MELAG Vacuklav 41-B User manual

Technical Manual
Autoclave
Vacuklav®41-B
Vacuklav®43-B
For Users and Service personnel
Please read the accompanying User Manual before you start operation of the auto-
clave. The instructions contain important safety precautions. Make sure to keep the
Technical Manual together with the User Manual near the autoclave. The instructions
are part of the product.

Foreword
This manual has been created for the autoclaves Vacuklav®1-B and Vacuklav®3-B. They are identical except
for their chamber depth and device depth.
The device name "autoclave" is used to designate the steam sterilizers Vacuklav®1-B and Vacuklav®3-B.
You also receive a User Manual with the autoclave. It contains important safety instructions and information
which you need to operate the autoclave. Read the User Manual completely through in proper sequence before
beginning operation with the autoclave.
This Technical Manual includes declarations of conformity, suitability statements and recommendations, instruc-
tions for setting up, installing and initial start-up of the autoclave including installation record, extended technical
information on the software and hardware and other technical data. The Technical Manual is meant for inter-
ested persons and service personnel.
Technical Manual Vacuklav®41-B Vacuklav®43-B
MELAG Medical Technology Berlin
Valid for Vacuklav®1-B, Vacuklav®3-B
as of software version 2. x
1st Edition February 2008
Responsible for the contents: Engineering Department
MELAG Medical Technology
Geneststraße 7-10
10829 Berlin
Germany
E-Mail: [email protected]
www.melag.de
© 2008-2009 MELAG Berlin
Document: Melag_Vacuklav_ 1B_ 3B_TechnicalManual_v3.doc/Revision: 3 – 09/1793
Subject to technical changes

Technical Manual Vacuklav®1-B and Vacuklav® 3-B
CONTENTS
Foreword.......................................................................................................................................................... 2
Device views......................................................................................................................................................... 8
Device views Flex-Display...................................................................................................................................9
Installation and setting up................................................................................................................................... 10
Removal from the packaging............................................................................................................................ 10
Space requirements.........................................................................................................................................11
Installation surface............................................................................................................................................ 11
Electrical connection......................................................................................................................................... 11
One-way drain..................................................................................................................................................11
Supply with → feed water.................................................................................................................................12
Installation material........................................................................................................................................... 12
Installation examples........................................................................................................................................13
Record of installation and setting up.................................................................................................................16
Please send to MELAG....................................................................................................................................16
Device and installation data.............................................................................................................................. 16
Executed work..................................................................................................................................................17
Contrast setting................................................................................................................................................19
Counter readings.............................................................................................................................................. 19
Program modifications...................................................................................................................................... 19
Electromagnetic compatibility............................................................................................................................20
Electromagnetic environment...........................................................................................................................20
Recommended protective distances................................................................................................................21
Technical tables................................................................................................................................................... 22
Precision and drift behaviour – Sensors .......................................................................................................... 22
Precision and drift behaviour –
Measuring chain...............................................................................................................................................22
Precision and drift after 1 year.......................................................................................................................... 22
Precision and drift after 5 years........................................................................................................................22
Nominal value tolerances.................................................................................................................................23
Pressure-Time-Charts......................................................................................................................................2
Quality of the → feed water..............................................................................................................................26
Frequently asked questions (FAQ) about the software....................................................................................27
Checklist for software update.............................................................................................................................42

Technical Manual Vacuklav®1-B and Vacuklav®3-B
Konformitätserklärung im Sinne der EG Richtlinie über Medizinprodukte 93/ 2/EWG
Declaration of conformity
Vacuklav®41-B
Vacuklav®43-B
In accordance with the EC- guidelines for medical devices 93/ 2/EWG
Manufacturer: MELAG oHG
Address: Geneststraße 9 – 10
10829 Berlin
Country: Germany
Product: Steam sterilizer (autoclave)
Type of device: Vacuklav®1-B/Vacuklav®3-B
Classification: Class 2a
We herewith declare that the above designated product conforms to the following
guideline:
Appendix I of the EC-guidelines for medical devices 93/42/EWG.
Notified body: DEKRA Certification Services GmbH
Handwerkstraße 15
70565 Stuttgart
No. of registration: 50199-Z2-00
The cited medical device is designated with the CE sign since 25-06-200 .
Berlin, 12-03-2008
..............................................
General Management
Quality – Made in Germany
Evidence Based Sterilization www.melag.de
2

