Ohmeda Giraffe OmniBed User manual

GiraffeTM OmniBedTM
Operator’s Manual

This Product will perform in conformity with the description thereof contained in this
operating manual and accompanying labels and/or inserts, when assembled, operated,
maintained and repaired in accordance with the instructions provided. This Product must
be checked periodically. A defective Product should not be used. Parts that are broken,
missing, plainly worn, distorted or contaminated should be replaced immediately. Should
such repair or replacement become necessary, Ohmeda Medical recommends that a
telephone or written request for service advice be made to the nearest Ohmeda Medical
Regional Service Center. This Product or any of its parts should not be repaired other
than in accordance with written instructions provided by Ohmeda Medical and by
Ohmeda Medical trained personnel. The Product must not be altered without the prior
written approval of Ohmeda Medical’s QualityAssurance Department. The user of this
Product shall have the sole responsibility for any malfunction which results from improper
use, faulty maintenance, improper repair, damage, or alteration by anyone other than
Ohmeda Medical.
CAUTION wU.S Federal law restricts this device to sale by or on the order of a licensed medical
practitioner.
AAAA 12345 This alpha character indicates the year of patient probe manufacture
and when the serial number was assigned; “A” = 1997, “B” = 1998, “C” = 1999,
“D” = 2000, etc. “I” and “O” are not used.
User Responsibility

Table of Contents
i i
6600-0342-000 08/08/01
General Information
Incubator/Warmer Operating Modes...................................................................1-1
Control Modes.....................................................................................................1-2
Controls and Displays .......................................................................................................1-2
Incubator Operation ............................................................................................1-3
Warmer Operation...............................................................................................1-4
Operating Mode Transition..................................................................................1-5
Accessory Options ............................................................................................................1-5
Alarms .............................................................................................................................1-11
Options Alarms................................................................................................................1-15
Alarm Table.....................................................................................................................1-16
Cable Connections and Mechanical Controls .................................................................1-18
Checkout Mechanical Checks ...........................................................................................................2-1
Controller Checks..............................................................................................................2-2
Accessory Checks.............................................................................................................2-3
Operating the OmniBed
Basic Operating Procedure ...............................................................................................3-1
Operation as an Incubator.................................................................................................3-3
Operation as a Warmer.....................................................................................................3-4
Attaching the Skin Temperature Probe .............................................................................3-4
Additional Operating Procedures ......................................................................................3-5
Accessing the Baby.............................................................................................3-5
Raising the Canopy.............................................................................................3-6
Raising and Lowering the Bed ............................................................................3-6
Tilting the Bed .....................................................................................................3-6
Using the X-ray Tray ...........................................................................................3-7
Administering Oxygen .........................................................................................3-7
Installing wristlets ................................................................................................3-7
Using the humidifier ............................................................................................3-8
Thermalink option...........................................................................................................................3-9
Using the In-Bed Scale .....................................................................................3-10
Mounting Accessories ..................................................................................................................3-11
Gas Cylinder Holder.....................................................................................................................3-12
Tubing Management Arm.............................................................................................................3-12
Oscillator Port Cover ....................................................................................................................3-12
Maintaining the OmniBed
Repair Policy .....................................................................................................................4-1
Maintenance schedule ......................................................................................................4-2
Disassembling the OmniBed for complete cleaning..........................................................4-3
Cleaning and disinfecting individual components .............................................................4-5
Accessories
Appendix
Specifications ................................................................................................................................ A-1
Performance.................................................................................................................................. A-2

Table of Contents
ii ii
6600-0342-000 08/08/01
wWarnings
Before using the OmniBed, read through this entire manual. As with all medical equip-
ment, attempting to use this device without a thorough understanding of its operation may
result in patient or user injury. This device should only be operated by personnel trained in
its operation under the direction of qualified medical personnel familiar with the risks and
benefits of this type of device. Additional precautions specific to certain procedures
are found in the text of this manual.
Complete the Checkout procedures in this manual before putting the unit into operation. If the
unit fails any portion of the checkout procedure it must be removed from use and repaired.
Do not use the OmniBed in the presence of flammable anesthetics; an explosion hazard
exists under these conditions.
Always disconnect the power before performing service or maintenance procedures
detailed in this manual. Apply power only if you are specifically instructed to do so as part
of the procedure.
Thoroughly air dry the OmniBed after cleaning it with flammable agents. Small amounts of
flammable agents, such as ether, alcohol or similar cleaning solvents left in the incubator
can cause a fire.
wCautions
Only competent individuals trained in the repair of this equipment should attempt to
service it as detailed in the Service Manual (6600-0343-000).
Detailed information for more extensive repairs is included in the service manual solely for
the convenience of users having proper knowledge, tools and test equipment, and for
service representatives trained by Ohmeda Medical.
Note: A Note provides additional information to clarify a point in the text.
CAUTION: A Caution statement is used when the possibility of damage to the equipment
exists.
WARNING: A Warning statement is used when the possibility of injury to the patient or
the operator exists.
Symbols
wAttention! Consult the manual for more information.
mType B equipment.
yFunctional Earth Terminal
xProtection Earth Terminal
Alarm Silence
∏Alternating Current
ot surface, do not touch
MAX Maximum
European Union Representative
Definitions

