15
SZX10
Problem Cause Remedy Page
1. The left and right fields of view
do not coincide.
The interpupillary distance is
adjusted improperly.
Adjust it correctly. 10
The diopter is adjusted incorrectly. Adjust it correctly. 10
The left and right eyepieces are
different.
Use a pair of matched eyepieces. 22
2. The field of view is obscured
partially or illuminated unevenly.
The aperture iris diaphragm is
stopped down excessively.
Open the aperture iris diaphragm. 8
The trinocular observation tube
and/or intermediate attachments
are installed improperly.
Mount them properly.
21
The light path selector knob is in
an intermediate position.
Set it correctly to the desired
position. 13
3. The monitor image is cut off
partially.
The light path selector knob is
pulled out incompletely.
Pull it out all the way. 13
4. Dirt or dust is visible in the field
of view.
Dirt/dust on the specimen. Clean thoroughly. 3
Dirt/dust on eyepiece. Clean thoroughly. 3
5. Details of observed image are
solid.
The aperture iris diaphragm is
stopped down excessively.
Open the aperture to a proper
diameter. 8
6. Visibility of observed image is
poor.
· Image is not sharp.
· Contrast is insufficient.
The objective is tilted. Screw it correctly until it stops. 21
The objective is dirty. Clean thoroughly.
3
The top and/or bottom lenses of
the microscope body are dirty
7. Zooming causes defocusing of
the observed image.
The eyepiece diopter is adjusted
improperly.
Adjust it correctly. 10
The focus adjustment is in
accurate.
Adjust focus at a high
magnification. 10
8. The coarse focus adjustment
knob does not rotate smoothly
The rotation tension of the knob is
set too high.
Decrease the rotation tension to
an optimum level. 7
9. The microscope body drops
spontaneously, causing the
focusing to be deviated during
observation.
The rotation tension of the coarse
focus adjustment knob is set too
low.
Increase the rotation tension to an
optimum level. 7
The microscope body dropped
spontaneously because the weight
exceeded 10 kg.
Use a focusing assembly
withstanding heavy loads. (SZX2-
FOF: 2.7 to 15 kg. SZX2-FOFH: 8
to 25 kg.)
16
5Troubleshooting Guide
Under certain conditions, performance of this unit may be adversely affected by factors other than defects. If a problem
occurs, please review the following list and take remedial action as needed. If you cannot solve the problem after
checking the entire list, contact Olympus.