Ohmeda Ohio 2000 Series User manual

I
and Maintenance Manual
Ohio@ Infant Warmer
Systems
All Models

11109/93
A
!
CAUTION: Federal law in U.S.A. and Canada restricts this device to
sale by or on the order of a licensed medical practitioner.
6600-0194-000
User Responsibility
This Product will perform in conformity with the description thereof
contained in this operating manual and accompanying labels and/or
inserts, when assembled, operated, maintained and repaired in accor-
dance with the instructions provided. This Product must be checked
periodically. A defective Product should not be used. Parts that are
broken, missing, plainly worn, distorted or contaminated should be
replaced immediately. Should such repair or replacement become neces-
sary, Ohmeda recommends that a telephone or written request for
service advice be made to the nearest Ohmeda Regional Service Center.
This Product or any of its parts should not be repaired other than in
accordance with written instructions provided by Ohmeda and by
Ohmeda trained personnel. The Product must not be altered without the
prior written approval of Ohmeda’s Quality Assurance Department. The
user of this Product shall have the sole responsibility for any malfunction
which results from improper use, faulty maintenance, improper repair,
damage, or alteration by anyone other than Ohmeda.

11/09/93
i
3/0peration
3-1
3.1 Control Panel Operation 3-l
A. Displays 3-l
B.
Indicator Lights 3-2
C. Switches 3-2
D. Alarms 3-4
3.2
Manual and Servo Mode Operation
3-7
3.3
Elapsed Timer Operation 3-11
6600-0194-000
Model Descriptions v
Definitions vii
Precautions ix
Warnings ix
Cautions xiii
l/Description l-l
1.1
General l-l
1.2 Support Structure l-l
1.3 Heater Assembly l-l
1.4 Control Unit l-2
1.5 Bed Platform l-2
1.6
Bassinet l-2
1.7
Accessories l-3
S/Setup and Checkout Procedure
2-1
2.1 Setup 2-l
2.2 Mechanical Checkout Procedure 2-l
A. Overall Appearance 2-l
B.
Heater Assembly Rotation 2-2
C. Mechanical Checks 2-2
D.
Warmer/Bassinet Interlock (model 3500 only)
2-3
E.
Warmer/Bassinet Unlock 2-3
F.
Accessory Checks 2-3
G. Wall Mount Checkout 2-4
2.3 Control Unit Checkout Procedure 2-4
A. Control Unit Check 2-5
B.
Elapsed Timer Check 2-6
C.
Observation Light Check 2-6
D.
Raise and Lower Bed Switch Check (Elevating Models
Only) 2-6
E.
Interlock Switch Check 2-7
F.
Power Failure, Memory and Battery Test (all models except the
2001 International) 2-7

waII
mounted units A-13
Pre-Installation Preparation
A- 13
Warmer Installation A-14
Warranty W-l
Installing
CaII
specifications A-12
Thermalink
Option Specifications A-10
Serial data A-10
Nurse
(EMC)
A-6
Mechanical (Without Accessories) A-6
3500 Bassinet A-7
Accessories A-8
Radiant Energy Distribution A-10
Alarms
A-4
Environmental Specifications
A-5
Electromagnetic Compatibility
4KXeaning
and Disinfecting 4-1
4.1
Cleaning 4-l
4.2
Wood Surfaces (model 3500 only)
4-2
4.3
Reusable Skin Temperature Probe
4-2
Appendix A-l
Temperature Conversion Chart A-l
Servo Mode Algorithm A-2
Infant Warmer System Specifications A-2
Electrical A-2
Controller A-3
Call
checkout 3-22
Rotating Drawer Option 3-22
CaII
System Interface
3-21
Nurse
ThermaLink
Options 3-20
Using the Serial Data Interface
3-21
Using the Nurse
Plate Mounting and Releasing Procedure
3-19
C.
ECMO Adapter 3-20
Table of Contents
3.4
Bassinet Operation (model 3500 only) 3-11
3.5
Bed Platform Operation 3-12
3.6
Side Panel Operation 3-13
3.7
X-ray Procedures 3-14
3.8
Oxygen Administration 3-16
3.9
Gas Cylinder Installation and Operation 3-16
3.10
Mounting Accessories 3-18
3.11
3.12
A.
Mounting and Releasing Procedure
3-19
B.
Adapter

