MDL64 ABS Technical Manual
For Tech Support Manuals and How To Videos go to: pascale.com/resource-library/ or email us at: sales@pascale.com
Configuration Calibration Menus
CAL 20 Calibration setting entry point. Use to enter this menu and enter selections
Full capacity of the scale
Standard capacities are 2, 5, 10, 20, 50, 100, 150
and 200 lbs.
Displayed resolution or
count by, rounded up to
nearest 1, 2 or 5
Default entry is scale capacity divided by 5,000
(NTEP –Legal for trade configuration). In some
applications and environments up to 20,000
divisions of capacity may be possible non legal
for trade
Zero Range ‑Input the Zero Range in
% of full scale. The amount of weight
the scale can Zero.
1, 2, 3, 4, 5, 6, 7, 8, 9, 10,
11, 12, 13, 14,
Select the primary weighing unit by
keying in a number :1 = lb*, 2 = kg,
3 = g, 4 = oz t, 5 = lb t, 6 = g, 7 = dwt,
8 = oz, 9 = c, 10 = oz f, 12 = l,
11 = ml, 13 = tons, 14 = lb –oz
Stable, First, Unstbl,
ntEP, Auto, Prn-1
Select whether the scale will respond to a data
output/print request when stable, first (positive)
stable, any time (unstable), or NTEP.
Auto: Data output/print when stable and min 10
grads above zero, prints again with min 25 grad
change from last print. Does not need to return
to zero data output/print again.
Prn-1: Single stable data output/print, must
return to zero to data output/print again.
Measurement Canada legal for trade overrange
configuration: Select YES (9d) or NO (105%) *
Zero tracking value entered as a
percent of display resolution. Entering a
0.25* equals 25% of one display
graduation. “0” disables the zero
tracking feature.
Stable/Motion configuration in grads/sec.