Rice Lake IQ plus 710 User manual

64505
IQ plus®710
Digital Weight Indicator
Version 2.1
Installation Manual
Discontinued Product

Discontinued Product

Contents
1.0 Introduction.................................................................................................................................. 1
1.1 Operating Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Front Panel Display. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Indicator Operations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.1 Toggle Gross/Net Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.2 Toggle Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.3 Zero Scale. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.4 Acquire Tare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.5 Remove Stored Tare Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.6 Print Ticket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.7 Display or Change Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.8 Display or Change Setpoint Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.9 Turn Setpoint On or Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
1.3.10 Display or Clear Accumulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.0 Installation ................................................................................................................................... 4
2.1 Unpacking and Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.2 Enclosure Disassembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3 Cable Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3.1 Load Cells. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2.3.2 Serial Communications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.3.3 Digital I/O . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.3.4 Analog Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.4 Analog Output Module Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.5 Enclosure Reassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.6 Board Removal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.7 Battery Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2.8 Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.0 Configuration ............................................................................................................................. 14
3.1 Configuration Methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.1 Revolution Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.2 EDP Command Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.1.3 Front Panel Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2 Menu Structures and Parameter Descriptions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.2.1 Configuration Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2.2 Format Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.2.3 Calibration Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.2.4 Serial Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.2.5 Program Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.2.6 Print Format Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.2.7 Setpoints Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.2.8 Digital Input Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
3.2.9 Analog Output Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
3.2.10 Version Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.0 Calibration ................................................................................................................................. 37
4.1 Front Panel Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
4.2 EDP Command Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
4.3 Revolution™Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
4.4 More About Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Technical training seminars are available through Rice Lake Weighing Systems.
Course descriptions and dates can be viewed at www.ricelake.com/training
or obtained by calling 715-234-9171 and asking for the training department.
© 2010 Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America.
Specifications subject to change without notice.
Rice Lake Weighing Systems is an ISO 9001 registered company. Version 2.1, August 2007
Discontinued Product

ii IQ plus 710 Installation Manual
4.4.1 Adjusting Final Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.4.2 Zero Deadload A/D Counts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
4.4.3 Calculating the Span Coefficient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.0 EDP Commands.......................................................................................................................... 41
5.1 The EDP Command Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.1.1 Key Press Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.1.2 Reporting Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1.3 The RESETCONFIGURATION Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1.4 Parameter Setting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.1.5 Normal Mode Commands. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
5.1.6 Batching Control Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
5.2 Saving and Transferring Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.2.1 Saving and Printing Indicator Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
5.2.2 Downloading Configuration Data from PC to Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.0 Print Formatting ......................................................................................................................... 49
6.1 Print Formatting Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.2 Default Ticket Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.3 Customizing Print Formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.3.1 Using the EDP Port. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.3.2 Using the Front Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
6.3.3 Using Revolution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.0 Truck Modes .............................................................................................................................. 53
7.1 Using the Truck Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.1.1 Modes 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.1.2 Modes 3, 4, 5, and 6 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
7.1.3 Single-Transaction Tare Weights and IDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.0 Setpoints .................................................................................................................................... 55
8.1 Batch and Continuous Setpoints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.2 Batching Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.2.1 Example 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
8.2.2 Example 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.3 Batching Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
9.0 Macro Programming.................................................................................................................. 60
9.1 Using the Macro Submenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
9.2 Macro Programming Examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
9.2.1 Example 1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
9.2.2 Example 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
10.0 Appendix .................................................................................................................................... 65
10.1 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
10.1.1 Displayed Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
10.1.2 Using the XE EDP Command . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
10.2 Status Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
10.3 TARE and ZERO Key Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
10.4 Data Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
10.4.1 Continuous Output Serial Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
10.4.2 Demand Output Serial Data Format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . 68
10.4.3 RS-485 Data Formats. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
10.5 Digital Filtering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
10.6 Conversion Factors for Secondary Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
10.7 Analog Output Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
10.8 Test Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
10.9 Software Revision Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
10.10 Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
IQ plus 710 Limited Warranty ............................................................................................................... 75
Discontinued Product