1-1 6600-0342-000 08/08/01 1-1
General Information
Incubator/Warmer operating modes
The GiraffeTMOmniBedTM combines the features of an incuba-
tor and a radiant warmer. When used as an incubator, a fan
and heater beneath the bed circulate warm air through the
closed infant compartment. The baby can be accessed
through portholes or doors. When you require even greater
access to the baby, the canopy can be raised and the com-
partment walls can be lowered. As the canopy raises, doors
open to expose a focused radiant heating element inside the
canopy that maintains the baby’s temperature like a warmer.
The OmniBedallows the full range of clinical procedures
without the disruptive intervention of transferring the baby from
one care bed to another.
CI.31.067
CI.31.001

1-2 6600-0342-000 08/08/01 1-2
General Information
Control Modes
The OmniBedhas two control modes when used as either an
incubator or a warmer. With the compartment closed, the
OmniBedcontrols the inside temperature comparing air
temperature readings from the probe on the compartment wall
to a setting you enter using the control panel (air mode) or
from a probe attached to the baby’s skin (baby mode). With
the canopy raised, the OmniBedcontrols radiant heater output
from a heater power percentage setting that you enter using
the control panel (manual mode) or by using temperature
readings from a probe attached to the baby’s skin (baby
mode).
Controls and Displays
The controls and displays, located at the top of the uprights at
the head of the bed, can be divided into two areas: on the left
are the buttons and numeric temperature displays that control
the temperature regulation of the infant, on the right are the
graphics display screen and control knob that control all the
accessory options.Temperatures can be shown in Celsius or
Fahrenheit degrees: the default setting is Celsius.
Temperature Regulation
At start up, the OmniBed prompts you to select one of the
control modes. If the unit was used in the last 2 hours, it will
ask if you would like to retain the settings you used previously.
If you are using the OmniBed as an incubator (canopy closed)-
you select baby or air mode. If you are using the OmniBed as
a warmer (canopy open)- you select baby or manual mode.
If you start with the canopy open, or open the bed during the
first minutes after start up, and you do not select a control
mode, the graphics screen will show the “Warm Up Mode”
message and the radiant heater will automatically run at 100%
power. When ten minutes has elapsed and you still have not
selected a control mode, the graphics display will show the
“Preheat Zone”message and the radiant heater will then run at
the preheat level; the default preheat level setting is 25%
power.
1. Alarm Light
2. Alarm Silence Button
3. Options Control Knob
4. Graphics Screen
5. Options Displays and Controls
6. Temperature Regulation Controls
7. ControlModeButtons
8. Temperature/Power controls
9. Numeric Displays
10.Power Failure Indicator
25
%
50
%
75
%
100
%
37
c
?
Options
Done/Exit
1
3
2
4
5
6
7
8
9
10
CI.31.007
1. Greater than 37°C Indicator
2. Greater than 37°C Button
3. Baby Mode Indicator
4. Baby Mode Button
5. Manual Mode Indicator
6. Manual Mode Button
7. Air Mode Indicator
8. Air Mode Button
9. Boost Air Curtain Button
10.Boost Air Curtain Indicator
11.Decrease Temperature/Power
12.Increase Temperature/Power
37
c
23
4
5
6
7
8
9
10
11
12
1

1-3 6600-0342-000 08/08/01 1-3
General Information
Incubator operation
When operating as an incubator, select the control mode by
pushing the Air Mode Button at the bottom or the Baby Mode
Button at the top of the temperature regulation controls.
Air Control Mode
TheAir Mode Indicator lights when air mode is selected. Two
temperature values are shown in the lower third of the display:
the value shown in the larger numerals at the top is the air
temperature measured by the compartment probe, the value
shown below in smaller numerals is the air set temperature. The
set temperature can be adjusted using the temperature/power
buttons located at the left of the mode buttons. Temperature
settings from 20°to 37°C can be entered in 0.1°C increments.
Temperatures from 37°to 39°C can only be entered after
pressing the >37°C Button.
Note: When operating the OmniBed as an incubator in the air
mode with a probe attached to the infant’s skin, the infant’s
temperature will also be shown in the baby temperature
section at the top of the display.
Baby Control Mode
The Baby Mode Indicator lights when the baby mode is selected.
Two temperature values are shown in the upper third of the
display: the value shown in larger numerals at the top is the
temperature measured by the probe on the infant’s skin, the
value shown below in smaller numerals is the baby set tempera-
ture, which can be adjusted by using the temperature/power
buttons located at the left of the mode buttons. Temperature
settings from 35°to 37°C can be entered in 0.1°C increments.
Temperatures settings from 37°to 37.5°C can only be entered
after pressing the >37°C Button.
Note: When operating the OmniBed as an incubator in the baby
mode, the compartment temperature will also be shown in the air
temperature display section at the bottom of the display.
Boost Air Curtain
If the OmniBed is being operated as an incubator, and the
doors will be open for an extended length of time, pushing the
Boost Air Curtain Button increases fan speed and improves
open door thermal performance.
An indicator on the button lights when the BoostAir Curtain is
activated. Running the fan at higher speed, however, will raise
the noise level in the infant compartment. To deactivate the
BoostAir Curtain, push the button a second time to resume
Whisper Quiet™mode. The BoostAir Curtain will automatically
shut off after twenty minutes and the indicator light will go out.
The button must be pressed again to continue for another twenty
minutes at high speed.
Note: The fan speed may increase automatically due to other
requirements without pushing the Boost Air Curtain button, and
in these cases the button indicator will not light.
1. Air Set Temperature
2. Air Temperature
3. Air Mode Button
25
%
50
%
75
%
100
%
CI.31.008
12
3
25
%
50
%
75
%
100
%
CI.31.009
1. Baby Mode Button
2. Baby Set Temperature
3. Baby Temperature
1
2
3
CI.31.010