Technical Manual Vacuklav®1-B and Vacuklav®3-B
Konformitätserklärung im Sinne der EN 13060 für Dampf-Klein-Sterilisatoren („Bescheinigung“)
Declaration of conformity
Vacuklav®41-B
Vacuklav®43-B
Declaration of conformity according to EN 13060 for small steam sterilizers
Manufacturer: MELAG oHG
Address: Geneststraße 9 - 10
10829 Berlin
Country: Germany
Product: Steam sterilizer (autoclave)
Type of device: Vacuklav®1-B/Vacuklav®3-B
Classification: Class 2a
We herewith declare that the above designated product conforms to the general re-
quirements set forth in the EN13060 Standard and has successfully passed the
Type tests according to EN 13060
for fulfilling requirements on a device with
„Class B“ – Sterilization program
(B-method)
Berlin, 12-03-2008
..............................................
General Management
Quality – Made in Germany
Evidence Based Sterilization www.melag.de
3

Technical Manual Vacuklav®1-B and Vacuklav®3-B
Eignungsbeleg nach den Empfehlungen der KHI am RKI (vom April 2006)
Certificate of Suitability
Vacuklav®41-B
Vacuklav®43-B
According to the recommendations of the Commission for Hospital Hygiene and In-
fection Prevention at the Robert-Koch Institute (of April 2006)
Manufacturer: MELAG oHG
Address: Geneststraße 9 - 10
10829 Berlin
Country: Germany
Product: Steam sterilizer (autoclave)
Type of device: Vacuklav®1-B/ Vacuklav®3-B
Classification: Class 2a
We herewith declare that the above designated sterilizer is suited for sterilization of
▪Solid instruments (wrapped and unwrapped)
▪Porous goods (wrapped and unwrapped)
▪Hollow bodies-instruments Type A (wrapped and un-
wrapped)
▪Hollow bodies-instruments Type B (wrapped and un-
wrapped)
Instructions on load quantities and loading variants are set forth in the User Manual
and must be observed.
Be sure to observe the manufacturer's instructions for medical devices intended for
sterilization according to DIN ISO 1766 .
Berlin, 12-03-2008
..............................................
General Management
Quality – Made in Germany
4

Technical Manual Vacuklav®1-B and Vacuklav®3-B
Evidence Based Sterilization www.melag.de
5

Technical Manual Vacuklav®1-B und Vacuklav®3-B
Geeigneter Prüfkörper nach den Empfehlungen der KHI am RKI (vom April 2006)
Suitable test body
Vacuklav®41-B
Vacuklav®43-B
According to the recommendations of the Commission for Hospital Hygiene and In-
fection Prevention at the Robert-Koch Institute (of April 2006)
Manufacturer: MELAG oHG
Address: Geneststraße 9 - 10
10829 Berlin
Country: Germany
Product: Steam sterilizer (autoclave)
Type of device: Vacuklav®1-B/Vacuklav®3-B
Classification: Class 2a
We herewith declare that the following test system is suited for testing the above
cited sterilizer:
Product: Helix-Test body according to EN 867-5
Type designation: MELAcontrol®
Berlin, 12-03-2008
..............................................
General Management
Quality – Made in Germany
Evidence Based Sterilization www.melag.de
6

Technical Manual Vacuklav®1-B and Vacuklav® 3-B
Device views
Front side
(1)
(2)
(3)
( )
(5)
(7)
(8)
(9)
(10)
Protective cap for slot of the →CF card
Control panel
Door (pivot opens to the left)
Protective cap for manometer to display the pressure of
the →double jacket steam-generator
Cap for emergency opening of the door
Mains switch (concealed, accessible from the side)
Ethernet-1-Data connection*
Reset button overheating protection*
Front foot of the unit (adjustable)
*behind covering
Back side
(11)
(12)
(13)
(1 )
(15)
(16)
(17)
(18)
(19)
(20)
(21)
(22)
(23)
Tank filler cap
Ethernet-2 data connection
Ethernet-3 data connection (can be retrofitted)
Power cable
Safety valve chamber
Sterile filter
Safety valve double jacket
Emergency overflow
One-way drain
Connection pressure discharge
Purified feed water inlet (for MELAdem®, swivelling
threaded fitting for hose ∅8x1, alternatively straight)
Back foot of the unit
Cooler
Fig. 1:Front- and rear view
7

