Olympus U-AW User manual

AX8247
INSTRUCTIONS
U-AW
This instruction manual is for the Olympus Motorized Attenuator Wheel Model U-AW.
To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this
motorized attenuator wheel, we recommend that you study this manual thoroughly before operating
the microscope.
Retain this instruction manual in an easily accessible place near the work desk for future reference.
MOTORIZED ATTENUATOR WHEEL
Optical Microscope Accessory

In accordance with European Directive on Waste Electrical and Electronic Equipment, this
symbol indicates that the product must not be disposed of as unsorted municipal waste, but
should be collected separately.
Refer to your local Olympus distributor in EU for return and/or collection systems available in
your country.
This product is a part of the system product which complies with the CE marking.
Please refer to the instruction manual of your system product for the safety instructions related to the CE
marking.
For Korea only
A급 기기 (업무용 방송통신기자재)
이 기기는 업무용(A급) 전자파적합기기로서 판 매자 또는 사용자는 이 점을
주의하시기 바라 며, 가정외의 지역에서 사용하는 것을 목적으로 합니다.

CONTENTS
IMPORTANT -- Be sure to read this section for safe use of the equipment. -- 1-2
SAFETY PRECAUTIONS ....................................................................................................................................1
1Handling Precautions..........................................................................................................................................................................2
2Maintenance and Storage..............................................................................................................................................................2
1 SYSTEM CHART ...............................................................................................................................3
2 ASSEMBLY.............................................................................................................................................6
1Installation in a Reflected Light Illumination System..............................................................................................6
2Installation in a Transmitted Light Illumination System.........................................................................................8
3 OPERATION..........................................................................................................................................10


1
U-AW
IMPORTANT
If the device is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the
device may also be damaged. Always use the device as outlined in this instruction manual.
The following symbols are used to set off text in this instruction manual.
: Indicates a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury or damage to the equipment or other property. It may also be used to
alert against unsafe practices.
} : Indicates commentary (for ease of operation and maintenance).
CAUTION
The motorized attenuator wheel can be installed in either a reflected light or
transmitted light illumination system. It can adjust the illumination brightness in six
steps (transmittances: 100%, 25%, 12.5%, 6%, 1.5% and 0%) with motorized operation.
SAFETY PRECAUTIONS
1. Before connecting the motorized attenuator wheel, be sure to set the main switch of the control box (BX3-CBH,
U-CBM, IX3-CBH, IX3-CBM) to “\” (OFF) or unplug its power cord.
2. Be careful not to insert your finger or let a foreign object enter inside the area in which the motorized attenuator
wheel is installed.
3. If this product is used in combination with laser products classified above CLASS 1, this instruction manual is not
effective. Follow the instruction manuals of the laser products.
Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equipment
in the safest possible manner.
Symbol Explanation
Indicates that the surface becomes hot, and should not be touched with bare hands.
Caution labels
Caution labels are placed at parts where special precaution is required when handling and using the system. Always
pay attentions to the caution labels.
Positions of caution
labels
· Motorized attenuator wheel
(U-AW)
If the caution label becomes dirty or is peeled off, contact Olympus for replacement.
[Caution against high temperature]

2
1. This device is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe impact.
2. Do not disassemble this system. Doing so could cause failure of the system.
3. Do not use the system where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations.
4. When this instrument is used, the image peripheries may be obscured when a high-power objective is used in
superwidefield observation (FN 26.5).
1Handling Precautions
1. Do not use organic solvents, which cause painted and plastic parts to deteriorate. Do not use organic solvents
to clean device components other than the glass components. To clean them, use a lint-free, soft cloth slightly
moistened with a diluted neutral detergent.
2. Before disposing of this product, be sure to follow the regulations and rules of your local government.
2Maintenance and Storage

3
U-AW
*1 The interface cable should be U-IFCBL15 (15 cm) or U-IFCBL200 (2 m) depending on the connected module.
*2 If the cable of the U-LH100-3 is too short, use the optional U-RMT extension cord.
If the interface cable contacts the lamp housing or its surroundings, the cable may melt and cause an
electric shock hazard. Distribute cables at enough distances from the lamp housing.
CAUTION
1SYSTEM CHART
BX63
BX53
BX63
BX53
Extension adapter
for lamp housing
U-LHEAD
Extension adapter
for lamp housing
U-LHEAD
Motorized
attenuator wheel
U-AW
Motorized
attenuator wheel
U-AW
Motorized
attenuator wheel
U-AW
LH100 adapter
U-LH100ADP
Control box
BX3-CBH
Control box
U-CBM
Interface cable *1
Interface cable *1
Transmitted light
lamp housing
U-LH100-3 *2,
U-LH100L-3, etc
Transmitted light
lamp housing
U-LH100-3 *2,
U-LH100L-3, etc
Reflected light
lamp housing
U-LH100HG,
U-LH100HGAPO,
U-LH75XEAPO
Reflected light
lamp housing
U-LH100HG,
U-LH100HGAPO,
U-LH75XEAPO

