manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Oakton
  6. •
  7. Measuring Instrument
  8. •
  9. Oakton pHTestr 50 User manual

Oakton pHTestr 50 User manual

Operating Instructions
pHTestr®50 and 50S Pocket Testers
Getting Started
The pH tester has been factory calibrated and usually works well out of the
box. However, after extended periods of non-use, it is best to ensure that the
sensor is hydrated. Fill the cap with electrode storage solution or pH buffer
and soak the sensor as needed to get a stable pH reading. A brief rinse with
deionized (DI) water is OK, but avoid soaking or storing in deionized water as
this will shorten pH electrode life. Prior to taking the measurements, periodic
calibration with certified standards is recommended for best accuracy.
Buffer Set Selection
Select either USA (pH 4.01, pH 7.00 and pH 10.01) or NIST (pH 4.01, pH 6.86,
and pH 9.18) to match your calibration standard values.
1. Press ON/OFF ( ) to power on.
2. Press MENU/
v
to enter setup window. Press HOLD/ to select Settings.
The display shows Buffer and Backlight.
3. Press HOLD/ to select Buffer. The display shows USA and NIST.
4. Press HOLD/ to select USA or scroll down by pressing MENU/
v
to toggle
between the two buffer standards.
5. The display shows the selected buffer with a .
Backlight Settings
1. Press MENU/
v
to enter setup window. Press HOLD/ to select Settings.
The display shows Buffer and Backlight.
2. Scroll down by pressing MENU/
v
to toggle between Buffer and Backlight.
Press HOLD/ to select Backlight. The display shows ON and OFF.
3. Scroll down by pressing MENU/
v
to toggle between ON and OFF.
Backlight ON increases readability in low-light conditions.
4. Press HOLD/ to select the desired backlight option. The display shows
the selected backlight option with a .
Temp Set
Reset
Backlight
hold
NIST
menu
USA
hold
Settings Buffer USA
hold
Settings
Temp Set
Reset
Buffer
hold
Backlight
ON OFF
hold
Backlight OFF
hold
menu
Temperature Settings
1. Press MENU/
v
to enter setup window. Scroll down by pressing MENU/
v
to select Temp Set. Press HOLD/ to select Temp Set. The display shows
Set °C/°F and Temp Cal.
2. Press HOLD/ to select Set °C/°F. Display automatically shows °C and ° F.
3. Scroll down by pressing MENU/
v
to toggle between °C and °F.
4. Press HOLD/ to select the desired temperature unit.
Settings
Reset Temp Cal
hold
°C
°F
menu
°C
hold
Temp Set
Set °C/°F
hold
Settings
Reset
hold
Temp Cal
23.4 °C
23.4 °C
menu
23.4°C
hold
Temp Set
Set °C/°F
Temp Cal
hold
Temperature Calibration
1. Press MENU/
v
to enter setup window. Scroll down by pressing MENU/
v
to highlight Temp Set. Press HOLD/ to select Temp Set. The display
shows Set °C/°F and Temp Cal.
2. Scroll down by pressing MENU/
v
to toggle between Set °C/°F and Temp
Cal. Press HOLD/ to select Temp Cal.
3. The lower display shows the current measured temperature reading based
on the last set offset and the upper display shows the current measured
temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the
temperature sensor to stabilize.
5. Press MENU/
v
to adjust the temperature value, then press HOLD/ to
confirm the new temperature value of the solution.
Note: To exit this program without confirming the calibration, press CAL/ESC.
CAL
Place the tester in
the buffer to start or
press to
escape.
cal
esc
cal
esc
Measurement
1. Press ON/OFF ( ) to power the tester on if needed.
2. Dip the sensor in about 2 cm to 3 cm into the sample. If using the supplied
cap, fill to the "MAX" line. Stir and let the reading stabilize. The timer icon
will blink during this time. Once the reading is stable, will appear.
