Milli-Q EQ 7000 User manual

User Manual
Milli-Q®EQ 7000

Table of contents
INTRODUCTION 1
System overview 2
LEDs overview 4
QUICK START 5
Exit screensaver 5
Dispense ultrapure water 5
Water Quality values 6
Dispense archiving 6
View & manage Alerts and Alarms 7
Navigate the screen interface 8
CONSUMABLE MENU 10
SETTINGS MENU 11
Manager access & password 12
System conguration 13
HMI conguration 13
Accessories 14
Alarm set points 15
Connectivity 15
Date Time & Language 16
Units 17
INFORMATION MENU 18
System 18
History 18
Flow schematic 19
Troubleshooting 19
Contacts 19
MAINTENANCE MENU 20
Depressurization 20
Consumable installation wizards 20
Cleaning & Sanitization 22
UV lamps 22
SELF HELP GUIDE 24
Depressurize system 24
Turn system o 24
Inaccurate volumetric dispensing 24
High TOC values 24
Low owrate 25
IPAK & other cartridges installation 25
REQUIREMENTS AND SPECIFICATIONS 26
Water specications 26
System specications 27
Dimensions and weight 28
Recycling 29
Ordering Information 30
LEGAL INFORMATION & WARRANTY 31

1
Congratulations!
Thank you for buying a Milli-Q® water purication system.
Milli-Q® EQ 7000 produces ultrapure water from a puried water source. Installation of this
product should be performed by a qualied service representative with access to qualied
installation documentation.
This user manual is a guide for use during the normal operation and maintenance of a
Milli-Q® EQ 7000 water purication system. It is highly recommended to fully read this manual
and understand its content before handling the water purication system.
System identication
System Catalog
number Voltage Frequency
Milli-Q®EQ 7000 with local dispenser ZEQ7000T0 100-240 Vac ±10% 50-60 Hz ±2Hz
Milli-Q®EQ 7000 with remote dispenser ZEQ7000TR 100-240 Vac ±10% 50-60 Hz ±2Hz
Manufacturing site:
Millipore SAS, 67120 Molsheim, France
For more information on your Milli-Q system, please call your local representative or visit our
website www.emdmillipore.com (North America) or www.merckmillipore.com (Rest of the World).
Intended use
The Milli-Q® EQ 7000 is intended to produce ultrapure water from a puried water source
primarily for use in a variety of laboratories worldwide.
The product is designed to produce ultrapure water with specic characteristics (refer
to the requirements and specications section), provided that it is fed with water
quality within specications and it is properly maintained as required by the supplier.
Merck KGaA does not warrant the product for any specic application. It is up to the user
to determine if the quality of the water produced matches their expectations, ts with
norms/legal requirements and to bear responsibility resulting from the usage of the water.
The product is not intended to produce: water for injection, water for dialysis, sterile water for
irrigation or injection, bacteriostatic water for injection, sterile puried water in containers, and
sterile water for injection in container or ingestion. The product is not intended to be used in
explosive environments according to ATEX Directive – equipment & protective systems intended
for use in potentially explosive atmospheres. In addition the product is not intended as a Medical
Device, including In-Vitro Devices.
INTRODUCTION

Introduction 2
Mounting options of the system
The Milli-Q® EQ 7000 oers 2 congurations:
Benchtop solution and Wall-mounted solution
Description of the system
The Milli-Q®EQ 7000 manages the production and the distribution of ultrapure water from a
puried water source such as a tank or a pressurized loop. It consists of three main units:
• The water purication unit manages the production of ultrapure water.
• The Q-POD unit manages the dispensing with Quick info for the performance of the system
through LEDs.
• The HMI unit provides the user interface.
Puried water either from a pressurized loop or a storage tank enters the purication unit.
It is rst puried by the IPAK Meta®polishing cartridge composed of ion-exchange resins,
removing the bulk of the ions. The puried water then ows to the UV 185nm oxidation lamp.
Organic molecules are partially oxidized, resulting in ions. These can then be captured in the
next stage when the puried water enters the IPAK Quanta®polishing cartridge, composed of
ion-exchange resins and synthetic activated carbon. The IPAK Meta and IPAK Quanta polishing
cartridges always need to be replaced together to ensure optimal purication down to trace
levels of contaminants. Throughout the purication steps, quality parameters are monitored
with resistivity and temperature sensors and the TOC (total oxidizable carbon) indicator. The
TOC indicator gives an indication of the total organic content measured in ppb at the point of
distribution.
System overview
Milli-Q® EQ 7000 consists of 3 main parts:
Q-POD® unit Water purication unit HMI
(Point Of Dispense) (Human Machine Interface)