11/09/93
V
IWS
6600-0194-000
2001 IWS (international)
3000

11109/93
grain
finish standard.
vi
6600-0194-000
4300,440O
and 5000 have a larger size bed than the model 3300
and are intended for surgical procedures and long term care in the Nursery
and L&D Rooms. Both the 5000 and 4400 have an elevating pedestal to
raise and lower the bed-to-floor height. The model 4400 has a narrower
foot print than the 5000 warmer. The model 4300 is a non-elevating
version of the 4400.
IWS Model Options
Free Standing
Wall mounted,
heater head only
Wall mounted heater
with rails
Wall mounted heater
with rails and bad
Free standing with
integral bed
Elevating pedestal
Standard Heater/Bed
2000 3000 3500
Series Series Series
3000
3500t
3050
Large Heater/Bed
4000 5000
Series Series
4000
3100
3150
2001
3300 4300
4400 5000
t Detachable bassinet and wood
.
The 3050 is a heater head only, the 3100 is a heater and dove tail
rails, and the 3150 includes a heater, rails and integral bed.
The model 3300 has an integral bed for infants and is intended for proce-
dures and long term care in the Nursery and L&D.
The model 3000, 3500 and 4000 are free standing Warmers which can be
used over a variety of infant bassinets in the general nursery, over post-
operative patients, during patient feeding, and in any other application
where controlled radiant heat is required.
The 3500 System also has a detachable Bassinet for infants, and is in-
tended for procedures and long term care in the Nursery, L&D, LDR and
LDRP Rooms.
The model
Model Descriptions
All models of the Ohmeda Infant Warmer System provide a controlled
source of radiant heat for infants and pediatric patients. The control sys-
tem uses a microprocessor and provides both manual and servo modes of
operation (except for the 2001 International, which is manual mode only).
The model 2001 International has an integral infant bed and is a manual
mode only warmer designed for short term attended care in the OR and
L&D.
The wall mounted 3050, 3100 and 3150 models are available to accommo-
date specific architectural requirements in the NICU, L&D, LDR and LDRP
room

11109/93
ix
AWarnings
Operator Safety
Before using the Ohmeda Infant Warmer System, read this entire
manual. Attempting to use this device without a thorough understand-
ing of its operation may result in patient or user injury. This device
should only be operated by personnel trained in its operation and under
the direction of qualified medical personnel familiar with the risks and
benefits of this type of device.
Overloading the shelves can affect the stability of the unit. Limit the load
to 20 lbs. (9 kg) per instrument shelf, mounted to a single upright, and
50 lbs. (23 kg) per monitor shelf, mounted between the uprights.
Limit the load of accessories to 50 pounds (23 kg) per side on the Warmer
to ensure stability. Accessories should not be mounted more than
56 inches (142 cm) above the floor. For models 3000 and 3500, limit the
load of accessories to 20 pounds (9 kg) maximum per side mounted no
more than 44 inches (112 cm) above the floor.
Limit the load placed on the x-ray cassette tray to 5 lbs (2.2 kg) to avoid a
tipping hazard.
Overloading the drawers can affect the stability of the unit. Limit the load
to 10 lbs. (4.6 kg) per drawer.
Do not use the Warmer in the presence of flammable anesthetics; a
possible explosion hazard exists under these conditions.
Do not touch the protective grill under the radiant heater or the top of the
heater assembly. These surfaces may be hot and a burn could result.
Due to the increased height of units with the ECMO option installed, a
tipping hazard may exist if tip loading occurs. Limit the total accessory
load to 50 lbs. (23 kg), no more than 25 lbs (11 kg) per side.
Use caution when rotating the cabinet to avoid damage to the drawers or
possible injury. Always ensure the drawers are fully closed before rotating
the cabinet.
Disconnect power to the Warmer and allow the heater rod to cool before
cleaning to avoid the possibility of a burn.
Never oil or grease oxygen equipment. Oils and grease oxidize readily,
and in the presence of oxygen, will burn violently. Vat-Kote is the only
oxygen service lubricant recommended (Ohmeda Stock No. 6700-0092-
200) if the use of a lubricant is specified.
Disconnect the Warmer power cord and allow the unit to cool before
replacing the alarm or observation lights.
On elevating models, check for proper clearance above the Warmer and
below the bed surface before raising or lowering the bed.
Enclosing the heater assembly inside a cabinet may prevent proper
ventilation and may create a fire hazard. If the heater is enclosed in a
cabinet, it must be equipped with a power cut-off device that prevents
operation while the cabinet is closed.
6600-0194-000