CAUTION: Testing dry samples is not recommended and
can lead to sensor damage. Soils must be wet and free from
particulates that may scratch the glass sensor. Excessive
force into dry samples can cause glass breakage.
3. Note the pH value or press HOLD/ to freeze the reading.
To release the reading, press HOLD/ again.
4. Press ON/OFF ( ) for 3 seconds to turn off tester. If you do not press a button
for 8.5 minutes, the tester will automatically shut off to conserve batteries.
User Reset
Reset to the user’s default settings by using the User Reset function. Buffer
selection and temperature user calibration are not affected by the User Reset
function.
1. Press MENU/
v
to enter setup window. Scroll down by pressing MENU/
v
to
select Reset. Press HOLD/ to select Reset. The display shows User Reset
and Factory Reset.
2. Press HOLD/ to select User Reset. The display automatically shows
No and Yes.
3. Scroll down by pressing MENU/
v
to toggle between No and Yes.
4. Press HOLD/ to confirm either No or Yes. The display shows the
User Reset option with a .
Settings
Temp Set Factory Reset
hold
No
Yes
menu hold
User Reset
Reset
hold
pH Calibration
Calibration should be done regularly, preferably once a week or more if using
the tester daily. You can calibrate up to three points using standards from
either the USA or the NIST buffer groups. For best results, use a
minimum of two pH buffers such as pH 4 and pH 7.
1. Press ON/OFF ( ) to power the tester on if needed.
2. Dip sensor in about 2 cm to 3 cm into the pH standard buffer solution.
If using the supplied cap, fill to the "MAX" line. For best results, allow
the reading to stabilize if needed before proceeding.
3. Press the CAL/ESC key to enter calibration mode. The CAL indicator will be
displayed. The upper display will show the unadjusted factory default read-
ing while the lower display will search for the corresponding value from the
selected buffer group.
Note: All testers have dual display during calibration mode. To abort
calibration, press CAL/ESC.
4. The timer icon will blink while the reading stabilizes (allow 1 to 2 minutes).
Once the reading is stabilized, the timer stops blinking. The tester will auto-
matically confirm the calibration value and return to the measurement win-
dow when successful.
5. Rinse electrode and repeat with other buffers if necessary.
• Aquaculture
• Aquariums
• Boilers
• Car washes
• Ecology
Applications
• Food sectors
• Hydroponics
• Labs
• Sanitation plants
• Steam generators
• Spas
• Studies
• Water and wastewater
treatment
• Water quality testing
in pools and more
Factory Reset
Reset the pH calibration to the factory reset default by using the Factory Reset
function.
1. Press MENU/
v
to enter setup window. Scroll down by pressing the
MENU/
v
to select Reset. Press HOLD/ to select Reset. The display shows
User Reset and Factory Reset.
2. Scroll down by pressing the MENU/
v
to toggle between the resets. Press
HOLD/ to select Factory Reset. The display automatically shows No and Yes.
3. Scroll down by pressing MENU/
v
to toggle between No and Yes.
4. Press HOLD/ to confirm either No or Yes. The display shows the Factory
Reset option with a .
Settings
Temp Set
User Reset
hold
No
Yes
menu hold
Reset
Factory Reset
hold
CAL
pH
7.45
7.01
25.0°C
Key Info
Scroll
Accept
EXIT
MENU
HOLD
CAL
Key Info
Scroll
Accept
EXIT
MENU
HOLD
CAL
Key Info
Scroll
Accept
EXIT
MENU
HOLD
CAL
Key Info
Scroll
Accept
EXIT
MENU
HOLD
CAL Key Info
Scroll
Accept
EXIT
MENU
HOLD
CAL
Key Info
Scroll
Accept
EXIT
MENU
HOLD
CAL
pH
pH
7.01
25.5°C
pH
pH
4.01
25.5°C
1065O100_35634-15,-25.indd 1 9/5/2017 3:01:06 PM
