REED R8085 User manual

Noise
Dosimeter
Instruction
Manual
R8085
www. .com information@itm.com1.800.561.8187

2
Table of Contents
Introduction ................................................................................................ 3
Product Quality........................................................................................... 4
Safety ......................................................................................................... 4
Features...................................................................................................... 4
Specications..........................................................................................5-7
Comparison of Frequency Weightings and Tolerance Limits................. 7
Time Weightings (Fast & Slow) Characteristics...................................... 7
Included...................................................................................................... 7
Instrument Description............................................................................... 8
Display Description .................................................................................... 9
%DOSE.................................................................................................. 9
Sound Pressure Level (SPL)................................................................... 9
Operating Instructions.........................................................................10-11
Power On/Off ....................................................................................... 10
Sound Pressure Level (dBA/dBC Mode).............................................. 10
Data Hold ............................................................................................. 11
Recording Maximum and Minimum Readings..................................... 11
Backlighting ......................................................................................... 11
What is Dose Noise Exposure?...........................................................11-15
Noise Exposure (%DOSE) - Test Duration Mode ............................12-13
Noise Exposure (%DOSE) - Programmable Time Mode.................13-15
%DOSE Testing Notes: ........................................................................ 15
Data Logging.......................................................................................15-16
Sound Pressure Level (SPL) Data Recording Modes........................... 15
SPL Manual Data Logging Mode ...............................................15-16
SPL Automatic Data Logging Mode ............................................... 16
Setup Mode.........................................................................................16-21
Setting the Time and Date (dAtE)......................................................... 17
Setting the Data Logging Sampling Rate (SP-t)................................... 17
continued...
www. .com information@itm.com1.800.561.8187

3
Enabling/Disabling Auto Power OFF (PoFF) ...................................17-18
Setting the Criterion Level in dB (SEt LC)............................................ 18
Setting the Threshold in dB (SEt LT) .................................................... 18
Setting the Exchange Rate in dB (SEt ER)........................................... 18
Checking Memory Space (SPC) .......................................................... 19
Clearing Stored Data (CLrE)................................................................. 19
Verify Set Time, Date and Sampling Rate ............................................ 19
Calibration Procedure .....................................................................19-20
Data Stream Output ............................................................................. 20
3.5mm Terminal to RS-232 Serial Connection..................................... 21
Battery Replacement................................................................................ 21
Applications.............................................................................................. 21
Accessories and Replacement Parts ....................................................... 22
Appendix A: USA Regulations - OSHA Regulation
Conversion Charts.................................................................................... 22
Appendix B: Canada Regulations - Occupational Health and Safety
Regulations, Occupational Exposure Limits ............................................ 23
Appendix C: Noise Exposure Limits, by
Canadian Jurisdiction............................................................................... 24
Appendix D: Regulations Conversion from "Percent Noise Exposure"
or "DOSE" to "8-Hour Time-Weighted Average Sound Level" (TWA)...... 25
Product Care ............................................................................................ 25
Product Warranty ..................................................................................... 26
Product Disposal and Recycling .............................................................. 26
Product Support....................................................................................... 26
Introduction
Thank you for purchasing your REED R8085 Noise Dosimeter. Please
read the following instructions carefully before using your instrument. By
following the steps outlined in this manual your meter will provide years of
reliable service.
www. .com information@itm.com1.800.561.8187

4
Product Quality
This product has been manufactured in an ISO9001 facility and has been
calibrated during the manufacturing process to meet stated product
specications. If a certicate of calibration is required please contact the
nearest authorized REED distributor or authorized Service Center. Please
note an additional fee for this service will apply.
Safety
Never attempt to repair or modify your instrument. Dismantling your
product, other than for the purpose of replacing batteries, may cause
damage that will not be covered under the manufacturer's warranty.
Servicing should only be provided by an authorized service center.
Features
• Noise Dosimeter, Sound Level Meter and Data Logger
• High accuracy of ±1.0dB meets Type 2 standards
• Designed to meet noise survey requirements in compliance with OSHA
and CCOHS regulations
• Personal sound exposure meets IEC 61252, ANSI S1.25
• A & C frequency weighting
• Fast & Slow time weighting
• Lightweight and compact design
• Adjustable Criterion Level, Exchange Rate and Threshold
• USB interface with Windows®compatible software to retrieve stored
data (sold separately)
• Easy-to-read backlit LCD display
• Data hold and Min/Max hold
• Tripod mount for long-term monitoring
• Low battery indicator and auto shut off
www. .com information@itm.com1.800.561.8187

