Quincy lab 140E Series User manual

140E & 180E DIGITAL MODELS
Enter/Exit Menu
Menu Scroll
Increase/Decrease Setpoint
(PRESS)
Enter/Exit Menu
Menu Scroll
Increase/Decrease Setpoint
(PRESS)
Quincy Lab, Inc. 109. Shore Dr. Burr Ridge, Illinois 60527 1-800-482-HEAT (4328)
Quincy Lab, Inc.
Model 12E Incubator
POWER
DIGITALMICROPROCESSOR
140 SERIES
Switch
opening
Issue Rev A 0122
Copyright Quincy Lab Inc. 2022
Model 10-180E
DIGITAL MICROPROCESSOR
Quincy Lab, Inc.
Model 10E Incubator
POWER
Switch
opening
140 SERIES
Model 12-140E
Interior Dimensions
INCHES W x H x D
(CM) W x H x D
Exterior Dimensions
INCHES W x H x D
(CM) W x H x D
12x10x10 18x16x12 18x16x1212x10x10
13x15x11 19x21x13 13x15x11 19x21x11
Standard Electrical
31x25x25 46x41x30 46x41x3031x25x25
33x38x28 48x53x33 33x38x28 48x53x33
SPECIFICATIONS
Common Unit Construction
Exterior: Powder-Coated Steel Interior: Aluminum
Insulation: Fiberglass Door: 140E: Acrylic, 180E: Steel Insulated
Thermo-control: PID Microprocessor Heater: Resistive-Tubular Incoloy
VOLTS / WATTS 115 / 120* 115 / 235* 115 / 270* 115 / 385*
Cubic Foot Capacity
.7 ft32.0 ft3.7 ft32.0 ft3
Common Unit Specifications
Operating Environment: Indoor use, altitude to 6,500 ft. (2,000m) Installation Category II,
Pollution Degree 2, ambient temperature 10oC/50oF to 35oC/95oF,
80% RH maximum.
Storage Temperature: -10oC/14oF to 70oC/158oF, 70% RH maximum.
Approvals: Underwriter's Laboratory Listed, Laboratory Equipment, C/UL
United States/Canadian. E212550 (115VAC models only)
Compliance: UL Standard 61010-1, IEC 61010-1, 2nd Edition.
Weight (lbs) 19 lbs 33 lbs 19 lbs 33 lbs
Temperature Range
Ambient + 2
o
C to 62
o
C
Ambient + 3
o
C to 94
o
C
uincy
ab
General Purpose Incubators
OPERATING MANUAL
Model Series 140E & 180E
°
37.0
37.0
°
37.0
37.0
°
37.0
37.0
°
37.0
37.0
°
37.0
37.0
°
37.0
37.0
MODEL
10-140E
MODEL
12-140E
MODEL
10-180E
MODEL
12-180E
PAGE 1

The unit is ready for your immediate use. All control parameters, calibration,
and tuning has been done at the factory, no adjustments are necessary.
Push the illuminated power button. All LEDs on the temperature control will
light up and display the current chamber temperature (white LEDs), the set
temperature (green LEDs), and the active output yellow indicator light (OUT).
The set temperature is constantly displayed in the lower right-hand corner of
the display. To change the set temperature, simply press the up arrow key or
the down arrow key until the desired set temperature is reached. (FIG 4)
The temperature control is set at the factory to read in 1/10 degree F, or
Fahrenheit units. To make a change see: Menu Level Functions (page 3).
Once the unit reaches the desired set-point, allow the incubator to cycle
for 20 minutes for the temperature to become fully stable.
Upon each initial powering-up, the control may typically overshoot the set
temperature by 1 or 2 degrees, especially if the temperature set point is
close to the operating ambient temperature. After equilibrium is achieved
the control will hold set temperature within the model specified tolerance.
• Use only a grounded outlet that is rated for your model's electrical requirement.
•
• Conduct periodic maintenance as required.
PAGE 2
above the unit .
Plug the unit into a grounded outlet that is rated for your model's
electrical requirement.
Minumum
above the
unit and are easy to install and adjust using the wire shelf brackets. To
I
Do NOT modify the provided plug. Do NOT use an extension cord.
Use an individual branch circuit for your incubator.
label on the back of the unit.
Align bracket ends with holes located inside the ovens walls
Insert shelf bracket ends into the hole and press down on the bracket
as shown in . Repeat proccess on the opposite side of the wall.
Place shelf on top the the brackets.
Read Operating Instructions Thoroughly Prior to Operation
Safety PrecautionsSafety Precautions

