
PrimaX 6
Safety Regulations
GB
must be exercised so that the electrolyte does not contact skin, clothing or circuitry otherwise
personal injury (burns) and/or equipment damage may result.
• The only absolute method to ensure proper overall operation of the device is to check it with a
known concentration of the gas for which it has been calibrated. Consequently, calibration checks
must be included as part of the routine inspection of the system.
• As with all devices of these types, high levels of, or long exposure to, certain compounds in the
tested atmosphere could contaminate the sensor. In atmospheres where the device may be
exposed to such materials, calibration must be performed frequently to ensure that the operation
is dependable and display indications are accurate.
• The device must not be painted. If painting is done in an area where a device is located, care must
be exercised to ensure that paint is not deposited on the sintered metal flashback arrestor in the
gas sensor inlet, if so equipped. Such paint deposits would interfere with the gas diffusion process.
• Use only genuine MSA replacement parts when performing any maintenance procedures
provided in this manual. Failure to do so may seriously impair instrument performance. Repair or
alteration of the device, beyond the scope of these maintenance instructions or by anyone other
than an authorised MSA service personnel, could cause the product to fail to perform as designed.
• The device is designed for applications in hazardous areas under atmospheric conditions.
• For correct measurements, the combustible gas sensors require an oxygen concentration greater
than 10 Vol%. Oxygen enriched atmospheres, greater then 21 Vol%, can affect the measurement
and the electrical safety of the device.
• The response time of the device will be increased by significant dust deposits on the Sensor.
Checks for dust deposits must be done at regular intervals.
• Catalytic combustible gas sensors may produce low or zero response to combustible gas after
exposure to substances as Silicon, Silane, Silicate, Halide and compounds containing Fluorine,
Chlorine, Iodine or Bromine.
• Catalytic sensor: After exposure of gas above the measuring range, the sensor has to be imme-
diately calibrated / adjusted, independent of the calibration interval. In the case of an adjustment
the sensitivity of the sensor has to be rechecked again after 24 hours.
• In case of very high flammable concentrations >100%LEL the device is able to lock all outputs
(LOC). This function shall be used for standalone applications according to ATEX requirements.
• Catalytic sensor: If the PrimaX is operated in combination with a control unit and (LOC) is deacti-
vated on the PrimaX, the control unit has to ensure the latching after the measuring range was
exceeded.
• Catalytic sensor: Before a reset of an over range (LOC) indication is done it shall be verified that
the gas concentration is below full scale.
• Toxic gases: If the measuring range is exceeded, sensor life time can be reduced or the sensor
might need a recovery phase.
• Used sensors have to be disposed of in an environmentally compatible way.
• At signal underrange between -1.25 % and -10 % of the measuring range the 4 - 20 mA output is
set to 3,8 mA. At signal underrange of more than -10 % of the measuring range the 4 - 20 mA
output is set to the error state (2 mA). To avoid a delayed alarm activation, a calibration should be
done when the output current is set to 3,8 mA and "LO" is shown on the display persistently.
• The lowest range which is suitable for the alarm thresholds should be selected for toxic gases.
• To compensate for possible deviations during measurement of flammable gases a calibration
factor of 1.05 has to be applied when using the CalGard for remote calibration (see chapter 8.3
"Remote Calibration")