Morsø 2B Classic 2020 User manual

CERTIFICATION TEST REPORT
Report no.:
300-ELAB-2526-EPA Rev4 110821
Page 1 of 49
Init.: JSA/
Order no.: 946206
No. of appendices: 35
Requested by:
Company: Morsø Jernstøberi A/S
Address: Furvej 19
Postcode/town: DK-7900 Nykøbing Mors
Country: Denmark
Web: www.morsoe.com
Product:
Wood heater type: Morsø 2B Classic 2020
Sample:
Receipt at DTI, Aarhus: 1. September 2020
Test period:
Date of testing: 2.-4. September 2020
Procedure
Testing of a wood heater in accordance with DTI method “ELAB-PP-BR-15” based on
a relevant selection of standards and methods:
ASTM E2515-11
Yes
ASTM E3053-17 (Cordwood)
Yes
US EPA Method 28R in combination with ASTM E2780-10 (Cribwood)
No
CSA B415.1-10
Yes
EPA Communication on alternative method for Cordwood testing
Yes
Result:
The stove/ meets the requirements of NSPS §40 CFR Part 60.
Remarks:
See paragraph 2 - Remarks.
Terms:
Accredited testing was carried out in compliance with international requirements, and the
general terms and conditions of The Danish Technological Institute. The test results apply to
the tested products only. This test report may be reproduced in extract only if the laboratory
has approved the extract in writing. Danish Technological Institute is an EU Notified Body with
identification number 1235 and DIN Certco test laboratory, PL 168.
Issued:
11. August 2021, Danish Technological Institute, Aarhus, Stoves&Boiler test lab
Signature:
Jes Sig Andersen Max Bjerrum
Senior Specialist Quality Assurance
Test reg. no. 300
Jes Sig Andersen
2021.08.11 11:06:34 +02'00'

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 2 of 49
300-ELAB-2526-EPA Rev4 110821
1. Introduction.........................................................................................................................4
General ...........................................................................................................................4
Report revision history overview .........................................................................................4
Revision 1 changes ...........................................................................................................4
Revision 2 changes ...........................................................................................................4
Revision 3 changes ...........................................................................................................4
Revision 4 changes ...........................................................................................................5
List of Issues found (ref. Sanzhez email of June 10. 2021).....................................................5
Scope of testing................................................................................................................8
Site.................................................................................................................................8
Participants ......................................................................................................................8
DTI staff ..................................................................................................................................8
Client staff ...............................................................................................................................8
Test specimen ..................................................................................................................8
Description of the wood heater ...........................................................................................9
2. Aging prior to testing .......................................................................................................... 10
3. Summary of test results......................................................................................................11
Test schedule ................................................................................................................. 11
Main results ................................................................................................................... 11
Summary of the CS+HF and LF tests the 02-09-2020.......................................................... 11
Duct traverse data 2. September 2020 ..............................................................................15
Summary of the CS+HF and MF tests the 03-09-2020.........................................................16
Duct traverse data the 3. September 2020......................................................................... 19
Summary of CS+HF test the 4-09-2020............................................................................. 20
Duct traverse data the 4. September 2020......................................................................... 22
Anomalies...................................................................................................................... 22
4. High Fire net fuel consumption and burn rate calculations .......................................................23
Summary of test results .................................................................................................. 24
ADEC 18 AAC 50.077 Compliance (Fairbanks proof) ............................................................24
CSA HF1 PM report (#1) ..................................................................................................25
CSA HF1 CO, HO, EFF report (#1).....................................................................................26
CSA LF report (#2) ......................................................................................................... 27
CSA HF2 PM report (#3) ..................................................................................................28
CSA HF2 CO, HO, EFF report (#3).....................................................................................29
CSA MF report (#4)......................................................................................................... 30
CSA HF3 PM report (#5) ..................................................................................................31
CSA HF3 CO, HO, and EFF report (#5)............................................................................... 32
Weighted avg. calculation (LF#2, HF2#3, MF#4, HF3#5).....................................................33
Test facility conditions ..................................................................................................... 34
Fuel properties ............................................................................................................... 34
Summary of test fuel load properties................................................................................. 35
5. Test accomplishment .......................................................................................................... 36

