page 8
1.5 Printout of the log _______________________________________________________
The autoclave has a serial interface jack (21) for
connection of the logging printer MELAprint 40®(or-
der no. 289). Power is also provided to the
MELAprint 40®through another connection (20) from
the autoclave.
When the printer is connected, a log containing the
process-relevant parameters will be automatically
printed out after each sterilization or disinfection cy-
cle (see example of log at the right).
It is normal here that time metering will be briefly in-
terrupted once or several times during the first sec-
onds after beginning of the sterilization period, dur-
ing heating of the pressure vessel up to sterilization
temperature.
At the bottom of the printout there is a line without
explain text with service information.
In addition, the last log to be printed out is always
permanently stored in the system memory (even if
power fails). To print out the last log again, press the
„SELECT“ button (15) and hold it down for at least
3 seconds.
In the case of a fault, a log will also be printed out to
indicate the time, the temperature, and the cause of
the fault. Section 8 below contains a list of the fault
numbers and their meanings.
1.6 Water-one-way system ___________________________________________________
A normal circulating water system, used in most
autoclaves where the steam is condensed in the wa-
ter reservoir again, has an important disadvantage.
The instruments must be absolutely free from dirt
and chemicals before sterilization/disinfection, oth-
erwise not only the autoclave but also the in-
struments will be damaged and become dirty. How-
ever, in daily practice, the instruments are prepared
insufficiently. This is critical especially for the sterili-
zation of turbines and contra-angles which have to
be oiled before sterilization. Oil, dirt and other
chemicals are getting into the apparatus and will
stay there because of the circulating system. This
can lead to spotty corrosive instruments and often to
a failure of the autoclave.
For this reason the MELAquick 12® uses the
one-way system: all water and steam used for
the process, is discharged into a waste water
container (23).
1.7 No start of program if water quality is inadequate _____________________________
You have purchased a high-quality German auto-
clave product. It is designed to ensure a long life of
service, even if the operator makes certain mis-
takes. For this reason, the MELAquick 12® has an
automatic warning system which prevents it from be-
ing operated if, for example, the operator forgets to
use distilled water and fills it with normal tap water
(which would damage it by the formation of boiler
scale). The MELAquick 12® must be filled with dis-
tilled or demineralized water (aqua dest / dem) in
accordance with DIN/EN 13060 or VDE 0510.
Please see Section 5.3 on details on water quality
and on the suppliers of the required water.
In order to prevent the instruments or the autoclave
from being damaged from formation of boiler scale
or corrosion, the autoclave has an integrated system
for testing of the quality of the water in the tank (con-
ductivity):
During the first 2 seconds after start of the cycle, the
autoclave automatically tests the quality of the water
in the water storage tank (52). If the water quality is
somewhat poorer than specified in DIN EN 13060,
but is still sufficiently good, then the lamp
„WATER“ (9) will flash to warn the operator. The
program continues to run, however. If the water
quality is not adequate, the lamp „WATER“ (9) will
flash, and the lamp „FAULT“ (10) will glow. In
such case, in order to protect the instruments and
the autoclave, the program will not start. The opera-
tor must replace the water in the tank by better (for
example, demineralzed) water and press the
„START“ (7) button again.
------------------------------------------
MELAquick 12
------------------------------------------
Date : 23.09.1994
Time : 10:37
Charge No.: 4
------------------------------------------
Program :
Sterilization 134°C
------------------------------------------
Program step Temp. Time
[°C] hh:mm:ss
Preheat Temperaure 132 10:37:39
Start 122 10:37:39
Sterilization start 134 10:39:25
Sterilization stop 134 10:39:27
Sterilization runs 134 10:39:27
Sterilization end 135 10:43:26
End of program 127 10:43:56
------------------------------------------
Program has run
succesfully !
------------------------------------------
Sterilization time : 04 min 00 s
Sterilization temp.: MIN 134°C, MAX 135°C
total Time : 06 min 17 s
------------------------------------------
8 2 09940933 3.22 C1 V1 DS1 PS0