Omega HH127 User manual

omega.com
e-mail: [email protected]
For latest product manuals:
omegamanual.info
HH127
TC Data Logger Thermometer
Shop online at
User’s Guide
MADE IN TAIWAN

Servicing North America:
U.S.A.: Omega Engineering, Inc., One Omega Drive, P.O. Box 4047
ISO 9001 Certified Stamford, CT 06907-0047 USA
Toll Free: 1-800-826-6342 TEL: (203) 359-1660
Canada: 976 Bergar
Laval (Quebec), Canada H7L 5A1
Toll-Free: 1-800-826-6342 TEL: (514) 856-6928
For immediate technical or application assistance:
U.S.A. and Canada: Sales Service: 1-800-826-6342/1-800-TC-OMEGA®
Customer Service: 1-800-622-2378/1-800-622-BEST®
Engineering Service: 1-800-872-9436/1-800-USA-WHEN®
Mexico: En Español: 001 (203) 359-7803 FAX: (001) 203-359-7807
Servicing Europe:
Benelux: Managed by the United Kingdom Office
Toll-Free: 0800 099 3344 TEL: +31 20 347 21 21
Czech Republic: Frystatska 184
733 01 Karviná, Czech Republic
Toll-Free: 0800-1-66342 TEL: +420-59-6311899
France: Managed by the United Kingdom Office
Toll-Free: 0800 466 342 TEL: +33 (0) 161 37 29 00
Germany/Austria: Daimlerstrasse 26
D-75392 Deckenpfronn, Germany
Toll-Free: 0 800 6397678 TEL: +49 (0) 7059 9398-0
United Kingdom: OMEGA Engineering Ltd.
ISO 9001 Certified One Omega Drive, River Bend Technology Centre, Northbank
Irlam, Manchester M44 5BD England
Toll-Free: 0800-488-488 TEL: +44 (0)161 777-6611
OMEGAnet®Online Service Internet e-mail
omega.com [email protected]
It is the policy of OMEGA Engineering, Inc. to comply with all worldwide safety and EMC/EMI
regulations that apply. OMEGA is constantly pursuing certification of its products to the European New
Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA accepts no liability for any
errors it contains, and reserves the right to alter specifications without notice.
WARNING: These products are not designed for use in, and should not be used for, human applications.

i
Introduction:
HH-127 is a hand held digital thermometer which uses a microprocessor.
The HH-127 has USB and RS232 ports.
Accessories:
DC-1.5V/4 AAA batteries
Rubber boot
K-type thermocouple x 2 pcs
Users manual
PC on-line accessories:
Temp Monitor software users manual
IrDA-USB and IrDA-RE232 transmission cable
Temp Monitor software CD

Warning
To avoid electrical shock or personal injury, follow these gui delines:
Before using the thermometer inspect the case. Do not use the thermometer if it appears damaged.
Look for cracks or missing plastic. Pay particular attention to the insulation around the connectors.
Disconnect the thermocouple from the thermometer before opening the case.
When the battery mark displays ( ), or you hear a short buzzing sound, please replace the battery
immediately.
Do not operate the thermometer around explosive gas, vapor, or dust.
Do not apply more than the rated voltage, as marked on the thermometer, either between the
thermocouple, or between any thermocouple and earth ground.
Caution
A Caution identifies conditions and actions that may damage the meter or the equipment under test.
Use the proper thermocouples, function, and range for your thermometer.
Do not attempt to recharge the batteries.
To prevent explosion, do not throw batteries into a fire.
Follow local laws or regulations when disposing of batteries.
Match the + and - polarities of the battery with the battery case.
1
•
•
•
•
•
•
•
•
•
•

Names of parts:
①Housing ⑥IrDA output port
②LCD display ⑦T1- thermocouple input
③Function control keys ⑧T2- thermocouple input
④Battery seat ⑨Screw hole
⑤Reset key ⑩Battery cover
T2
T1
①
②
③
④
⑤
⑥
⑦ ⑧
⑨
⑩
+
-
+
-
2