MTS - Manual Temperature Setting (pHTestr 50S only)
The MTS range is 0°C to 60.0°C (32.0°F to140.0°F). User reset will set
temperature to default value 25°C or 77°F.
1. Press MENU/
v
to select setup window. Scroll down by pressing MENU/
v
to select Temp Set. Press HOLD/ to select Temp Set. The display shows
Set °C/°F and Temp Cal.
2. Scroll down by pressing MENU/
v
to toggle between Set °C/°F and
Temp Cal. Press HOLD/ to select Temp Cal.
3. The lower display shows the current measured temperature reading based
on the last set offset and the upper display shows the current measured
temperature reading based on factory default calibration.
4. Dip the tester into a solution of known temperature and allow time for the
built-in temperature sensor to stabilize.
5. Press MENU/
v
to adjust the temperature value or press the HOLD/ to
confirm the calibrated value as new temperature value of the solution.
Note: To exit this program without confirming the calibration, press CAL/ESC
before the automatic confirmation takes place.
Settings
Reset
hold
Temp Cal
23.4 °C
23.4 °C
menu
23.4 °C
hold
Temp Set
Set °C/°F
Temp Cal
hold
HOLD Function
This feature helps to freeze the display reading for a delayed observation.
holdhold
pH
4.01
23.5 °C
PH pH
4.01
23.5 °C
PH
Press HOLD
to release
Sensor Maintenance
1. Before measuring soil pH with the pHTestr 50S, wet your soil sample with
distilled water and ensure that the soil is free of particulates. Even though
the pHTestr 50S is rugged, hard surfaces such as stones and pebbles can
still cause breakage.
2. Rinse the sensor with sensor storage solution after each measurement.
Care has to be taken not to damage the sensor’s glass electrode especially
while rinsing the pHTestr 50S penetrating tip electrode.
3. In aggressive chemicals, dirty or viscous solutions, and solutions with
heavy metals or proteins, take readings quickly and rinse electrode
immediately afterward. For the pHTestr 50S, the remnants of the
semisolid samples on the penetrating electrode can be removed by
rubbing it with some table salt and then rinsing. Mild detergent and
warm water can be used to wash the penetrating electrode clean.
4. For best results, keep pH electrode stored upright in protective storage
cap with electrode storage solution to keep the sensor hydrated. For
pHTestr 50S, keep tip moistened in sleeve with electrode storage solution.
Never store your pHTestr in deionized water for extended periods. Use
deionized water for rinsing only.
Specifications pHTestr 50 pHTestr 50S (spear tip)
pH
pH range –1.00 to 15.00 pH –1.00 to 15.0 pH
Resolution 0.01 pH 0.01 pH
Relative accuracy ± 0.01 pH ± 0.01 pH
Calibration points Up to 3 points Up to 3 points
Buffer set
standard selection USA 4.01/7.00/10.01
NIST 4.01/6.86/9.18 USA 4.01/7.00/10.01
NIST 4.01/6.86/9.18
Calibration window ±1.00 pH ±1.00 pH
Calibration type Point to Point Point to Point
Temperature
Temperature range 0 to 60°C / 32 to 140.0°F 0 to 60°C / 32 to 140.0°F
Temp. resolution 0.1°C / 0.1°F 0.1°C / 0.1°F
Temperature
accuracy
From 0 to 50ºC (±0.5ºC / ±0.9ºF
+ 1 LSD); From 50 to 60ºC
(±1.0ºC / ±1.8 ºF + 1 LSD) NA
Temperature
compensation Yes (Automatic Temperature
Compensation) Yes (Manual Temperature
Compensation)
General
Display Graphics, dot matrix 80 x 100 pixel Graphics, dot matrix 80 x 100 pixel
Auto off 8.5 minutes (from last key press) 8.5 minutes (from last key press)
Reset User / Factory User / Factory
Power requirement Four AAA 1.5 V batteries Four AAA 1.5 V batteries
Battey life >250 hours >250 hours
Waterproofing IP67 IP67
Ambient operating
temperature 5 to 45°C / 41 to 113°F 5 to 45°C / 41 to 113°F
