Ohio Medical Moblvac III-cs Specifications

CONSTANT SUCTION
UNIT
OPERATOR/MAINTENANCE
M A N U A L
Ohio Medical Corporation
1111 Lakeside Drive
Gurnee, Illinois 60031-4099
1-800-662-5822/847-855-0800
Fax: 847/855-6218
www.ohiomedical.com

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
2
The purpose of the Moblvac®III-cs is to perform all routine constant suctioning procedures. These
may include tracheal and oral aspiration, wound and thoracic drainage. The Moblvac®III-cs is also
ideal for use within the G.I. lab as well as for backup suction in the O.R. and E.R. departments.
The Ohio Medical Moblvac®III-cs is a constant suction unit designed for use in the hospital, surgery
center, and physicians offices. Because of the Moblvac’s constant suction capacities are driven by a
fan cooled, rotary carbon vane pump. It comes equipped with a disposable exhaust filter, an in-line
suction filter with 14” of tubing, a disposable 1200cc collection canister with mounting bracket, and a
preventative maintenance kit.
This manual covers Moblvac®III-cs in the following sections:
1. INSTRUCTIONS FOR USE................................................................... Page 3
2. MAINTENANCE ................................................................................. Page 6
3. PERFORMANCE TEST & ADJUSTMENTS........................................ Page 9
4. TROUBLESHOOTING .......................................................................... Page 10
5. DEVICE SPECIFICATIONS ................................................................. Page 11
6. DETAILED DRAWINGS (120 VAC).................................................... Page 12
7. DETAILED DRAWINGS (220 VAC).................................................... Page 16
If, after reading this manual you require additional information, please contact Ohio Medicals’
Customer Service Department at 800-662-5822 or your local Ohio Medical distributor.

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
3
1. INSTRUCTIONS FOR USE
PRIOR TO INITIAL USE
Upon receiving your new Moblvac®III-cs, perform the following initial tests to ensure that your unit is
in good working order and that no damage has occurred during shipment.
SET UP INSTRUCTIONS
Your Moblvac®III-cs has been shipped with some minor assembly required. A 3/32” Allen wrench is
the only tool required.
1. Visually inspect all components for physical damage that may have occurred during shipping.
2. After removing all components from their cartons, insert the backpole assembly into the back
of the Moblvac®III-cs base.
3. Align the two “collection mounting brackets” located at the top of the backpole assembly to-
wards the front of the unit.
4. Tighten the 10-32 set screw located at the back of the base assembly with a 3/32” Allen
wrench.
5. Connect the tubing from the bottom of the backpole to the safety overflow trap assembly.
6. Plug the power cord into an electrical outlet. Depress the “POWER” switch and listen to verify
that the pump starts.
7. Plug the vacuum port located at the top of the backpole and adjust the vacuum level by turning
the vacuum regulator knob. Verify that the vacuum gauge reflects a charge in vacuum level
while turning the regulator knob.
8. Depress the “POWER” switch to turn the unit OFF (O).
9. Place a collection canister bracket (ring) in the slide mounting bracket of the backpole assembly or
pedestal stand and insert a collection canister.
IMPORTANT: This device is for use only by persons properly trained in medical suction
techniques and in the operations of suction equipment. Improper use could cause injury.
Thoroughly read this operations manual to familiarize yourself with the Moblvac®III-cs
before using the device.
DANGER: POSSIBLE EXPLOSION HAZARD IF USED IN THE PRESENCE OF
FLAMMABLE ANESTHETICS.

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
4
The Moblvac®III-cs will use any collection device. However, if you are not using an Ohio Medical
1200cc disposable collection canister, make sure that the collection device that is being used is
equipped with a safety overflow mechanism to protect the pump from accidental overflow.
Ensure that you have the appropriate collection device and suction tubing. Depending on the proce-
dure, ensure that you have the appropriate Chest Drainage Unit, Nasal Gastric tube, aspirating tip, or
wound drain for patient use. The Moblvac®III-cs is now ready to place into service.
OPERATING INSTRUCTIONS
Verify that a clean bacteriostatic exhaust filter, in-line suction filter, collection canister, and the neces-
sary tubing are installed on the Moblvac®III-cs.
ORAL, NASAL AND TRACHEAL ASPIRATION
1. Place a clean collection canister in the upper bracket, located on the backpole assembly of your
Moblvac® III-cs. Ensure that tubing is connected from the “VACUUM PORT ASSEMBLY” of
the Moblvac®III-cs to the “vacuum port” of the collection canister and from the “patient port”
of the collection canister to the aspirating tip that is being used.
2. Open the vacuum port located at the top of the backpole assembly and place your finger over
the outlet.
3. Turn the Moblvac®III-cs ON.
3. Adjust the vacuum to the desired vacuum level by using the “VACUUM REGULATOR”
Knob.
5. Open the vacuum port located at the top of the backole and proceed with the suction procedure.
PLEURAL DRAINAGE
Moblvac®III-cs can be utilized with any “Disposable Chest Drainage Unit” available on the market.
Some chest drainage units come with a built in flow adjustment valve and some do not. The Ohio
Medical Needle Valve Assembly is available for those units that do not offer this feature.
Follow all manufacturers’ directions on the Chest Drainage Unit for its setup and use.
Chest Drainage Units Without Built In Flow Adjustment Valves.
Place the Chest Drainage Unit in the “optional” Ohio Medical Chest Drainage Unit Bracket. Ensure
that tubing is connected from the “VACUUM PORT ASSEMBLY” of the Moblvac III-cs to the Ohio
Medical Needle Valve Assembly which is mounted in the collection canister bracket of the backpole.
Connect tubing from the Needle Valve Assembly to the vacuum port tubing of the Chest Drainage
Unit.