Display Elements:
①
Choose T1.T2 thermocouple types ⑫
Carry out record function
②
Main display ⑬
20 minutes auto power off
③
Main, secondary display reading is T1 or T2 ⑭
Examine recorded reading
④
Secondary display ⑮
Five sections of battery power instruction
⑤
Carry out TC- 0°C calibration function ⑯
Temperature unit
⑥
Main, secondary display maximum reading ⑰
Main display shows time's "year"
⑦
Main, secondary display minimum reading ⑱
Third display shows time's "hour:minute", or
"minute:second"
⑧
Main display carry out alarm function ⑲
Third display
⑨
Main display carry out relative reading , third
display shows relative reference reading
⑳
Third display shows “T1-T2” reading
⑩
Carry out set function Secondary display shows time’s “day”, ”month”
⑪
Hold display reading
Battery power indicator and replacement:
The battery power is expressed by five sections of batteries marks. ( )
When the battery mark shows ( ) and you hear a short buzzing sound, this indicates the battery
power is weak, please immediately close the unit and replace with 4 AAA batteries to ensure the
measuring reading is accurate.
When battery power is lower than normal operation, this will automatically stop the thermometer
operation, and the main display will show BATT and the third display will show LO. Please immediately
turn off the unit and replace the 4 AAA batteries.
.
To replace the batteries please use a screwdriver to open battery cover.
C
H
T1-T2
F
d:m m s
:
h m
:
y
①
②
③
④
⑤
⑥ ⑦
⑧
⑨
⑩
⑪
⑫
⑬
⑭
⑮
⑯
⑰
⑱
⑲
⑳
K J T E R S N
3
1
21
•
•
•
•

Keys:
Power key: turn on/off thermometer. When not canceling auto power off function, after 20 minutes the
meter will auto power off. Initial temperature unit
is (°C), when you power on meter next it will keep
the last power off state.
dual inputs: secondary display shows T2
measuring value and third display shows T1-T2
value.
Temperature unit key: Choose temperature unit to be Centigrade (°C) or Fahrenheit (°F).
Maximum/minimum key: Main, secondary display show (T1, T2) maximum/minimum value at same
time, third display shows (T1- T2) real time
readings.
C
F
MAXMIN
C
F
C
F
MAXMIN
MAXMIN
MAXMIN
4

Hold readings key: Hold LCD display readings.
Press once: Cancels hold readings function.
Press for 2 seconds: Cancels hold readings
function and cancels auto power off function.
Relative readings key: Main display shows (T1 or T2 readings - relative readings), third display shows
relative readings.
When not setting up relative readings: When
pushing key the main display shows (T1 or
T2) readings as the relative readings. Third
display is the stored relative value.
When setting up relative readings: Pushing the
key, the third display shows set relative
reading.
Note: Set up operation for relative readings, please
consult the explanation for setting up relative
Alarm function key: Main display shows (T1 or T2) reading. If (T1 or T2) are greater than or
smaller than set up critical high value or low
value of alarm, the buzzer will send out alarm
sound continuously.
Note: Set up operation of alarm limit value,
please consult the explanation for setting up
alarm critical value on page 12.
5
readings on page 11.

Adjustment key of thermocouple 0°C:Main display shows (T1 or T2) readings, use and main display
(T1 or T2) thermocouple of the same type,
imports input of main display 0°C (32°F) of
standard temperature value, adjust the cold
junction compensation of third display, enables
main display to show 0°C (32°F), then press
key finish adjusting. The adjusting range is about
0°C ~50°C (32°F~122°F) of third display cold
junction compensation, exceed range show Err
and unable to adjust, need to press key to
give up and left. The cold junction compensation
adjusting (cancelled) in power off, while starting
the meter again will recover the normal cold
junction compensation.
Press once: Enter adjust manually and third
display shows the cold junction compensation,
enable main display shows 0°C (32°F).
>Press to move right one-figure that adjusted.
>Press to move left one-figure that adjusted.
>Press to add one-figure upward that adjusted.
>Press to minus one downward that adjusted.
Press for 2 seconds: Adjust once automatically the
cold junction compensation of the third display,
Make main display shows 0°C (32°F). If need to
adjust again, enter again after press key to
leave.
ADJ
TC O C
C
F
MAX
M
IN
ADJ
TC O C
ADJ
TC O C
ADJ
TC O C
6
TC O °C
ADJ
TC O °C
ADJ
TC O °C
ADJ
°C
°F
MAXMIN