1-4 6600-0342-000 08/08/01 1-4
General Information
Warmer Operation
When operating as a radiant warmer, select the control mode
by pushing the Manual Mode Button in the middle or the Baby
Mode Button at the top of the controls.
Manual Control Mode
The Manual Mode Indicator lights when the manual mode is
selected. The power setting is indicated by the bar graph LED
in the middle third of the display. The percentage of full power
can adjusted in 5% increments using the temperature/power
buttons located to the left of the mode buttons.
Note: when operating the OmniBed as a warmer in the
manual mode with a probe attached to the infant’s skin, the
infant’s temperature will also be shown in baby temperature
section at the top of the display.
Baby Control Mode
The Baby Mode Indicator lights when baby mode is selected.
Two temperature values are shown in the upper third of the
display: the value shown in larger numerals at the top is the
temperature measured by the probe on the infant’s skin, the
value shown below in smaller numerals is the baby set tem-
perature. The set temperature can be adjusted by using the
temperature/power buttons located to the left of the mode
buttons. Temperature settings from 35°to 37°C can be entered
in 0.1°C increments. Temperature settings from 37°to 37.5°C
can only be entered after pressing the >37°C Button.
Note: when operating the OmniBed as a warmer in the baby
mode, the bar graph in the middle of the display will also
indicate the percentage of power at which the heater is
currently running.
Operating with Two Temperature
Probes
The OmniBed is equipped with two patient temperature probe
jacks for co-bedding of twins, or to monitor a single baby’s
temperature from two anatomical sites. They are identified as
probe 1 and probe 2 on the jack panel behind the bed. Tem-
perature readings for probe jack 1 are displayed in Baby
Temperature at the top of the numeric displays. Temperature
readings from probe jack 2 are displayed in the upper right
corner of the graphic display labeled Probe 2.
The OmniBed will only Baby Control from a probe attached to
jack 1. When both jack 1 and jack 2 are used, the unit will not
operate in the Baby Control mode.
CI.31.011
1. Manual Mode Button
2. % Radiant Heater Power
25
%
50
%
75
%
100
%
1
2
1. Baby Mode Button
2. Baby Temperature
3. Baby Set Temperature
25
%
50
%
75
%
100
%
CI.31.012
1
2
3
?
Options
Done/Exit
Probe 2
36.8
Alarm messages displayed here
1 :3 am
6/23/Name Birth: 6/21/ Wed
1. Probe 2 Temperature
1
CI.31.013

1-5 6600-0342-000 08/08/01 1-5
General Information
Operating Mode Transition
If the canopy is raised while the OmniBed is being operated as an incubator, it
automatically switches to operation as a radiant warmer. If it was in the baby
mode as an incubator, after the canopy is raised it will continue to operate in
the baby mode, using the same baby set temperature you entered earlier.
If the canopy is raised while the OmniBed is being operated in the air mode as
an incubator, it will operate in the manual mode as a warmer. The radiant
heater will activate at the preheat level (the factory set preheat level is 25%;
this can be adjusted on the service screen), and the OmniBed will prompt you
to select a heater power percentage.
If the canopy is lowered while the OmniBed is being operated as a radiant
warmer, it automatically switches to operation as an incubator. If it was in the
baby mode as a warmer, after the canopy is lowered it will continue to operate in
the baby mode, using the same baby set temperature you entered earlier.
If the canopy is lowered while the OmniBed is being operated in the manual
mode as an warmer, it will operate in the air mode as an incubator. The
compartment heater and fan will start at the last air set temperature saved
since power-off, or at the preheat temperature of 33°C, and the OmniBed will
prompt you to select an air set temperature.
Accessory Options
The OmniBed may be equipped with a number of accessory options.
Menu icons of these options are shown on the graphics display at the
right of the control area. Pressing the knob activates the screen. Rotating
the control knob either direction scrolls through each menu icon; highlight-
ing each icon as it is selected. Pressing in the control knob selects the
highlighted option and brings up its menu screen.
?
Options
Done/Exit
Probe 2
36.8
Alarm messages displayed here
1 :3 am
6/23/Name Birth: 6/21/ Wed
1. Help
2. Timer
3. Trending
4. Comfort Zone
5. Setup/Customize
6. Scale
7. Day
8. Time*
9. Date*
10. Baby Information
11. Humidity
12. Options
11
234
5
6
7
1
8
12
CI.31.013
9
10 * The date and time may be set using the
service screen; see the service manual.