*1109/936600-0194.000
&
5 cm)will result in incorrect operation and may affect the patient’s
condition.
Do not place any accessories or any other objects directly over the bed
surface. This may block radiant heat and lead to cooling of the infant.
Do not place items on top of the heater assembly. Items placed on top of
the heater assembly can fall and injure the patient, prevent adequate
ventilation of the heater assembly, and may pose a fire hazard.
Do not perform the Checkout Procedures (Mechanical and Control Unit)
while a patient occupies the Warmer.
Complete the “Checkout Procedures” section of this manual before
putting the unit into operation. If the warmer fails any portion of the
checkout procedures it must be removed from use and repaired.
Regularly inspect the bed side panel latching mechanism, and the bed-
side locking mechanism on the model 3500, to ensure proper operation.
In the service position the strength of the wall mounted unit hinge
bracket is reduced. Never place a patient in the bed when the heater is in
the service position. Never leave the unit unattended in the service
position or with a pin removed.
Check for proper clearance above the Warmer and below the bed surface
before raising or lowering the bed.
Inspect all patient connected tubes or wires before and after moving or
tilting the bed. Tilting or moving the warmer bed up or down can pull on
tubing or leads connected to the patient. This may disconnect tubes or
leads, restrict gas or liquid flow, or move sensors out of position.
Prolonged exposure to the light emitted by the observation lamp in this
unit may harm the unprotected eyes of the infant. For safety, cover the
infant’s eyes.
Do not use the Warmer system if the system failure alarm is activated.
Remove the unit from service and refer to qualified personnel for repair.
Radiant energy can adversely affect blood components. When using
intravenous tubing systems for delivery of blood components to patients
occupying a warmer, shield any tubing with aluminum foil.
When using a radiant warmer, change the patient’s diapers frequently.
Radiant energy causes more rapid urine evaporation, and may lead to
inaccurate urine diagnostic test/ analysis and inaccurate weight mea-
surements.
X
f
2 inches
(69
Precautions
Keep hands clear of the hinge area when installing a wall mounted
heater assembly. A pinch hazard exists.
The 3050 heater assembly weighs approximately 30 lbs (14 kg). Proper
installation may require two people. Due to the weight of the 3100 and
3150 units, approximately 75 lbs. (34 kg), proper installation will require
two people.
Patient Safety
Bed-to-heater spacing which differs from the specified 27

11/09/93
xi
50/pkg)
6600-0194-000
photo-
therapy lamp usage, etc.) or the patient condition requiring heater
adjustments in response to these changes. In the servo mode, the infant
warmer automatically adjusts heater output to maintain the desired skin
temperature, reducing (but not eliminating) the need to monitor the
patient and make adjustments to the equipment.
Use of electrosurgical units or other electrical field radiating equipment
can affect the operation of the warmer. Keep the patient probe lead as far
away as possible from electrosurgical cables. Do not allow excess electri-
cal cables to be laid on the bed platform. Use of electrosurgical units or
other instruments which radiate electrical fields can cause indirect
heating, by several tenths of a degree of the skin temperature probe due
to absorbed electrical energy. When using these devices near the radiant
warmer, operate the Infant Warmer in manual mode for maximum safety.
The use of phototherapy equipment may raise the patient’s temperature.
Radiant warmers increase an infant’s insensible water loss. Take appro-
priate measures to maintain the patient’s fluid balance while caring for
them in a radiant warmer.
Use only the Reusable Ohmeda skin temperature probe (Stock No. 0208-
0697-700) and heat reflective patches (Stock No. 0203-1980-300,
responsibihty
for detecting changes in the environment (drafts, direct sunlight,
Precautions
Do not install chest drainage tubes in the Tubing Organizer Rear Side
Panel.
Install tubing in the appropriately sized holes. Use of inappropriate holes
may cause kinking, pinching or restriction of flow through the tubes, and
may interfere with the proper operation of therapy equipment.
Do not lower the Tubing Organizer Rear Side Panel with tubing attached.
Lowering can pull on the tubing, causing the tubes to dislodge from the
patient.
Do not move the warmer by pushing or pulling on the bed side panels.
This action may lead to the deterioration and breakage of the compo-
nents which form a safety barrier around the infant.
Ensure that the bedside panels are locked in position when a patient
occupies the bed. Blankets or other foreign objects may prevent the
latches from fully engaging.
Do not leave the patient unattended when the side panels are lowered.
Never place an infant on the X-ray cassette tray.
Do not place any foreign objects on the warmer bed or in the under bed
cavity while performing X-ray procedures. Incompatible materials in the
path of the X-ray may adversely affect the quality of the X-ray image. Use
of mattress or bedding materials other than those supplied by Ohmeda
should be evaluated by a Neonatologist or Radiologist.
Do not leave the patient unattended when using the warmer. Check the
patient’s temperature regularly to ensure the comfort and the safety of
the patient. If the warmer is used for an extended time, it is recom-
mended that the servo mode of operation be used. When an alarm is
silenced, close monitoring of the patient’s condition is required.
Use the servo mode unless the manual mode is specifically prescribed.
While both modes require patient monitoring, the manual mode requires
constant attention. In the manual mode, you must take the