Introduction 3
The Q-POD dispenser is detachable from its support. Its dispensing ‘wheel’ delivers water at 3
dierent ows: Low - Medium - High, up 2 L/min.
At the outlet of the dispenser, the nal purication is performed by the application-specic POD-
Pak. The large 7” touchscreen on the HMI oers a wide range of applications for users. It also
has a USB port to easily export data.
Note: High ow reaches above 2L/min when attached to a pressurized loop
When not in active use, the water within the system will recirculate for three minutes every hour.
This is to maintain a clean environment and to avoid contamination. The system should not be
turned o as this will stop the periodic recirculation taking place.
System ow schematics
The system can be fed either by a pressurized loop or a storage tank.
1Inlet strainer 9Strainer
2Inlet valve 10 UV lamp 185mm
3Feed Conductivity Cell (option) 11 Intermediate resistivity cell
4Internal Pressure Regulator (option) 12 IPAK Quanta®Cartridge
5Pressure Sensor 13 Product resistivity cell
6Pump 14 Point Of Dispense (POD)
73-way valve 15 Recirculation valve
8IPAK Meta®cartridge
1 2 34 5
6
7
8
9 10
11
12 13
14
15

Introduction 4
LEDs overview
On the Q-POD® unit, 3 LEDs indicate the dierent states of the system:
1LED on the Q-POD arm 3Water Quality LED
2System Power LED 4System Alarm / Alert LED
1
2
3
4
LEDs description
LED Meaning
2
The system is powered on
STEADY: The water quality ts the requirements (water
quality values are below alarm setpoints)
BREATHING: The system is in recirculation mode
A water quality alarm is raised
An alert or alarm is raised
An high importance alarm is raised: the system is stopped
for safety reason
Note:
- The 3 LEDs light up when powering the system on.
- The 3 LEDs are breathing in case of maintenance (see page 20)

5
QUICK START(after installation performed by a qualied Field Service Representative)
Exit screensaver
If the setting is active and there has been a certain time of user inactivity, the screensaver will
be displayed on the HMI touchscreen.
Screensaver
Tap anywhere on the screen or initiate a manual dispense by pressing the dispenser wheel to
exit the screensaver.
Home screen
Status bar
Home Menu
Dispense ultrapure water
It is a good practice to always recirculate water before dispensing.
To do this, press on the recirculation icon, this also refreshes the water quality
parameters:
• Product resistivity
• TOC
• Temperature
There are several options to dispense ultrapure water:
Manual Dispensing
To start dispensing in low ow, turn the dis-
pense wheel one notch counter-clockwise.
+
To start dispensing at full ow, press down on
the wheel.
Adjust owrate Stop
To increase the ow rate keep turning the
wheel counter-clockwise until full ow is
reached. To decrease the ow rate, turn the
wheel clockwise.
Low ow Full ow
_
+
To stop any ongoing dispensing, press down
on the wheel or rotate the wheel clockwise
until no more water ows from the dispenser.
Recirculate

Quick start 6
Volumetric Dispensing (100mL to 10L)
Enables a user to auto dispense a pre-selected volume.
Note: When connected to a storage tank, always make sure there is enough water
present before dispensing. This information is always displayed in the status bar:
Select the volume to dispense, press ok to
validate. The input value can be customized,
go to page 14 Settings > System conguration
Press the Play button to start dispensing the
pre-selected volume.
The system will automatically stop once the requested volume is reached.
Water Quality values
When the dispensing is nished, the resistivity and temperature are instantly displayed on
the left side of the HMI. Ninety seconds later a new TOC value at the point of distribution is
indicated.
The corresponding LED lit green if the water quality te the requirements.
If the calculated TOC value is from 0 to 5
ppb, ≤ 5 ppb will be displayed
If the calculated TOC value is from 6 to 10 ppb,
≤ 10 ppb will be displayed
For values upper to 10 ppb, a rounded value will be displayed.
Dispense archiving
Water quality values of each dispense are automatically archived and can be retrieved in the
history report section available in the information menu. Go to page 10 Information > History
for further information.

