Operating Instructions
POWER ON/OFF
Pressthe”PWR”buttontoturnthemeteron,
pressandholdfor2secondstoturnthemeter
off.
Whenthemeterispoweredon,theLCDdisplays
theWBGTmodeandthepreviouslyselected
“IndoororOutdoor”WBGTmeasurement
mode.Shortlyafter,abeepwillsoundandthe
instrumentwillbegintakingmeasurements.
MAX/MIN/AVG
Pressthe MAX/MIN buttontocyclethedisplay
throughtheMaximum,Minimum,Average
measurementsofAmbientTemperature,Global
TemperatureandRelativeHumiditysincethe
instrumentwaspoweredon.
PressandholdtheMAX/MINbuttonuntil“CLr”
appearsonthebottomofLCD.Theinstrumentwill
automaticallyreturntonormalmeasurementmode
oncethereadingshavecleared.
INDOOR/OUTDOOR Temperature MODE SELECTION
Wheninthe WBGTmode,pressandholdtheUNITbuttontotoggle
betweenindoororoutdoortemperature.Theinstrumentautomatically
calculatesthemeasurementusingthefollowingparameters:
a. OutdoorformulaWBGT=0.7Tw(wetbulbtemp)+0.2Tg(globe
temp)+0.1Td(airtemp.)
b. IndoorformulaWBGT=0.7Tw(wetbulbtemp)+0.3Tg(globe
temp)
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com