Tethys D60 - Tethys T60 - Tethys T45 SHORT-FORM MANUAL
4COMMISSIONING INSTRUCTIONS
Once the appliance has been installed correctly, it must be prepared for use:
Set the current date and time.
Fill up with regenerating salt.
Add detergent and neutralising agent and any other chemicals.
Run program “16 Service”, preferably twice in succession, to operate the peristaltic pumps and prime the
detergent inlet lines correctly.
4.1 USING THE WATER SOFTENER
The appliance is equipped with a water softener which removes the substances that cause hardness from the inlet
water.
fig. 7 –The salt tank is accessible with the door open; filling the softener tank with regenerating salt.
The softener tank contains about 1 kg of coarse salt. The salt tank is in the bottom of the inside of the instruments washer
chamber. Remove the bottom basket, unscrew the cap of the tank by turning it anticlockwise and pour in the salt using the funnel
supplied; take care not to spill salt in the tank. Remove any salt residues before screwing the cap back on.
ATTENTION
The first time the instrument washer is used, a litre of water must be poured into the tank with the salt. Whenever the
tank is filled, always take care to ensure that the cap is firmly closed. The mixture of water and detergent must not get
into the salt tank, as this would impair operation of the regeneration resins.
Only use coarse salt for domestic dishwashers. Do not use cooking salt.
Fill up with salt when necessary before starting a washing program, so that any excess salt solution will immediately be
removed; if salt water is left to stand inside the tank, it may cause corrosion. Carry out a prewash cycle to prevent this if
necessary.
4.2 USING DETERGENT AND NEUTRALISING AGENT
The appliance is equipped with detergent dispensing appliances, depending on the model in use (refer to the technical data of the
relevant model).
1. Pump P1, dispensing of neutral or weak alkaline liquid detergent
2. Pump P2: dispensing of acid neutralising agent.
3. Pump P3: always optional, never installed by default. Auxiliary detergent pump.
Key: feature installed, optional feature, - feature not available for installation
ATTENTION
If the can level sensors are not fitted: check the levels of product in the cans / bottles to ensure that programs
are not performed without detergent or neutralising agent.
During installation, or when a completely empty drum of product is replaced, run program no. 16 SERVICE to
charge the system with liquid. The pipeline from the drum to the pump must be filled to ensure that the product is
dispensed properly during subsequent washing cycles.
4.2.1 LIQUID DETERGENT INTAKE SYSTEM
A - Configuration with LEVEL SENSOR (standard on Tethys D60, optional on Tethys T60, Tethys T45).
1. Suction nozzle with integral level sensor and tapered rubber support for fitting onto the can.
2. Silicone hose for connection between intake pipe and peristaltic pump.
3. Detergent intake filter, fitted directly on the suction nozzle pipe.
B - STANDARD configuration for Tethys T60, Tethys T45:
1. one stainless steel intake pipe, for drawing in detergent from the can;