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
5
OR
Chest Drainage Units With Built In Flow Adjustment Valves.
Place the Chest Drainage Unit in the “optional” Ohio Medical Chest Drainage Unit Bracket. Ensure
that tubing is connected from the “VACUUM PORT ASSEMBLY” of the Moblvac®III-cs to the vac-
uum port tubing of the Chest Drainage Unit.
1. Open the vacuum port located at the top of the backpole assembly and place your figer over the
outlet.
2. Turn the Moblvac®III-cs ON.
3. Adjust the Moblvac®III-cs to a low vacuum setting by using the “VACUUM REGULATOR”
knob.
NOTE: The vacuum level provided to the chest cavity is regulated by the Chest Drainage Unit. Set-
ting the Moblvac to a low vacuum level ensures that you are producing sufficient vacuum to properly
operate the chest drainage unit.
4. Open the vacuum port located at the top of the backpole. Adjust the Ohio Medical needle
valve or Chest Drainage Unit (per the manufacturers’ instructions) as necessary and proceed
with the suction procedure.
REPROCESSING AND CLEANING INSTRUCTIONS
1. Discard all contaminated parts after any suctioning procedure. These components may include
the collection canister, disposable chest drainage unit, in-line filter, exhaust filter, and all suc-
tion tubing.
2. Wipe the surface of the unit clean with a mild antiseptic and a clean soft cloth.
3. Place a new collection canister, suction tubing, and filters with the Moblvac®III-cs.
4. Check the air inlet filter at the back of the unit to see that it is clear of any dirt. If cleaning is
necessary simply “pop off” the filter cover so the filter can be vacuumed or, if the dirt is exces-
sive, washed in a mild antiseptic and thoroughly rinsed in water. After the filter is completely
dried, reinstall the filter and snap the cover back into place.
5. Inspect the overflow trap assembly for any evidence of an accidental overflow. If an overflow
has occurred, use the following guidelines.

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
6
OVERFLOW - Aspirant has contaminated the overflow trap assembly. Replace and/or clean
parts according to procedures described in OVERFLOW CLEANING PROCE-
DURE on page 6.
SEVERE OVERFLOW - Aspirant has contaminated the overflow trap assembly and the vac-
uum port tubing. Replace and/or clean external parts according to procedure described in
OVERFLOW CLEANING PROCEDURE below. In addition, open the unit to assess the ex-
tent of the internal over flow.
OVERFLOW CLEANING PROCEDURE
NOTE: THESE PROCEDURES ARE ONLY REQUIRED WHEN AN OVERFLOW
HAS OCCURRED.
1. Discard all external contaminated tubing.
2. Clean the overflow trap assembly by unscrewing the cap and disassembling the various compo-
nents. Wash all parts with mild antiseptic, dry, reassemble and verify that the float is in the
down position and the cap seals tightly on the glass jar.
3. Replaces all discarded tubing.
4. If the vacuum port tubing was contaminated, open the unit and inspect the following parts for
contamination (See MAINTENANCE section below for opening the unit):
All Internal Tubing: If contaminated, replace.
Regulator Assembly: If contaminated, clean according to procedure on page 7.
Pump Assembly: If contaminated, clean according to procedure on page 8.
2. MAINTENANCE
Tools required to service the Moblvac III-cs.
(1) Pliers (1) Phillips screwdriver
CAUTION: NEVER DISASSEMBLE THE MOBLVAC III-cs WHEN THE POWER CORD
IS CONNECTED TO AN ELECTRICAL OUTLET.
(1) 11/32” nut driver (1) 5/64” Allen wrench
(1) Adjustable wrench (1) 1/16” Allen wrench
(1) Flathead screwdriver