Choose types key of thermocouple: To choose types of thermocouple.
−>Press to move right one -figure that
Press button.
thermocouple types chosen.
−>Press to move left one -figure that
thermocouple types chosen.
T1
、
、、
、
T2 exchange key
:
::
:
Main, secondary displays are exchanged (T1, T2) measurement readings.
Look over key of the perpetual calendar: Main display shows year, secondary display shows month, third
display shows hour: minute.
T1 T2
Press to enter.
C
F
C
F
T1 T2
T1 T2
7

The data record key: The data record is divided into real time recording and scheduled recording on
page 13. Set up the recording interval time from instructions on page 16.
To start recording in real
time press the button.
interval set quantity of data
Data logging starts at the
points appears in the third
display.
To stop recording press the
button.
Review logged data pen
instructions on page 9.
When in the scheduled
record mode the following
electricity settings are
available.
Mode to save electricity is under order setting record: Need to set up ( INTV) interval time ≥ 10 seconds.
Can increase the service time of order setting record.
press power key once>Enter mode 1 of save electricity
,
,,
,
symbol will glimmer, MCU will enter
the sleep state, will start MCU when the measurement recorded. Press power key again finishes mode
of save electricity 1.
press power key for 2 seconds>Enter mode 2 of save electricity,
,,
,symbol and REC will
glimmer,
,,
,MCU will enter the sleep state and close the peripheral drive circuit, will start MCU and drive
circuit when wanting to measurement record. If unable to finish the function of order setting record
under mode of save electricity 2, press power key 2 seconds to finish order setting record function.
H
H
8

Press buton to review recorded data: Main, secondary display shows recorded data.
Record data on (T1, T2), third
display shows the log of recorded
data, if it shows OU the memory
is already full.
>press once to look over
each logged data, press and hold
to look over the recorded data
forward fast.
>press once to look over
each logged data, press and hold
to look over the recorded data
backwards fast.
>Press to move left one
-figure that adjusted.
>Press to move right one
-figure that adjusted.
>
>>
>Press once to look
over record time that the
record loggers, main display
shows year, secondary
display shows month, date,
third display shows hour:
minute. Press or key
each time at this moment,
can circulate and change
third display shows hour:
minute or: second.
>Press key first and then press key, carry out clear memory function, press key
again, get back to measuring state.
Back light: Open/close LCD back light, will close automatically after opening one minute.
Function key of reset: Take a cross screwdriver, unlock after pressing Reset key under the battery cover,
the system will produce reset movement.
No press power key: The system will be produced reset.
except 0°C of cold junction compensation of the
thermocouple that set up, will all resume for the initial
state and enter and set up the function of the perpetual
calendar.
Press power key: The system is produced reset, clear
0°C of cold junction compensation of the thermocouple set
up and give out two sounds of buzzing, resume for the
initial state and enter and set up the function of the
perpetual calendar.
Note: Set up operation method of perpetual calendar,
please consult set up the perpetual calendar explanation.
MAXM I N
C
F
MAXM I N
MAXM I N
MAXMI N
MAXMIN
MAXMIN
SET
MAX
M
IN
C
F
MAX
MIN
SET
SET
9

Set up the options function key: To enter the choice of the set up function required, carry out the set up
function.
Press once: Confirms
selection of the chosen function.
Press for 2 seconds: Leaves set
up function.
>Press once to move
right one-figure of the set up
function chosen.
>Press once to move left
one-figure of the set up function
chosen.
SET
SET
C
F
10
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F
°C
°F

