Bio-logic®AuDX PRO FLEX®
QUICK START GUIDE
Getting started
Re-initialize pump unit
Place probe tip into the red probe cavity for the adult probe tip or the blue
cavity for the infant probe tip
Turn the device on
Tap the tools icon
Select (tap) Functional Checks
Select Pump Unit Test
The red lines should be straight lines with no or just little oscillation or stairs
effect.
The green line should ideally be horizontal (no pressure drop).
With probe and cavity connected, some pressure drop is acceptable. If just
testing the device with pressure outlet closed, there should be no pressure
drop.
• Perform pump unittest
• After test is done, quickly tap on the graph 4 times
• Get “service menu”
• Press the ‘x’ button