8TUA60/TUA70
1 Installation
Installation of this product should be carried out by a specialist technician (by MORITA CORP. or by a company
designated by us).
The installation must be in accordance with the supplied installation control standard.
●There is a danger of accidents and faults due to unforeseen circumstances.
Always ensure that this product is fixed to the floor using an appropriate method (stated in the supplied installa-
tion control standard).
●If not properly fixed, there is danger that the product will topple over and cause an accident or fault.
Do not use a power supply other than one of the stipulated power supply frequency [Hz], voltage [V], and permis-
sible current [A] (or power consumption).
Always ensure that a dedicated power supply circuit is provided for each unit, avoid sharing the power supply with
other appliances and avoid a common/branching power supply.
●There is a danger of accidents and faults such as fire/electric shock etc.
Do not locate the unit where it is likely to get wet or in a location that gets wet.
●There is a danger that a short circuit or overheating may occur and cause a fault or accident such as fire/electric
shock.
Do not install in a location where chemicals are stored or where gas is generated.
●There is a danger of accidents and faults such as fire/electric shock etc.
Do not position the power cord, ground wire, or foot control cable in a location where damage may easily occur
(door, passageway etc.).
●There is a danger of accidents and faults such as a short circuit/electric shock etc.
Do not install the unit in an unstable location, such as on a slope or on a site subject to vibrations or impacts.
●This may cause a fault.
Do not install the unit in a location where there is a possibility of adverse eects occurring due to atmospheric
pressure, temperature, ventilation, sunlight, dust, salinity, sulfur/ozone in the air, etc.
●This may cause a fault.
Do not install this unit in the vicinity of equipment that generates electromagnetic waves such as communication
equipment, elevators etc.
●This may cause a malfunction.
Keep heaters away from the device. Also, make sure that it is not exposed to warm air directly.
●Heat softens and expands the tube, causing breakage or other damage.
Do not install this unit in a location which does not meet the working environment conditions stipulated below.
①temperature: 10-35°C
②RH: 30-75% (without condensation)
③Air pressure: 700-1,060 hPa
●This may cause a fault.
1.1 Installation
1.2 Preparation
Use tap water to supply the water line of the main unit.
●The use of cleansers, industrial water, unhygienic water etc. may damage the patient's health. Moreover, this
may damage the pipes inside the main unit and may cause a fault.