11109/93
6600-0194-000
observa-
Xii
time will allow simultaneous depletion of both
cylinders with no reserve oxygen supply available.
Use only one cylinder gasket per yoke. Use of more than one gasket could
cause leakage of the cylinder gas
Do not leave gas cylinder valves open if the pipeline supply is in use.
Pressures from both oxygen cylinders may become equal, and if simulta-
neously used, cylinder supplies may be depleted, leaving no reserve
supply in case of pipeline failure.
The patient probe is not isolated from earth ground. Any additional
equipment used with the Ohmeda Infant Warmer System must comply
with UL 544, CSA 22.2, IEC 601 and VDE 750.
The computer or RS-232 monitor’s user program must continuously check
the data link. The program should constantly verify connection to the
warmer controller and check for updated data.
Remote monitoring does not replace the need for direct patient
retrolen-
tal fibroplasia (retinopathy or prematurity). It is probable that even
concentrations of 40% or less oxygen (formerly considered safe) could be
dangerous to some infants. Therefore, arterial blood gas measurements
are extremely important for regulation of the concentration of inspired
oxygen when an oxygen-enriched environment is considered necessary.
(See current edition of “Standards and Recommendations for Hospital
Care of Newborn Infants” prepared by the Committee of Fetus and
Newborn of the Academy of Pediatrics.)
Do not leave both cylinder valves open. Leaving both oxygen cylinder
valves open at the same
50/pkg)
to monitor the patient’s skin temperature. Use of
other manufacturer’s probes may affect the accuracy of warmer opera-
tion and the electrical safety of the patient.
The skin temperature probe should be located on the patient’s skin in an
area which is directly in the path of the radiant heat. It should not be
attached to an area which is shielded from the radiant heat or between
the patient and the mattress. Large temperature gradients and very long
servo response times will result from improper probe placement.
Rectal temperatures must never be used to servo control a patient’s
temperature.
Intimate contact between the skin temperature probe tip and the
patient’s skin must be maintained for accurate skin temperature mea-
surement. Underheating or overheating may result from poor contact
between the skin temperature probe and the patient. Verify that the skin
temperature probe is securely attached to the patient at least once every
half hour.
In the servo mode, verify that the patient temperature probe is securely
attached to the patient at least once every half hour. A dislodged probe
may not trigger an alarm. If the probe becomes dislodged, the warmer
can over or under heat the infant.
Oxygen concentrations higher than 40% can increase the risk of
lO/pkg;
Stock No.
6600-0196-700.
Precautions
or the disposable probe (Stock No. 6600-0208-700,

Xlll
11/09/93
-
excessive solution causes damage to internal
components.
Use of cleaning/disinfecting solutions containing chemicals that are not
listed in the table on page 4-2 (i.e. alcohol, acetone, etc.), or chemicals in
greater concentrations than those listed, may damage the patient probe
or other material being cleaned.
Do not autoclave or gas sterilize the skin temperature probe. Do not
immerse the probe in liquid cleaner. Avoid placing excessive strain on
the probe lead. Always remove the probe by grasping the plug at the
panel. Do not pull on the probe lead. These precautions will help avoid
damage to the probe.
Do not autoclave or gas sterilize the mattress.
Only competent individuals trained in the repair of this equipment should
attempt to service it as detailed in the service manual. The Service
Manual provides detailed information solely for use by individuals having
proper knowledge, tools and test equipment, and for service representa-
tives trained by Ohmeda.
. .
6600-0194-000
ACautions
To prevent the drawers from opening unintentionally while moving the
3500 Bassinet, move it from the front only.
On elevating models, do not continue to run the motor at the upper and
lower limit positions; equipment damage may result.
Yoke check valves are not intended to provide a leak-free seal; always
use a yoke plug and a fresh cylinder gasket to seal an unused cylinder
port.
Open cylinder valves S-L-O-W-L-Y to avoid damaging the regulators.
Use cleaning solution sparingly on a cloth when cleaning the Warmer. Do
not saturate the unit
Precautions
tion by qualified medical personnel.
If you connect the Nurse Call output to a system which uses a normally
open connection, a disconnected Nurse Call cable will not trigger an
alarm.
When installing a wall mounted unit, to achieve adequate structural
strength all four holes of each mounting bracket must engage tracks.
Should the hole saw miss a track do not proceed. Consult the project
engineer for further direction.

