Checking Fiber OWL 4 Series Optical Power
Meters for Proper Operation
USING A LASER OWL SINGLEMODE LASER SOURCE
STEP 1 - Connect Fiber OWL 4 and Laser OWL
Connect the Fiber OWL 4 and Laser OWL together with a yellow
singlemode patch cable as shown below.
STEP 2 - Power ON the Fiber OWL 4
Press to power on the
Fiber OWL 4.
After a few seconds, the
OPERATING MODE screen will
appear on the Fiber OWL 4
display, with SIMPLE METER
being highlighted as shown at right.
STEP 3 - Enter SIMPLE METER
Press on the Fiber OWL 4
to enter SIMPLE METER.
The di splay sho uld rea d
‘UNDER’ as shown at right.
By default, the wavelength will
be set to ‘850nm’and the measurement units will be set to
‘dBm’.
NOTE: it is recommended to thoroughly clean and inspect all
patch cord connectors before making any connection.
F2
XXXXXXXXXXXXXXXXXX82
XXXXXXXXXXXXXXXXXX98%
UNITS ZERO WAVEXXX XXXX
XXXXXXXXXXXXXXXX850nm
XXXXXXXXXXXXXXXXXXdBm
_____________________
UNDER
STEP 4 - Set Fiber OWL 4 to 1310nm
Press on the Fiber OWL 4.
The wavelength display will
show ‘1310nm’ as shown at
right.
STEP 5 - Power ON the Laser OWL
Press to power on the Laser OWL.
The 1310nm indicator LED
should now be lit as shown at
right.
STEP 6 - Check optical power
The power reading displayed on
the Fiber OWL 4 will probably
not be exactly -10.00 dBm as
shown at right, but should be
fairly close.
Repeat this procedure for each
additional patch cable that
needs testing.
STEP 7 - Replace the patch cable
The easiest way to troubleshoot low power levels is to try another
patch cable. Over time, patch cables can wear out or become
damaged the more they are used for optical loss testing.
Replacing the patch cable usually fixes the problem.
STEP 8 - Clean patch cables and optical ports
If the problem still exists after patch cable replacement, there
may be some debris, such as dust, dirt, or finger oil, that has
collected on the connector endface or in the equipment optical
ports.
Thoroughly clean and inspect the optical ports of the Fiber OWL
4 and the Laser OWL, as well as the fiber connector endfaces,
according to the brochures that have been included with this kit.
Several cleaning cycles may be required.
If the power level is still below -11.00 dBm, even after a thorough
cleaning and inspection and patch cable replacement, contact
OWL technical support at 262-473-0643 for more information.
NOTE: this test may also be run
at 1550nm by setting the power
meter and source to 1550nm.
NOTE: the following steps are only required if the optical power
reading on the Fiber OWL 4 is -11.00 dBm or below.
F3 F3 F3
XXXXXXXXXXXXXXXXXX82
XXXXXXXXXXXXXXXXXX98%
UNITS ZERO WAVEXXX XXXX
XXXXXXXXXXXXXXX1310nm
XXXXXXXXXXXXXXXXXXdBm
_____________________
UNDER
XXXXXXXXXXXXXXXXXX82
XXXXXXXXXXXXXXXXXX98%
UNITS ZERO WAVEXXX XXXX
XXXXXXXXXXXXXXX1310nm
XXXXXXXXXXXXXXXXXXdBm
_____________________
X X10 00
LASER OWL
connected to
1310nm port
TEST PROCEDURES
CHECK SINGLEMODE REFERENCE CABLES
WITH A LASER OWL SINGLEMODE LASER SOURCE
LIGHT SOURCE CONNECTOR
PORT NOTES:
– CONNECTOR TYPE MAY VARY
– DO NOT INSERT ANGLED
PHYSICAL CONTACT (APC)
CONNECTOR
X XXNEXT SELECT
x XCERTIFICATION METER
XXXX XTECH SUPPORT
xx- X -OPERATING MODE
_____________________
XXXX XSIMPLE METER