1-6 6600-0342-000 08/08/01 1-6
General Information
Help: Select the Help icon (question mark) to bring up the
“Help”screen. The screen will present a menu of topics on
how your OmniBed unit works. “Alarms”list all the alarm
messages on the left and more detailed descriptions of the
alarm and how to react to a given alarm on the right. “Controls”
identifies the various button and switches on the control panel.
“Operation”lists basic procedures on the left and step by step
instructions on the right. Select Done/Exit to return to the
option icon screen.
If you select the Help icon during an alarm condition, the
description of that particular alarm will automatically appear on
the screen; it is not necessary to scroll through the help menu
options to find it.
?
Alarms
Controls
Operation
Done/Exit
1 :3 am
6/23/Name Birth: 6/21/ Wed
ALARMS
High Air Temp
Low Air Temp
Baby Cold
Baby Hot
Air Temp> 38C
Air Temp> 4 C
down
Done/Exit
BABY COLD
1. C under BABY SET TEMP
Temp probe poorly attached to baby
Reheating a very cold baby
BABY SET TEMP too low
Adjust and/or monitor to assure baby
temp increases.
DONE/EXIT
Baby Cold-Check Probe 1
1 :3 am
6/23/Name Birth: 6/21/ Wed
OPERATION
Start Up
Incubator Use
Warmer Use
Done/Exit
1. Preheat incubator in AIR MODE
appropriately for patient. Use
COMFORT ZONE screen to calculate
an appropriate temperature.
Select temperature using the
up or down switches. Press >37C
button to adjust above 37C.
2. If BABY MODE desired, attach sensor
to baby and plug into panel.
3. Switch to BABY MODE and select
control temperature desired.
CONTINUE
DONE/EXIT
1 :3 am
6/23/Name Birth: 6/21/ Wed
Alarm messages displayed here
CONTROLS
37
cBaby Control
Override
Increase Set Point
Manual Control
Decrease Set Point
Boost Air Curtain
Air Control
1 :3 am
6/23/Name Birth: 6/21/ Wed
CI.31.021
CI.31.082
CI.31.023
CI.31.024

1-7 6600-0342-000 08/08/01 1-7
General Information
Timer: Select the Timer icon (clock face) to bring up the
“Timer”menu screen. Select Start/Stop to start timer; select it
again to stop the timer. Scroll down and select Reset to return
the timer to zero. Select the Apgar Tone symbol (musical note)
to activate or deactivate a tone after one minute and every 5
minutes after the timer was started. Select Done/Exit to return
to the option icon screen.
Note: If the OmniBed was powered up as a warmer, this audio
tone will sound as a default; if the unit was started as an
incubator, the audio tone will be off as a default.
Trend: Select the Trending icon (graph) to bring up the “Trend-
ing”screen. The graph will plot temperatures from two patient
skin probes (if they are installed in the jacks), the air tempera-
ture read at the compartment probe, and the set temperature
entered during the time period. The graph also shows what
control modes the OmniBed was in during the trended period,
and heater power percentage if it was operated as a radiant
warmer. By selecting “Humid”in the “Display”option the
“Modes”graph is replaced by a graph plotting the percent
relative humidity read by the compartment humidity sensor.
Turn the knob to scroll through 2, 8, 24, or 96 hours and press
it to select the desired trending time period.
Note: The trending screen does not retain data that is not
currently shown on the display. For example, if you have
selected that only 2 hours of data be trended, after the display
has been running for 3 hours, the first hour of data is lost.
?
Options
Done/Exit
Timer
Start/Stop
Reset
Tones
Done/Exit
O:OO
1 :3 am
6/23/Name Birth: 6/21/ Wed
?
Options
Done/Exit
1 :3 am
6/23/Name Birth: 6/21/ Wed
Trending
Pat temp 1:
Pattemp 2:
Air temp:
Set Temp:
Display: Mode/Humid
Time Span: 2 hr
Done/Exit
11:3
C
36
37
35
34
12: 12:3 13:
Rad Power
Baby
Air
Manual
1 :3 am
6/23/Name Birth: 6/21/ Wed
Trending
Pat temp 1:
Pattemp 2:
Air temp:
Set Temp:
Display: Mode/Humid
Time Span: 2 hr
Done/Exit
11:3
C
36
37
35
34
12: 12:3 13:
8
6
4
2
Humidity
%
1 :3 am
6/23/Name Birth: 6/21/ Wed
CI.31.025
CI.31.026
CI.31.068
CI.31.069