Quick start 7
View & manage Alerts and Alarms
Alerts give you advanced warning when maintenance is required and Alarms notify you when
the system has encountered a technical issue.
Yellow LED is displayed when an alert is raised.
Red LED is displayed when an alarm is raised.
Press on the alert or the alarm
1
icon
present in the bottom bar to display the alert(s)
or/and alarm(s) which are active.
Note: The notication number indicates how many active alarms are present. When there are
no active alerts or alarms, the icon disappears.
Select a specic alert or alarm to get further
information and better understand the root cause.
Advice is provided in the Information tab to help
the user troubleshoot.
If the issue is not resolved, select the Contact
tab to nd the technical support hotline number
for further assistance.
To protect the system and user applications, high-
importance alarms will automatically stop the
system from dispensing water.
Note: These are displayed on the screensaver to
alert the user when the system is not in active
use.
Alerts provide advanced warning of any
maintenance actions that will be required. These
can be “closed” to be reminded again in 24h or
can be “cleared” permanently.
An Order tab will be present containing ordering
information when applicable.

Quick start 8
Navigate the screen interface
The HMI 7” touchscreen allows the user to control and monitor the water system. All
sub-menus and applications can easily be reached in a few clicks or swipes, similar to a
smartphone.
HOME SCREEN
MAIN MENUS
Home
- Water quality information
- Dispensing function
- Alerts & Alarms
(when applicable)
Menu
- Consumables
- Information
- Settings
- Maintenance
- Login
Go to
Consumables
Menu
Go to
Settings
Menu
Go to
Information
Menu
Go to
Maintenance
Menu
Go to
Manager access

Quick start 9
Apps
Indicates how many screens are present and where you are.
Swipe left & right anywhere on screen to switch between screens.
Pops-up
When you click in a zone of text, the
keypad will automatically appear.
To remove the keypad click on or
on any zone outside the keypad.
Note: When in a menu or app level screen, 1 minute of inactivity generates a timeout back
to the home screen.
When in a pop-up level screen, there is no timeout, therefore the user remains on this screen
until the pop-up is manually closed.
•••
When there
are multiple
lines, a
scroll bar is
present to
swipe up &
down

Consumable Menu 10
CONSUMABLE MENU
Viewing Consumables Status
Color icons and color LEDs allow to instantly view the status of all installed consumables.
Press on the
•••
icon to get the
consumable ID, ordering information and a
description of the purication technology.
Note: Certicates of quality are now available online www.mymilliqconsumables.com
All cartridges have been designed to be easily replaced by a user. A shortcut is available to the
replacement wizard by clicking on the “replace pack” button in the pop-up.
Green:no action required
Yellow: notice for a maintenance action
Red: maintenance required!
LEDs

11
Dierent settings are available to customize the water purication system operation.
Depending on the user prole (end-user operator / system manager) some of the settings will
have either read-only or read & write access .
SETTINGS MENU
Settings End-user System
Manager
System
conguration
Lab close
Recirculation Read only Read only
Safety stop Read only
Foot pedal Read only
Screen saver Read only
Tank/Loop setting Read only Read only
Water sensor Read only
Feed quality measure Read only
Alarm relay activation Read only
Alarm relay conguration
Custom Volume
HMI
conguration
Brightness
Buzzer Read only
Alarm
conguration
Product resistivity Read only
Product TOC Read only
Inlet strainer clean Read only
IPAK Polishing cartridges
Millipak
Millipak Gold
LC-Pak
VOC-Pak
EDS-Pak
LC-Pak 0,1µm
0,1µm Pak
Connectivity Local network Read only
Date, Time & Lan-
guage
Time zone Read only
Date Read only
Time Read only
Language Read only
Password Activate / Change password
Units Resistivity / Conductivity Read only
Temperature Read only
Tank volume Read only
Temperature compensation Read only
Pressure Read only

Settings 12
Available apps:
°C / °F
Note: Dispensing while in this menu is not available.
Manager access & password
Login as a Manager
1. Press on the menu button from the Home page.
2. Press on Login.
3. Enter system manager password.
To display & see the password while typing, click on the “Show code” checkbox.
Once logged-in, the manager icon is displayed in the top left corner.
If a manual logout is not performed, the system manager prole will be automatically logged
o after 1 hour.
Note: By default this parameter is activated. The default password is PASS
If the system manager password is forgotten, contact the technical support hotline.
Log o as a Manager
1. Press on the menu button .
2. Press on “Logout”.
Modify system manager settings
This can only be done when logged-in as a
manager.
In the Password app:
1. Deactivate/re-activate the manager password
through the password activation slider button.
2. If required, change the password.
Important! Deactivating the system manager prole will allow any user user to have read &
write access on all settings.
System
Conguration
Date, Time &
Language Password
Alarm
Conguration Connectivity Units
HMI
Conguration