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 3 of 49
300-ELAB-2526-EPA Rev4 110821
Remarks ........................................................................................36
Start-up operation ..........................................................................36
Sampling arrangement ....................................................................................................36
Fluepipe and chimney configuration................................................................................... 37
6. Sampling methods.............................................................................................................. 38
Particulate extraction system............................................................................................ 38
Calculation of PM emission ............................................................................................... 38
7. Quality assurance............................................................................................................... 38
Instrument calibration .....................................................................................................38
Logger data.................................................................................................................... 38
Morsø’s Quality Assurance Plan......................................................................................... 38
8. Documentation material ...................................................................................................... 39
9. Remarks ........................................................................................................................... 40
Internal correction of gasmeters ....................................................................................... 40
Unitary DTI Cribwood and Cordwood spreadsheet ............................................................... 40
Request of restriction of the air valve action....................................................................... 40
Duplicated High fire calculations .......................................................................................40
10. Discussion of Results .......................................................................................................... 40
11. Main results....................................................................................................................... 41
12. Test details........................................................................................................................42
Pre-conditioning.............................................................................................................. 42
Data, HF1 (#1) test run 2nd of September 2020..................................................................43
Data, LF (#2) test run 2nd of September 2020 .................................................................... 44
Data, HF2 (#3) test run 3rd of September 2020 .................................................................. 45
Data, MF (#4) test run 3rd of September 2020....................................................................46
Data, HF3 (#5) test run 4th of September 2020 .................................................................. 47
13. Test equipment.................................................................................................................. 48
14. Annexes............................................................................................................................ 49

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 4 of 49
300-ELAB-2526-EPA Rev4 110821
General
This report concerns testing of a free-standing cast iron wood heater, type Morsø 2B Classic 2020
The certification test report holds 35 annexes. From this 2. revision of the report, do not any longer
distinguish between CBI and non-CBI reports. The same unitary report applies for all purposes, meaning
only one version of the report is issued.
Please find the full list of annexes in chapter 14.
Figures are stated in European notation, with a comma as the decimal separator and period as the
thousand’s separator.
Report revision history overview
Release
Report number
Number of pages
(report only)
Issued date
initial
300-ELAB-2526-EPA (CBI and non-CBI)
45 (non-CBI)
18. Nov. 2020
Rev 1
300-ELAB-2526-EPA Rev 1 211221 (CBI
and non-CBI)
45 (non-CBI)
21. Jan. 2021
Rev 2
300-ELAB-2526-EPA Rev 2 220621
48
22. June 2021
Rev 3
300-ELAB-2526-EPA Rev 3 250621
49
25. June 2021
Rev 4
300-ELAB-2526-EPA Rev 4 110821
49
11. August 2021
Revision 1 changes
•New issue date 21. January 2021 and revision # raised to Rev 1
•Clause 3.5 amended with the laboratory’s declaration of test appropriateness, starting with “Based on the
evidence above”
•Clause 3.6 amended with the laboratory’s declaration of test appropriateness, starting with “Based on the
evidence above”
•Clause 3.8 amended with the laboratory’s declaration of test appropriateness, starting with “Based on the
evidence above”
•Annex 35; new annex holding the laboratory hand notes
Revision 2 changes
•New issue date 22. June 2021 and revision # raised to Rev 2
•Clause 1.1 General amended with notice of abandoning of distinguishing between a CBI and a non-CBI report.
Now one unitary report applies for all purposes.
•New clause 1.3 added, holding the EPA list of findings
•Clause 3.2, further elaboration of the reason for discarding of the HF1 test due to a clogged filter
•Clause 4.2, Note added on discarding of the HF1 results
•Report page count raised from 45 to 48
Revision 3 changes
•New issue date 25. June 2021 and revision # raised to Rev 3
•Clause 7.3, the test string “and the action of the air controls” was added
•Revised Annex 26, Assembly drawings showing in detail how the lower restriction of the air controls is
accomplished, notably drawing 2B-148 (Primary Air Controls) showing the bushing which stops the spinner
from being closed below three quarters of a turn. Further drawing 2B-511 (Main assy drawing) displaying an
explanatory note on the air controls extreme settings, the minimum being 0,75 turn which resembles an
effective open valve area of 212 sq mm.
•Revised annex 31a, US Users’ manual, on page 9 (printed manual pages 14-15) providing information in a
highlighted frame on the nature of the lower air controls restriction and a warning of NOT to tamper with the
physical stop.
•Revised annex 31b, CAN(Fr) Users’ manual, the same warning

