovingpartswithyourngers
QTY. Quantityofproductinthepacking
Nonsterile
Expirydate
Productiondate
Sterilizedwithethyleneoxide
Disposable
Donotuseiftheconfectionisdamaged
Theproductismadewithcomponentsthatdonotcontainlatex

58 Manualofuseandmaintenance
2 IDENTIFICATION DATA
Aprecisedescriptionofthemodelandindicationoftheserialnumberofthedevicewillenableprompt
andeffectiveanswersfromourPost-SalesService.
AlwaysrefertothisdatawhencontactingMectron’sTechnicalAssistanceService.
2.1 Piezosurgery Medical device identication
EachdeviceisprovidedwithanIDtag(Pic.1)whichgivesthetechnicalcharacteristicsandtheserial
number.ThisIDtagisplacedonthebackofthedevice.Furtherinformationiscontainedinparagraph
12.1“Technicaldata”.
2.2 Piezosurgery Medical handpiece identication
TheserialnumberofthePiezosurgeryMedicalhandpieceisengravedonthehandpiece’smetalring
(Pic.2-Ref.A)inthefollowingform:NNNAANNN.
Pic.2
A
Pic.1
Manufacturer
Mectron S.p.A.
Via Loreto 15/a
16042 Carasco -GE- Italy
401 AB 000 made in Italy
PIEZOSURGERY Medical
100-240 V~ 200 VA - 50/60 Hz.
Intermittent operation UL 60601-1 CSA C22.2
No. 601-1 - E113015

Manualofuseandmaintenance
EN
59
3 DELIVERY
3.1 Delivery of the device
Thedevicecontainselectroniccomponentswhichcouldbedamagedifknockedabout,evenwhen
properlypacked.Thereforetransportandstockingmustbehandledwithparticularcare.
Donotputonecartonontopofanothertoavoidcrushingtheonesunderneath.
AllgoodsforwardedbyMECTRONhavebeencarefullycheckedbeforesending.
Thedeviceisdeliveredopportunelyprotectedandpacked.
Onreceiptofthedevice,checkforanyeventualdamageoccurredduringtransportandifso,issue
acomplainttothehaulingrm.
3.2 Standard list of materials supplied
ThePiezosurgeryMedicaldevicecomeswithaseriesofaccessories:
1xDeviceunit(Pic.3-Ref.A).
1xFootswitchwithcableandpinjack(Pic.3-Ref.G).
2xPeristalticpump(Pic.3-Ref.D)
1xElectricsupplycable(Pic.3-Ref.H).
2xBagsupportpole(Pic.3-Ref.F).
1xOperatinginstructionsmanual.
1xCleaningandSterilizinginstructionsmanual
Thislistmayvaryduringpromotionalcampaigns.
3.3 Accessories
ThefollowingaccessoriesaresoldseparatelyfromthePiezosurgeryMedicaldevice.
Sterileaccessories(disposable):
- Inserts(Pic.3-Ref.B)
- PiezosurgeryMedicalIrrigationKit(Pic.3-Ref.I).
Nonsterileaccessories(multipurpose):
- PiezosurgeryMedicalhandpiecescompletewithcord(Pic.3-Ref.E)andconnectorprotective
cap.
- PiezosurgeryMedicaldynamometrickeys(forxingtheinsertstothehandpiece)(Pic.3-Ref.C).

