2 Installation requirements
7
Dimensions Vacuklav 24 B+ Vacuklav 30 B+
Min. clearance to the rear E 5 cm
Free area with a fully-opened door F 62.5 cm 58 cm
Min. clearance to the top G 5 cm
*) including a safety combination in accordance with EN1717
Maintain the side clearances when installing the steam sterilizer.
The steam sterilizer works with a cooler on the rear of the device. The function and life-span of the steam
sterilizer can be compromized if heat dissipation via the cooler is restricted in any way. The steam sterilizer
may only be installed if sufficient air circulation can be guaranteed.
Additional space requirement for the feed water supply
Additional space is required for a storage container or a water treatment unit. It is also necessary to
guarantee free access to the hoses and cables leading from the steam sterilizer to the water treatment
unit.
Space
requirements
MELAdem 40 MELAdem 47 Storage
container
Water treatment unit Pressure tank
Width 32 cm 39 cm -- 21 cm
Height 35 cm 47 cm 51 cm 38 cm
Depth 15 cm 15 cm -- 23 cm
Diameter -- -- 24 cm --
On-site requirements
Mains supply
Implement the following safety measures when dealing with the cable and power plug:
uNever damage or alter the power plug or cable.
uNever bend or twist the power cable.
uNever remove the plug by pulling on the power cable. Always take a grip on the plug.
uNever place any heavy objects on the power cable.
uNever run the power cable over areas in which it could become trapped (e.g. doors or windows).
uNever lead the cable along a source of heat.
uNever use any nails, paper fasteners or similar objects to fix the cable.
uShould the power plug or cable suffer damage, switch off the device. The power cable or plug should
only be replaced by authorized technicians.
On-site requirements of the mains connection
Property
Electricity supply 220-240 V, 50/60 Hz
Max. voltage range 207-253 V
Building fuses separate power circuit with 16 A fuse, 30 mA RCD protection (to guarantee
continued practice operation during steam sterilizer malfunction)
Length of power cable*)1.35 m
Other Additional socket for the MELAprint 42/44 log printer etc.
*) Comply with the specifications in the connection diagram.
The mains socket must be freely accessible after set up so that the steam sterilizer can be disconnected
from the electricity supply at any time.