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 5 of 49
300-ELAB-2526-EPA Rev4 110821
•Revised annex 34, Manufacturers’ QA plan, showing on page 4 where and how to
check the action of the air controls, again emphasizing that the minimum primary air
controls setting is 0,75 turn corresponding to a gap between the spinner and the
door on approximately 1 mm
•Report page count raised from 48 to 49
Revision 4 changes
•New issue date 11. August 2021 and revision # raised to Rev 4
•New clause 1.3 added, describing the report changes from the initial (final) report to the Rev 1 report
•New clause 1.4 added, describing the report changes from the Rev 1 to the Rev 2 report
•New clause 1.5 added, describing the report changes from the Rev 2 to the Rev 3 report
•New clause 1.6 added, describing the report changes from the Rev 3 to the Rev 4 report
List of Issues found (ref. Sanzhez email of June 10. 2021)
List of Issues Found
Issue
Applicable
Method/Rule
Section
Notes
How the
manufacturer/test
lab addressed the
issue
Clause
# or
page #
where
to find
addres-
sing of
irregu-
larities
1. Page 32 "For
the low fire
test, after
ignition the
air valve is
throttled to 3
quarters of a
revolution
open, and for
the medium fire
test, after
ignition the air
valve is
throttled to 1
and one
quarter of a
revolution
open."
Need to document
that the stove cannot
burn any lower than the
¾ turn as described in
the low burn test.
See additional
notes below.
40 CFR 60.537
(2) All
documentation
pertaining to the
certification test
used to obtain
certification,
including the full
test report and raw
data sheets,
laboratory
technician notes,
calculations, the
test results for all
test runs, and
discussions of the
appropriateness
and validity of all
test runs, including
runs attempted but
not completed. The
retained
certification test
documentation
must include, as
applicable, detailed
discussion of all
anomalies,
whether all burn
Manual says
that the low
fire is fixed and
cannot be
adjusted
further.
This means
that it was
tested at the
lowest possible
burn rate. Is
that
assertion
correct?
Need
clarification in
the TR about
the ¾ turn for
the low burn
setting.
There is a fixed stop in
place on production models
that prevents the air
controls from being closed
more that as tested (3/4 of
turn), which results in a
linear measurement of
1mm between the spinner
and the frame. The
appliance cannot be turned
any lower down than what
was tested. This restriction
is shown in detail in
drawing 2B-148, page 7 of
annex 26 (assembly
drawings) and further in
drawing 2B-511, page 2 of
annex 26
(reply clarified the 21/6
2021 by JSA)
Annex 26
(assembly
drawings),
page 2, the
main assy
drawing 2B-
511 and on
page 7,
primary air
controls
assy,
drawing 2B-
148
(21/6 JSA)

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 6 of 49
300-ELAB-2526-EPA Rev4 110821
rate categories
were properly
achieved, any data
not used in the
calculations and,
for any test runs
not completed, the
data that were
collected and the
reason that the
test run was not
completed. The
retained
certification test
also must include
documentation
that the burn
rate for the low
burn rate
category was no
greater than the
rate that an
operator can
achieve in home
use and no
greater than is
advertised by the
manufacturer or
retailer.
2. Page 36,
Discussion of
Results doesn't
document run
appropriateness
The basis for
determining
whether or not
the appliance has
complied with
this requirement
is based on 40
CFR
60.533(b)(5).
Per the
requirement,
"Documentation
must include
discussion of
each test run and
its
appropriateness
and validity, and
must include
detailed
discussion of all
anomalies,
whether all burn
rate categories
were achieved,
any data not
used in the
calculations and,
for any test runs
not completed,
the data
collected during
Revised
discussion.
The test run summaries
(sections 3.3, 3.5, and 3.7)
have been edited to state
that the runs were
determined to be
appropriate.
(reply modified the 21/6 by
JSA –former page
numbers deleted)
End of
section 3.3
on page 14;
End of
section 3.5
on page 18;
End of
section 3.7
on page 21
(21/6 JSA)

