
16 820i Installation Manual
3.1.2 Serial Command Configuration
The serial command set can be used to configure the
820i indicator using either a personal computer,
terminal, or remote keyboard. Like Revolution, serial
command configuration sends commands to the
indicator serial port; unlike Revolution, serial
commands can be sent using any external device
capable of sending ASCII characters over a serial
connection.
Serial commands duplicate the functions available
using the indicator front panel and provide some
functions not otherwise available. Serial commands
can be used to simulate pressing front panel keys, to
configure the indicator, or to dump lists of parameter
settings. See Section 9.0 on page 79 for more
information about using the serial command set.
3.1.3 Front Panel Configuration
Use the CONFIG submenu under the SCALES menu
to configure A/D scales. For example, in an indicator
with a single-channel A/D card installed in Slot 1, the
Scale Configuration display will show the A/D listed
(Slot 1 Channel 1) under the AVAILABLE A/D’s column. Use
the left navigation key to select the A/D, then press the
center softkey, Add. The A/D is then moved to the
Associated A/D’s column. If no other A/D’s are listed
in the AVAILABLE A/D’s column, the center softkey
changes to Done, as shown in Figure 3-2. Press Done to
exit the Scale Configuration display.
See Section 10.3 on page 93 for information about
configuring serial scales.
05/24/2006 02:22PM
Change
Type Done
CONFIG SCALE 1
SCALE 2
Scale 1
Scale 2
Scale 3
SCALES ASSOCIATED A/D’sAVAILABLE A/D’s
Slot 1 Channel 1
Figure 3-2. Scale Configuration Display
3.1.4 Multi-Range and Multi-Interval Scales
The 820i supports multi-range and multi-interval scales
of either two or three ranges or intervals.
Multi-range scales provide two or three ranges, each
extending from zero to the maximum capacity
specified for the range, that can specify different scale
intervals (graduations). The scale interval changes as
the applied weight increases but does not reset to lower
range intervals until the scale returns to zero.
Multi-interval scales divide the scale into two or three
partial weighing ranges, each with different scale
intervals. The scale interval changes with both
increasing and decreasing loads applied.
To configure a multi-range or multi-interval scale, use
the SPLIT parameter to select 2RNG or 3RNG (for
multi-range scales), or 2INTVL or 3INTVL (for
multi-interval scales). Selecting a SPLIT value other
than OFF allows specification of decimal point, display
divisions, and maximum capacity for each range or
interval.
The SPLIT parameter is used to enable multi-range or
multi-interval. The SPLIT parameter is in the SCALES
menu, see Figure 3-4, and Table 3-2. After setting the
SPLIT parameter, the Format menu selection will
change as shown in Figure 3-6, and Table 3-4.
If using streaming with multi-range or multi-interval,
the stream must be set to Custom in Revolution. The
Tokens for Secondary and Tertiary Units must be set to
L or K to match the Primary, refer to the Serial Menu,
Tokens Parameter, in Section 3.2.2 on page 30. They
can be set using Revolution or through the front panel.
In multi-range, each range has its own capacity and
display division, extending from zero. The scale
display division will increase at the entered range
capacities, either two or three ranges. Once the range
has increased to the next level, the display division will
remain in new range until the scale returns to zero. The
tare value can be taken in any range.
For example,
Range 1 is 0 - 3000 x 1 lb.
Range 2 is 0 - 10,000 x 5 lb.
In multi-interval, the scale has one capacity, which is
segmented into weighing intervals, either two or three
intervals, each with different display division sizes. As
the weight value exceeds an interval or set interval, the
display division will increase, as the weight falls below
an interval or set interval, the display division will
decrease. The tare can only be taken in the first
interval.
For example,
Range 1 is 0-30 x 0.01 lb.
Range 2 is 30 - 60 x 0.02 lbs.
3.1.5 Total Scale Configuration
The output of A/D scales, serial scales, or iQUBE
systems can be configured to function as a total scale.
Once configured and calibrated, the total scale can be
used as a source for other system functions, including
streaming, setpoints, print formatting, and analog
output.