l-l
11/09/93
3000,3100,3500, and 4000 the proper
bed-to-heater spacing is indicated by a label located on the right alumi-
num upright (as viewed from the front).
6600-0194-000
(69+5
cm) will result in incorrect operation and may affect the
patient’s condition.
For your convenience, on Model
f
2 inches
A
!
WARNING: Bed-to-heater spacing which differs from the specified 27
(69+5
cm) beneath the heater assembly.
f
2 inches
RS-
232 serial data is provided for use with customized research software or
compatible vital signs monitors. The Nurse Call connection lets you
integrate the Warmer into your current or future remote alarm systems.
1.2 Support Structure
The support structure provides excellent stability for the radiant heater
and optional accessories. On free standing units, the support structure
consists of the base assembly and the uprights with an integral rail
system. The rails provide a means for mounting accessories and ancillary
equipment to the unit. Refer to Section 1.7 for a description of the acces-
sory items.
1.3 Heater Assembly
The warmer is designed to warm infants on a bed surface located 27
inches
ThermaLink feature simplifies charting and monitoring.
In this section
1.1
1.2
1.3
1.4
1.5
1.6
1.7
General l-l
Support Structure
Heater Assembly
Control Unit l-2
Bed Platform l-2
l-l
l-l
Bassinet (3500 Series) l-2
Accessories l-3
1.1 General
All models of the Ohmeda Infant Warmer System provide a controlled
source of radiant heat for infants and pediatric patients. The control
system uses a microprocessor and provides both manual and servo
modes of operation (model 2001 International is manual mode only). The
patient temperature, control temperature, and elapsed time displays are
digital for ease of viewing. The control panel includes a one hour elapsed
timer with time displayed in minutes and seconds. The timer features
optional audible tones for use during Apgar scoring. A complete audio
and visual alarm system is included on the control panel.
The optional

11/09/93
tiltable bed provides a smooth, dampened motion to
avoid disturbing the patient. * A chest drainage hanger attached to the
bed and rear cross bar extrusion provides a convenient location to hang
chest drainage devices. *
1.6 Bassinet (3500 Series)
The bassinet unit can be detached and rolled away from the model 3500
warmer unit by pulling up on the locking lever knob located at the top
rear of the bassinet (see Figure l-l). The bassinet includes:
1.
Transparent locking side panels which fold down for easy patient
access and are removable for cleaning.
l
Not available on the model 3500 wanner
l-2
6600-0194-000
tiltable
bed platform allows Trendelenburg and Fowler positioning. The hydrau-
lic system for the
l/Description
1.4
The heater assembly consists of a radiant heater, parabolic reflector,
observation light, and a visual alarm light. The parabolic reflector focuses
radiant energy on the bed surface, minimizing energy loss due to scatter-
ing and providing an even field of radiant heat over the bed surface. The
observation light provides intense light for procedures. The alarm light is
located on the end of the heater assembly for ease of viewing. The entire
heater assembly rotates to the side for X-ray procedures and for observa-
tion lamp replacement.
Control Unit
The control unit contains the electronic circuits and controls used to
operate the radiant heater and the observation light. The control unit
performs regular self checks during its operation including failure diag-
nostics.
Either manual or servo mode of operation may be selected. In the manual
mode of operation, select the level of radiant heat output as indicated by
the % power display on the control panel. The control circuit then main-
tains the selected level of radiant heat. The manual mode has a preheat
setting which allows the Warmer to be preheated. In the servo mode of
operation, select the patient’s control temperature. A skin temperature
probe is used to monitor the patient skin temperature. The control
system modulates the radiant heat to maintain the patient at the se-
lected control temperature. The patient’s skin temperature is continu-
ously displayed.
Alarms activate to alert the operator of a low or high patient tempera-
ture, a skin temperature probe failure, a power failure, equipment failure
or a check patient prompt.
1.5 Bed Platform
Warmers with beds include a mattress and transparent side panels. The
side panels fold down for easy access to the patient and can be removed
for cleaning. The X-ray cassette tray (located in an opening beneath the
bed when installed) pulls out for insertion of X-ray cassettes and allows
X-rays to be taken without moving the patient. * Markings located along
the side panels allow easy location of the cassette in the X-ray tray
relative to the position of the patient on the warmer bed.*
The

