4 Method of operation
Do not place objects or bring fingers, feet, or footwear near to movable parts such as the seat and backrest hinge of the
main unit. (refer to the List of Precautions)
• There is a risk that items will be caught during operation and accidents such as injury and damage/faults may occur.
Do not place heavy objects on the edges of the seat/backrest/headrest or on the light/tray etc. and do not apply exces-
sive forces to these areas.
Avoid behavior such as standing/jumping on the main unit.
• There is a danger of the main unit toppling over or incurring damage and causing an accident or fault.
Do not spill liquid inside the main unit and do not insert easily combustible items and metals.
• If the unit continues to be used as is, there is a danger of accidents and faults such as fire/electric shock etc.
Do not place a heat source such as a heater near the main unit.
• There is a risk of deformation and damage to the main unit and rupture of the piping due to residual air pressure.
Observe the following points during operation of the main unit and refer to the “List of precautions” in the next section
to ensure safety.
1) Apart from doctors and assistants, no-one should operate the main body without permission (as a mischievous prank
etc.)
2) Keep an eye on patients (especially infants and the elderly)
3) Do not allow patients to adopt a dangerous posture (put their legs on either side of the seat, stand on the seat, kneel
on the seat, sit while wearing footwear such as heels and slippers etc.)
4) Do not let movement/posture exert a heavy burden on patients (such as a sudden headrest operation for patients with
a cervical vertebrae disorder or adopt a supine posture while suering back pain etc.)
5) Do not allow hands, parts of the body/belongings etc. to come close to the movable parts of the main unit and vicin-
ity (do not allow the patient to grasp the tray arm, place hands at the base of the backrest, sit with a phone in their
pocket, etc.)
6) Do not allow more than one patient on the seat (holding a child etc.)
7) Do not let anyone unrelated to treatment enter the surrounding area (such as a child approaching/touching the unit)
8) Do not move the small tray etc. close to the surrounding walls/windows/equipment/appliances etc. (collision may
occur during operation etc.)
9) Do not behave in a way considered to be dangerous
10) Do not operate the chair in a situation that is considered to be dangerous
• There is a danger of accidents causing injury etc., damage/faults occurring.
If there is a risk of injury or a collision during automatic operation of the main unit, carry out the following stop procedures
immediately.
(A) Step on the foot pedal
(B) Operate the manual operation lever (either up, down, upright, or reclined)
(C) Turn the main switch OFF
• There is a danger of accidents causing injury etc., damage/faults occurring.
If a patient uses a pacemaker, do not simultaneously use the ultrasonic scaler, electrocautery scalpel and root
canal length measuring instrument.
•
The pacemaker may malfunction and damage the patient’s health.
Always be sure to turn the main switch OFF when using an electrocautery scalpel near and around the main unit. Re-
frain from using devices generating electromagnetic waves such as cell phones etc.
• Malfunctions arise due to electromagnetic waves and an accident/fault may occur due to unforeseen circumstances.