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 7 of 49
300-ELAB-2526-EPA Rev4 110821
the test run and
the reason(s)
that the test run
was not
completed and
why."
3. Page 54-56,
Annex 3 says,
"Sequence of
images from
the tests, the
5th February
2020"
but report indicates
test was 9/2/2020.
Date just
needs to be
changed in the
test report.
This was a typo that has
been corrected.
Annex 3
cover page;
Annex 3
header on
pages 2-4
(21/6 JSA)
4. Page 484 –
doesn't include
proper use of
air controls and
proper low fire
operation.
In manufacturer
instructions to the lab
they included things
such as turning controls
3 ¼ for fully open etc.
40 CFR 60.536
(g)(3)(iii) Proper
use of air controls,
including how to
establish good
combustion and
how to ensure
good combustion
at the lowest burn
rate for which the
heater is
warranted.
Would
recommend
manufacturer
modifying text
to give more
description on
home use
instructions on
operating the
air controls.
User manual instructions
have been updated to
provide more clarity on
various operational settings
including images for better
guidance
(reply modified and
clarified the 21/6 by JSA –
former page number
deleted)
Annex 31a
Users’
manual (US)
page 9;
Annex 31b
Users’
manual
(CAN) page
9
(21/6 JSA)
5. Did not see raw
data sheets for
filter weighing
etc.
6. Did not see raw
data sheets for
dry gas meter
for each test
run.
40 CFR 60.533
(b)(5)
Need filter
weight raw
data sheets
and raw
technician
notes. Submit
data.
Technician notes have been
added to the report, see
Annex 35
Gas meter values are
recorded pre- and post test
only, in-test monitoring of
sample rates and
proportionality are done via
calibrate flow meters, data
from these flow meters are
shown on raw data sheets.
Calibration certificates by
external Force Technology
to be seen in Annex 15,
page 34-36 (main train –
instrument id 144236
tracible to DTIs’ instrument
database
And further page 37-39
(split train –instrument id
144239)
(reply amended the 21/6
by JSA –Flow metering
details added and former
page number deleted)
Annex 35 in
its entirety;
Annex 21
HF1 Logger
data, page
7-11 (main
train flow
vel. in the
right most
column,
labelled Flow
H) and
pages 12-16
(split train
flow vel. in
the left most
column,
labelled Flow
D)
And similarly
in annexes
22-25 for
the
remaining
tests
(21/6 JSA)

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 8 of 49
300-ELAB-2526-EPA Rev4 110821
7. HF1 was
omitted and we
see the
description in
the anomalies
section which
we find
acceptable.
8. HF1 in other
sections
(tables) needs
the same note
stating that it
was omitted
due to a
clogged filter.
If the TR is
being modified,
then please
make the
appropriate
changes to
justify the
exclusion of
HF1.
A statement in
each table that
says:
"HF1 was
excluded due
to a clogged
filter."
There are no fewer than 6
references to the fact that
HF1 was excluded due to
filter issues, including all
the relevant summary
tables (3.1, 3.2, 4.1, 4.2,
and 11)
As well as the individual
run summary section
(3.3), where else would
you have us include this
information?
(reply amended the 21/6
by JSA –the statement
“HF1 was excluded due to
a clogged filter” was
amended to the
explanatory note
underneath clause 3.2
‘Main results’)
Clause 3.2
on page 10;
Clause 3.3
on page 14
(bottom);
Clause 11
on page 40
(21/6 JSA)
Scope of testing
The appliance was tested to demonstrate compliance with the NSPS 2020 limits, using the ASTM E3053
ALT-125 Cordwood test method broadly accepted by the administrator. Please find the letter of acceptance
enclosed in annex 1.
Site
Testing was accomplished by Danish Technological Institute, Kongsvang Allé 29, DK-8000 Aarhus C,
Denmark in accordance with DTI’s accredited EPA test procedure ELAB-PP-BR-15.
The test procedure is amended in annex 14.
Participants
DTI staff
Testing in the laboratory was accomplished by:
-Jes Sig Andersen, Senior Specialist (lead)
-Rene Lyngsø Hvidberg, Senior specialist (trainee)
Client staff
The test work was witnessed by:
-Frank Juel Nielsen, Morsø Jernstøberi A/S
Test specimen
The stove was manufactured by:
Company: Morsø Jernstøberi A/S
Address: Furvej 6