Set up relative value: When setting up relative value the third display will show 0 at beginning of set up. Press
, , , , keys to select value
and highlight , then use
then press .
Set up range is limited to the highest
and lowest temperature range that each
thermocouple can be measured. Adjust
to the set up value that you want then
press key to finish set up.
Can set up the range in every type
thermocouple:
K-TYPE:
::
:-200°C to 1372°C
(-328°F to 2501°F)
(-346°F to 2192°F)
J-TYPE:
::
:-210°C to 1200°C
(-418°F to 752°F)
-250°C to 400°C
(-346°F to 1832°F)
-210°C to 1000°C
(-238°F to 2372°F)
-150°C to 1300°C
(-32°F to 3212°F)
0°C to 1767°C
T-TYPE:
E-TYPE:
::
:
R- S-TYPE
:
N-TYPE:
>Press to move right one-figure that adjusted. >Press to move left one-figure that adjusted when in
Press key then press key, third display shows CLEA to clear set relative
value, press key again to finish clearing.
Rel value set mode.
Set up compensation value of cold junction of 0°C of thermocouple: Use and main display (T1 or T2)
thermocouple of the same type, from main
display (T1 or T2) inputs 0°C (32°F) temperature
value, adjust third display shows the cold junction
compensating value, enable main display shows
0°C (32°F). Press key again to finish setting up.
third display shows the cold junction
compensating value and the adjusting range is
about 0°C~50°C (32°F~122°F), will show Err and
unable to continue adjusting when exceeding the
range, need to press key to give up leaving.
The set up cold junction compensating value need
to use Reset add key to clear the set up value.
>Press to move right one-figure that adjusted.
>Press to move left one-figure that adjusted.
>Press to add one-figure upward that adjusted.
>Press to minus one-figure downward that
adjusted.
Note: Operation of Reset, please consult Reset function explanation.
MAXMIN
C
F
or
or
or
SET SET
SET
SET
C
F
MAX
M I N
SET
SET
SET
C
F
MAX MIN
SE
T
MAXMIN
C
F
SET
SET
SET
SET
S
ET
C
F
C
F
MAXMIN
11

12
Set up critical value of alarm: Set up main display shows (T1 or T2) critical value level of the alarm, secondary
display shows high critical value of the alarm and
third display shows low critical value of the alarm.
Begun from the set up value directly when there is
set up critical value the alarm. While use not
choosing type function of thermocouple or
temperature unit to choose function and use not
setting up critical value of alarm, the high, low
critical value of alarm shows that can be
measured for every type thermocouple highest
and minimum temperature value, adjust to you
want and press key again to finish setting up.
Can set up the range in every type thermocouple:
K-TYPE:
::
:-200 to 1372°C (-328 to 2501°F)
-210 to 1200°C (-346 to 2192°F)
-250 to 400°C (-418 to 752°F)
-210 to 1000°C (-346 to 1832°F)
-150 to 1300°C (-238 to 2372°F)
J-TYPE:
::
:
T-TYPE:
::
:
E-TYPE:
::
:
0 to 1767°C (32 to 3212°F)
:
R、
、、
、S-TYPE
N-TYPE:
::
:
>Press to add one-figure upward that adjusted.
>Press to minus one-figure downward that adjusted.
>Press to move right one-figure that adjusted.
>Press to move left one-figure that adjusted.
The sections set up of order setting record: Divide into the sections of order setting record (circulation of
every day) and (circulation of every year). Can only choose one to set up, every kind can be set up for 9
sections at most.
Choose circulation of every day or circulation of
every year: Can press key to switch over and
choose circulation of every day or circulation of
every year under not finishing the set up in any
section.
To delete the set up sections of circulation of every
day: Press key then press key, when main
display shows CLEA then press key to finish
deleting.
C
F
MAXMI N
SET
SET
MAXMIN
S
ET
SET
C
F
SET
SET