11/09/93
l-3
.
4.
Three drawers which open from the front (or from the side in an
optional configuration) for storing equipment or supplies. The top
drawer is provided with a key lock.
5. Front locking casters.
Figure l-l
Operating the
bassinet
locking lever knob
1.7 Accessories
The rail mounting system is the basis for most of the accessories for the
Warmer. This patented design consists of a dovetail shaped aluminum
extrusion and a positive locking mounting block. Mounting blocks are
attached to various accessories for mounting to the rail system. The
mounting blocks are produced in two standard lengths and are machined
to accommodate each specific accessory.
Accessories which mount on a single upright are interchangeable be-
tween all model Warmers.
Accessories can only be mounted to the inside dovetail rail on each
upright of the 3500 Warmer.
6600-0194-000
l/Description
2. A mattress.
3.
A tiltable, positive-lock bed platform which allows Trendelenburg
and Fowler positioning at 4” and 8“ tilt positions

11/09/93
2-l
!
Mechanical Checkout Procedure
WARNING: Before using the Ohmeda Infant Warmer System, read this
entire manual. Attempting to use this device without a thorough under-
standing of its operation may result in patient or user injury.
WARNING: Do not perform the Checkout Procedures (Mechanical and
Control Unit) while a patient occupies the warmer.
WARNING: Complete the “Checkout Procedures” section of this manual
before putting the unit into operation. If the warmer fails any portion of
the checkout procedures it must be removed from use and repaired.
A.
1.
2.
Overall Appearance
Disconnect the power cord from the ac power source for the mechani-
cal checks portion of this procedure.
Check the overall appearance of the Infant Warmer/ Bassinet System.
There should be no obvious damage.
6600-0194-000
A
!
A
!
In this section
2.1 Setup 2-l
2.2 Mechanical Checkout Procedure 2-l
A.
Overall Appearance 2-l
B.
Heater Assembly Rotation 2-2
C. Mechanical Checks 2-2
D.
Warmer/Bassinet Interlock (model 3500 only)
2-3
E.
Warmer/Bassinet Unlock (model 3500 only)
2-3
F.
Accessory Checks 2-3
G. Wall Mount Checkout 2-4
2.3 Control Unit Checkout Procedure 2-4
A.
Control Unit Check 2-5
B.
Elapsed Timer Check 2-6
C.
Observation Light Check 2-6
D.
Raise and Lower Bed Switch Check (Elevating Models
Only) 2-6
E.
Interlock Switch Check 2-7
F.
Power Failure, Memory and Battery Test (all models except the
2001 International) 2-7
2.1 Setup
Refer to the setup instructions shipped with the warmer for initial unpack-
ing and setup of the unit after shipment. For model 3050, 3100 and 3150
wall mounted units, refer to the mounting instructions in the Appendix.
After removal from the shipping containers, inspect the Ohmeda Infant
Warmer System and all accessory items for any signs of damage which
may have occurred during shipment. File a damage claim with the ship-
ping carrier if damage has occurred. Also confirm the presence of all
accessory items or factory installed options as listed on the packing slip.
2.2
A

11/09/93
detented positions.
WARNING Overloading the drawers can affect the stability of the
unit. Limit the load to 10 lbs. (4.6 kg) per drawer.
4.
For the model 3500 bassinet, remove the keys from the top drawer and
lock it. Check that the drawer is securely held closed.
2-2
6600-0194-000
A
!
3.
4.
For the model 3500, separate the warmer from the bassinet.
For units with casters, check that all casters are in firm contact with
the floor and that the warmer is stable and moves freely.
For the model 3500, check that all six casters on the warmer move freely.
Note: On the model 3500, it is possible that the two center casters may
not be in contact with the floor at all times.
5.
Lock the two front casters and check that the warmer is held in place.
On the model 3500, lock the two rear casters and check that the
warmer is held in place.
6.
For the model 3500, place the bassinet on a level surface. Check that
all four casters are in firm contact with the floor and that the Bassinet
moves freely.
7.
For the model 3500, lock the two front casters and check that the
Bassinet is held in place.
8.
Examine the power cord for damage. Replace the power cord if dam-
age is evident.
9.
Examine the unit for objects placed on top of the heater assembly.
WARNING: Do not place any accessories or other objects directly
over the bed surface. This may block radiant heat energy and lead to
cooling of the infant.
WARNING: Do not place items on top of the heater assembly. Items
placed on top of the heater assembly can fall and injure the patient,
prevent adequate ventilation of the heater assembly, and may pose
a fire hazard.
B. Heater Assembly Rotation
Rotate the Heater Assembly to the side and then back to the normal
position. Check for a smooth rotation.
C. Mechanical Checks
1.
Check the operation of the bed sides. The bed sides should operate
smoothly.
WARNING: Regularly inspect the bed side panel latching mecha-
nism, and the bedside locking mechanism on the model 3500, to
ensure proper operation.
2.
Check the operation of the tilt mechanism. Verify that the bed plat-
form operates smoothly and locks in normal, Trendelenburg and Fowler
positions.
3.
If the drawer package is installed, check that all the drawers open and
close freely. On units with the rotating drawer package, make sure the
drawers rotate from front to 90” left and right and firmly stop in the
A
!
A
!
A
!
2/Setup
and Checkout Procedures