Introduction 1
About This Manual
This manual is intended for use by service technicians responsible for installing and servicing IQ plus®710
digital weight indicators.
This manual applies to indicators using Version2.1 of the IQ plus 710 software. See Section 10.9 on page 73 for
a summary of software changes included in this release.
Configuration and calibration of the indicator can be accomplished using the indicator front panel keys, the EDP
command set, or Version 3.0 or later of the Revolution™configuration utility (Revolution III). See Section 3.1 on
page 14 for information about configuration methods.
Some procedures described in this manual require work inside the indicator enclosure. These procedures are
to be performed by qualified service personnel only.
Authorized distributors and their employees can view or download this manual from the Rice
Lake Weighing Systems distributor site at www.ricelake.com.
The Operator Card included with this manual provides basic operating instructions for users of the IQ
plus 710. Please leave the Operator Card with the indicator when installation and configuration are
complete.
1.0 Introduction
The IQ plus 710 is a single-channel digital weight
indicator housed in a NEMA 4X/IP66-rated stainless
steel enclosure. The indicator front panel consists of a
29-button keypad with a large, seven-digit,
14-segment, vacuum fluorescent display,
two-character dot-matrix annunciator field, and a
sixteen-character dot-matrix prompt field. Features
include:
• Drives up to eight 350Ω or sixteen 700Ωload
cells
• Supports 4- and 6-wire load cell connections
• Eight configurable digital inputs
• Eight digital outputs
• Electronic data processing (EDP) port for full
duplex RS-232 or RS-485 communications at
up to 19200 bps
• Printer port for full duplex RS-232 and
output-only 20 mA current loop
communications at up to 19200 bps
• Optional analog output module provides 0–10
VDC or 4–20 mA tracking of gross or net
weight values
• Optional Remote I/O Interface for
communication with PLC®and SLC™
controllers using the Allen-Bradley®Remote
I/O networks1
• Optional interface for Profibus®DP network
communications2
• Available in 115 VAC and 230 VAC versions
The IQ plus 710 is NTEP-certified for Classes III and
III L at 10,000 divisions. See Section 10.10 on
page 74 for more information about NTEP and OIML
certifications and Measurement Canada approval.
1.1 Operating Modes
The IQ plus 710 has three modes of operation:
Normal mode
Normal mode is the weighing mode of the
indicator. The indicator displays gross, net, or tare
weights as required, using the secondary display
to indicate scale status and the type of weight
value displayed. Once configuration is complete
and a legal seal is affixed to the back of the
indicator, this is the only mode in which the IQ
plus 710 can operate.
Setup mode
Most of the procedures described in this manual
require the indicator to be in setup mode,
including configuration and calibration.
To enter setup mode, remove the large fillister
head screw from the enclosure backplate. Insert a
screwdriver or a similar tool into the access hole
and press the setup switch once. The indicator
display changes to show the word CONFIG.
Test mode
Test mode provides a number of diagnostic
functions for the IQ plus 710 indicator. Like setup
mode, test mode is entered using the setup switch.
See Section 10.8 on page 72 for more information
about entering and using test mode.
Caution
1. Allen-Bradley®, PLC®, and SLC™are trademarks of
Allen-Bradley Company, Inc., a Rockwell International
company.
2. Profibus®is a registered trademark of Profibus
International.
Discontinued Product

2IQ plus 710 Installation Manual
Figure 1-1. IQ plus 710 Front Panel
1.2 Front Panel Display
Figure 1-1 shows the IQ plus 710 front panel keys and
the key functions assigned in normal mode.
The IQ plus 710 display is divided into three areas
(see Figure 1-2):
• The primary display consists of seven large,
14-segment digits used to display weight data.
• A two-digit annunciator shows the units
associated with the displayed value: lb=pounds,
kg=kilograms, oz=ounces, T=short tons,
t=metric tons, LT=long tons, g=grams,
GN=grains. When the units configured are troy
pounds or troy ounces, the word troy is shown
in the secondary display area in addition to the
lb or oz annunciator. The units can also be set to
NONE (no units information displayed).
• The two-digit annunciator also displays
whether the indicator is in numeric entry (NE)
or alpha entry (AE) mode for some
operations.
The 16-digit secondary display is used to display the
weighing mode (Gross/Brutto or Net) and status
indicators, including standstill ( ) and center of
zero ( ).
The symbols shown over the keys in Figure 1-1
(representing up, down, enter, left, right) describe the
key functions assigned in setup mode. In setup mode,
the keys are used to navigate through menus, select
digits within numeric values, and
increment/decrement values. See Section 3.1.3 on
page 15 for information about using the front panel
keys in keys in setup mode.
Figure 1-2. IQ Plus 710 Front Panel Display Areas
N
13*."3:%*41-":
4&$0/%"3:%*41-":
%*(*5
"//6/$*"503
GPS6/*54
BMQIBFOUSZ"&
OVNFSJDFOUSZ/&
Discontinued Product