Menu Level Functions Guide
Enter / Exit MENU MENU Scroll Decrease Increase
To navigate the controller's menu, please refer to the easy-to-use Menu Guide below: Menu setting changes can be done
quickly with our 5-button digital microprocessor. Your incubator's controller allows you to easily do the following:
Changes digit
cursor on set temperature
Set the operating set-point temperature
Select between degrees Fahrenheit or Centigrade
Calibrate your unit to your independent temperature sensing device.
M E N U G U I D E
To change the SP (set-point)
Press the or arrow key to adjust
To change the UNIT from °F to °C
Temperature Units
This parameter allows the controller to work and display the temperature on the specified unit. The value on the parame-
ter D-U can be adjusted to run in °F or °C.
The factory default operational D-U is set to run in °F.
62.0
62.0
Temperature Set-point
This is the target temperature value which the unit will operate under normal conditions. The value is located on the lower
right corner (green LEDs) and can be adjusted to your model’s specified temperature range.
F
D-U
Press and hold for (3) seconds
Press or to toggle between temperature units.
Set-point
Temperature
Units
Your incubator’s temperature offset value has been adjusted at the factory so that the temperature displayed on the
controller (white LEDs) reads the same as your unit's interior chamber temperature when measured against a NIST
(National Institute of Standard and Technology) traceable temperature box.
If an offset needs to be performed in the field to adjust the temperature displayed on the controller to match your indepen-
dent temperature probe, follow the sequence shown below:
0.0
CNS
Press once and press once to see CNS.
Adjust CNS using arrow key.
Press once to return to temperature display.
Temperature Offset
To calibrate control to independent probe/sensor
Auto-tune your Incubator for maximum efficiency
Lock the set-temperature
Temperature
Offset Value
PAGE 3

Control prompts will only display when a fault or alarm condition exists.
Control Self Diagnostics
Indicates Input Error
ALARM Codes "S.ERR" & "- - - -"
or
Check to make sure Thermocouple
wiring is connected securely
Indicates Internal Circuit Error
ALARM Code "E333"
Turn Controller OFF and On. If problem
remains, replace Controller
Indicates Internal Memory Error
ALARM Code "E111"
Turn Controller OFF and On. If problem
remains, replace Controller
ELECTRONIC FAIL SAFETY ALARM
62
65
If the incubator exceeds its maximum allowable internal temperature, the safety
alarm in the controller ( represents alarm condition) will cut power to the heating
elements until the warmer is manually reset. It is important to allow the incuba-
tor’s temperature to drop below its maximum operating temperature before
performing a reset to the unit.
To reset unit, power OFF and ON to clear alarm status. If problem remains, please
contact technical support for assistance.
How to lock controller’s Set-Point
The controller can be locked so that no changes can be made to the incubator’s set-point. Follow the steps below to lock
your controller.
Press the and HOLD together for (3) seconds.
0
OAPT
Change OAPT to (3) using arrow key.
Press and HOLD together for (3) seconds to
return to temperature display.
To unlock controller simply change the OAPT value back to zero (0).
Setting the OAPT setting to (3) will lock all parameters on the controller.
Factory default OAPT is set to (0).
The control's P.I.D values have been derived from a performed tuning at the factory using the AT-1 option which allows for
optimal performance and faster response time. You may, however, choose to tune your incubator to your specific application
and/or most used set-point temperature by selecting and performing one of the tuning options shown below.
Auto-tune
At-1 (for 40% Auto-tune) - The 40% Auto-tune (At-1), will stabilize the oven temperature quicker and with less 'overshoot',
but will be somewhat less precise.
At-2 (for 100% Auto-tune) - The 100% Auto-tune (At-2) will take longer to stabilize oven temperature but will be more
precise. This option takes longer to complete the Auto-tune process.
To auto-tune your incubator
Press once to see AT
Press or to toggle between AT-1 or AT-2
Select tune option, press once to begin auto-tuning.
OFF
AT
Auto-tune
Option
NOTE: Yellow light next to the word TUNE
indicates auto-tune is in progress.
Set the set-point to which the incubator will Auto-tune.
PAGE 4