Relative humidity 5% to 85% noncondensing 5% to 85% noncondensing
Storage temp. –20 to 60°C / –4 to 140°F –20 to 60°C / –4 to 140°F
Storage humidity 5% to 85% noncondensing 5% to 85% noncondensing
Ordering Code Product Description
35634-15 pHTestr 50 pocket tester with case, lanyard, and batteries
35634-25 pHTestr 50S spear-tip pocket tester with case, lanyard, and batteries
35634-17 Replacement sensor module for pHTestr 50
35634-27 Replacement sensor module for pHTestr 50S
35634-09 Replacement sensor cap for pHTestr 50 only
09376-00 Replacement alkaline batteries; AAA, 1.5 V. Pack of 12
17101-45 NIST-traceable calibration with data for pH testers
Sensor Replacement
You can replace the sensor module at the fraction of the cost of a new tester.
When the tester fails to calibrate or gives fl uctuating readings in calibration
standards, you need to change the sensor.
1. With dry hands, grip the ring with sensor facing you. Twist the ring
counterclockwise. Save the ring for later use.
2. Pull the old sensor module away from the tester.
3. Align the four tabs on the new sensor module so that they match the
four slots on the tester.
4. Gently push the module onto the slots to sit it in position. Push the smaller
O-ring fully onto the new sensor module. Push the other O-ring over the
module and thread sensor ring into place by firmly twisting clockwise.
Note: It is necessary that you recalibrate your tester prior to measurement
after a sensor replacement.
Replacing the Batteries
The pH tester uses four AAA
1.5 V batteries.
1. To remove the battery cover,
See Figure 1. Clear the front
catch and then the back
catch, before sliding the
cover off.
2. To remove the battery plate,
push the center tab towards
the front of the tester as
shown in Figure 2. Once
unlocked, remove the plate
to access the batteries.
3. Invert the tester upside down
to remove the batteries. Each
side uses two AAA batteries.
Orient each battery with
positive terminal facing
downward.
4. To lock the battery plate, align
the small tabs (Figure 3) into
the guide ribs on the hous-
ing and then press down.
See Figure 4.
Fig. 3: Align tabs Fig. 4: Push down to lock
Fig. 5: Battery plate unlocked Fig. 6: Battery plate locked
Fig. 2: Push to unlock
Fig. 1: Removing battery cover
1. Press HOLD/ button to
freeze the measurement. 2. Press HOLD/ again to
release the measurement.
Warranty
This instrument is supplied with a warranty against manufacturing defects for
a period of one year from the date of purchase.
Return of Items
Authorization must be obtained from your distributor before returning items
for any reason. When applying for authorization, please include information
regarding the reason the item(s) are to be returned. We reserve the right to
make improvements in design, construction and appearance of products without
notice. Prices are subject to change without notice.
Self-Diagnostic Messages
Batteries are weak and need replacement soon.
stable error Appears when calibration is attempted but the reading is not yet stable. Wait for the
reading to stabilize or manually confirm the calibration by pressing enter.
buffer error The buffer is outside of the calibration range.
slope error The 2nd or 3rd calibration point is not within 80% to 120% slope range.
over range The reading is above the measuring range of tester.
under range The reading is below the measuring range of tester.
Accessories
1065O100_MAN_35634-15,-25 August, 2017
www.4oakton.com
Key Info
Scroll
Accept
EXIT
MENU
HOLD
CAL
pH
4.01
23.5 °C
PH
pH
4.01
23.5 °C
PH
Press HOLD
to release
1065O100_35634-15,-25.indd 2 9/5/2017 3:01:08 PM

