
13 September, 2018 g
Preface
Maintenance
Inspecting the
system
Routinely check the instrument for exterior damage.
Follow your medical facilities safety guidelines.
Decontamination Decontamination that can be performed by the operator is limited to cleaning and
disinfecting the device. Any maintenance inside the device must be performed by
qualified service personnel only.
Cleaning
instructions
Regular cleaning maintenance should be performed according to frequency of
use of the device. Always observe your local hygiene authority’s guidelines, and
the following points below:
•Disconnect the mains power before cleaning the equipment.
•Clean the equipment surfaces with a clean, slightly damp cloth with a mild
detergent (e.g., Wet Wipes®), and wipe it dry.
•Make sure that no liquids enter the device at push buttons and other openings
in the enclosure.
Do not use cleaning detergents, or cleaning agents based on sol-
vent, silicon-based, abrasive and/or flammable substances.
Disinfection
procedure
When disinfecting the equipment, if required, surgical spirit (70%), a Chlorine
(1000ppm), or ethanol (70%) detergent may be used. Carefully follow the
disinfectant manufacturer’s instructions for use, and the steps under the Cleaning
Instructions section above.