Chamber Loading
For
The unit can operate
2°C above room temperature but temperature stability will be degraded.
non-typical ambient condions can be done by performing an
PAGE 5
lower heat-shield, is designed to travel up the sidewalls and can have a slightly elevated temperature from setpoint and the
rest of the chamber.
This will show by non-uniform processing and long
It is important to note that large trays placed on lower shelves prevent enough heat to rise within the chamber.
Probes can be inserted at
the top port.

140E & 180E DIGITAL MODELS
Maintenance
To clean interior and exterior surfaces, use a damp cloth with or without an all-purpose cleaner. Avoid commercially available
oven cleaners. The acrylic door should only be cleaned using a lint-free cloth, with or without water. Paper towels can mar
the surface of the acrylic door. Use of any commercial cleansers on the acrylic door will cause crazing and cracking of the
.
PAGE 6
Unit not turning ON
when switch is in the
ON position
1. Check if the unit tripped a GFCI outlet or fuse. Try a different
power outlet connection before moving to number 2 on this list.
2. Check (red) fuse holder in the back of the unit for missing
fuse or broken fuse. Replace as needed.
3. Check all wire connections in the electrical area (bottom) and
make sure there are no loose or disconnected wires in the unit.
Incubator not heating
or over heating.
2.Turn unit OFF and ON to clear (SUB/OUT) Alarm, if the
problem persists contact Quincy Lab. for further assistance.
Common Replacement Components Common Additional Equipment
Digital P.I.D Controller 10 Series Wire Shelf
For a complete list of replacement components, part and additional
equipment, visit us on the web at www.quincylab.com
1. Tripped GFCI power outlet
2. Damaged or missing fuse
3. Disconnected / loose wires.
2. The unit overheated and alarmed
PART # 101-1230 PART # 101-1000
S.S Relay
PART # 401-1235
PROBLEM POSSIBLE CAUSE WHAT TO DO BEFORE CALLING TECH. SUPPORT
1.Make sure the setpoint temperature is set correctly on the
lower right corner and is above the stated 2°C (140 series) or 3°C
(180 series) above room temperature.
1. Set temperature not set correctly or
set too low.
Quincy Lab, Inc. warrants to the original purchaser that this product will be free from defects in mate-
rial and workmanship under normal use throughout the warranty period. The standard warranty period
for this instrument is twenty four (24) months from date of shipment. The instrument warranty is supple-
mented with a three year warranty on the heating element. Please refer to your invoice or shipping
documents to determine the active warranty period. This warranty covers parts & labor (labor at factory only)
and shipping cost for replacement parts.
Limited Warranty
Email:
Voice:
Fax:
800-482-4328
773-622-2282
Quincy Lab, Inc.
109 Shore Dr.,
Burr Ridge, Illinois 60527
on the web at
• If a liquid is spilled inside the unit, disconnect it from the power supply and have it checked by a
competent person.
spilled on or inside the unit.
3. Incorrect OFFSET (CNS)value 3.Verify that the Temperature Offset (pg.3) is between -10°F
to10°F. Any value higher than this may affect the heating
process. Bring value down to zero, and perform a temperature
calibration. Scan QR-code above for additional information.
10 Series Shelf Support
PART # 101-1001
12 Series Wire Shelf
PART # 101-3000
12 Series Shelf Support
PART # 101-3001
SPACE LEFT
INTENTIONALLY BLANK
This manual suits for next models
5
Table of contents
Other Quincy lab Accessories manuals
Popular Accessories manuals by other brands

Baumer
Baumer OADM 21I6480/S14F manual

Krowne
Krowne BS36L series Installation and operation instruction manual

Vertiv
Vertiv Liebert Xtreme Density Series System Design Manual

amora
amora calliope instruction manual

DOL
DOL 26SCR Series Technical user guide

TrueLife
TrueLife AquaFloss Station O300 Ozone user manual