This manual suits for next models

1

Other Oakton Measuring Instrument manuals

Oakton CON 10 User manual

Oakton

Oakton CON 10 User manual

Oakton 35630-00 User manual

Oakton

Oakton 35630-00 User manual

Oakton pHTutor User manual

Oakton

Oakton pHTutor User manual

Oakton CON 400 User manual

Oakton

Oakton CON 400 User manual

Oakton 35612-series User manual

Oakton

Oakton 35612-series User manual

Oakton pH 300 User manual

Oakton

Oakton pH 300 User manual

Oakton PH550 User manual

Oakton

Oakton PH550 User manual

Oakton 35700-00 User manual

Oakton

Oakton 35700-00 User manual

Oakton WD-35606-00, WD-35606-05 User manual

Oakton

Oakton WD-35606-00, WD-35606-05 User manual

Oakton 35618 Series User manual

Oakton

Oakton 35618 Series User manual

Oakton pH 10 Series User manual

Oakton

Oakton pH 10 Series User manual

Oakton ION 5 Acorn Series User manual

Oakton

Oakton ION 5 Acorn Series User manual

Oakton pH 5+ User manual

Oakton

Oakton pH 5+ User manual

Oakton 450 Series User manual

Oakton

Oakton 450 Series User manual

Oakton PC 700 User manual

Oakton

Oakton PC 700 User manual

Oakton ION 6 User manual

Oakton

Oakton ION 6 User manual

Oakton EcoTestr pH1 User manual

Oakton

Oakton EcoTestr pH1 User manual

Oakton 700 Series User manual

Oakton

Oakton 700 Series User manual

Oakton WD-35427-10 User manual

Oakton

Oakton WD-35427-10 User manual

Oakton CON 700 User manual

Oakton

Oakton CON 700 User manual

Oakton 35616-00 User manual

Oakton

Oakton 35616-00 User manual

Oakton Acorn TEMP 4 User manual

Oakton

Oakton Acorn TEMP 4 User manual

Oakton T100WL User manual

Oakton

Oakton T100WL User manual

Oakton CON550 User manual

Oakton

Oakton CON550 User manual

Popular Measuring Instrument manuals by other brands

Doris Schmithals Messwerkzeuge WP 201 066ECO user manual

Doris Schmithals Messwerkzeuge

Doris Schmithals Messwerkzeuge WP 201 066ECO user manual

CEMB N200 Use and maintenance instruction manual

CEMB

CEMB N200 Use and maintenance instruction manual

finder 7M.38.8.400.0112 user manual

finder

finder 7M.38.8.400.0112 user manual

BASETech BTTH-1014 operating instructions

BASETech

BASETech BTTH-1014 operating instructions

Michell Instruments Easidew Online user manual

Michell Instruments

Michell Instruments Easidew Online user manual

Brouwland pH110 manual

Brouwland

Brouwland pH110 manual

Tractel dynafor LLZS Series Operating and maintenance instructions

Tractel

Tractel dynafor LLZS Series Operating and maintenance instructions

Diablo 5000B Getting started manual

Diablo

Diablo 5000B Getting started manual

Nissei TX-1202 user manual

Nissei

Nissei TX-1202 user manual

DOLD IP 5883 Installation / monitoring technique

DOLD

DOLD IP 5883 Installation / monitoring technique

twintex TI512 Operation manual

twintex

twintex TI512 Operation manual

DataCan 109684 user manual

DataCan

DataCan 109684 user manual

UEi C75 instruction manual

UEi

UEi C75 instruction manual

Ecotech Serinus 30 instruction manual

Ecotech

Ecotech Serinus 30 instruction manual

PROPOINT 8319683 user manual

PROPOINT

PROPOINT 8319683 user manual

ProAir FGP Series manual

ProAir

ProAir FGP Series manual

Delta-T Devices WET150 Kit quick start guide

Delta-T Devices

Delta-T Devices WET150 Kit quick start guide

red lion PAXLPT manual

red lion

red lion PAXLPT manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.