1-8 6600-0342-000 08/08/01 1-8
General Information
Comfort zone: Select Comfort Zone icon (thermometer and
face) to bring up the “Comfort Zone”screen.
Important: The comfort zone screen is only a reference
information screen used for calculating suggested temperature
ranges* during air control mode from data you supply. You
must manually enter the temperature settings using the
buttons on the temperature regulation controls.
Push the knob to select weight then enter the patient’s weight
range in grams by turning the knob in 100 gram increments.
Push the knob to enter the weight value.
Scroll to Gest age.
Push the knob to select Gest age then turn the knob to find the
patient’s gestational age in weeks.
Push the knob to enter the gestational age value and scroll
down to Post natal age.
Push the knob to select Post natal age then turn the knob to
find the age in days. Push the knob to enter the postnatal age.
The suggested incubator air set temperature range will now
appear in degrees Celsius. If you wish to use this suggested
temperature setting, enter it using temperature control buttons
to the left of the screen.
Select done/exit to return to the option menu screen.
*
Adapted from Sauer, Dane and Visser, Archives of Disease in Childhood, 1984. 59, 18-22.
?
Options
Done/Exit
1 :3 am
6/23/Name Birth: 6/21/ Wed
?
Options
Done/Exit
Comfort Zone
Weight
Gest age
Post natal age
Done/Exit
Suggested Temp Range
34.4 - 35.4 C
14 -15 grams
34 wks
1 day
1 :3 am
6/23/Name Birth: 6/21/ Wed
CI.31.027
CI.31.028

1-9 6600-0342-000 08/08/01 1-9
General Information
Setup/Customize: Select Setup/Customize icon (wrench) to
bring up the setup screen.
Select “Temp Units”to change temperature displays from the
default Celsius setting to Fahrenheit.
Select “Alarm Volume”to choose from 4 volume levels.
Select “Man Temp Alarm”to choose from 3 settings: choose “Off”
to deactivate the alarm, choose “.5”to set the alarm to activate
when a 0.5°C difference is read between a set temperature and
the baby probe temperature, or choose “1.0”to set the alarm to
activate when a 1.0°C difference is read between a set tempera-
ture and the baby probe temperature. “Off”is the default setting.
The “Man Temp Alarm”can be used in either the warmer Manual
mode or the incubator Air mode. However, to select a set tem-
perature you must first select the Baby mode, input the set point,
then select the Manual orAir modes.
This optional alarm provides you audiovisual notification in the Air
control mode if the skin probe temperature drifts outside the
range you select. For example, if you select 0.5°C on the setup
screen, and then set 36.5°C as the skin temperature threshold in
the Baby mode before switching to the Air mode, the alarm will
sound if the baby’s skin temperature falls below 36°C or rises
above 37°C while in theAir mode. If this alarm is not required, or
if the skin temperature probe is not being used, “Man Temp
Alarm”should always be set to “Off”on the setup screen.
Select “Patient Alarm”to choose between 2 settings: choose
“.5””to set the alarm to activate when a 0.5°C difference is
read between a set temperature and the baby probe tempera-
ture, or choose “1.0”to set the alarm to activate when a 1.0°C
difference is read between a set temperature and the baby
probe temperature.”1.0”is the default setting. The “Patient
Alarm”is used in either the incubator or warmer Baby modes.
Select “Canopy Pedal”to disable or enable foot pedals that
raise the canopy. When the pedal is disabled, the canopy can
still be raised or lowered with the buttons on either side at the
head of the bed.
Select “Elevate”to disable or enable the foot pedals that raise
or lower the bed height.
Select “Baby Info”to bring up the Baby Info screen. By turning
the knob until the desired character appears and then pushing
the knob, you can enter up to a 9-character name. Blanks and
periods can also be selected. Selecting “”allows you to go
back and change the previous character. Selecting “”
enters the letters you have selected for names that are less
than 9 characters long. By selecting the month, day and year
numbers, you can enter the baby’s birth date.
These settings will be saved for 2 hours after power down
Select Done/Exit to return to the option icon screen.
Note: Aservice screen allows you to change preheat
% power, alarm criteria, disable the canopy control pedal and
other default settings. See the service manual for a description
of this screen and how it may be accessed.
?
Options
Done/Exit
Alarms:
Setup
Temp Units C
Alarm Volume
Man Temp Alarm Off
Patient Alarm 1. C
Canopy Pedal Enable
Elevate Enable
Baby Info
Done/Exit
1 :3 am
6/23/Name Birth: 6/21/ Wed
CI.31.031