60 Manualofuseandmaintenance
Pic.3
A
B C
D E
F G I
H

Manualofuseandmaintenance
EN
61
4 INSTALLATION
4.1 Safety measures to apply in the installation phase
DANGER: Interference with other devices.
Il Piezosurgery Medical could emit frequencies which interfere with other devices in the vicinity.
Installthe PiezosurgeryMedical device at an appropriate distance fromthe lifesupport devices.
Beforeoperatingalwayscheckthecorrectfunctioningofalldevices.
DANGER:
Theelectricsystemofthepremisesinwhichthedeviceisinstalledandusedmustconformtothe
lawsinforceandtherelativeprescriptionsrelativetoelectricalsafety.
DANGER:
Donotinstallthedeviceinplaceswherethereisariskofexplosion.Thedeviceisnotdesignedto
operateinenvironmentsinwhichthereareammableatmospheres(anestheticmixtures,oxygen,
etc.).
DANGER:
Installthedeviceinaplacewhereitisprotectedfrombeingknockedaboutand/orfromaccidental
sprayofwaterorotherliquids.
DANGER:
Donotinstallthedeviceaboveornearsourcesofheat.
Whileinstalling,makesurethereisadequateaircirculationaroundthedevice.Donotcovertheslits
foraircirculationplacedonthebackofthescreen.
WARNING:
DonotexposethedevicetodirectsunlightorUVsourcesoflight.
WARNING:
Thedevicemaybetransportedbutitisnecessarytohandlewithgreatcarewhenitismoved.During
transport position the footswitch o the ground in such a manner that it can only be intentionally
activatedbytheoperator.
WARNING:
Assoonasthesterilizationcycleiscompletedandbeforeconnectingthehandpiecetothebodyof
thedevice,makesuretheelectricalcontactsofthehandpiececordconnectorareperfectlydry.If
necessary,drythecontactsbyblowingcompressedairformedicaluse.
WARNING:
Donotusepointedorsharpobjectsonthelcdscreen.ThiscoulddamagetheTouchScreenorthe
screenitself.Alightpressureofthenger,eveninalatexglove,allowstheoperatortochoosethe
desiredsettings.
WARNING:
Donotletthedevicebodyorthefootswitchtogetwet.Ifliquidweretoseepintothedevicebodyor
footswitch,thiscouldcausedamage.

62 Manualofuseandmaintenance
4.2 First installation
WARNING:
Removealltheaccessoriesbeforeextractingthedevicebody.
NOTE:AfterunpackingthePiezosurgeryMedicaldevice,keepthecartonsandpackingmaterial.If
thedeviceneedstobesentfromoneplaceoanotherthepackingmaterialpurposelydesignedfor
thisdevicewillprovidesuitableprotection.
In order to guarantee a perfect performance of the device, this must be installed by a Mectron
authorizedtechnician.Thedevicemustbeinstalledinasuitableandspaciousplace.
Thetechnicianmustcarryoutthefollowing5points:
1. Unpackthedeviceasshowninthefollowingpicture(Pic.4).
7
123
45
6Pic.4

Manualofuseandmaintenance
EN
63
2. Installthe2peristalticpumpsasfollows:
-Slightly push the peristaltic pump’s head towards the corresponding base (placed on the
device)(Pic.5).
Pic.5
-Gentlyrotatethepump’sheadbackwardsandforwardstomatchthetwoparts.
-Rotateclockwiseuntilthepump’sheadblockswitha“CLICK”andsettlesinaverticalposition
(Pic.6).
CLICK!
Pic.6
-Gentlypullthepump’sheadtomakesurethepumpispositionedcorrectly.
3. Explainto the user the safetymeasures tobe considered for the installationto beexecuted
correctly.
4. Fillintheinstallationform(includingthepurchaser’sdata).
5. SendthecompiledinstallationmoduletoMectronS.p.A.inordertoguaranteethetraceabilityof
thedeviceandtheactivationoftheguarantee.

64 Manualofuseandmaintenance
5 ASSEMBLY, CONFIGURATION AND USE
5.1 Safety measures to take before using
DANGER: Qualied and specialized personnel.
The device must be used exclusively by specialized personnel opportunely trained such as the
surgeon.
Improper use may cause injury to patients and/or users and damage to the components of the
device.
DANGER: Use.
Onlyusethedeviceforthepurposeforwhichitwasdesigned(seeparagraph1.4“Use”).Failureto
complywiththisprescriptionmaycauseseriousinjurytothepatient,theoperatoranddamage/faults
tothedevice.
DANGER:
DANGER:
Checkthelevelofphysiologicalsolutioncontainedinthesolutionbag.
Replacethesolutionbagwithanewonebeforeitiscompletelyempty.
DANGER: Contraindications.
DonotusethePiezosurgeryMedicaldeviceonpatientswithcardiacstimulators(pace-makers)or
otherelectronicimplantdevices.
This prescription concerns also the operator. In case of doubt, contact the manufacturer of the
implantdevice.
DANGER: Interference with other devices.
Theelectrosurgicallancetmayinterferewiththecorrectfunctioningofthedevice.
DANGER: Interference with other devices.
Il Piezosurgery Medical could emit frequencies which interfere with other devices in the vicinity.
InstallthePiezosurgeryMedicaldeviceatanappropriatedistancefromlifesupportdevices.Before
operatingalwayscheckthecorrectfunctioningofalldevices.
DANGER: Check the state of the device before using for treatment.
- Alwayscheckthatthereisnowaterbeneaththedevice.
- Beforeusingfortreatment,checkthatthedeviceisfunctioningproperlyandthattheaccessories
areefcient.
Ifanomaliesin function aredetected,do not starttreatment.If the anomaliesregard the device,
contactMectron’sauthorizedtechnicalassistance.
DANGER: Clean and sterilize the products supplied.
Allproducts(neworrepaired)aredeliveredun-sterilized.
Beforeusingfortreatmentalltheseproductsmustbecleanedandsterilizedscrupulouslyfollowing
theinstructionscontainedinthe“InstructionsforCleaningandSterilizing”manual.
DANGER: Clean and sterilize before using.
Afterusing,allreusableaccessories(handpieceanddynamometrickey)mustbecarefullycleaned
and sterilized each time before using again, as indicated in the “Instructions for Cleaning and
Sterilizing”manualsuppliedtogetherwiththePiezosurgeryMedicaldevice.
DANGER: Clean and sterilize before surgery.
Makesurethatallreusableaccessories(handpieceanddynamometrickey)havebeencleanedand
sterilizedbeforeusing.Donotuseiftheyhavenotbeensterilized.
DANGER:
OnlyuseoriginalMectronpowercables,accessoriesandspares.