Introduction 3
1.3 Indicator Operations
Basic IQ plus 710 operations are summarized below:
1.3.1 Toggle Gross/Net Mode
Press the GROSS/NET key to switch the display mode
from gross to net, or from net to gross. If a tare value
has been entered or acquired, the net value is the gross
weight minus the tare. If no tare has been entered or
acquired, the display remains in gross mode.
Gross mode is indicated by the word Gross (or
Brutto in OIML mode) on the secondary display; net
mode is indicated by the word Net.
1.3.2 Toggle Units
Press the UNITS key to switch between primary and
secondary units. The units identifier is shown to the
right of the primary display. Troy ounces and troy
pounds are indicated by the word troy on the
secondary display.
1.3.3 Zero Scale
1. In gross mode, remove all weight from the
scale and wait for the standstill annunciator
().
2. Press the ZERO key. The center of zero
() annunciator lights to indicate the
scale is zeroed.
1.3.4 Acquire Tare
1. Place container on scale and wait for the
standstill annunciator ( ).
2. Press the TARE key to acquire the tare weight
of the container.
3. Display shifts to net weight and shows the
word Net on the secondary display.
To display the current tare value, press the DISPLAY
TARE key.
1.3.5 Remove Stored Tare Value
1. Remove all weight from the scale and wait for
the standstill annunciator ( ).
2. Press the TARE key (or, in OIML mode, the
ZERO key). Display shifts to gross weight and
shows the word Gross on the secondary
display.
1.3.6 Print Ticket
1. Wait for the standstill annunciator ( ).
2. Press the PRINT key to send data to the serial
port.
1.3.7 Display or Change Time and Date
To display the date, press the TIME/DATE key once;
press TIME/DATE a second time to display the time.
To set the date, press the TIME/DATE key once. Use the
numeric keypad to enter the date, then press the
ENTER key. The date must be entered in the date
format configured for the indicator: MMDDYY,
DDMMYY, or YYMMDD.
To set the time, press the TIME/DATE key twice. Use
the numeric keypad to enter the time in 24-hour
format, then press the ENTER key.
1.3.8 Display or Change Setpoint Value
To display a setpoint value, use the numeric keypad to
enter the setpoint number, then press the SETPOINT
key. Or, you can display a setpoint value by pressing
the SETPOINT key a number of times equal to the
setpoint number. For example, to display the value of
setpoint 4, press the SETPOINT key four times.
To change the setpoint value, display the current
value, then use the numeric keypad to enter the new
value and press the ENTER key.
NOTE: Some indicator configurations may not allow setpoint
values to be changed through the front panel or may require a
password to display or change the setpoint value.
1.3.9 Turn Setpoint On or Off
To turn a setpoint off at the front panel, use the
numeric keypad to enter the setpoint number, then
press the SETPOINT key (or, press the SETPOINT key a
number of times equal to the setpoint number). With
the correct setpoint displayed, press CLEAR to turn the
setpoint off.
To re-enable a setpoint on that has been turned off at
the front panel, press the SETPOINT key until the
correct setpoint is displayed, then press ENTER to turn
the setpoint back on.
NOTE: Some indicator configurations may not allow setpoints
to be turned off through the front panel or may require a
password to turn the setpoint on and off.
1.3.10 Display or Clear Accumulator
If the accumulator function is enabled, the current net
weight is added to the accumulator each time the
indicator performs a print operation.
• To display the current accumulator value,
press the ACCUM key.
• To clear the accumulator, press ACCUM to
show the current value, then press the CLEAR
key twice to reset the accumulator.
Discontinued Product