1-10 6600-0342-000 08/08/01 1-10
General Information
Scale: Available if the in-bed scale is installed. Select the
Scale icon (balance scale) to bring up the scale menu screen.
Select Weigh to start the weighing procedure.An audio prompt
sounds and the screen will prompt you to lift the baby from the
bed surface while the scale zeros. It then tells you to replace
the baby on the bed and calculates the infant’s weight. Select
Trend to plot a series of weights on a graph. Select Modify to
change to smaller increments on the Y axis, and so enlarge
the differences shown between weights plotted on the graph.
Select Done/Exit to return to the option icon screen.
?
Options
Done/Exit
1 :3 am
6/23/Name Birth: 6/21/ Wed
?
Options
Done/Exit
85
%
85
%
set
Humidifier
1 :3 am
6/23/Name Birth: 6/21/ Wed
CI.31.032
CI.31.033 CI.31.029
?
Options
Done/Exit
1 :3 am
6/23/Name Birth: 6/21/ Wed
Weigh
Enter
Modify
Done/Exit
8000
6000
4000
2000
0
gms
1 :3 am
6/23/Name Birth: 6/21/ Wed
CI.31.030
Servo-Humidity: Select the Humidity icon to bring up the
humidity menu screen. The actual relative humidity measured
inside the infant compartment is displayed in large numerals.
The set relative humidity is shown in smaller numerals. Adjust
the relative humidity percentage to the desired level by push-
ing the knob to highlight the setting. Turn the knob to increase
or decrease the humidity setting then push the knob to enter
the new setting. Exit the menu. The actual and set relative
humidity will remain displayed. To deactivate the humidifier, set
the humidity level to “off”.

1-11 6600-0342-000 08/08/01 1-11
General Information
Alarms
An alarm message appears across the top graphics display
screen identifying the various alarm conditions. In addition, the
alarm light at the top of the control panel housing activates and
an audio alarm sounds.
The audio tone varies with the alarm’s priority: an alternating
two tone alarm indicates a critical alarm that triggers automatic
heater shutdown; an intermittent single tone alarm is used for
other alarms. The alarm light remains on continuously for over
temperature and system failure alarms; it blinks for all other
alarms.
With the exception of the system failure and power failure
alarms, all audio alarms can be silenced by pressing the alarm
silence button located above the graphics screen. At the end of
the alarm silence period the alarm reactivates if the alarm
condition has not been resolved. If two or more alarm condi-
tions occur at the same time, the audio signal for the highest
priority alarm will sound while all the alarm messages will
alternately appear at the top of the graphics screen. If you
select the Help icon during an alarm condition, the description
of that particular alarm will automatically appear on the screen.
1. Alarm Light
2. Alarm Silence Button
3. Alarm Display
?
Options
Done/Exit
Probe 2
36.8
High Air Temperature
1 :3 am
6/23/Name Birth: 6/21/ Wed
?
Options
Done/Exit
Probe 2
36.8
Low Air Temperature
1 :3 am
6/23/Name Birth: 6/21/ Wed
High Air Temperature
Active only when the canopy is closed and the unit is in the air
mode, the High Air Temperature alarm triggers when the
temperature measured by the compartment probe exceeds the
air temperature you set previously by more than 1.5°C. The
audio alarm is suspended for ten minutes when you push the
alarm silence button. This alarm usually occurs when you have
just lowered the air set temperature, or when you have
changed from baby to air control mode. Monitor the unit
closely to assure the air temperature decreases.
Low Air Temperature
Active only when the canopy is closed and the unit is in the air
mode, the Low Air Temperature alarm triggers when the
temperature measured by the compartment probe falls more
than 3.0°C below the air temperature you set previously. The
audio alarm is suspended for a ten minute period when you
push the alarm silence button. This alarm usually occurs when
the doors are open without using Air Curtain Boost, when the
canopy has just been lowered, or when both doors are open at
the same time. Monitor the unit closely to assure the air
temperature increases.
Note: The Low Temperature alarm is suspended for 60 minutes
when the unit is first turned on, for 30 minutes after the canopy is
closed, and 15 minutes after the set point is increased.
CI.31.035 CI.31.034
?
Options
Done/Exit
Alarm messages displayed here
123
CI.31.054