5
Specifications
Measurement Type: Sound Pressure Level: dB
Personal Noise Exposure: %Dose
Measuring Range(s): 35 to 130dB
Low: 35 to 80dB
High: 80 to 130dB
Auto: 35 to 130dB
Dose: 70 to 130dB
Dynamic Range: 50dB (in each range)
Accuracy:
31.5Hz ±3.0dB, 63Hz ±2.0dB, 125Hz ±1.5dB,
250Hz ±1.5dB, 500Hz ±1.5dB, 1kHz ±1.0dB,
2kHz ±2.0dB, 4kHz ±3.0dB, 8kHz ±5.0dB
Resolution: 0.1dB
Response Time: Approx. 1 second
Frequency Range: 31.5Hz to 8kHz
Frequency Weighting: A, C
Time Weighting: Fast/Slow (125mS and 1 second)
Microphone: 0.5" (12.7mm) electret condenser
Display: 4-digit LCD
Backlit Display: Yes
Data Hold: Yes
Min: Yes
Max: Yes
Alarm Indicators: Over (visual on-screen)
Datalogging Capabilities: Yes
Real-Time Clock and
Date Stamp: Yes
Selectable Sampling Rate: Yes (1, 2, 5, 10, 30, 60, 120, 300,
600, 1800, 3600 seconds)
continued...
www. .com information@itm.com1.800.561.8187

6
Selectable Dose
Measurement Rate: Yes (0, 1, 5, 10, 15, 30 minutes/
1, 2, 4, 8, 10, 12, 24 hours)
Auto Shut-off: Yes (after 10 minutes/off)
Tripod Mountable: Yes
Low Battery Indicator: Yes
Power Supply: 3 x AAA batteries
Output: Yes, RS-232/USB (optional)
PC Connectivity: USB Cable (R8085-USB, optional)
Software: PC Software (R8085-PC, optional)
Software OS Compatability: Windows XP/Vista/7/8/10
Product Certications: CE, IEC61252, ANSI S1.25 for Noise Dose
and Sound Exposure Meters,
IEC 61672 Class 2/ANSI S1.4 Type S
for Sound Level Meters
Operating Temperature: 32 to 122°F (0 to 50°C)
Storage Temperature: 14 to 140°F (-10 to 60°C)
Operating Humidity Range: <85%
Dimensions: 4.2 x 2.3 x 1.0" (107 x 58 x 25mm)
Weight: 4.3oz (122g)
continued...
www. .com information@itm.com1.800.561.8187

7
Comparison of Frequency Weightings and Tolerance Limits
Frequency
Hz
A Weighting
Characteristics
C Weighting
Characteristics
Tolerance
(IEC 61672
class 2)
31.5 -39.4dB -3dB
±3.0
dB
63 -26.2dB -0.8dB
±2.0
dB
125 -16.1dB -0.2dB
±1.5
dB
250 -8.6dB 0dB
±1.5
dB
500 -3.2dB 0dB
±1.5
dB
1K 0dB 0dB
±1.0
dB
2K +1.2dB -0.2dB
±2.0
dB
4K +1dB -0.8dB
±3.0
dB
8K -1.1dB -3dB
±5.0
dB
Time Weightings (Fast & Slow) Characteristics
Time Weighting
Characteristics
Max Response
Reference
continuous Signal
Tolerance
(IEC 61672 class 2)
F (Fast) -1.0dB
±1.0
dB
S (Slow) -4.1dB
±1.0
dB
Included
• Dosimeter
• Microphone with Clip
• Batteries
• Wrist Strap
• Carrying Case
www. .com information@itm.com1.800.561.8187

8
Instrument Description
3
4
5
14
13
12
9
6
7
8
10
15
16
R8085
Noise Dosimeter
11
ESC
HOLD RUN/
PAUSE
DUR/BEG TIME
SET LOG
ENTER
FUNC.
UNLOCK
1. LCD Display
2. HOLD Button
3. Power/ESC/Backlight Button
4. Up/Duration/Begin/Fast/
Slow/dB Range Button
5. FUNCTION/SET Button
6. Microphone Input Jack
7. RUN/PAUSE Button
8.
Down/A/C/TIME Check Button
9. ENTER/LOG Button
10. Wrist Strap Anchoring Point
11. RS-232 Output Jack
12. Tripod Mounting Screw
13. Battery Cover
14. Battery Cover Screw
15. Microphone
16. Microphone Plug
www. .com information@itm.com1.800.561.8187