Technical Manual Vacuklav®1-B und Vacuklav®3-B
Device views Flex-Display
Front side
(1)
(2)
(3)
( )
(5)
(6)
(7)
(8)
(9)
(10)
Protective cap for slot of the →CF card
Control panel
Door (pivot opens to the left)
Protective cap for manometer to display the pressure of
the →double jacket steam-generator
Cap for emergency opening of the door
Mains switch (concealed, accessible from the side)
Ethernet-1-Data connection*
Reset button overheating protection*
Front foot of the unit (adjustable)
*behind covering
Back side
(11)
(12)
(13)
(1 )
(15)
(16)
(17)
(18)
(19)
(20)
(21)
(22)
(23)
(2 )
Tank filler cap
Ethernet-2 data connection
Ethernet-3 data connection (can be retrofitted)
Power cable
Safety valve chamber
Sterile filter
Safety valve double jacket
Emergency overflow
One-way drain
Connection pressure discharge
Purified feed water inlet (for MELAdem®, swivelling
threaded fitting for hose ∅8x1, alternatively straight)
Back foot of the unit
Cooler
8

Technical Manual Vacuklav®1-B and Vacuklav® 3-B
Installation and setting up
Warning Please read the User Manual of the autoclave and the water treatment
unit before setting up the autoclave.
Warning
If the autoclave has already been put into operation, please observe the
following before setting up again in case of any eventual transport:
■ To transport the autoclave, for instance if the practice moves to a new
location, wait until the pressure indicator on the manometer display
shows null bar after the autoclave is switched off. Be sure to check
the pressure indicator on the manometer before moving the autoclave
■ To transport the unit over longer distances, an →authorised person
must prepare the autoclave beforehand according to instructions and
completely empty the →double jacket steam generator. In order to do
so, you can use the Draining program in the SPECIAL menu.
Non-compliance can lead to severe burns and damages to the autoclave
as well as malfunctions.
The autoclave sterilizes on the basis of the →fractionated vacuum. A
membraneous pump is used to create the especially deep vacuum. Thus
the autoclave can be immediately put into operation without additional in-
stallation work, apart from providing the necessary power supply. For the
optional connection of the one-way drain and an external water treatment
unit please follow the water main and general information for the proper in-
stallation which must be observed when setting up the device.
Removal from the packaging
Unpack autoclave Lift the autoclave out of the carton with the carrying strap.
Remove carrying strap To remove the carrying strap, unscrew the four screws on each side of the
housing. Retighten these screws firmly without the washers. Keep the car-
rying strap and the washers.
After switching on After switching on the autoclave and before initial start-up, open the door
and remove the trays and the accessories from the chamber.
To do this you must go into door mode:
▪ Press the Door button on the Welcome start screen; afterwards the
door can be opened.
Notice!
After 5 seconds of inactivity, the display automatically switches back
to the main menu. The autoclave then begins to suck in the feed wa-
ter needed for the sterilization, which triggers an error message!
9

Installation and setting up
Space requirements
The space requirements for the autoclave corresponds
to its dimensions plus at least 5 cm on the side and 10
cm at the back side. The autoclave should be freely ac-
cessible from above, for easy filling of the storage tank
and to ensure proper ventilation.
The autoclave works with a cooler located at the rear of
the device.
The function and service life can be impaired if the heat
dissipation over the cooler is restricted.
Therefore we do not recommend incorporating the auto-
clave; this is only possible if sufficient air circulation is
ensured.
Dimensions in parentheses: Vacuklav®3-B
Fig. 2: Dimensions Vacuklav®1-B/ 3-B
For →Feed water supply In case of external feed water supply there is additionally required space
for the water treatment unit MELAdem0 or MELAdem7.
Installation surface
Flat and level Place the autoclave on a flat and level surface.
Load capacity The Vacuklav®1-B weighs without load and without feed water 59 kg.
The Vacuklav®3-B weighs without load and without feed water 66 kg.
Electrical connection
Power socket 230 V, 50 Hz, connected load 3 00 VA, separate fuse protection 16, pro-
tection from leakage current 30 mA
Power cable The power cable is 1.35 m long.
Log printer
ELAprint®42
If you want to connect a log printer to the autoclave, you need another
socket for its power supply.
One-way drain
Wall drain or siphon drain A wall drain, nominal DN 0 or a siphon drain (sink drain) is required for
the one-way drain.
Waste water hose A 1 m wastewater hose (MELAG Art. No. 39180) can be ordered from
MELAG to directly connect the autoclave to the waste water, i.e. the pres-
sure discharge goes directly into the siphon of the domestic water supply.
One-way overflow hose on
the siphon
If you close the one-way overflow hose of the autoclave at the siphon of
the domestic water supply, you need a 2 m wastewater hose (Art. No.
36585) which you can also order from MELAG.
The drain must be located at least 30 cm beneath the autoclave and be in-
stalled dip-free with continuous descent.
10