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 9 of 49
300-ELAB-2526-EPA Rev4 110821
Postcode/town: DK-7900 Nykøbing Mors
Country: Denmark
The stove weighs 98 kg.
The stove is not equipped with a catalyst. The landscape type of firebox is deeper that its wide,
accommodation up to half a meter wood logs.
Variants: At the time of testing there was no known variants of the 2B Classic 2020 wood heater.
Description of the wood heater
Appliance Manufacturer: Morsø Jernstøberi A/S
Wood Stove Model: 2B Classic 2020
Type: Freestanding, radiant-type wood fired room heater.
WOOD HEATER DESCRIPTION
Materials of Construction: The unit is constructed primarily of cast iron with a stainless-steel secondary
combustion air supplying baffle. The firebox is lined with molded vermiculite firebricks. The feed door has
a 145 mm by 158 mm glass panel and one 870 mm by 8 mm glass fiber gasket. The ash door is sealed
with one 570 mm by 8 mm glass fiber gasket.

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 10 of 49
300-ELAB-2526-EPA Rev4 110821
Air Introduction System: Air enters the firebox through a spin-draft located at
the front of the appliance at the top of the fuel-loading door. Secondary air enters
the appliance through the upper back and supplies a three-step, tiered hollow baffle.
Combustion Control Mechanisms: The combustion air inlet is controlled by a spin draft on the fuel-
loading door.
Combustor: N/A.
Internal Baffles: A stainless steel baffle with a ceramic blanket is mounted in the upper portion of the
firebox. The flame path is forced to the front of the firebox where it travels up through the opening
between the baffle and primary air manifold.
Other Features: N/ A.
Flue Outlet: The 6’’ diameter flue outlet is located at the top of the unit.
Firebox volume: 0,01943 m3or 0,6826 ft3 calculated in SolidWorks CAD simulator. Please find the firebox
drawing showing details on the firebox dimensions amended in Annex 33
The stove had been aged more than 50 hours of operation prior to the certification test while pre-testing
at Morsø including:
•The 21st of August 10 hours
•The 24th of August 10 hours
•The 26th of August 10 hours
•The 27th of August 10 hours
•The 28th of August 11 hours
Please find the set of pre-conditioning data in annex 2.

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 11 of 49
300-ELAB-2526-EPA Rev4 110821
Test schedule
The full certification test comprises three HF test runs, one MF test run and one LF test run.
Date
Test I
Test II
Remarks
2-9-2020
HF test
LF test
Both tests valid, however the HF filters clogged
3-9-2020
HF test
MF test
Both tests valid
4-9-2020
HF test
none
Valid HF test
Main results
Please see also the full set of test results in table of chapter 11 and with more decimals in section 4.10
*) The HF1 test is discarded of in the weighted average emission calculation. HF1 was excluded due to a
clogged filter. Please find details in the test summary below in section 3.3
Summary of the CS+HF and LF tests the 02-09-2020
Data logger files Id: 2020-09-02_08:43:10 (CS+HF) and 2020-09-02_14:01:25 (LF)
12:19:15
Ignition of the Cold Start part test using the gas torch for 45 seconds. The air valve
is set in position 100% open (full action = 3 and one quarter revolution open).
0.604 kg of kindling (10% moisture DB) and 0.913 kg start-up fuel (19,7% moisture
DB) was entered
12:20:00
Ignition is over, the door is kept ajar.
12:24:00
The door is closed, and the air valve is maintained in its fully open position
12:47:05
End of the Cold start at 550 grams of embers, which value is taken down. The
embers are evened out
12:47:10
Loading of the High Fire fuel load; 4 logs of 3,155 kg beech wood (19.1% moisture
DB)
12:47:50
End of loading time after 40 seconds.
The door was closed right away, and the air valve was maintained in its fully open
position, being the High Fire setting allowing maximum output rate
13:19:15
Change of the filter holder arrangement in the split extraction train at the hour at
gas meter reading 8856,9 normal litres
13:40-
13:50
The flow rate of the main train needs repetitive adjustments, the filter has taken on
to clogging
13:50:45
PM sampling in the main train is suspended and the pair of filters extracted
13:53:50
PM sampling in the main train is resumed on a fresh pair of filters
14:00:30
End of High Fire test cycle at 300 grams of embers (net). The combined bed of
ember mass at the end of the High fire test hence amounted 550+300 = 850 g.
The bed of embers was evened, and the air valve kept in its fully open position
14:06:00
Loading of the Low Fire fuel load; 5 logs of in total 3,840 kg of firewood (19,5%
moisture DB) at platform scale reading 675 grams
14:06:50
End of loading time after 50 seconds has elapsed; the door is closed right away
14:15:35
The air valve is throttled to its final position being ¾ of a revolution open according
to the 15% of the fuel load mass combusted criteria
15:06:20
Change of the filter holder arrangement in the split extraction train at the hour at
gas meter reading 9577,91 normal litres
19:19:48
The Low Fire test is over at platform scale reading of 675 grams
Burn rate kg dry matter/hour
Emission grams/hour
#1*
HF1 2-9
1,92
2,3
#2
LF 2-9
0,61
0,1
#3
HF2 3-9
2,05
1,7
#4
MF 3-9
0,92
0,3
#5
HF3 4-9
2,33
1,1
Weighted average
0,5