11/09/93
2-3
the
load of accessories to 50 pounds (23 kg) per
side
on the Warmer to ensure stability. Accessories should not be
mounted more than 56 inches (142 cm) above the floor. For models
3000 and 3500, limit the load of accessories to 20 pounds (9 kg)
maximum per side mounted no more than 44 inches (112 cm) above
the floor.
WARNING: Due to the increased height of units with the ECMO
option installed, a tipping hazard may exist. Limit the total acces-
sory load to 50 lbs. (23 kg), no more than 25 lbs. (11 kg) per side.
6600-0194-000
.
CAUTION: To prevent the drawers from opening unintentionally
while moving the detached bassinet, move it from the front only.
Accessory Checks
Perform these checks if they are applicable.
1.
Check that all accessories are mounted securely and that the load
limits are not exceeded.
2.
Check that all gas accessories are installed and operating properly
(refer to Section 3.9).
3.
Where applicable, perform the checkout procedures detailed in the
Operation and Maintenance Manuals for the accessories.
WARNING: Limit
.
With the locking pin handle (located at the rear of the bassinet)
raised, pull the bassinet forward until it is fully detached from the
warmer
A
!
1.
2.
3.
4.
5.
6.
E.
1.
2.
F.
Lock the 3500 warmer’s two rear casters.
Align the rear of the bassinet with the warmer’s base.
Push the bassinet into the alignment/locking track of the warmer’s
base.
Continue pushing the bassinet until its locking pin drops into the
socket at the end of the alignment track. There should be an audible
click when the pin drops into position.
Lightly push and pull the bassinet to verify that the two units are
securely held together.
Unlock the two rear casters on the warmer and check that the com-
bined unit moves smoothly.
Warmer/Bassinet Unlock
(model 3500 only)
Verify that the two bassinet front casters are unlocked and the two
warmer rear casters are locked
A
!
A
!
2Betup
and Checkout Procedures
D. Warmer/Bassinet Interlock
(model 3500 only)

11/09/93
6600-0194-000
A
!
WARNING: Complete the “Checkout Procedures” section of this manual
before putting the unit into operation. If the warmer fails any portion of
the checkout procedures it must be removed from use and repaired.
2-4
A
!
WARNING: Do not perform the Checkout Procedures (Mechanical and
Control Unit) while a patient occupies the warmer.
raise/lower
are only present on elevating models.
alarm
are not present on the model 2001
international control panel. Bed
*
Note: Servo mode indicator, mode switch, and probe failure
*
Indicator
Elapsed
Time
Start/Hold
and Reset
I
Bed
Controls
Light
Alarm Alarm Mode
Increase
Apgar
On/Off Lights* Silence Switch* Decrease
Tones
Switch On/Off
Mode
Apgar
Indicators
I
I
II
A
!
1.
Verify that the warmer is rigidly secured to the wall and the heater
assembly is level.
Note: To
access the control unit and display module for service
procedures, either hinge pin may be removed, allowing the warmer
to pivot away from the wall.
2.
Check that both hinge pins are in place and fully inserted with the
pin heads at the top of the hinge.
WARNING: In the service position the strength of the hinge
bracket is reduced. Never place a patient in the bed when the
heater is in the service position. Never leave the unit unattended
in the service position
or with either pin removed.
2.3 Control Unit Checkout Procedure
Figure 2-2
Control Panel
Model Patient Control Heater
Elapsed
Number Temperature Temperature Power % Time Display
2/Setun
and Checkout Procedures
G. Wall Mount Checkout
Figure 2-1
Wall mount pin
insertion