Installation and setting up
Supply with → feed water
Feed water one-way system Since the contamination of feed water in autoclaves using a water circula-
tion system regularly leads to early damages to the autoclaves and instru-
ments, the autoclave works in the gentle yet effective one-way water sys-
tem. This means that it uses fresh purified feed water for every sterilization
run. The autoclave gets the feed water either from a water storage con-
tainer which the practice team has refilled e.g. with distilled water from the
MELAdest®65. Or it gets the feed water fully automatically from the water
treatment unit MELAdem®0 or MELAdem7.
Quality of
feed water
The quality of the →distilled or demineralized feed water for the steam
generation must be at least VDE 0510.
Provide feed water supply
from internal storage tank
Before the initial start-up of the autoclave, the right chamber of the internal
storage tank of the autoclave must be filled with feed water up to the max-
imum marking.
Notice!
At the initial start-up of the autoclave, the unit needs to be filled once with
2.5 L to fill the double jacket. For this reason we recommend you to refill
the right chamber of the internal storage tank afterwards immediately.
Installation material
Additional material, that can be ordered ELAG Art.
No.
Water connection-Set MELAdem®0 09033
Water connection-Set MELAdem®7 0903
Water connection-Set consisting of (contained in Art. No. 09033 and 0903 : 25655
Feed water-inlet fixtures for hose ∅6x1 (replaces pipe fitting for hose ∅8x1) 211 0
Hollow screw for feed water inlet fixtures 53 30
2x Cu seal 1/8“ for Art. No. 53 61 2360
2x Cu seal 1/ " for Art. No. 53 71 (replace by twisting Art. No. 53 71) 36060
Pressure discharge fitting (1/ "-fast screw connection for hose (∅8x1 (screw on at the device) 53 71
One-way drain hose PTFE-pipe ∅8/6 –cut to size 39180
One-way drain hose (overflow hose, 2m) 36585
Surface-mounted siphon with double hose nozzle 37 10
Double hose nozzle with anti-flooding flap for connection to existing sink siphon 37 00
Water tap 3/ “ (with safeguard combination) 37310
additional tap with return flow inhibitor and pipe aerator
(to attach to an available angle valve)
58130
Leak monitor (water stop) with shut-off valve and probe 01056
Safety combination, consisting of return flow inhibitor and pipe aerator according to EN 1717 9600
Table 1: Installation material
11

Installation and setting up
Installation examples
Example 1 – Standard scope of delivery Direct of →feed water from the internal
storage tank
Wastewater is collected in the wastewa-
ter chamber of the storage tank
No water connection necessary
The autoclave is supplied with feed water
directly over a hose from the internal stor-
age tank. Thus no additional water connec-
tion is required aside from the power sup-
ply line. An integrated float-actuated level
switch in the device notifies a lack of feed
water. A program can be started only after
water is refilled.
The used feed water (wastewater) is col-
lected in the wastewater chamber of the in-
ternal storage tank and later drained off
manually. A float-actuated level switch in
the wastewater chamber also notifies that
the wastewater chamber is full.
Fig. 3: Direct feed water supply from the internal storage tank
(1) Tank filler cap
(2) Sterile filter
(3) Safety valve chamber
( ) Safety valve double jacket
(5) Emergency overflow
(6) One-way drain
(7) Connection pressure discharge
(8) Feed water inlet fitting
(9) Power cable
(10) Cooler
(11) Ethernet-3-Data connection (can be retrofitted)
(12) Ethernet-2-Data connection
Example 2 – ELAdem®40 Supply of →feed water from the Ion
exchanger ELAdem®40
Pressure discharge is directly connec-
ted with siphon
12