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
7
To access the internal components of the Moblvac®III-cs, remove the two (2) Phillips head screws lo-
cated in the rear of the unit near the back assembly. Tilt the cover forward.
REVENTATIVE MAINTENANCE
Preventative maintenance is recommended every six months. It is up to the user’s discretion to clean
the unit more often after frequent use or to lengthen the schedule if use is infrequent.
A Preventative Maintenance Kit is available which contains all necessary items and instructions for
performing the preventative maintenance procedure.
VACUUM REGULATOR REPLACEMENT
1. Tilt the shroud assembly forward off of the Moblvac®III-cs base (make note of the orientation
of all tubing and the regulator body).
2. Disconnect all three (3) tubing connections from the regulator body.
3. Use a small flathead screwdriver to loosen the set screw on the vacuum regulator knob. Un-
screw and remove the vacuum regulator knob from the control panel.
4. Remove the outside locknut that secures the regulator to the control panel. The regulator can
now be removed. (Please note: the depth of the inside locknut on the shaft of the regulator
body. This determines the height of the regulator body in the control panel).
5. To mount the new regulator, first remove the knob from the body by loosening the set screw on
the vacuum regulator knob (Use a flathead screwdriver).
6. Mount the new regulator body to the control panel making sure the fittings are in the same ori-
entation as the original, by re-installing the outside locknut.
7. Insert the regulator knob fully onto regulator adjusting shaft.
8. Tighten the set screw until regulator knob is secure to shaft and cannot be pulled off by hand.
9. Reconnect all tubing.
VACUUM REGULATOR CLEANING
1. Remove the vacuum regulator assembly as described in the VACUUM REGULATOR RE-
PLACEMENT section above.

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
8
2. Discard all tubing
3. Open the regulator by removing the two screws on the cap, using a phillips screwdriver.
4. Disassemble black cartridge from unit by unscrewing it from the brass adjustment screw and
cap sub-assembly and remove two o-rings.
5. Clean the body, adjusting screw and cap, fittings and O-rings in a mild soap solution or isopro-
pyl alcohol and dry all parts completely. Do not wash black cartridge.
6. Apply a light coating of a silicone based lubricant (i.e.: Dow Corning 111) to the O-rings and
reassemble back onto the black cartridge.
7. Reassemble the vacuum regulator and remount the regulator in the Moblvac®III-cs.
8. Attach new vacuum tubing.
PUMP REPLACEMENT
1. Disconnect the three (3) pump wires from the PCB and two (2) wires from the capacitor.
2. Disconnect the tubing from the fittings on the pump.
3. If necessary, loosen the mounting bracket for the capacitor and remove the capacitor.
4. Using an 11/32” nut driver, remove the four (4) nuts from the base of the pump. Remove
pump.
5. To mount replacement pump, insert the pump onto the four (4) vibration bumper posts and se-
cure the pump in place.
6. Connect the pump wires to the PCB and the capacitor.
7. Reconnect all tubing.
8. Proceed to PERFORMANCE TEST & ADJUSTMENTS on page 9.
PUMP CLEANING
NOTE: The heart of the Moblvac®III-cs is a rotary carbon vane pump. It is NOT recommended that
the pump be disassembled for routine cleaning. However, if performance has been affected by the
pump becoming contaminated with aspirant, or if an overflow problem has occurred, the following

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
9
procedure should be performed.
1. Remove the pump assembly from the Moblvac III-cs as described above in the PUMP
REPLACEMENT section.
2. .Remove the three (3) pump head screws located on the pump head.
3. Remove the cover plate, shim, wear plate and the four (4) vanes.
4. Wash all exposed areas with an isopropyl alcohol solution.
Before you reassemble the pump make certain that all components are completely dry. Rust is likely
to form if any moisture is present.
5. Replace the vanes, wear plate, shim, and cover plate.
6. Replace and equally tighten the three (3) pump head screws.
7. Reinstall the pump assembly.
3. PERFORMANCE TEST AND ADJUSTMENT
It is recommended that you verify performance or the Moblvac III-cs after:
- An overflow.
- Preventative maintenance.
- Any maintenance.
To verify pump & regulator operation:
1. Plug the power cord into an electrical outlet. Turn the unit ON and listen to verify that the
pump starts.
2. Open the vacuum port located at the top of the backpole assembly and place your finger over
the outlet.
3. Adjust the vacuum level by turning the vacuum regulator knob. Verify that the vacuum gauge
reflects a change in vacuum level while turning the regulator knob. Also verify that you feel
vacuum at your finger tip.
4. Turn the unit OFF.
The following steps will verify the flow and the vacuum specifications:
1. Connect a suction flow measuring device to the vacuum outlet of the Moblvac®III-cs.
2. Turn the unit ON.