Manualofuseandmaintenance
EN
65
DANGER:
Donotactivatethedeviceifthehandpieceisfaultyorbroken.Immediatelyreplacethehandpiece.
DANGER:
Iftheperistalticpumpisfaulty,usetheotherchannel.
DANGER: Inserts with water irrigation.
Checktheirrigationreleasebeforeandduringuse.Makesuretheliquidcomesoutoftheinsert’s
point.
Donotusethedeviceifthephysiologicalsolutionisnotreleasedorthepumpisfaulty.
DANGER: Injury to people.
Makesurethecablesdonothinderthefreepassageofpeople.
DANGER: Injury to people.
Whentheperistalticpumpapisopen(Pic.12-Ref.C)donotactivatethePiezosurgeryMedical
footswitch.Themovingpartscouldinjuretheoperator.
WARNING:
Donotusepointedorsharpobjectsonthelcdscreen.Thiscoulddamagethetouchscreenorthe
screenitself.Alightpressureofthenger,eveninalatexglove,allowstheoperatortochoosethe
desiredsettings.
WARNING: Contraindication.
Do not use to treat metal or ceramic prostheses. The ultrasound vibrations could cause the de-
cementationoftheprostheses.
DANGER:
Beforesurgery,makesuretheinsertiscorrectlyxedtothehandpiece.
DANGER:
Checkthattheinsertxedtothehandpieceisthesameastheoneselectedonthedisplay,otherwise
anoptimalperformancemaynotbeobtained.
DANGER:
ThePiezosurgeryMedicaldeviceisandeviceusedinbonesurgery.Howeveraprolongedcontact
and/or excessive force of the inserts on soft tissues should be avoided because these could
causeburnsand/ornoncutlesions.Useparticularcautionwhenusingsharpinserts.Aprolonged
mechanicaluseofsharpinsertscouldcutthroughthesofttissues.Whenusinginproximityofsoft
tissues/nerves(forexampletheperineviumoftheperipheralnervoussystemortheduramaterof
thecentralnervoussystem)itishighlyrecommendedtonishthecutwithabluntdiamondinsertin
ordertominimizethepotentialriskofdamagingthesetissues.

66 Manualofuseandmaintenance
5.2 Assembly and switching on the system
Beforeoperatingitisnecessarytoassemblethesystemasdescribedinthefollowingsteps:
5.2.1 Additional earth connection
- Ifrequired,insertadditionalearthingtothecorrespondingconnector(Pic.7-Ref.A).
5.2.2 Power supply cable connection
- Insertthepowersupplyconnectorintheconnectorplacedonthebackofthedevice(Pic.7-
Ref.B)andthentothewallsocket.
Pic.7
C
AB
D
E
Pic.8
A
B