4IQ plus 710 Installation Manual
2.0 Installation
This section describes procedures for connecting load
cells, digital I/O, and serial communications cables to
the IQ plus 710 indicator. Instructions for field
installation of the analog output option and
replacement of the CPU board are included, along
with assembly drawings and parts lists for the service
technician.
• Use a wrist strap to ground yourself and protect
components from electrostatic discharge (ESD) when
working inside the indicator enclosure.
• This unit uses double pole/neutral fusing which could
create an electric shock hazard. Procedures requiring
work inside the indicator must be performed by
qualified service personnel only.
• The supply cord serves as the power disconnect for the
IQ plus 710. The power outlet supplying the indicator
must be installed near the unit and be easily accessible.
2.1 Unpacking and Assembly
Immediately after unpacking, visually inspect the IQ
plus 710 to ensure all components are included and
undamaged. The shipping carton should contain the
indicator with attached tilt stand, this manual, and a
parts kit. If any parts were damaged in shipment,
notify Rice Lake Weighing Systems and the shipper
immediately.
The parts kit contains the items listed below:
• Capacity and identification labels.
• Two 8-32NC x 7/16 fillister head screws (PN
30623). These screws occupy the holes above
and on either side of the setup screw on the
indicator backplate (see Figure 2-5 on page 8).
• Ten 8-32NC x 3/8 machine screws (PN 14862)
for the indicator backplate (see #29 in
Figure 2-9 on page 13).
• Twelve bonded sealing washers (PN 45042) for
backplate screws included in the parts kit.
• Four cord grip reducing glands (PN 15664).
• Four rubber bumpers (“feet”) for the tilt stand,
PN 42149.
• 6-position screw terminal (PN 70599) for
connector J1; two 7-position screw terminals
(PN 70600) for connectors J4 and J12; two
10-position screw terminals (PN 70601) for
connectors J7 and J8.
• One kep nut (PN 14676) for grounding cable
shield wires against the backplate.
2.2 Enclosure Disassembly
The indicator enclosure must be opened to connect
cables for load cells, communications, digital inputs,
digital outputs, and analog output.
The IQ plus 710 has no on/off switch.
Before opening the unit, ensure the power
cord is disconnected from the power
outlet.
Ensure power to the indicator is disconnected, then
place the indicator face-down on an antistatic work
mat. Remove the screws that hold the backplate to the
enclosure body, then lift the backplate away from the
enclosure and set it aside.
NOTE: With the backplate removed, check the battery to
ensure it has not been jarred loose from the battery holder
(B1) in shipping.
2.3 Cable Connections
The IQ plus 710 provides five cord grips for cabling
into the indicator: one for the power cord, three to
accommodate communications, digital I/O, and
analog output cables, and one metal cord grip for the
load cell cable. The three nylon cord grips come with
a plug installed to prevent moisture from entering the
enclosure. Depending on your application, remove the
plug from any cord grip that will be used and install
cables as required.
Except for the power cord and load cell cables, all
cables routed through the cord grips should be
grounded to the ground stud on the indicator
backplate.
2.3.1 Load Cells
To attach cable from a load cell or junction box, use
the following procedure to cable through the metal
cord grip:
1. Disconnect indicator from power source.
2. Place indicator face-down on an antistatic work
mat. Remove screws that hold the backplate to
the enclosure body.
3. Loosen cord grips, then route cables through the
stainless steel cord grip. Determine cable length
required to reach connector J1. Mark the cable at
the inside edge of the cord grip (see Figure 2-1
on page 5).
Caution
Warning
Discontinued Product

Installation 5
Figure 2-1. Mark Length of Cable at Inside of Cord Grip
4. Remove cable from cord grip. Strip insulation
from the marked position to the end of the cable.
5. For cables with braided shielding, cut the shield
at the same location as the insulation, then strip
an additional 15 mm of insulation from the cable
to allow the shield to contact the inside of the
metal cord grip (see Figure 2-2).
Figure 2-2. Braided Shield Cabling
For cables with foil-type shielding, strip an
additional 15 mm of insulation from the cable.
Cut the foil shield at a location about 15 mm
inside the enclosure, then fold the foil shield
back on the cable (see Figure 2-3). Ensure silver
(conductive) side of foil is turned outward for
contact with the metal cord grip.
Figure 2-3. Foil Shielded Cabling
6. Cut the shield wire just inside the enclosure (see
Figure 2-3). Shield wire function is provided by
contact with the metal cord grip.
7. Route cables through cord grip so cable shield
contacts cord grip fitting as shown in Figures
2-2 and 2-3. Tighten cord grip.
8. Next, remove connector J1 from the board. The
connector plugs into a header on the board (see
Figure 2-4). Wire the load cell cable from the
load cell or junction box to connector J1 as
shown in Table 2-1.
9. If using 6-wire load cell cable (with sense
wires), remove jumpers JP1 and JP2 before
reinstalling connector J1. For 4-wire installation,
leave jumpers JP1 and JP2 on.
10.Finish installation using cable mounts and ties to
secure cables inside of indicator enclosure.
When connections are complete, reinstall connector J1
onto the header so that it snaps securely into place.
Use two cable ties to secure the load cell cable to the
inside of the enclosure.
Setting the Load Cell Compensation Jumper
The load cell compensation jumper JP5 (near the
transformer on the CPU board; see Figure 2-4) must be
set ON for load cells with unbalanced bridges. The
compensation jumper has the effect of lowering the
positive excitation voltage. Uncompensated unbalanced
load cells can cause instability or calibration errors.
For RL1040 and RL1042 load cells, set the
compensation jumper as follows:
• RL1040 load cells: jumper OFF
• RL1042 load cells: jumper ON
For other load cell types, use the following procedure
to determine the correct jumper position;
1. Disconnect load cell from indicator and use
an ohmmeter to measure the following:
• +EXC to +SIG, +EXC to –SIG
• –EXC to +SIG, –EXC to –SIG
Measured values between the excitation line
and each of the signal lines should be within
2–3Ω.
2. If the +EXC measurements are Š 5% larger than
the –EXC measurements, set the compensation
jumper in the ON position. If the +EXC
measurements are < 5% greater (or are less) than
the –EXC measurements, set the jumper in the
OFF position.
Mark Position
on Cable
Braid
Foil (silver side out)
Shield wire (cut)
J1 Pin Function
1+SIG
2–SIG
3+SENSE
4–SENSE
5+EXC
6–EXC
For 6-wire load cell connections,
remove jumpers JP1 and JP2.
Table 2-1. J1 Pin Assignments
Discontinued Product

