Oakton 700 Series User manual

Keys Functions
Powers the meter on and off. Upon power on, the meter
automatically begins in the mode that was last used. Calibration
and memory values are retained even if meter is unplugged.
Toggle between available measurement modes; pH/Temp,
mV/Temp, or ppm/mV (Ion 700 only).
Also used to switch from pH to Temp during pH calibration
mode. Press and hold for 5 seconds to enter SETUP mode.
Toggles between measurement and calibration modes.
In SETUP mode, returns user to the measurement mode.
MI (Memory Insert) stores values into memory.
Increase value. Scroll up in SETUP mode.
MR (Memory Recall) recalls values from memory
Decrease value. Scroll down in SETUP mode.
Freezes measured reading. Press again to resume live reading.
Confirms calibration values in CAL mode.
Confirms selections in SETUP mode.
View recalled values in memory mode.
nLCD Annunciators
Quick Guide
Bench 700 Series
pH/mV/Ion/Conductivity/TDS/Dissolved Oxygen/ºC/ºF
nKeypad Functions
Part of Thermo Fisher Scientific
The 700 series meters can retain up to 100 points into memory for later retrieval.
1. In the measurement mode, press to insert the measured value into memory.
The stored memory location value (StO) is briefly displayed.
2. To recall data from memory, press . The location of the most recent stored data is displayed
first. Press or to select the location of the desired data, then press to accept.
3. Press to return to the stored data location. Press to return to measurement mode.
For more information on our products, please contact our channel partner or visit our websites listed below:
Eutech Instruments Pte Ltd Oakton Instruments Eutech Instruments Europe B.V.
Blk 55, Ayer Rajah Crescent, 625 E Bunker Ct, P.O. Box 254, 3860 AG Nijkerk,
#04-16/24, Singapore 139949 Vernon Hills, IL 60061, USA The Netherlands Wallerstraat 125K,
eutech@thermofisher.com info@4oakton.com 3862 CN Nijkerk, The Netherlands
www.eutechinst.com www.4oakton.com eutech@thermofisher.com
www.eutechinst.com
Automatic
temperature
compensation
is active
Setup
mode
Measured
value is held
Current
measurement
unit / mode
of operation
Electrode error
when appears
with ERR
Primary
reading
Secondary
reading
During pH CAL,
secondary reading is
the auto recognized pH
buer standard value
Measured
value is stable
Error or invalid
key pressed
Prompts user
to select buer
During setup,
select auto hold
ON, ready ON or
Ready OFF
Measurement
mode
Calibration
mode
Memory mode or
when data is stored
into memory
GETTING STARTED
STORING AND RECALLING DATA
nIon Calibration (Ion 700 Only)
The available ION calibration values are 0.10, 1.0, 10.0, 100.0, and 1000 parts per million (ppm).
Choose a minimum of 2 consecutive values for calibration and prepare the corresponding ion
calibration solutions. For best results always begin with your lowest standard value, followed by
the next lowest standard.
1. Connect the ISE and press as needed to select ppm. Note: The primary reading will show
“- - - -” without a stored ion calibration. The secondary value is the corresponding mV reading
of the ISE.
2. Dip the ISE into your lowest standard solution value and stir. Press when the secondary
reading is stable.
3. Press or to match the primary display to your corresponding ion calibration
value (0.10, 1.0, 10, etc).
4. Press to accept or to abort. The primary display will show the next highest calibration
standard value.
5. Rinse the ISE then dip into the corresponding calibration standard and stir.
6. Press to accept or to cancel. The mV/decade slope value will be displayed briefly if the
calibration is successful. “SLP Err” indicates calibration for the current point was not successful.
7. Revert to step 3 to calibrate additional points or press as needed to return to measurement
mode at any time.
14
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com