13
The order setting of circulation of
every day: A day 24 hours from
(00: 00: 00) 0 hour 0 minute 0
second to (23: 59: 59) 23 hours 59
minutes 59 seconds, at most can
set up 9 sections from 1 to 9 that
shows in main display. Symbol
START represents sets up the
beginning record time of section.
Symbol CLOSE represents sets
up the end record time of section.
Need from small to big when to
set up sections of order setting, At
the same time the set up time of
each section can not overlap and
repeat. It will be unable to finish
setting up to violate the above
rule of setting up. After finishing
sections of set up order setting
record, after carrying out, unless
the capacity of memory has been
To start order setting scheduled Data Logging press and
hold (REC) for 2 seconds. Display will show REC, START
and CLOSE.
START or CLOSE will only appear during the section
presently active. When complete data can be viewed by
using review mode on page 9.
already full or to cancel the
function of order setting,
repeated execution that will be
incessant every day.
>Press to add one-figure
upward that adjusted.
>
Press to minus one-figure
downward that adjusted.
>
Press to move right one
-figure that adjusted. When the
-figure of adjusted in hour:
minute (??: ??) that the most
right–figure, press again turn
into: second (: ??).
>Press to move left one
-figure that adjusted. When the
-figure of adjusted in: second (:??)
that the most left–figure, press
again turn into hour: minute
(??: ??).
>Press once to confirm set
up, press for 2 seconds to leave
the set up function.
C
F
MAXMIN
S
ET
SET
SET
SET
SET
SET
SET
SET
SET
SET
SET
°C
°F
°C
°F
°C
°F
MAXMI N
MAXMI N

14
Inquire about set up
sections of circulation of
every day: Press key
then press key, main
display glimmers shows
sections of order setting
inquired, wish to leave
the inquire function,
press key then press
key.
>Press once to
inquire the set up sections
will add 1, , there is no
function when the last
one set up section.
>Presses once to
inquire the set up sections
will minus 1, there is no
function when the first set
up section.
>
>>
>At start time (START) that inquire the set up sections to press once, will turn into stop time (CLOSE).
Press once in stop time (CLOSE), the third display shows hour: minute (??: ??) will turn into: second
(: ??).
>
>>
>At stop time (CLOSE) that inquire the set up sections to press once, will turn into start time (START).
Press once in start time (START), the third display shows hour: minute (??: ??) will turn into: second
(: ??).
To delete the set up sections of circulation of every
year: Press key then press key, while main
display shows CLEA then to press key to
finish deleting.
S
ET
MAXMI N
C
F
C
F
C
F
MAXMIN
SET
SET
SET
SET SET
SET
MAXMIN
SET
C
F
MAXMIN
SET
SET
SET
SET
M AX MI N

15
The order setting of circulation of
Scheduled Data Logging
every year: One year is divided
into 12 months, A day 24 hours
from (00: 00: 00) 0 hour 0 minute
0 second to (23: 59: 59) 23 hours
59 minutes 59 seconds, at most
can set up 9 sections from 1 to 9
that shows in main display.
Symbol START represents sets
up the beginning record time of
section. Symbol CLOSE
represents sets up the end record
time of section. Need from small
to big when to set up sections of
order setting, At the same time
the set up time of each section can
not overlap and repeat. It will be
unable to finish setting up to
violate the above rule of setting
up. After finishing sections of set
up order setting record, after
carrying out, unless the capacity
of memory has been already full
or to cancel the function of order
setting, repeated execution that
will be incessant every year.
>Press to add one-figure
upward that adjusted.
>Press to minus one-figure
downward that adjusted.
>Press to move right one
-figure that adjusted. When the
-figure of adjusted in hour:
minute (??: ??) that the most
right–figure, press again turn
into: second (: ??).
>Press to move left one
-figure that adjusted. When the
-figure of adjusted in: second (:??)
that the most left–figure, press
again turn into hour: minute
(??: ??).
>Press once to confirm set up, press for 2 seconds to leave the set up function.
C
F
MAXMIN
SET
SET SET SET
SET
SET
SET
SET SET
SET
SET
°C
°F
°C
°F
°C
°F
MAXMIN
MAXMIN