Installation and setting up
Connection of the ion exchanger
MELAdem0 with MELAjet®. The simplest
installation is to connect the MELAdem0
directly to the purified feed water inlet of
the autoclave and thus generate deminer-
alized water from normal tap water. The
spray pistol MELAjet® is for the final rinsing
of the instruments with demineralized wa-
ter before sterilization.
Fig. : Feed water-supply from the Ion exchanger MELAdem0
(1) Safety combination, consisting of return flow inhibitor and pipe aerator according to EN 1717
(2) MELAjet® (MELAG Art. No. 30300 – optional)
(3) MELAdem®0
( ) One-way drain
(5) Feed water inlet fitting (swivelling threaded fitting for hose ∅6x1, alternatively straight)
(6) Inlet hose feed water
(7) Feed water filter MELAdem®
(8) Wall drain (available on-site)
(9) Double hose nozzle with anti-flooding flap (MELAG Art. No. 37 00 – in the surface-mounted-siphon
(contained in Art. No. 37 10)
(10) Pressure discharge hose (MELAG Art. No. 39180)
(11) Water tap (available on-site)
(12) Leak monitor (water stop) with shut-off valve and probe (MELAG Art. No. 01056 – optional),
recommended with fixed water connection
(13) Power cable
(1 ) Cooler
(15) Ethernet-3 Data connection (can be retrofitted)
(16) Ethernet-2 Data connection
Before you upgrade from the standard version to one with a water treatment unit empty both chambers
of the internal storage tank!
The feed water inlet fixture (5) on the autoclave must be exchanged with the hollow screw (MELAG Art. No.
53 30) with feed water inlet fixture for hose ∅6x1 (MELAG Art. No. 211 0) because of the other hose diameter.
Likewise exchange the two Cu seals (MELAG Art. No. 2360) – in the water connection set). The used feed water
(wastewater) is conducted under pressure directly over the one-way drain hose (MELAG Art. No. 39180) into the
siphon of the domestic water supply. An automatic one-way-overflow is also possible and is described in example
3.
13

Installation and setting up
Example 3 – ELAdem®47 Supply of →feed water from the re-
verse osmosis unit ELAdem
47
Drain via one-way drain and double-
chamber wastewater trap
Connection of the reverse osmosis unit
MELAdem®7, which can be directly at-
tached to the purified feed water inlet of
the autoclave. Other water treatment units
with corresponding water quality can also
be connected after consultation with
MELAG.
Fig. 5: Feed water-supply from the reverse osmosis unit MELAdem7
(1) Tank filler cap
(2) Safeguard combination, consisting of a return flow inhibitor and pipe aerator according DIN EN 1717
– optional
(3) One-way drain
( ) Feed water tap
(5) Feed water inlet fitting (swivelling threaded fitting for hose ∅6x1, alternatively straight)
(6) Feed water filter MELAdem®
(7) Double-chamber waste water trap (MELAG Art. No. 26635)
(8) Wall drain (NW 0 – available on-site)
(9) Double hose nozzle with anti-flooding flap (MELAG Art. No. 37 00 – contained in the surface-mounted si-
phon with Art. No. 37 10
(10) Concentrate drain MELAdem®7
(11) Inlet hose feed water
(12) One-way overflow hose (MELAG Art. No. 36585)
(13) MELAdem®7
(1 ) Water tap (available on-site)
(15) Power cable
(16) Leak monitor (water stop) with shut-off valve and probe (MELAG Art. No. 01056 – optional), recommended
with fixed water connection
(17) Ethernet-3 Data connection (can be retrofitted)
(18) Ethernet-2- Data connection
Installation as in example 2 – MELAdem®0. The wastewater drainage is automated here. The one-way overflow
hose (12) is connected to the one-way drain of the autoclave and connected over the double hose nozzle (9) with
the double-chamber wastewater trap (7). In the case of direct connection of the water treatment unit to the do-
mestic water supply, we recommend the additional Installation of the leak monitor (water stop valve) with shut-off
valve and probe (16).
14

Fax + 9(0)30 75791199 Telephone + 9(0)30 7579110
Record of installation and setting up
Please send to MELAG.
Dear Madam/Sir,
MELAG Medical Technology
Geneststraße 7 - 10
10829 Berlin
within the scope of Quality Assurance we are oblig-
ated, in cooperation with you, the responsible spe-
cialist MELAG dealer, to install this MELAG auto-
clave at the operator's in accordance with good en-
gineering practice and provide instructions in its
use.
Please copy, fill out and sign this form and send it
to us after the autoclave has been successfully in-
stalled.
The returned form is prerequisite for the
ELAG factory warranty.
Device and installation data
We, hereinafter referred to as specialist ELAG dealer, have today installed, instructed the personnel and
performed the initial start-up of the autoclave Vacuklav® 41-B/43-B as described below:
Specialist dealer (name, address, stamp)
……………………………………………………………….
Operator (name, address/ stamp)
……………………………………………………………….
E-mail-address for software update
……………………………………………………………….
E-mail-address for software update
……………………………………………………………….
Serial No. Vacuklav® 1-B/ 3-B
……………………………………………………………….
Remarks
………………………………………………………………
Please place checkmark where applicable Date
………………………………………………………………
OFirst installation OSubsequent
installation
The following persons were present during the instructions for use:
Name, first name (please print in block letters) Signature
From the practice/ clinic: ………………………………………………………...... …………………………………….
………………………………………………………….. …………………………………….
………………………………………………………….. …………………………………….
………………………………………………………….. …………………………………….
Name, first name (please print in block letters) Signature
Instructing technician ………………………………………………………….. …………………………………….
From the company: ………………………………………………………….. …………………………………….
15