Manualofuseandmaintenance
EN
67
5.2.3 Footswitch - stirrup opening
WARNING:
ThefootswitchisspecicallydesignedtobeusedonlywiththePiezosurgeryMedicaldevice.Only
useoriginalPiezosurgeryMedicalfootswitchasotherwisedamageormalfunctionmayoccur.
- Rotatethemetalstirrupuntilitisxedinverticalposition(Pic.8-Ref.A).
- Rotateandmanuallyfastenthelateralclampstosecurethestirrupinthecorrectposition(Pic.8
-Ref.B).
5.2.4 Connecting the footswitch connector to the body of the device.
- Insert the connector of the footswitch cable into the corresponding connector placed on the
backofthedeviceandmarkedwiththespecialsymbolasindicatedinPic.9.Todisconnectthe
footswitchfromthedevicebody,holdtheconnector,pressthereleasetabandpullback.
WARNING:
Donotattempttotwistorrotatetheconnectorofthefootswitchcablewhileinsertingorremoving.
Twistingmaydamagetheconnector.
WARNING:
Toavoiddamagingthefootswitchcable,connectand/ordisconnectalwaysholdingonlytheconnector.
WARNING:
Alwayscheckbeforeandafterusethatthefootswitch,thecableanditsconnectorarenotdamaged.
Ifdamageisdetecteddonotusethefootswitch.
Pic.9

68 Manualofuseandmaintenance
5.2.5 Insertion of the poles for supporting the physiological solution
- Insertthepolesforsupportingthephysiologicalsolutionintheappropriateholesinthedevice
(Pic.10).
Pic.10
5.2.6 Switching on the device
- Switchonthedevicewiththemainswitchplacedonthebackofthedevice(Pic.7-Ref.C)making
surethefootswitchisnotpressedduringthisprocedure.Thedevicethusswitcheson.

Manualofuseandmaintenance
EN
69
5.2.7 Choice of language
Aftertheentiresystemhasbeenchecked,thefollowingdisplayappears:(Pic.11).
Thersttimethedeviceisswitchedon,thepresetlanguageistheEnglishlanguage.Ifadifferent
languageisrequired:
- Pressthe“LANGUAGE”button(Pic.11-Ref.A).
- Selecttheagcorrespondingtothedesiredlanguage.
- Press“RETURN”:theselectedlanguagewillthenbethepresetlanguagewhichwillremainthe
sameuntilanewselectionismade.
NOTE:Thelanguagemaybemodiedonlywhenthethereisnohandpieceinsertedinthedevice.
Pic.11
A
5.2.8 Setting up the irrigation circuit
A. OPERATIONS TO CARRY OUT IN STERILE CONDITIONS:
DANGER:
ThePiezosurgeryMedicalIrrigationKitisguaranteedforasingleuse.Segregateanddisposeofthe
kitaccordingtothelawsinforceregulatinghospitalwaste.
DANGER:
BeforeusingthePiezosurgeryMedicalIrrigationKit,checktheintegrityofthesterilepackingand
carefullyinspecttheproducttoexcludeanyeventualdamageorfaults.Donotusetheirrigationkitif
thepackinghasbeentamperedwithordamaged.
Theirrigationkitlosesitssterileconditionifthepackingisbroken or damaged. If the packing is
damaged,disposeofthekit.Donotattempttosterilizeorreusethekit.

70 Manualofuseandmaintenance
DANGER: Contraindications.
CheckthatthePiezosurgeryMedicalhandpiecehasbeencleanedandsterilizedbeforeusing.Ifit
isn’t,donotuse.
DANGER:
ThePiezosurgeryMedicalhandpiece(neworserviced)isnotsterilewhenitissuppliedandmust
thereforebecarefullycleanedandsterilized(withtheautoclave)beforeusing.
DANGER:
AftersterilizingthePiezosurgeryMedicalhandpieceintheautoclave,waituntilithascooleddown
beforeusing.
WARNING:
Whenthesterilizationcycleiscompleteandbeforeconnectingthehandpiecetothedevicebody,
makesuretheelectricalcontactsofthehandpiece-cordconnectorareperfectlydry.Ifnecessary,dry
thecontactsbyblowingcompressedairformedicaluse.
DANGER:
After using, each handpiece-cord must be thoroughly cleaned and sterilized before using again,
followingtheinstructionscontainedinthe “CleaningandSterilizingInstructions”manualsupplied
withthePiezosurgeryMedicaldevice.
DANGER:
Alwayscheck,beforeandafterusing,thatthehandpieceanditscordarenotdamaged.Ifdamageis
detected,donotusethehandpiece-cord.
Pic.12
A
B
C
D
Other manuals for Piezosurgery
2
Table of contents
Other mectron Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Cardionics