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 12 of 49
300-ELAB-2526-EPA Rev4 110821
Post
Check
The mass combusted during the last 30 minutes of the LF test was 736,6-675,2 =
61,4 grams, thus rendering the LF test valid with respect to the fire out criteria of no
less the 50 g combusted during half an hour.
Please find the corresponding sequence of images in annex 3
Course of combustion and residual mass over the last 30 minutes of the Low Fire test

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 13 of 49
300-ELAB-2526-EPA Rev4 110821
CS and HF1 (#1) proportional sampling rate variation as per ASTM E2515 clause 9.8.1

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 14 of 49
300-ELAB-2526-EPA Rev4 110821
LF (#2) proportional sampling rate variation as per ASTM E2515 clause 9.8.1
Evidence of filter temperature maintained in-between 26,7-32,2 οC (80-90 F) on the 2nd of September as
per Alt-125 letter clause 1) requirement modifying the ASTM E2515
20
22
24
26
28
30
32
12:19
12:30
12:42
12:53
13:05
13:16
13:28
13:39
13:51
14:02
14:14
14:25
14:37
14:48
15:00
15:11
15:23
15:34
15:46
15:57
16:09
16:20
16:32
16:43
16:55
17:06
17:18
17:29
17:41
17:52
18:04
18:15
18:27
18:38
18:50
19:01
19:13
CS+HF+LF Filter temperatures 02-09-2020 [ºC]
Ambient Main train Spilt train 1. hour Split train remaining time
Main train filters clogging

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 15 of 49
300-ELAB-2526-EPA Rev4 110821
EPA
Surface
Surface
Surface
Surface
Surface
Mean
Flue gas
temperature
temperature
temperature
temperature
temperature
temp
Time
Temperature
(οC)
Top (οC)
Rear (οC)
Right side
(οC)
Left side (οC)
Bottom (οC)
(οC)
12:19:15
23,3
24,3
24,0
25,9
25,4
24,7
24,9
Evidence of cold starting conditions on day one, the 2nd of September 2020 to ASTM E3053 clause 8.5.1
(Ambient temp was 24,0 degr C at 12:19:15 hours)
Based on the evidence presented above and, DTI declares test runs #1 (Cold start and High fire 1 test)
and #2 (Low fire test) was determined to be conditional and appropriate. Because the filter clogged and
sampling had to be suspended temporarily during test run #1, we exclude test run #1, meaning it will
not be used for the average result calculations. Test run #2 however is deemed valid in any respect.
Please also see section 3.9 Anomalies for further discussion of the test run #1 filter clogging incident.
Duct traverse data 2. September 2020
Measuring points
Start
Intermed.
Center
Intermed.
Last
ref. ASTM E2515
6,70%
25%
50%
75%
93,30%
[relative]
Fig 6
12,7
37,5
75,0
112,5
137,3
[mm]
First direction (X)
21,3
23,3
31,9
25,9
18,2
Pd [Pa]
Angular direction (Y)
25,6
32,6
32,8
28,9
20,4
Pd [Pa]
Pitot tube
dynamic
pressure
Duct
tempera-
ture
Pitot tube
static
pressure
30,8
Pd [Pa]
24,3
Temp [°C]
48,8
Ps [Pa]
Resulting pitot factor Fp: 0,92