1-12 6600-0342-000 08/08/01 1-12
General Information
Baby Hot - Check Probe 1
Active whenever the unit is in the baby mode, the Baby Hot
alarm activates when the temperature measured by the probe
on the baby skin is 1.0°C above the baby temperature you set
previously (this can be adjusted to 0.5°C using the Setup/
Customize screen). Monitor the patient carefully and check the
attachment of the probe on the baby’s skin. The audio alarm is
suspended for a ten minute period when you push the alarm
silence button. This alarm can also be activated in the Manual
and Air modes. See setup screen.
Baby Cold - Check Probe 1
Active whenever the unit is in the baby mode, the Baby Cold
alarm activates when the temperature measured by the probe
on the baby skin is 1.0°C below the baby temperature you set
previously (this can be adjusted to 0.5°C using the Setup/
Customize screen). Monitor the patient carefully and check the
attachment of the probe on the baby’s skin. The audio alarm is
suspended for a ten minute period when you push the alarm
silence button. This alarm can also be activated in the Manual
and Air modes. See setup screen.
Air Temperature Greater Than... Alarms
These alarms warn of abnormally high temperatures read by
the compartment probe in all control modes and automatically
shut off the heater. They are not self resetting; once the alarms
are activated they must be manually reset by pushing the
alarm silence button, even if the alarm condition is resolved.
Computer independent circuitry continuously measures and
compares the compartment temperature with factory set
levels. Visual and audible alarms are activated and the heater
shuts off if the compartment temperature exceeds these preset
levels, independent of the software.
The functionality of all other alarms is continually checked by
the microprocessor software during normal operation. If a fault
occurs in any of this circuitry, an indicator lights and a tone is
sounded.
Additionally, a computer independent “watchdog”circuit will
reset the computer, turn off the heater, and activate the alarms
in the event of a microprocessor failure or software error.
Air Temperature >38 C
Active only when the canopy is closed and the set temperature
is below 37°C, the Air Temp >38°C alarm activates in the air
mode when temperatures greater than 38°C are read by the
compartment probe. This alarm may occur when you have just
lowered the air set temperature, or when you have changed
from baby to air control mode. Monitor the unit closely to
assure the air temperature decreases.
?
Options
Done/Exit
Baby Hot- Check Probe 1
1 :3 am
6/23/Name Birth: 6/21/ Wed
?
Options
Done/Exit
Baby Cold- Check Probe 1
1 :3 am
6/23/Name Birth: 6/21/ Wed
?
Options
Done/Exit
Air Temperature > 3 C
1 :3 am
6/23/Name Birth: 6/21/ Wed
CI.31.036
CI.31.037
CI.31.038

1-13 6600-0342-000 08/08/01 1-13
General Information
Air Temperature > 40 C
In the air mode, when the air temperature is set above 37°C by
using the >37°C key, the Air Temp >40°C alarm activates when
temperatures greater than 40°C are read by the compartment
probe.
In the baby mode, the Air Temp >40°C alarm activates when-
ever temperatures greater than 40°C are read by the compart-
ment probe. Check for external heat sources on the baby
(sunlight, phototherapy lamp, etc.)
Disconnected Baby Probe 1
Active in all modes, the Disconnected Baby Probe alarm
activates when the baby skin probe jack is unplugged at the
jack panel. Check that jack is fully inserted. Check that the
Man Temp Alarm has not been set in error.
Fan Failure
Active only when the canopy is closed, the Fan Failure alarm
triggers when the fan is missing or malfunctioning and there is
inadequate air flow. The heater is automatically turned off.
Remove the baby from the unit. Service required.
Fan Always in High Speed
Active only when the canopy is closed, the Fan Always in High
Speed alarm activates when the heater temperature sensor
malfunctions.
Air Probe Disconnect
Active only when the canopy is closed, the Air Probe Discon-
nect alarm activates when the compartment air temperature
probe is disconnected. The heater is automatically turned off.
Confirm connection of compartment probe to unit.
Air Probe Failure
Active only when the canopy is closed, the Air Probe Failure
alarm activates when the compartment air temperature probe
malfunctions. The heater is automatically turned off. Remove
the baby from the unit. Service required.
In Transition- Heat Off
If the canopy is stopped or is prevented from reaching it’s fully
raised or fully closed position, this alarm activates. All heat is
shut off until the canopy reaches its travel limit.
?
Options
Done/Exit
Air Temperature > 40 C
1 :3 am
6/23/Name Birth: 6/21/ Wed
?
Options
Done/Exit
Disconnected Baby Probe
?
Options
Done/Exit
Fan Failure
?
Options
Done/Exit
Fan Always in High Speed
?
Options
Done/Exit
Air Probe Disconnect
?
Options
Done/Exit
Air Probe Failure
?
Options
Done/Exit
In Transition- Heat Off
CI.31.039
CI.31.040
CI.31.041
CI.31.042
CI.31.043
CI.31.044
CI.31.045