Settings 13
System conguration
Lab close
When leaving the lab for long periods of inactivity, this mode can be activated to save energy
and reduce wear of system components. Recirculation is reduced to once a day. When it is 24
hours prior to resuming lab activity, the system automatically resumes a recirculation every
hour, ensuring it is ready for use.
1. Activate Lab close via the slider button.
2. Enter a Start and End date (minimum of 3 days). Lab close mode will be initiated at 00:01
on the selected date.
Safety ow stop
A precautionary measure that stops a Q-POD dispensing after having continuously delivered
water for a certain duration.
Adjust to the preferred time by using the arrows or click on the box to access the keypad.
By default the timer is set to 15 minutes and is active. This setting will not impact volumetric
dispensing functions.
Screensaver
This is the maximum duration of inactivity on the HMI before the screensaver starts.
1. Activate / deactivate via the slider button.
2. Adjust to the preferred time by using the arrows or click on the box to access the keypad.
By default the screensaver is active and the timer is set to 5 minutes.
Custom Volume
This is the volume you can customize to be able to dispense with the Volumetric Dispensing
function.
The volume can be set from 100 to 25000 mL in 100 mL increments.
HMI conguration
This app allows a user to congure all parameters specic to the HMI.
Screen brightness
Adjust to the preferred brightness from 1 to 7 using the arrows or click on the box to access
the keypad and type-in the value.
Buzzer
The POD can emit a sound when alerts/alarms are triggered. This can be activated or
deactivated using the slider button. By default, the sound is deactivated.

Settings 14
Accessories
Water sensor
A water sensor can be connected to the system unit to stop
the system from producing water in case of water spillage.
Foot pedal
The pedal is connected to the system.
In the HMI conguration app, activate the foot pedal via the
slider button.
Dispense using the foot pedal
1. Press once and release to dispense in full ow.
2. Press once to stop the ow dispensing.
Feed water conductivity sensor
A conductivity sensor can be installed inside the system to
monitor the feed water quality entering the system from the
loop. If bad water quality is detected, an alarm that stops
water production will appear, informing the user that the
system is at risk and to check the feed water quality. This
will protect the downstream purication technologies in case
of a problem. This accessory is not meant to be used with
deionised feed waters.
Inlet Pressure Regulator
This accessory will allow to regulate the pressure of the
water entering the system from the loop. This will protect the
downstream elements from high water pressure. It is highly
recommended to use this regulator instead of an external
pressure regulator on the feed water tubing.
Alarm relay
This accessory will allow to relay all, or a selection of
alarms to a remote device.
Select the alarm(s) that will activate the relay.

Settings 15
Alarm set points
Adjust the alarm set points by either clicking on the arrows or alternatively, click in the box to
access the keypad and directly type-in the value.
Water quality
• Product resistivity/conductivity
• Product TOC
Inlet strainer clean
This is only applicable when set-up in Loop conguration. It is recommended to clean the inlet
strainer once a year to avoid clogging. Depending on the feed water and its particle concentra-
tion, the inlet strainer clean frequency can be adapted.
Cartridge lifetime
Cartridges should be changed when the system alerts the user.
Saturated pack will start to let ion and organic contaminants through. This is called breakthrough
and often starts with traces that cannot be detected by on-board monitoring. Some technologies
will clog, creating a back pressure which can result in ow-rate issues.
In validated settings, to adjust lifetimes according to existing SOPs, click on the applicable
lters and adjust accordingly.
• IPAK polishing cartridges
• POD-Paks
Connectivity
The system oers the possibility to be connected to a laptop with a xed IP address or to your
local network (DHCP protocol/xed IP address) via an Ethernet port.
The IP address of the system can be seen by swiping the screen top to bottom.