9
Display Description
%DOSE
1. Test Duration, Start
(BEG), Stop (END)
Indicators
2. Locked Keypad
Indicator
3. Test Duration
Time Value
4. Time Stamp
5. Time Weighting
Indicator
6. Unit of Measure
7. Frequency Weighting Indicator
8. Recording Status Indicator
9. %DOSE Measurement
Sound Pressure Level (SPL)
1. dB Range Indicator
2. Time Weighting
Indicator
3. Unit of Measure
4. Frequency Weighting
Indicator
5. Sound Level
Measurement
6. Max/Min Recording
Indicator
3
4
9
5
6
7
8
BEG
END
DUR
FAST
%DOSE
m.s.
A
3
4
5
FAST
A
dB
6
REC
www. .com information@itm.com1.800.561.8187

10
Operating Instructions
Power On/Off
Turn the meter on by pressing the POWER button. To turn the meter off,
press and hold the POWER button for 2 seconds.
Sound Pressure Level (dBA/dBC Mode)
1. At start up, the meter's default settings are "Auto Range",
"A Frequency Weighting" and "Fast Time Weighting".
2. Press and hold the button to select the desired Response Time. If
the sound source consists of short bursts, set the response to FAST
(125ms). To measure average sound level, select SLOW (1 second).
3. Press and hold the button to select the Frequency weighting.
Select "A" Weighting for a general noise sound level, or "C"
Weighting for measuring high level noise. "C" Weighting is a
standard weighting of the audible frequencies commonly used for
the measurement of Peak Sound Pressure Level. If the "C" Weighted
level is much higher than the "A" Weighted level, then there will be a
large amount of low-frequency noise.
4. Press the button to scroll through the three dB range types. There
are two manual dB ranges to select from and an auto range. The
manual ranges are described as follows:
Low (as indicated by "L"): 30 to 80dB range
High (as indicated by "H"): 80 to 130dB range
5. The meter can be held comfortably in one hand, placed in a pocket
or positioned on a tripod for long-term monitoring.
6. Attach the microphone with the provided clip to an area where it will
be exposed to sound levels (i.e. shirt collar).
Note: Set the cable in a way that does not hinder the working movements
of the user. Direct contact with strong winds or blowing air on the
microphone may cause measurement errors.
continued...
www. .com information@itm.com1.800.561.8187

11
Data Hold
1. While taking a measurement, press the HOLD button to freeze the
current readings on the display.
2. While in this mode a "HOLD" symbol will appear.
3. Press the HOLD button again to resume normal operation.
Note: When the Data Hold feature is active all buttons except the POWER
button are disabled.
Recording Maximum and Minimum Readings
1. Press the RUN/PAUSE button to enter recording mode as indicated
by "REC" on the LCD. The meter will now begin recording maximum
and minimum readings.
2. While in recording mode:
A) Press the RUN/PAUSE button once and the maximum value
will appear on the display as indicated by "REC MAX".
B) Press the RUN/PAUSE button again and the minimum value
will appear on the display as indicated by "REC MIN".
C) To exit recording mode and resume normal operation, press
and hold the RUN/PAUSE button for approx. two seconds.
Note: When in recording mode the POWER button is disabled and the
meter cannot be turned off.
Backlighting
Press the BACKLIGHT button to turn the LCD Backlight on. It will
automatically turn off after approx. 5 seconds.
The R8085 is designed to calculate and display an individual's noise exposure in
%DOSE on the LCD display. All stored readings that are captured during the modes
described in this manual can be accessed, and exported to Excel for further analysis by
using the optional PC software (R8085-PC) and USB cable (R8085-USB).
What is Dose Noise Exposure?
The unit of measure, %DOSE, is used to quantify noise exposure that is
measured during an employee's work shift. 100% dose is the maximum
allowable noise exposure in accordance with OSHA, CCOHS, MSHA,
DOD, ACGIH, and ISO standards. Most standards specify what specic
Criterion Level, Exchange Rate, Response Time, and Frequency weighting
the dosimeter should be set up with in order to meet these
specic standards.
continued...
www. .com information@itm.com1.800.561.8187