npH Calibration
For best results, periodic calibration with known accurate standards is recommended. The 700
series meters can be calibrated with up to 5 buffers. The non-volatile memory retains all calibration
values upon meter shut down.
The following calibration standards are automatically recognized;
To eliminate temperature errors associated with the pH electrode, attach the automatic
temperature compensation (ATC) probe for best accuracy.
1. Press as needed to select pH.
2. Dip the pH and ATC electrodes into pH buffer and press . The secondary display will lock on
the appropriate buffer value. Provide stirring for best results.
When the READY indicator appears, press to accept. The primary reading will flash briefly
before the secondary display begins scrolling the remaining available buffers.
3. Rinse the pH and ATC electrodes then dip into the next pH buffer. The secondary display will
lock on the appropriate buffer value. When the READY indicator appears, press to accept.
The primary reading will flash briefly then display the percent efficiency (slope) before the
secondary display begins scrolling the remaining available buffers.
4. To calibrate another buffer repeat step 3 or press to return to the measurement mode.
nTemperature Calibration
The thermistor sensor used for automatic temperature compensation and measurement is
both accurate and stable. Temperature calibration is recommended upon electrode
replacement, whenever the temperature reading is suspect, or if matching against a certified
thermometer is desired.
1. Connect the temperature probe to the meter and place into a solution with a known accurate
temperature such as a constant temperature bath.
Note: To adjust the manual temperature compensation (MTC) value, do not connect the
temperature probe.
2. Press as needed to select pH or mV/R.mV.
3. Press followed by . The primary display shows the measured temperature while the
secondary display shows the factory default temperature.
4. Adjust the temperature using or . Press to accept or to cancel. The meter
allows an adjustable maximum value of ± 5 ºC (or ± 0.9 ºF) from the factory default temperature.
nMillivolt (mV) Oset Adjustment
Oxidization Reduction Potential (ORP or Redox) is not a precise measurement, but is useful as a
relative indicator. mV offset adjustment makes readings comparable to a reference.
1. Connect an ORP electrode and press as needed to select mV (or R.mV).
2. Dip the ORP electrode into a solution with a known mV value and stir.
3. Press when the reading is stable. The primary display shows the relative millivolt value
(R.mV) while the secondary display shows the factory default mV value.
4. Adjust the R.mV value using or . Press to accept or to cancel.
The meter allows an adjustable maximum value of ± 150 mV from the factory default
mV value.
Note: When an offset has been stored successfully, R.mV replaces mV.
nMeter Connections
pH BNC connection for pH, ORP (Redox), or
Ion Selective Electrodes (ISE).
REF
Pin connection for half cell reference electrodes.
Requires separate half cell BNC electrode.
Note: REF is not commonly used and is not required.
ATC For Automatic Temperature Compensation probe.
DC Power supply.
CALIBRATION
USA buer group 1.68, 4.01, 7.00, 10.01, 12.45
NIST buer group 1.68, 4.01, 6.86, 9.18, 12.45
DC
ATC REF.
pH
32

npH Calibration
For best results, periodic calibration with known accurate standards is recommended. The 700
series meters can be calibrated with up to 5 buffers. The non-volatile memory retains all calibration
values upon meter shut down.
The following calibration standards are automatically recognized;
To eliminate temperature errors associated with the pH electrode, attach the automatic
temperature compensation (ATC) probe for best accuracy.
1. Press as needed to select pH.
2. Dip the pH and ATC electrodes into pH buffer and press . The secondary display will lock on
the appropriate buffer value. Provide stirring for best results.
When the READY indicator appears, press to accept. The primary reading will flash briefly
before the secondary display begins scrolling the remaining available buffers.
3. Rinse the pH and ATC electrodes then dip into the next pH buffer. The secondary display will
lock on the appropriate buffer value. When the READY indicator appears, press to accept.
The primary reading will flash briefly then display the percent efficiency (slope) before the
secondary display begins scrolling the remaining available buffers.
4. To calibrate another buffer repeat step 3 or press to return to the measurement mode.
nTemperature Calibration
The thermistor sensor used for automatic temperature compensation and measurement is
both accurate and stable. Temperature calibration is recommended upon electrode
replacement, whenever the temperature reading is suspect, or if matching against a certified
thermometer is desired.
1. Connect the temperature probe to the meter and place into a solution with a known accurate
temperature such as a constant temperature bath.
Note: To adjust the manual temperature compensation (MTC) value, do not connect the
temperature probe.
2. Press as needed to select pH or mV/R.mV.
3. Press followed by . The primary display shows the measured temperature while the
secondary display shows the factory default temperature.
4. Adjust the temperature using or . Press to accept or to cancel. The meter
allows an adjustable maximum value of ± 5 ºC (or ± 0.9 ºF) from the factory default temperature.
nMillivolt (mV) Oset Adjustment
Oxidization Reduction Potential (ORP or Redox) is not a precise measurement, but is useful as a
relative indicator. mV offset adjustment makes readings comparable to a reference.
1. Connect an ORP electrode and press as needed to select mV (or R.mV).
2. Dip the ORP electrode into a solution with a known mV value and stir.
3. Press when the reading is stable. The primary display shows the relative millivolt value
(R.mV) while the secondary display shows the factory default mV value.
4. Adjust the R.mV value using or . Press to accept or to cancel.
The meter allows an adjustable maximum value of ± 150 mV from the factory default
mV value.
Note: When an offset has been stored successfully, R.mV replaces mV.
nMeter Connections
pH BNC connection for pH, ORP (Redox), or
Ion Selective Electrodes (ISE).
REF
Pin connection for half cell reference electrodes.
Requires separate half cell BNC electrode.
Note: REF is not commonly used and is not required.
ATC For Automatic Temperature Compensation probe.
DC Power supply.
CALIBRATION
USA buer group 1.68, 4.01, 7.00, 10.01, 12.45
NIST buer group 1.68, 4.01, 6.86, 9.18, 12.45
DC
ATC REF.
pH
32