4
*1 The interface cable should be U-IFCBL100 (1 m).
*2 If the cable of the U-LH100-3 is too short, use the optional U-RMT extension cord.
If the interface cable contacts the lamp housing or its surroundings, the cable may melt and cause an
electric shock hazard. Distribute cables at enough distances from the lamp housing.
CAUTION
IX83
IX83
Extension adapter
for lamp housing
U-LHEAD
Extension adapter
for lamp housing
U-LHEAD
Motorized
attenuator wheel
U-AW
Plate
(Accessory of U-AW )
Motorized
attenuator wheel
U-AW
Control box
IX3-CBH
Interface cable *1
Interface cable *1
Transmitted light
lamp housing
U-LH100-3 *2,
U-LH100L-3, etc
Reflected light
lamp housing
U-LH100HG,
U-LH100HGAPO,
U-LH75XEAPO

5
U-AW
*1 The interface cable should be U-IFCBL200 (2 m).
*2 If the cable of the U-LH100-3 is too short, use the optional U-RMT extension cord.
If the interface cable contacts the lamp housing or its surroundings, the cable may melt and cause an
electric shock hazard. Distribute cables at enough distances from the lamp housing.
CAUTION
IX73
IX73
Extension adapter
for lamp housing
U-LHEAD
Extension adapter
for lamp housing
U-LHEAD
Motorized
attenuator wheel
U-AW
Plate
(Accessory of U-AW )
Motorized
attenuator wheel
U-AW
Control box M
IX3-CBM
Interface cable *1
Transmitted light
lamp housing
U-LH100-3 *2,
U-LH100L-3, etc
Reflected light
lamp housing
U-LH100HG,
U-LH100HGAPO,
U-LH75XEAPO

6
2ASSEMBLY
} The only tool required for the assembly is the Allen screwdriver provided with the microscope frame.
1Installation in a Reflected Light Illumination System
When using BX63/BX53
Loosen the U-LHEAD clamping screws a, and separate the U-LHEAD to 2 units: one for the frame and the other for
the lamp housing. One for the frame displays [FRAME] and the other for the lamp housing displays [ ].
Loosen the lamp housing clamping screws b(2 pcs.) on the fluorescence illuminator sufficiently using the Allen
screwdriver, fit the mount dovetail cof the U-LHEAD (frame side) and tighten the clamping screws b.
Attach the U-AW dto the U-LHEAD (frame side), and tighten the clamping screws a.
Loosen the clamping screws e(4 pcs.) of the U-AW, fit the U-LHEAD (lamp housing side) fand tighten the clamping
screws e(upper 2 pcs. only).
Attach the lamp housing gto the U-LHEAD (lamp housing side) fin the same way as above.
Insert one end of the interface cable into the connector hand insert the other end into the BX63F microscope frame
or control box (see “SYSTEM CHART”).
1
2
3
4
5
6
When using the BX53,
attach the motorized
attenuator wheel in this way
to avoid the thermal effect
from the lamp housing.
CAUTION
bcad
f
g
e
h

7
U-AW
Loosen the U-LHEAD clamping screws a, and separate the U-LHEAD to 2 units: one for the frame and the other for
the lamp housing. One for the frame displays [FRAME] and the other for the lamp housing displays [ ].
Loosen the lamp housing clamping screws b(2 pcs.) on the fluorescence illuminator sufficiently using the Allen
screwdriver, fit the mount dovetail cof the U-LHEAD (frame side) and tighten the clamping screws b.
Attach the U-AW dto the U-LHEAD (frame side), and tighten the clamping screws a.
Loosen the clamping screws e(4 pcs.) of the U-AW, fit the U-LHEAD (lamp housing side) fand tighten the clamping
screws e(upper 2 pcs. only).
Attach the lamp housing gto the U-LHEAD (lamp housing side) fin the same way as above.
Insert one end of the interface cable into the connector hand insert the other end into the control box IX3-CBH or
control box IX3-CBM (see “SYSTEM CHART”).
1
2
3
4
5
6
When using IX83/IX73
bca
d
g
e
h
f