12
According to OSHA's noise standard, specically, 29 CFR 1910.95, the
dosimeter must be set up to record noise exposure using the
following criteria:
• Exchange rate: 5dB
• Frequency weighting: A
• Response: Slow
• Criterion level: 90dBA (Hearing Conservation) or 90dBA (Administrative
and Engineering Controls).
• Threshold: 80dBA (Hearing Conservation) or 90dBA (Administrative and
Engineering Controls).
The Dosimeter allows for testing to begin immediately which is referred to
as a Test Duration (DUR) or with programmable beginning and end times
(BEG & END). Both modes are described in the following 2 sections.
Noise Exposure (%DOSE) - Test Duration Mode
1. Press the FUNC. button until you see SPL on the LCD.
2. Press and hold the button to select the desired Response Time.
3. Set the Criterion Level (Lc), Exchange Rate (ER), and Threshold Level
(Lt) (see Setup Mode section for details).
Note: At start up, the meter's default settings are Lc 90, Lt 70,
ER 3.
4. Press the FUNC. button to enter the %DOSE mode. By default, the
meter will be set to test duration (DUR).
5. Press and hold the button to set the duration time required
for testing.
6. Use the and buttons to select between the following
duration times:
24Hr 12Hr
10
Hr
8
Hr
4
Hr
2Hr 1Hr 30 minutes 15 minutes 10 minutes
5
minutes
1
minute
7. Press the ENTER button to conrm selection and return to the
main screen.
Note: Make sure you have enough memory to complete your noise survey
(see Clear Stored Data in Setup Mode section for details).
8. The meter can be placed in a pant or shirt pocket.
continued...
www. .com information@itm.com1.800.561.8187

13
9. Clip the microphone as close as possible to the user's ear
i.e. shirt collar.
Note: Set the cable in a way that does not hinder the user's ability to
perform their tasks.
10. Press the RUN/PAUSE button to begin measuring accumulated
noise exposure.
Note: When recording begins or is restarted all buttons except the
Backlight button are locked.
11. The elapsed duration timer will start automatically and "DUR" will
ash on the LCD display conrming test duration has begun.
12. To pause a recording, rst unlock the keypad by pressing the FUNC.
and ENTER buttons simultaneously and then press the RUN/PAUSE
button. The pause icon will appear conrming testing is now
paused while both the DUR icon will stop ashing.
13. Press the RUN/PAUSE button again to resume testing.
14. When the duration time is reached, "DUR" will stop ashing and the
nal accumulated noise exposure value is displayed.
15. To erase the previous accumulated %DOSE value and start a new
duration test, press the FUNC. and ENTER buttons simultaneously.
16. Press the FUNC. button to resume normal operation.
Noise Exposure (%DOSE) - Programmable Time Mode
This mode allows testing to begin and end at user selected times. It is
strongly advised to set the correct date and time before proceeding (see
Setting the Time and Date in Setup mode section for details).
1. Press the FUNC. button until you see SPL on the LCD.
2. Press and hold the button to select the desired Response Time.
3. Set the Sampling Rate, Criterion Level (Lc), Exchange Rate (ER), and
Threshold Level (Lt) (see Setup Mode section for details).
Note: At start up, the meter's default settings are Sampling
Rate 2 sec., Lc 90, Lt 70, ER 3.
4. Press the FUNC. button to enter the %DOSE mode. By default, the
meter will be set to test duration (DUR).
5. Press and hold the button to switch to programmable mode as
indicated by "BEG" on the LCD.
continued...
www. .com information@itm.com1.800.561.8187

