ThecurrentalarmvaluewillappearontheLCD.Pressthe“UP”and
“DOWN”buttonstomodifytheppmvaluebyincrementsof100.The
alarmrangesfrom100to9900ppm.Whenthedesiredalarmvalueis
displayed,pressthe“ENTER”buttontosave,orpressthe“ESC”button
to cancel.
WhenmeasuredCO2valuesareabovethesetlimitthemeterwillstart
tobeep.Pressanybutton(exceptforthe“POWER”button)tostopthe
beeping,ormovethemetertoalocationwheretheppmvalueisbelow
thealarmlimit.Ifthebeepingdoesnotstop,restartthemeter.
Temperature Unit
Tochangethetemperaturescalefrom°Cto°F,pressandholdthe“SET”
buttonfor2seconds.Pressthe“DOWN”buttonand“Unit”andP3.0will
appearontheLCD.Pressthe“ENTER”button,andmodifythetempera-
tureunitbypressingthe“UP”or“DOWN”buttons.Pressthe“ENTER”
buttontosavethesetting,orpressthe“ESC”buttontocancel.
CO2Calibration
Thismeteriscalibratedatastandard400ppmCO2concentration. It’s sug-
gestedtoperformmanualcalibrationregularlytomaintaingoodaccuracy.
Foramorethoroughcalibrationandformaintenanceforthisproduct,
pleasecontactREEDInstrumentsatinfo@reedinstruments.com.
DonotcalibratethismeterinanenvironmentwithunknownCO2con-
centrations.Manualcalibrationshouldbedoneinfreshoutdoorairthat
iswellventilatedandinsunnyweather.Turnonthemeterandholddown
the“CAL”and“MODE”buttonssimultaneously.“400ppm”and“CAL”will
blinkontheLCDindicatingyouhaveactivatedmanualcalibration,andwill
blinkforapproximately5minutes.Whentheblinkingstops,CO2calibra-
tioniscomplete,andthemeterwillautomaticallyswitchbacktonormal-
measuringmode.Toabortthecalibration,turnoffthemeteratanytime.
NOTE:Ensurethebatteriesarewithfullvoltageduringthecalibrationto
preventfrominterruptionorfailedcalibration.