Settings 16
The DHCP (Dynamic Host Conguration Protocol) is an “automatic conguration” of a device anytime
it connects to an IP Network. This “automatic conguration” is called allocation. The system is
automatically recognized and an IP address is assigned to it. and congured so that the networks
resources can be used. Check with your local IT resources which type of connection is best suited to you.
Change network connection settings
1. For a direct connection, change the network settings if necessary
(default IP address: 192.168.1.69).
2. For network connection, activate DHCP via the DHCP activation button.
Duplicate user interface in a browser
Once connected:
1. Go to your browser. For best browsing performance, Chrome® is recommended.
2. Enter the IP address of your system which can be found in the connectivity pop-up screen.
Note: The view from a laptop enables a user to view the HMI display remotely. Dispensing
operations cannot be conducted remotely for safety reasons.
Date Time & Language
The norm ISO®8601 has been used as a reference for a standardized way of presenting dates
and times.
Date
Calendar date representations are shown in the form “2020-07-06” [YYYY-MM-DD]. [YYYY]
indicates a four-digit year. [MM] indicates a two-digit month of the year, 01 through 12. [DD]
indicates a two-digit day of that month, 01 through 31.
Start by setting the year:
1. Click on the month & year title, this displays months and only the year as a title. Click again
on the year title to display years.
2. Select the year, this then displays months.
3. Select the month, this then displays the days.
4. Finally select the day to conrm the date setting.
Time ZONE
1. Select the continental or oceanic zone you are in on the left. This updates the right-hand
side with all the major cities within this time zone.
2. Select the appropriate city. If you have not found a city, make sure you are in the right time
zone.
The Milli-Q® system is now congured to display current local time and automatically updates
when daylight saving time applies.
Time
The time is set and displayed in a 24h clock format. A time of day is written in the 24-hour
notation in the form hh:mm (for example 14:23), where hh (00 to 23) is the number of full hours
that have passed since midnight, mm (00 to 59) is the number of full minutes that have passed
since the last full hour.
Language
There are 9 languages available: Chinese / English / French / German / Italian / Spanish /
Portuguese / Japanese / Russian

Settings 17
Units
Units can be formatted to local needs:
Resistivity / Conductivity MΩ.cm µS/cm
Temperature °C °F
Storage tank level L %
Pressure Bar / kPa / psi
Select and press on the unit of interest. The selected unit will appear in blue. Press “OK” to
conrm selection and exit the pop-up.
Temperature compensation mode
It is possible to show non-temperature compensated resistivity or non-temperature compensated
conductivity. Temperature compensation is a way of standardizing resistivity or conductivity to
measurements that would be seen if the water temperature was 25°C.
Select the preferred mode among the 3 modes of temperature compensation available:
TC1
By default TC1 mode is selected.
The resistivity or conductivity values are temperature compensated to 25°C.
The values are normalised. The system rmware eliminates small uctuations of
temperature compensated resistivity or conductivity due to the fact that these two
parameters are not measured exactly at the same time.
TC2
The actual temperature compensated resistivity or conductivity values are
displayed. The TC2 setting should be used in applications that require the detection
of trace ionic levels or when performing the verication of the system resistivity
meter operation with an independent calibrated resistivity meter. In some operating
conditions the feed water can be warmer or cooler than the water temperature
inside the system. As a result, this can cause small uctuations of the resistivity
and conductivity values. Resistivity values could uctuate between 18.0 Mohms.cm
and 18.4 Mohms.cm @25°C while the actual resistivity is 18.2 Mohms.cm @25°C.
NTC
Non Temperature Compensation. The temperature compensation is o. The displayed
resistivity or conductivity is not temperature compensated. The temperature of the
water is shown at the same time as the non-temperature compensated resistivity
or conductivity value.

18
INFORMATION MENU
This menu contains useful system information and provides a system status. Dispensing is
available while in this menu.
Available apps:
System
Find system details
Select System ID to nd the information on the
water system model and installation details.
History
2 user reports exist:
• Daily quality measures report
Each row represents a day (24h) and is dated. It contains daily averages for product
resistivity, temperature and TOC. Also included is the total volume dispensed during the day.
• Dispensing events report
Each row is a dispense operation. It is an archive of all the dispense reports.
To export data or preview a graph:
1. Select a time period of interest with a start and
end date.
2. A graph pre-view of the last 30 records is
available on the HMI screen.
3. Export the report in .ods (open document
source) format to visualize the full data and
integrate them into a data management system.
Note: A complete history report with all system activity is available. By default it is only
available when logged-in as the system manager. This can be modied by de-activating the
system manager prole in the settings menu, go to page 13 Settings > Manager access.
System
Contacts
History Troubleshooting
Flow schematic
Table of contents
Other Milli-Q Water Filtration System manuals
Popular Water Filtration System manuals by other brands

BGS technic
BGS technic 1806 instruction manual

Kenmore
Kenmore 625.344800 owner's manual

Pentair
Pentair GS-210RO-MF Installation and operating instructions

Diesel Performance Products, Inc.
Diesel Performance Products, Inc. FASS FF-1001 installation manual

SUNSET
SUNSET CAP1007S-KIT instruction manual

Eden
Eden 304S user manual