14
6. Press and hold the button to set the recording start time.
7. Use the and buttons to adjust the current date as indicated
by "dd".
8. Press the FUNC. button to conrm selection and switch to the hour
digit as indicated by "h".
9. Use the and buttons until the desired number value is reached.
10. Press the FUNC. button to conrm selection and switch to the
minutes digit as indicated by "m".
11. Use the and buttons until the desired number value is reached.
12. Press the ENTER button to save your selection and skip to the
recording end time parameter as indicated by "END" on the LCD.
13. Repeat steps 6 through 10 to set the recording end time.
14. Press the ENTER button to save your selection and return to the
main screen.
Note: Make sure you have enough memory to complete your noise survey
(see Clear Stored Data in Setup Mode section for details).
15. The meter can be placed in a pant or shirt pocket.
16. Clip the microphone as close as possible to the user's ear
i.e. shirt collar.
Note: Set the cable in a way that does not hinder the user's ability to
perform their tasks.
17. Press and hold the LOG button to enter automatic data logging
mode for SPL recordings as indicated by "LOGGER" on the LCD.
18. Press the RUN/PAUSE button to activate the programmed start and
end times.
19. The keypad will automatically lock and the meter will start measuring
once the programmed start time is reached.
20. When started, "DUR" will ash on the LCD display conrming
programmed testing has begun while "LOGGER" will ash every time
a SPL reading is stored at the user selected sampling rate.
21. To pause a recording, rst unlock the keypad by pressing the FUNC.
and ENTER buttons simultaneously and then press the RUN/PAUSE
button. The pause icon will appear conrming testing is now
paused while both the DUR and LOGGER icons will
stop ashing. continued...
www. .com information@itm.com1.800.561.8187

15
22. Press the RUN/PAUSE button again to resume testing.
23. When the end time is reached, both the DUR and LOGGER icons
will stop ashing and the nal accumulated noise exposure value is
displayed.
24. To end the current session, rst unlock the keypad again and press
and hold the ENTER button for 2 seconds.
25. Press the FUNC. button to resume normal operation.
%DOSE Testing Notes:
• Never shout into the microphone as this will affect the test results.
• Microphone should not come into contact with any objects during
recording to avoid erroneous test results.
• For best results, test over the course of several days and average
the %DOSE.
Data Logging
Sound Pressure Level (SPL) Data Recording Modes
Manual Data Logging: Manually log readings into the internal memory
(see Manual Data Logging Mode section for details).
Automatic Data Logging: Setup the meter parameters in order to
automatically log data into the internal memory.
Note: The meter can store up to 16,000 data points and will indicate
"FULL" on the LCD display when this amount is exceeded.
Please note to use the data logging feature for Sound Level Pressure (SPL) the optional
PC software (R8085-PC) and USB cable (R8085-USB) are required. With the software,
all stored readings that are captured can be downloaded and exported to Excel for
further analysis.
SPL Manual Data Logging Mode
In manual mode the LOG button is pressed to manually log a reading into
the internal memory at any time. In order to setup the meter for manual
data logging, follow steps 1 through 4 below.
1. Set the data logging sampling rate to "0" seconds.
2. Press and hold the LOG button to enter manual data logging mode.
3. Press the LOG button to save a reading to memory. "LOGGER" will
ash each time a data point is stored.
continued...
www. .com information@itm.com1.800.561.8187

16
4. To exit manual data logging mode, press and hold the LOG button
for 2 seconds to resume normal operation.
SPL Automatic Data Logging Mode
In automatic mode, the desired data logging sampling rate is set to 1, 2,
5, 10, 30, 60, 120, 300, 600, 1800 or 3600 seconds prior to recording.
In order to setup the meter for automatic data logging, follow steps 1
through 5 below.
1. Press and hold the LOG button to enter automatic data
logging mode.
2. The "LOGGER" will ash every time a reading is stored at the user
selected sampling rate.
3. To pause the data logging session press the LOG button and
LOGGER will stop ashing.
4. To resume press the LOG button again.
5. To end the current data logging session, press and hold the LOG
button for 2 seconds to resume normal operation.
Setup Mode
1. Press and hold the FUNC. button for 2 seconds to enter
Setup Mode.
2. Press the FUNC. button continuously to scroll through the
following parameters.
Parameter Description
dAtE Set the time and date
SP-t Set the data logging sampling rate
PoFF Enable or disable the auto-power off
SEt LC Set Criterion level in dB
SEt LT Set Threshold in dB
SEt ER Set Exchange rate in dB
SPC Check Memory space
CLrE Clear Stored Data
continued...
www. .com information@itm.com1.800.561.8187