11/09/93
2-5
“HH.H”.
6600-0194-000
35.O”C.
7.
Disconnect the skin temperature probe. Verify the following:
a.
The probe failure indicator light is lit.
b.
There is an alternating two tone alarm.
C.
The overhead alarm light is flashing.
d.
The patient temperature display flashes
0.5”C
by a qualified service
person).
6.
Press the decrease touch switch and verify that the minimum servo
control temperature attainable is
37.5”C.
Note:
A patient temperature alarm occurs if the difference between
the patient temperature and the control temperature is greater than
1°C (the difference can be adjusted to
36.5”C.
5.
Press the increase (
A
) touch switch and verify that the maximum
servo control temperature attainable is
“LL.L”
may occur here
if the skin temperature probe is below 30°C. Warm the probe with
your fingers or silence the alarm.
a.
The servo mode indicator is lit.
b.
An operator prompt tone sounds and the control temperature
display flashes
(v)
touch
switches to attain the high and low limits as indicated by the %
power display.
Note: Steps 3 through 8 do not apply to the model 2001 International.
3.
Connect the skin temperature probe to the warmer.
4.
Press the mode touch switch to place the warmer in the servo mode
and verify the following:
Note:
An alternating two tone alarm, a flashing overhead alarm light
and the patient temperature display flashing
2/Setup
and Checkout Procedures
A. Control Unit Check
1.
Connect the warmer power cord to an appropriate power source.
Refer to the rating plate on the Warmer for the proper voltage
needed. Switch the power On and verify the following on the Control
Panel (Figure 2-l):
a.
The alternating two tone audible alarm sounds and all displays
and indicators are lit for approximately two seconds.
Note:
During this time the controller also performs self check
functions. If the controller detects a failure, the alarm stays on
and service is required.
Note:
All alarms except system failure or power failure are pre-
ceded by a 30 second intermittent operator prompt tone.
b.
The manual mode indicator is lit.
c.
Operator prompt tones sound and the % power display flashes.
2.
Adjust the heat output with the increase (A) and decrease

11109/93
l/2
inches (98 cm) off the floor.
2-6
6600-0194-000
l/2
inches (118 cm) off the floor.
2.
Press the Lower Bed touch switch and verify that the bed lowers to a
minimum of 38
A
!
CAUTION:
Do not continue to run the motor at the upper and lower
limit positions; equipment damage may result.
1.
Press the Raise Bed touch switch and verify that the bed raises to a
maximum of 46
“0O:OO”.
If the elapsed timer is not used for approximately two
minutes, the display switches off.
C. Observation Light Check
Press the Light On/Off touch switch. Verify that the observation light
functions.
D. Raise and Lower Bed Switch Check
(Elevating Models Only)
“HH.H”.
9.
Switch to the manual mode and set the heat at 25% power.
B.
Elapsed Timer Check
1.
Press the Start/Hold touch switch to activate the elapsed timer.
Verify that the timer starts operating.
2.
3.
Press the Apgar tones On/Off touch switch for the Apgar tones.
Verify that the indicator light for the Apgar tones is extinguished.
Press the Apgar tones On/Off switch for the Apgar tones again.
Verify that the indicator light for the Apgar tones is lit.
4.
Press the Start/Hold touch switch. Verify that the present elapsed
time is held.
5.
Press the Start/Hold touch switch and verify that the timer updates
to the current elapsed time and the Apgar tones continue to sound at
the specified times (at 1 minute and at every 5 minute interval after
the elapsed timer has started).
6.
Press the Reset touch switch and verify that the timer indicates
“HH.H”.
e.
After one minute, the alternating two tone alarm sounds, the
overhead alarm flashes and the patient temperature display
flashes
2/Setup
and Checkout Procedures
8.
Press the alarm silence touch switch and verify the following:
a.
The probe failure indicator light is lit.
b.
The alternating two tone alarm is silenced.
c.
The overhead alarm light is lit.
d.
The patient temperature display indicates
This manual suits for next models
15
Table of contents
Other Ohmeda Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Ametek
Ametek Reichert ClearChart2 user guide

bort medical
bort medical MobiDig Extension quick guide

DENTSPLY
DENTSPLY inLab MC X5 operating instructions

Donjoy
Donjoy Shoulder Cradle Application guide

Mangar International
Mangar International Leglifter User instructions

Otto Bock
Otto Bock SiOCX TF 7T450-1 Instructions for use