6IQ plus 710 Installation Manual
Figure 2-4. IQ plus 710 CPU and Power Supply Board, Version 2.1
LOCATION OF INSTALLED ANALOG OUTPUT MODULE
100
76
75
51
50 26
25
1
U22
U23
U17
C96
JP4
J9
J6
1
U19
J15
3
6
7
8
9
1011
12
14
15
T1
J4
B1
J10
C100
C104
U16
EPROM
J8
U18
J7
GND
RS485-B
RS485-A
GND
+5V
+I
–I
7
6
5
4
3
2
1
SERIAL COMMUNICATIONS - 2
+20 mA
–20 mA
Prn RxD
Prn TxD
EDP RxD
GND
EDP TxD
7
6
5
4
3
2
1
SERIAL COMMUNICATIONS - 1
J12
–EXC
+EXC
–SENSE
+SENSE
–SIG
+SIG
654321
LOAD CELL CONNECTOR
F2F1
P2
JP3
JP1 JP2
J5
ANALOG OUTPUT
KEYPAD CONNECTOR
BLUE WIRE
BROWN WIRE
TO LINE FILTER
DI6
DI5
DI4
DI3
DI2
DI1
GND
7
6
5
4
3
2
1
10
9
8
GND
DI8
DI7
DIGITAL INPUT
DO6
DO5
DO4
DO3
DO2
DO1
GND
7
6
5
4
3
2
1
10
9
8
+5V
DO8
DO7
DIGITAL OUTPUT
4
3
2
1
J1
J1
JP5
LOAD
CELL
COMP
ON
OFF
A/D Converter
Discontinued Product

Installation 7
2.3.2 Serial Communications
To attach serial communications cables, remove
connector J4 or J12 from the board. Connector J4
provides connections for the EDP (Electronic Data
Processing) port, printer port, and 20 mA current loop
transmit signals; connector J12 provides RS-485
signals. Table 2-2 shows the pin assignments for
connectors J12 and J4.
Once cables are attached, reconnect J12 or J4 to the
header on the board. Use cable ties to secure serial
cables to the inside of the enclosure.
The EDP port supports RS-232 or RS-485
communications; the printer port provides active 20
mA output and full-duplex RS-232 communications.
Both ports are configured using the SERIAL menu.
See Section 3.0 on page 14 for configuration
information.
2.3.3 Digital I/O
Digital inputs can be set to provide many indicator
functions, including all keypad functions. Digital
inputs are active (on) with low voltage (0 VDC),
inactive (off) at 5 VDC. Use the DIG IN menu to
configure the digital inputs.
Digital outputs are typically used to control relays that
drive other equipment. Outputs are designed to sink,
rather than source, switching current. Each output is a
normally open collector circuit, capable of sinking
250 mA when active. Digital outputs are wired to
switch relays when the digital output is active (low, 0
VDC) with reference to a 5 VDC supply.
Up to eight relays can be mounted inside the flat front
enclosure; up to four relays can be mounted inside the
sloped front enclosure. Use the SETPTS menu to
configure digital outputs.
Table 2-3 shows the pin assignments for connectors J7
and J8.
2.3.4 Analog Output
If the optional analog output module is installed,
attach the output cable to connector J1 on the analog
output board. Table 2-4 lists the analog output pin
assignments.
Use the ALGOUT menu to configure and calibrate the
analog output module when cabling is complete. See
Section 2.4 for information about installing the analog
output module.
2.4 Analog Output Module Installation
To install or replace the analog output module, follow
the steps listed in Section 2.2 on page 4 for opening
the IQ plus 710 enclosure.
Mount the analog output module on its standoffs in
the location shown in Figure 2-4 on page 6 and plug
the module input into connector J5 on the IQ plus 710
board. Connect the output cable to the analog output
module as shown in Table 2-4, then reassemble the
enclosure (Section 2.5).
See Section 10.7 on page 71 for analog output
calibration procedures.
Connector Pin Signal
J4 1EDP TxD
2GND
3EDP RxD
4Printer TxD
5Printer RxD
6Printer –20 mA TxD
7Printer +20 mA TxD
J12 1Reserved
2Reserved
3+5V
4GND
5RS485-A
6RS485-B
7GND
Table 2-2. J4 and J12 Pin Assignments
Pin J7 Signal J8 Signal
1GND GND
2DI1 DO1
3DI2 DO2
4DI3 DO3
5DI4 DO4
6DI5 DO5
7DI6 DO6
8DI7 DO7
9DI8 DO8
10 GND +5V
Table 2-3. J7 and J8 Pin Assignments (Digital I/O)
Pin Signal
1+ Current Out
2– Current Out
3+ Voltage Out
4– Voltage Out
Table 2-4. Analog Output Module Pin Assignments
Discontinued Product