8
2Installation in a Transmitted Light Illumination System
When using BX63
Loosen the U-LH100ADP clamping screws aand b, and separate the U-LH100ADP to 3 units: [Front], [Middle] and
[Rear]. (The [Middle] unit is not used.)
Loosen the lamp housing clamping screw cof the microscope frame using the Allen screwdriver, fit the mount
dovetail dof the U-LH100ADP [Front] and tighten the clamping screw c.
Attach the U-AW eto the U-LH100ADP [Front] and tighten the clamping screws a.
Loosen the clamping screws f(4 pcs.) of the U-AW, fit the U-LH100ADP [Rear] gby aligning the positioning pin,
and tighten the clamping screws f(upper 2 pcs. only).
Attach the lamp housing hto the U-LH100ADP [Rear] gin the same way as above.
Insert one end of the interface cable into the connector iand insert the other end into the BX63F microscope frame
or control box (see “SYSTEM CHART”).
When U-AW is not mounted properly, U-AW may not be recognized in some cases. If it is not recognized,
mount U-AW properly.
1
2
3
4
5
6
CAUTION
This adapter (the part without
lens) is not used.
Positioning pin
c
d
egh
f
i
b
a

9
U-AW
When using IX83/IX73
Loosen the U-LHEAD clamping screws a, and separate the U-LHEAD to 2 units: one for the frame and the other for
the lamp housing. One for the frame displays [FRAME] and the other for the lamp housing displays [ ].
Loosen the lamp housing clamping screw b(4 psc.) of the U-AW and insert the plate provided with U-AW into the
adapter mounting area. Be sure to insert the plate by fitting its position so that the nail cof the plate enters in the
dent dof the microswitch.
Insert the U-LHEAD (lamp housing side) einto the adapter mounting area of the U-AW, and tighten the clamping
screw f(upper 2 pcs. only)
Loosen the illumination column clamping screws g(2 pcs.) using Allen screwdriver, insert the mount dovetail hof
the U-LHEAD (frame side) into the illumination column, and tighten the clamping screws g.
Attach the U-AW ito the U-LHEAD (frame side), and tighten the clamping screws a.
Secure the lamp housing jto the U-LHEAD (lamp housing side) ein the same way as above.
Insert one end of the interface cable into the connector kand insert the other end into IX83P2ZF/IX83P1ZF or the
control box IX3-CBM (see “SYSTEM CHART”).
When U-AW is not mounted properly, U-AW may not be recognized in some cases. If it is not recognized,
mount U-AW properly.
1
2
3
4
5
6
7
CAUTION
2– 3
b
f
c
e
d
g a
h
i
j
k
e

10
3OPERATION
The transmittance of the motorized attenuator wheel can be switched from the following modules.
System Using the Control Box BX3-CBH
· Touch-panel controller (provided with the BX63)
· Controller U-MCZ
System Using the Control Box U-CBM
· Hand switch U-HSCBM for CBM
System Using the Control Box IX3-CBH
· Touch-panel controller (provided with the IX83P1ZF or IX83P2ZF)
· Controller U-MCZ
System Using the Control Box M IX3-CBM
· Hand switch U-HSCBM for CBM
For detailed operating procedures, refer to the instruction manuals of the system modules.


AX8247 05
8F Olympus Tower, 446 Bongeunsa-ro, Gangnam-gu, Seoul, 06153 Korea
Distributed by
Manufactured by
48 Woerd Avenue Waltham, MA 02453, U.S.A.
Shinjuku Monolith, 2-3-1 Nishi-Shinjuku, Shinjuku-ku, Tokyo 163-0914, Japan
www.olympus-global.com
Table of contents
Other Olympus Microscope manuals

Olympus
Olympus BXFM Series User manual

Olympus
Olympus CX21LED User manual

Olympus
Olympus IX81-ZDC2 User manual

Olympus
Olympus SZX7 User manual

Olympus
Olympus IX70 User manual

Olympus
Olympus IX2-DSU User manual

Olympus
Olympus SZ-III User manual

Olympus
Olympus GX41 User manual

Olympus
Olympus BH2-5RE Installation instructions

Olympus
Olympus CH10 Operating instructions

Olympus
Olympus BX46 User manual

Olympus
Olympus bx40 User manual

Olympus
Olympus GX53 User manual

Olympus
Olympus SZX2-LTTR User manual

Olympus
Olympus X-Tr Operating instructions

Olympus
Olympus CX43-RFAB User manual

Olympus
Olympus BX51M User manual

Olympus
Olympus CX41 Installation guide

Olympus
Olympus CX31 User manual

Olympus
Olympus KHC User manual