17
3. Once the appropriate parameter has been selected follow the
associated instructions below.
Note: The meter automatically exits out of the Setup mode if no key is
pressed within 7 seconds.
Setting the Time and Date (dAtE)
1. Press the ENTER button when the dAtE parameter appears on
the LCD.
2. Use the and buttons to adjust the year as indicated by "YY".
3. Press the ENTER button to conrm selection.
4. Repeat steps 2 and 3 for month, day, hour, minute and seconds as
indicated by (mm./dd/h./m./s).
5. After each value has been selected and conrmed, it will
automatically skip to the next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation. The internal clock will keep accurate time
when the meter is powered off. When new batteries are installed the clock
will have to be reset.
Setting the Data Logging Sampling Rate (SP-t)
Follow steps 1 and 2 when "SP-t" appears on the LCD.
1. Use the and buttons to adjust the sampling rate between 0, 1,
2, 5, 10, 30, 60, 120, 300, 600, 1800, and 3600 seconds.
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
Enabling/Disabling Auto Power OFF (PoFF)
Follow steps 1 and 2 when "PoFF" appears on the LCD.
1. Use the and buttons to select between ON (enabled) or OFF
(disabled). With the Auto Power OFF feature enabled, the meter will
automatically switch OFF after 10 minutes of inactivity to preserve
battery life.
continued...
www. .com information@itm.com1.800.561.8187

18
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
Setting the Criterion Level in dB (SEt LC)
Follow steps 1 and 2 when "SEt LC" appears on the LCD.
1. Use the and buttons to adjust the criterion level between 80,
84, 85, and 90 dB.
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
Setting the Threshold in dB (SEt LT)
Follow steps 1 and 2 when "SEt LT" appears on the LCD.
1. Use the and buttons to adjust the threshold which can be set
anywhere between 70 and 90 dB.
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
Setting the Exchange Rate in dB (SEt ER)
Follow steps 1 and 2 when "SEt ER" appears on the LCD.
1. Use the and buttons to adjust the threshold which can be set
between 3, 4, 5 and 6 dB.
2. Press the ENTER button to conrm selection and skip to
next parameter.
Note: At any time, you can press the ESC button to exit the Setup mode
and resume normal operation.
continued...
www. .com information@itm.com1.800.561.8187

19
Checking Memory Space (SPC)
When "SPC" appears on the LCD, the upper part of the display will
indicate the remaining datapoints available to record.
Clearing Stored Data (CLrE)
Follow steps 1 through 5 when "CLrE" appears on the LCD.
1. Use the and buttons to select between Yes (Clear data) or
No (Abort).
2. Press the ENTER button to conrm selection.
Note: If "No" is selected, the meter will skip to the next parameter.
3. If "Yes" is selected, press the ENTER button again to re-conrm.
4. The meter will clear all stored data and automatically skip to the next
parameter when complete.
5. When all parameters have been set up, press the ESC button to exit
the Setup mode and resume normal operation.
Verify Set Time, Date and Sampling Rate
During normal operation press the TIME button to display the year,
month, day, hour, minutes, and selected sampling rate.
Calibration Procedure
In order to calibrate the R8085, an external calibrator that can provide a
94.0dB signal is required (REED R8090). If the meter has been exposed
to extreme conditions or has not been used in a long period of time the
meter should be calibrated by following steps 1 through 11 below.
1 Power on the meter.
2 Press the FUNC. button to ensure you are in SPL mode.
3. Select "80 - 100dB" (Hi) range.
4. Select "FAST" time weighting.
5. Select "A" weighting.
6. Insert Microphone into the sound calibrator's microphone adapter.
7. Power on the sound calibrator (R8090).
Note: If the measured value is not within "94 ±0.2" dB, adjust the meter's
calibration by entering calibration mode.
continued...
www. .com information@itm.com1.800.561.8187

20
8. Press and hold the HOLD and RUN/PAUSE buttons simultaneously
for approx. 3 seconds to enter calibration mode.
9. When in calibration mode, the current sound level measurement will
appear on the lower and upper part of the LCD display.
10. Press the and buttons to adjust the upper display until it
matches the output of the calibrator (94.0dB).
11. Once the calibration is complete, press ENTER button to save the
calibration value into the memory which will ash conrming it has
been saved.
12. Press the FUNC. button to exit the calibration mode and resume
normal operation.
Data Stream Output
Using Terminal application, a data stream can be viewed from the RS-232
output, Connect the RS-232/USB cable between the product and
terminal/PC and use the following setting to view the data stream.
Terminal Settings:
• Bits per second: 9600
• Data bits: 8
• Parity: None
• Stop bits: 1
41170100000715 = +71.5 dB
Reading
Decimal Place 0-3
Positive (+) = 0, Negative (-) = 1
dB =
17, % = 03
SPL = 1, Dose = 2
4
continued...
www. .com information@itm.com1.800.561.8187
Table of contents
Other REED Measuring Instrument manuals