Record of installation and setting up
Executed work
Please confirm the respective points with a checkmark after completion of the described task
Secure feed water supply (depending on installation variant
ORight feed water chamber of the internal storage tank filled up with feed water
OWater treatment unit MELAdem®0 or MELAdem®7 set up according to installation instructions
OIn case of installing a water treatment unit : Autoclave was set in the user menu on „water supply – extern“
OIt was pointed out that demineralized water with a minimum quality according to VDE 0510 must be used as
feed water, and the items to be sterilized must be clean and free of cleanser and disinfectant residues
Install wastewater connection (depending on installation variant)
OWastewater is collected in the internal storage tank (no wastewater connection)
OOne-way drain connected to the existing siphon of the domestic water supply or the double-chamber
wastewater trap from MELAG
Align the autoclave
For fault-free operation, the autoclave must be set up horizontally with the help of a water level located close to the
sterilization-chamber flange. Then depending on the type of autoclave, the front feet of the unit must be unscrewed
by about three or five rotations in order to give the autoclave a slight backwards tilt.
OVacuklav®1-B: front feet of the unit screwed out 3 rotations
OVacuklav®3-B: front feet of the unit screwed out 5 rotations
Conduct and record vacuum test with empty chamber and Universal-Program with 1.5 kg load (instru-
ments)
OVacuum test performed with empty cold chamber. Leakage: ……………..mbar/min
OUniversal-Program successfully executed with about 1.5 kg load (instruments)
OA printout of the program cycle was created with
the log printer MELAprint®2 and pasted into lower
fields provided for this purpose
ONo printer available – leakage rate and successful
run of the Universal-Program were recorded in the
fields provided for this purpose n
Check the time setting, if necessary reset the time according to instructions in the User anual
OTime is correct or correctly reset
Reset service meter according to separate instructions
OService meter is reset OService meter is not reset
Instructing the operating personnel and handing over the technical documentation
OInstruction on documenting with CF card to archive and print out the data has been concluded
OOperating personnel were instructed
OUser Manual was handed over OTechnical Manual was handed over
OChamber certificate was handed over
OMDD Certificate of Conformity was handed over
OWarranty Certificate was handed over
The device was properly installed by the above named technician
Owithout fault
Owith fault: ………………………………………………………………………………………..
........
Logs Vacuum test Universal program
Please paste and fold up logs here;
if no printer is available, then enter
leakage rate and successful comple-
tion of the Universal program.
16

17

Technical Manual Vacuklav®1-B and Vacuklav® 3-B
Contrast setting
You can set the contrast of the →Display in the SETUP menu. To do so, se-
lect the option Display Contrast.
Press Edit, so that reverse display appears.
Increase or reduce the contrast of the display by holding down the navigation
keys respectively.
Exit the SETUP menu by pressing MENU e it and SAVE, to permanently
save the contrast setting.
Counter readings
You can retrieve counter readings and other current technical data of the
autoclave in the INFO Menu Counter Readings.
To get to the menu, simultaneously press the keys 2 + 3.
Program modifications
The autoclave program cycles correspond to the practice-relevant require-
ments of fractionation, heating up, sterilization, pressure discharge, drying
and ventilation with their parameters of pressure, temperature and time.
Additional drying The standard function of additional drying provides an option for influencing
the program cycle.
Further changes to the program cycles are possible in individual cases and
within the scope of the guaranteed sterilization efficiency, but can only be ex-
ecuted by authorised persons. Please consult your specialist dealer or con-
tact MELAG.
18
This manual suits for next models
1
Table of contents
Other MELAG Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Lowenstein Medical
Lowenstein Medical prisma20C Service and repair manual

Roche
Roche cobas b 221 system Service manual

Braun
Braun Aesculap Aeos Instructions for use/Technical description

PathRight Medical
PathRight Medical RestoreX Instructions for use

FujiFilm
FujiFilm SonoSite Edge II user guide

Reison
Reison nordiska 10-396 user manual