Keys Functions
Powers the meter on and off. Upon power on, the meter
automatically begins in the mode that was last used. Calibration
and memory values are retained even if meter is unplugged.
Toggle between available measurement modes; pH/Temp,
mV/Temp, or ppm/mV (Ion 700 only).
Also used to switch from pH to Temp during pH calibration
mode. Press and hold for 5 seconds to enter SETUP mode.
Toggles between measurement and calibration modes.
In SETUP mode, returns user to the measurement mode.
MI (Memory Insert) stores values into memory.
Increase value. Scroll up in SETUP mode.
MR (Memory Recall) recalls values from memory
Decrease value. Scroll down in SETUP mode.
Freezes measured reading. Press again to resume live reading.
Confirms calibration values in CAL mode.
Confirms selections in SETUP mode.
View recalled values in memory mode.
nLCD Annunciators
Quick Guide
Bench 700 Series
pH/mV/Ion/Conductivity/TDS/Dissolved Oxygen/ºC/ºF
nKeypad Functions
Part of Thermo Fisher Scientific
The 700 series meters can retain up to 100 points into memory for later retrieval.
1. In the measurement mode, press to insert the measured value into memory.
The stored memory location value (StO) is briefly displayed.
2. To recall data from memory, press . The location of the most recent stored data is displayed
first. Press or to select the location of the desired data, then press to accept.
3. Press to return to the stored data location. Press to return to measurement mode.
For more information on our products, please contact our channel partner or visit our websites listed below: Automatic
temperature
compensation
is active
Setup
mode
Measured
value is held
Current
measurement
unit / mode
of operation
Electrode error
when appears
with ERR
Primary
reading
Secondary
reading
During pH CAL,
secondary reading is
the auto recognized pH
buer standard value
Measured
value is stable
Error or invalid
key pressed
Prompts user
to select buer
During setup,
select auto hold
ON, ready ON or
Ready OFF
Measurement
mode
Calibration
mode
Memory mode or
when data is stored
into memory
GETTING STARTED
STORING AND RECALLING DATA
nIon Calibration (Ion 700 Only)
The available ION calibration values are 0.10, 1.0, 10.0, 100.0, and 1000 parts per million (ppm).
Choose a minimum of 2 consecutive values for calibration and prepare the corresponding ion
calibration solutions. For best results always begin with your lowest standard value, followed by
the next lowest standard.
1. Connect the ISE and press as needed to select ppm. Note: The primary reading will show
“- - - -” without a stored ion calibration. The secondary value is the corresponding mV reading
of the ISE.
2. Dip the ISE into your lowest standard solution value and stir. Press when the secondary
reading is stable.
3. Press or to match the primary display to your corresponding ion calibration
value (0.10, 1.0, 10, etc).
4. Press to accept or to abort. The primary display will show the next highest calibration
standard value.
5. Rinse the ISE then dip into the corresponding calibration standard and stir.
6. Press to accept or to cancel. The mV/decade slope value will be displayed briefly if the
calibration is successful. “SLP Err” indicates calibration for the current point was not successful.
7. Revert to step 3 to calibrate additional points or press as needed to return to measurement
mode at any time.
14
99 Washington Street
Melrose, MA 02176
Phone 781-665-1400
Toll Free 1-800-517-8431
Visit us at www.TestEquipmentDepot.com
Table of contents
Other Oakton Measuring Instrument manuals

Oakton
Oakton pH 5+ User manual

Oakton
Oakton ORPTestr 5 User manual

Oakton
Oakton pH 10 Series User manual

Oakton
Oakton 35630-00 User manual

Oakton
Oakton 450 Series User manual

Oakton
Oakton EcoTestr pH1 User manual

Oakton
Oakton PD 300 User manual

Oakton
Oakton CON 6 User manual

Oakton
Oakton Ion 510 Series User manual

Oakton
Oakton WD-35610-00 User manual

Oakton
Oakton ION 6 User manual

Oakton
Oakton PCD 650 User manual

Oakton
Oakton 35700-00 User manual

Oakton
Oakton EC250 User manual

Oakton
Oakton CON 400 User manual

Oakton
Oakton 35612-series User manual

Oakton
Oakton 35618 Series User manual

Oakton
Oakton 35616-00 User manual

Oakton
Oakton 150 Series User manual

Oakton
Oakton PH550 User manual