
Mindray North America 800 MacArthur Blvd. Mahwah, NJ 07430 Tel: 201.995.8000 www.mindray.com 0002-08-1659 Rev. A
Page 12 of 19
11. Perform the following Internal Gas Connections Test:
a) Close and remove all gas cylinders from the A5.
b) Connect only the AIR line pressure hose to the A5 from the wall supply. Leave all other line pressure
hoses disconnected.
c) With the A5 powered ON, rotate the AIR flow control knob to ensure a continuous flow increase
throughout its full range.
d) Fully rotate the N2O flow control knob and verify that there is no flow.
e) Fully rotate the O2 flow control knob and verify that there is no flow.
f) Disconnect the AIR line pressure hose from the A5 and connect the O2 line pressure hose from the
wall supply, rotate the O2 flow control knob to ensure a continuous flow increase throughout its full
range (0 to 15 L/min).
g) Fully rotate the N2O flow control knob and verify that there is no flow.
h) Fully rotate the AIR flow control knob and verify that there is no flow.
i) Connect the N2O line pressure hose from the wall supply. With the O2 flow control knob fully opened,
rotate the N2O flow control knob to ensure a continuous flow increase throughout its full range. Skip
this step if N2O supply is not available
j) Fully rotate the AIR flow control knob and verify that there is no flow. Skip this step if N2O supply is
not available
k) Close all 3 flow control knobs and reconnect the AIR line pressure hose.
12. Perform the following VCV Adult Ventilation Mode Test:
a) Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
b) Set the mechanical Auto/Manual switch to AUTO.
c) Connect a Vent Tester to the breathing system. Unidirectional Vent Testers need to be connected to
the expiratory limb.
d) Set the ventilator controls to the following:
i. Ventilation Mode: VCV
ii. Vt: 600
iii. Rate: 8
iv. I:E: 1:2
v. Tpause: 10
vi. PEEP: Off
vii. Plimit: 50
e) Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp. rate and minute
volume values appear on the screen.
f) Verify the Tidal Volume display on the Vent Tester is within 12% (±72 mL) of the set value within
approximately 1 minute from the start of ventilation.
g) Verify the Tidal Volume display is within 14% (±84 mL) of the set value within approximately 1 minute
from the start of ventilation.
13. Prominently label the unit that the drive gas has been modified and the system may be used.
Please follow the steps below to convert the drive gas of Mindray A5 Anesthesia delivery systems from air to
oxygen.
1. The systems should be converted back to using oxygen as the drive gas by performing the above steps in
reverse.
2. Perform the Pneumatic Leak Tests per chapter 3.9 of the service manual