16
Inquire about set up
sections of circulation of
every year: Press key
then press key, main
display glimmers shows
sections of order setting
inquired, wish to leave the
inquire function, press
key then press key.
>Press once to
inquire the set up sections
will add 1, there is no
function when the last one
set up section.
>Presses once to
inquire the set up sections
will minus 1, there is no
function when the first set
up section.
>At start time (START) that inquire the set up sections to press once, will turn into stop time (CLOSE).
Press once in stop time (CLOSE), the third display shows hour: minute (??: ??) will turn into: second
(: ??).
>At stop time (CLOSE) that inquire the set up sections to press once, will turn into start time (START).
Press once in start time (START), the third display shows hour: minute (??: ??) will turn into: second
(: ??).
Set up interval time of record: The interval time of record, minimum interval time is one second, the biggest
interval time is 59 minutes and 59 seconds, adjust
to you want value and press key to finish
setting up to.
>Press to add one-figure upward that
adjusted.
>Press to minus one-figure downward that
adjusted.
>Press to move right one-figure that
adjusted.
>Press to move left one-figure that adjusted.
SET
MAXMIN
C
F
C
F
C
F
MAXM I N
SET
SET
SET
SET SET
C
F
MAXM I N
SET
SET
MAXMIN
SET
C
F
MAX
MIN
MAX
M
IN
SET
C
F

17
The time to set up of perpetual calendar: Main display shows year, the adjust range is from 2000 to
2099, secondary display shows date. month (xx.
xx), Third display shows hour: minute(24 hours
system). The adjust range from (00: 00) 0 hour 0
minute to (23: 59) 23 hours 59 minutes. Adjust to
set up value the effective date that you want then
to press key to finish setting up. If ineffective
time of date, need to adjust to time of effective
date could be left and set up.
>Press to add one-figure upward that
adjusted.
>Press to minus one-figure downward that
adjusted.
>Press to move right one-figure that
adjusted.
>Press to move left one-figure that adjusted.
General specification:
Display 4 1/2 digit crystal display, max display 19999
Polarity indicator No indicator in positive, negative indicator “-”
Overload indicator Positive overload show “OL”, negative overload show “-OL”
Low power indicator When five sections of batteries mark show( ), expressed battery power is
run low.
Power 4 AAA batteries
Auto power off
The pressed key use has not reached 20 minutes the battery power will be
turned off, hold down key for 3 seconds to cancel auto power off
function, and mark of display will be disappeared.
Reading renewal 4 times/sec
Data logger Every channel 9999 loggers at most (only for X3 = B, C)
General mode Approximately 250 hours/alkaline battery
Mode 1 of save
electricity
Approximately 3000 hours (interval = 15 minutes)/alkaline
battery
Battery life
Mode 2 of save
electricity
Approximately 7 months (interval = 15 minutes)/alkaline
battery
Operating temperature 0° ~ 50°C (32° ~ 122°F), <80% RH
-10° ~ 60°C (-4° ~ 140°F) <70% RH
164 L x 76 W x 32mm H
Storage temperature
Dimension
Weight Approximately 415g (include batteries)
C
F
MAX
MI N
SET
SET
MAX
M
IN
C
F
Table of contents
Other Omega Thermometer manuals

Omega
Omega OMEGASCOPE OS520 User manual

Omega
Omega OS880 Series User manual

Omega
Omega HH507R User manual

Omega
Omega OS950 Series User manual

Omega
Omega OS642F User manual

Omega
Omega OS642F-LS User manual

Omega
Omega OS449L-30 User manual

Omega
Omega OSP SERIES User manual

Omega
Omega hh12c User manual

Omega
Omega HH804U User manual