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
10
IV. TROUBLESHOOTING
Problem Cause Correction
Low or no vacuum on
running unit. 1. Regulator is turned all the way off. 1. Turn regulator knob clockwise to start
flow or increase vacuum.
2. An improper tubing connection or
crimped tube in the system. 2. Check all external vacuum parts for
crimped tubing. If still no vacuum
check all internal tubing connections.
3. Mechanical shut-off is activated in
either the overflow trap assembly or
the collection canister.
3. If mechanical shut-off has been
activated on a full canister, replace the
canister If the overflow trap assembly
has been activated in the safety
overflow jar. Follow Overflow Cleaning
Procedures on page 6.
4. Vacuum port is plugged. 4. Open vacuum port located at the top of
the backpole.
5. Collection canister improperly
installed or defective. 5. Check canister for any cracks. Verity
that all ports on the canister lid are
tight.
1. Unit is not plugged in. 1. Plug the unit into an outlet.
2. Faulty electrical connections. 2. Make sure that all wires are secured
tightly on the lugs and the lugs
themselves are secure on the
terminals.
3. Pump has seized. 3. Clean pump according to Pump
Cleaning on page 8.
4. The motor may be worn and cannot
deliver the torque required to operate
the pump, or the bearing is
damaged and is locking the rotor in
place.
4. Replace the pump.
5. Blown fuse(s). 5. Replace fuse(s). Also check #3 above.
1. Gauge is either not connected or is
faulty. 1. Check that tubing is properly
connected between vacuum regulator
and gauge.
2. Blockage in vacuum lines 2. Check that the 2-way solenoid is not
stuck in the pen position; or Check
Valve is not stuck closed; ofthat the
vacuum regulator is not completely
turned off.
High noise level. 1. Unit enclosure is not properly
closed. 1. Check that the unit is properly closed.
Verify that the outer flanges ofthe
edge trim are not stuck between the
enclosures.
2. Pump is running a high pitch. 2. Replace the vanes and clean the pump
or replace entire pump.
3. Loose fittings on the exhaust side of
the pump. 3. Check all fittings on the exhaust side
of the pump for any loose connection.
Pump does not turn
on when power
switch is depressed.
Gauge does not
register vacuum level.

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
11
5. DEVICE SPECIFICATIONS
PUMP
Rotary Carbon vane type.
PERFORMANCE
Vacuum Range: Up to 580mm Hg
Free Air Flow: 38 LPM minimum
CONTROLS
Vacuum Regulator: Rotary type on panel.
Vacuum Gauge: Calibrated in mm Hg
ELECTRICAL REQUIREMENTS
AC: 120V, 60Hz, 2A
220V, 50Hz, 2A
Fuses: Slow blow.
COLLECTION DEVICE
Canister: Disposable plastic with mechanical shutoff.
Capacity: 1200cc standard.
Tubing: 14” with bacterial filter.
PHYSICAL DIMENSIONS
Overall Height: 28”
Overall Width: 14”
OverallDepth: 16”
Weight: 32 lbs.
Warranty: 3 years
ISO 13485
FM 33489 MEDICAL EQUIPMENT LISTED 544U
MEDICAL EQUIPMENT CLASSIFIED BY UNDER-
WRITERS LABORATORIES INC. WITH RESPECT
TO ELECTRIC SHOCK, FIRE AND MECHANICAL
HAZARDS ONLY IN ACCORDANCE WITH UL
2601-1, AND C22.2 NO. 601.1
#9N56
220 vac

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
12
FIGURE 1

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
13
Figure 2

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
14
FIGURE 3

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
15

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
16
FIGURE 4

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
17
FIOGURE 5

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
18
FIGURE 6

Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, Il 60031-4099 (800)662-5822—Fax: (847) 855-6218
www.ohiomedical.com —– 550107 (Rev.5) 08/2005
19

Ohio Medical Corporation (800)662-5822—Fax: (847) 855-6218
www.aerosinstruments.com —– 550107 (Rev.5) 08/2005
1111 Lakeside Drive
Gurnee, IL 60031-4099
Phone: 800-662-5822/(847-855-0800)
Fax: 847-855-6218
http://www.ohiomedical.com
Table of contents
Other Ohio Medical Medical Equipment manuals

Ohio Medical
Ohio Medical Instavac I Installation instructions

Ohio Medical
Ohio Medical AMVEX CLICK STYLE User manual

Ohio Medical
Ohio Medical Push-To-Set PTS-ISU User manual

Ohio Medical
Ohio Medical PTS-ISU Series User manual

Ohio Medical
Ohio Medical care-e-vac 3 Installation instructions

Ohio Medical
Ohio Medical Push-To-Set PC2SA User manual

Ohio Medical
Ohio Medical MiniOX 3000 User manual

Ohio Medical
Ohio Medical PTS-ISU User manual