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 16 of 49
300-ELAB-2526-EPA Rev4 110821
Summary of the CS+HF and MF tests the 03-09-2020
Data logger file Id: 2020-09-03_08-30-28
11:34:00
Ignition of the Cold Start part test using the gas torch for 60 seconds. The air valve
is set in position 100% open (full action 3 and one quarter revolution). 0.603 kg of
kindling (10% moisture DB) and 0.897 kg start-up fuel (19,7% moisture DB) was
entered
11:39:45
Ignition is over, the door is closed. The air valve is maintained in its fully open
position
11:55:26
End of the Cold start at 550 grams of embers, which value is taken down. The
embers are evened out
11:55:30
Start loading of the High Fire fuel load using 4 logs of 3,190 kg firewood (19,3%
moisture DB)
11:56:25
End of loading time after 55 seconds.
12:01:25
The door was closed after 5 minutes acrac, and the air valve was kept in its fully
open position
12:34:00
Change of the split train filter holder arrangement at the hour, at gas meter reading
11828,3 normal litres
13:04:51
End of High Fire test cycle at 300 grams of embers (net), the combined bed of ember
masses at the end of the High fire test was hence 550+300 = 850 g.
The bed of embers was evened, and the air valve kept in its fully open position
Break
Pass time to meet the LF start of test criteria
13:14:40
At the platform scale reading of 675 grams, start loading of the Medium Fire fuel
load using 5 logs of in total 3,736 kg of firewood (19,4% moisture DB)
13:15:25
End of the loading time after 45 seconds and the air valve kept fully open to a start
and then slowly reduced stepwise.
13:19:30
At platform scale reading 4,25 kg, the air valve is set in its final MF position being 1
and a quarter revolution open
14:14:40
Change of the filter holder arrangement in the split extraction train at the hour, at
gas meter reading 12471,6 normal litres
16:38:18
The Medium Fire test is over at platform scale reading of 675 grams
Please find the corresponding sequence of images in annex 4
0
50
100
150
200
250
300
350
0
5
10
15
11:34 12:46
Fluegas temperature [C]
Weight [Kg]
Time [hh:mm]
High Fire 2 test (#3)
2020-09-03
Weight Flue T

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 17 of 49
300-ELAB-2526-EPA Rev4 110821
CS and HF2 (#3) proportional sampling rate variation as per ASTM E2515 clause 9.8.1
90
92
94
96
98
100
102
104
106
108
110
11:34 11:44 11:54 12:04 12:14 12:24 12:34 12:44 12:54 13:04
CS and HF2 proportional rate variation [%]
0
50
100
150
200
250
300
350
0
5
10
15
13:14 14:26 15:38
Flue gas temperatur [°C]
Weight [Kg]
Time [hh:mm]
Medium Fire test (#4)
2020-09-03
Weight
Flue T

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 18 of 49
300-ELAB-2526-EPA Rev4 110821
MF (#4) proportional sampling rate variation as per ASTM E2515 clause 9.8.1
Evidence of filter temperature maintained in-between 26,7-32,2 οC (80-90 F) the 3rd of September as per
Alt-125 letter clause 1) requirement modifying ASTM E2515
EPA
Surface
Surface
Surface
Surface
Surface
Mean
Flue gas
temperature
temperature
temperature
temperature
temperature
temp
Time
Temperature
(οC)
Top (οC)
Rear (οC)
Right side
(οC)
Left side (οC)
Bottom (οC)
(οC)
11:34:00
23,0
24,3
23,6
25,5
24,9
24,5
24,6
Evidence of cold starting conditions on day two, the 3rd of September 2020 to ASTM E2515 clause 9.8.1
(Ambient temp was 23,8 degr C at 13:34:00 hours)
20,0
22,0
24,0
26,0
28,0
30,0
32,0
11:33
11:42
11:50
11:59
12:07
12:16
12:24
12:33
12:41
12:50
12:58
13:07
13:15
13:24
13:32
13:41
13:50
13:58
14:07
14:15
14:24
14:32
14:41
14:49
14:58
15:06
15:15
15:23
15:32
15:40
15:49
15:57
16:06
16:14
16:23
16:31
CS+HF+MF Filter temperatures 03-09-2020 [ºC]
Ambient Main train Split train 1 hour Split train remaining time
End of HF test at 13:04:51
90
92
94
96
98
100
102
104
106
108
110
MF proportional rate variation [%]