1-14 6600-0342-000 08/08/01 1-14
General Information
Check Baby
Active only when the canopy is open, the Check Baby alarm activates
when the radiant heater has been running at 100% power for more
than twelve minutes. Check the attachment of the skin probe to the
baby and monitor the baby’s temperature closely.
When the unit is in the manual mode, the Check baby alarm activates
when the radiant heater’s preheat power percentage has been
exceeded for more than twelve minutes. The default power setting is
25%, but this can be adjusted on the Service Screen. See the service
manual for information about how this screen may be accessed.A
single alarm tone sounds.After fifteen minutes, a dual tone alarm
sounds, the heater is turned off automatically, and the alarm message
changes to “Check Baby- Heat Off.”
System Failure
WARNING ww
ww
wDo not use the OmniBed if the system
failure alarm is activated. Remove the
unit from service and refer to qualified
personnel for repair.
If an electrical failure is detected, the system failure alarm
automatically shuts off the heater and triggers a two tone audio
alarm that can not be silenced.
Bad Membrane Switch
One of the electronic switches behind the control panel buttons
has failed. This alarm can not be silenced.
Power Failure
If power to the unit is cut off, an audio alarm sounds and
a LED lights.
Service Alarms
The following alarm messages indicate a repair is required, but
the OmniBed can be used in another operating mode, the unit
need not be taken immediately out of service.
•Radiant Heater Failure- The OmniBed can stilled be used in
the closed bed incubator operating mode.
•Heater Doors Not Closed- The OmniBed can still be used in
the open bed warmer operating mode.
•Canopy Pedal Failure- Use the hand buttons to raise or
lower the canopy.
•Bed Up Pedal Failure- A switch in the pedal is shorted.
Continue to use at the current height.
•Bed Down Pedal Failure- A switch in the pedal is shorted.
Continue to use at the current height.
•Bed Heater Failure- The OmniBed can still be used in the
open bed warmer operating mode.
•Motor Drive Failure- The drive system for raising and lowering the
canopy or raising and lowering the bed has failed. Continue to
use the Omnibed in its current operating mode. The humidifier
will not operate when this alarm is activated.
?
Options
Done/Exit
Check Baby
1 :3 am
6/23/Name Birth: 6/21/ Wed
?
Options
Done/Exit
System Failure
A
CI.31.046
CI.31.047
CI.31.055
1
1. Power Failure Indicator

1-15 6600-0342-000 08/08/01 1-15
General Information
Option Alarms
Humidifier
Add Water- Water level low in humidifier reservoir. Add
distilled water.
Water Reservoir Not Engaged- Humidity water reservoir
missing or incorrectly installed.
Humidity Probe Failure- Probe failed; humidifier not operat-
ing. Check jack connection. Consider increasing air tempera-
ture or removing baby.
Humidity Failure- Heater failed; humidifier not operating.
Repair humidifier.
?
Options
Done/Exit
Add Water
?
Options
Done/Exit
Water Reservoir Not Engaged
?
Options
Done/Exit
Humidity Probe Failure
?
Options
Done/Exit
Humidifier Failure
1 :3 am
6/23/Name Birth: 6/21/ Wed
Scale
Exceeds Maximum Weight- Weight on scale platform is 8
kilograms or more. Check for other objects on scale
Scale failure- System failure; note error code and remove
scale from service.
?
Options
Done/Exit
Exceeds Maximum Weight
?
Options
Done/Exit
Scale Failure 3
CI.31.048
CI.31.049
CI.31.050
CI.31.051
CI.31.052
CI.31.053

1-16 6600-0342-000 08/08/01 1-16
General Information
Alarm Table
Temperature Regulation
Alarm Activation criteria Alarm Silence Audio Signal Mode Heater Status Inc Wmr
High Air Temp 1.5°C over AST 10 min 2 Air On On/Off
Low Air Temp 3.0°C under AST 10 min 2 Air On On/Off
Baby Cold 1.0°C under BST* 15 min 2 Baby On On/On
5 min >2°C1
Baby Hot 1.0°C over BST* 15 min 2 Baby On On/On
5 min >2°C1
Disconnected Disconected 2 min 2 Baby On On/On
Baby Probe at jack panel 1 (after 30 sec)
Air Temp >38°C >38°C 10 min 1 All Off On/Off
(AST <37°C)
Air Temp >40°C >40°C 10 min 1 All Off On/Off
(AST >37°C)
Fan Failure Fan missing 10 min 1 All Off On/Off
or not rotating
Fan Always in Heater none 2 All On On/Off
High Speed temperature
sensor failure
Air Probe Disconnection 10 min 2 All Off On/On
Disconnect of probe connector 1 (after 30 sec)
Air Probe Failure Failure 10 min 2 All Off On/On
of compartment air probe 1 (after 30 sec)
Power Failure Switch on, non-silenceable 1 All Off On/On
but no power 2 (after 4 sec)
System failure any non user non-silenceable 1 All Off On/On
correctable failure
Check Baby Radiant heater at 12 min 2 Baby Off On/On
100% for 12 min 1 (after 15 min)
Radiant heater % 12 min 2 Manual On Off/On
above preheat set 1 (after 15 min) Manual Off Off/On
In Transition- Canopy stopped 10 min 2 (after 30 sec) All Off On/On
Heater Off in transition 1 (after 2 min)
Table of contents
Other Ohmeda Medical Equipment manuals
Popular Medical Equipment manuals by other brands

ATS Dental
ATS Dental I TECH LITE Operation and maintenance instruction manual

Dräger
Dräger PC Link Quick reference guide

Roche
Roche CoaguChek XS System Training manual

Otto Bock
Otto Bock 50P10 Manu Immobil Instructions for use

BEARMACH
BEARMACH HawkEye PRO user manual

BLACKROLL
BLACKROLL Posture user manual