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 19 of 49
300-ELAB-2526-EPA Rev4 110821
Based on the evidence presented above, DTI declares test runs #3 (cold start and
high fire 2) and #4 (medium fire test) were determined to be as well appropriate as
valid.
Duct traverse data the 3. September 2020
Measuring points
Start
Intermed.
Center
Intermed.
Last
ref. ASTM E2515
6,70%
25%
50%
75%
93,30%
[relative]
Fig 6
12,7
37,5
75,0
112,5
137,3
[mm]
First direction (X)
24,9
30,8
33,2
29,3
23,3
Pd [Pa]
Angular direction (Y)
20,5
22,5
32,7
27,3
22,5
Pd [Pa]
Pitot tube
dynamic
pressure
Duct
tempera-
ture
Pitot tube
static
pressure
31,5
Pd [Pa]
26,5
Temp [°C]
48,4
Ps [Pa]
Resulting pitot factor Fp: 0,92

Report No.: 300-ELAB-2526-EPA Rev4 110821
Date: 11.08.2021
Initials: JSA/
Page 20 of 49
300-ELAB-2526-EPA Rev4 110821
Summary of CS+HF test the 4-09-2020
Data logger file Id: 2020-09-04_09-16-26
10:05:58
Ignition of the Cold Start part test using the gas torch for 55 seconds. The air valve
is set in position 100% open (full action 3 and one quarter revolution). 0.605 kg of
kindling (10% moisture DB) and 0.909 kg start-up fuel (19,5% moisture DB) was
entered
10:11:54
Ignition is over, the door is closed. The air valve is maintained in its fully open
position
10:30:40
End of the Cold start at 550 grams of embers, which value is taken down. The bed of
embers is
10:31:00
Start loading of the High Fire fuel load using 4 logs of 3,183 kg firewood (19,2%
moisture DB)
10:31:55
End of loading time after 55 seconds.
10:36:50
The door was closed after 5 minutes and 50 seconds acrac, and the air valve was
kept in its fully open position
11:05:58
Change of the split train filter holder arrangement at the hour, at gas meter reading
13924,5 normal litres
11:32:25
End of High Fire test cycle at 290 grams of embers (net), the combined bed of ember
masses at the end of the High fire test was hence 550+290 = 840 g.
0
50
100
150
200
250
300
350
0
5
10
15
10:05 11:17
Flue gas temperatur [°C]
Weight [Kg]
Time [hh:mm]
CS and HF3 test (#5)
2020-09-04
Weight
Table of contents
Other Morsø Stove manuals
Popular Stove manuals by other brands

EdilKamin
EdilKamin IDROPELLBOX Installation, use and maintenance

Cecilware
Cecilware EL-1 Specifications

Morso
Morso 2B Classic Installation and operating instructions

HearthStone
HearthStone STARLET owner's manual

Heta
Heta Scan-line 8 Operating and installation instructions

Breckwell
Breckwell Luxury P2700 Series owner's manual

EQUATION
EQUATION CUBE 6 INSTALLATION, USAGE AND MAINTENANCE MANUAL

Harman
Harman XXV owner's manual

HASE
HASE Lagos operating instructions

Lennox Hearth Products
Lennox Hearth Products BELLA Installation and operation manual

Broseley
Broseley Inset Ignite 6 Remote Installation & operating instructions

Quadra-Fire
Quadra-Fire GARN-MBK-IPI owner's manual