8IQ plus 710 Installation Manual
2.5 Enclosure Reassembly
Once cabling is complete, position the backplate over
the enclosure and reinstall the backplate screws. Use
the torque pattern shown in Figure 2-5 to prevent
distorting the backplate gasket. Torque screws to 10
in-lb (1.13 N-m).
Figure 2-5. IQ plus 710 Enclosure Backplate
2.6 Board Removal
If you must remove the IQ plus 710 CPU board, use
the following procedure:
1. Disconnect power to the indicator. Loosen
cord grips and remove backplate as described
in Section 2.2 on page 4.
2. Unplug connectors J1 (load cell cable), J4 and
J12 (serial communications), J7 and J8 (digital I/
O), J6 (keypad ribbon cable), and JP4 (setup
switch). If an analog output board is installed,
disconnect the analog output cable. See
Figure 2-4 on page 6 for connector locations.
3. Remove the standoff and three nuts from the
corners of the CPU board.
4. Cut the cable tie that holds the line filter load
wires to the enclosure.
5. Lift the board off of its spacers just far enough to
access the setscrews that secure the line filter
load wires at connector P2. Use a small
screwdriver to loosen the setscrews and
disconnect power to the board.
6. Remove the CPU board from the enclosure.
To replace the CPU board, reverse the above
procedure. Be sure to reinstall cable ties to secure all
cables inside the indicator enclosure.
2.7 Battery Replacement
The lithium battery on the CPU board maintains the
real-time clock and protects data stored in the system
RAM when the indicator is not connected to AC
power.
System RAM data includes prompts, truck ID storage,
and keyboard locks. This information is lost if the
battery loses power and the indicator is disconnected
from AC power. To prevent loss of data, do the
following:
• Periodically check the battery voltage and
replace when the voltage drops below 3.1 VDC.
The battery should last a minimum of one year.
Therefore, in critical applications, replace the
battery every twelve months.
• Use the Revolution™configuration utility or
EDP commands (see Section 5.2 on page 48) to
store a copy of the indicator configuration on a
PC before attempting battery replacement. If
any data is lost, the indicator configuration can
be restored from the PC.
Risk of explosion if battery is replaced with
incorrect type. Dispose of batteries per
manufacturer instruction.
Setup switch access screw
Fillister head screws
1
Torque backplate screws
to 10 in-lb (1.13 N-m)
23
4
56
7
8
9
10
11 16
12
13
14
15
Caution
Discontinued Product

Installation 9
2.8 Replacement Parts
Table 2-5 lists replacement parts for the IQ plus 710, including all parts referenced in Figures 2-6 through 2-10.
Ref
Number PN Description (Quantity) Figure
141397 Enclosure, sloped front (1) Figure 2-9 on page 13
41401 Enclosure, flat front (1)
241398 Enclosure backplate (1) Figure 2-6 on page 10
314626 Kep nuts, 8-32NC hex (13) Figure 2-9 on page 13
430375 Nylon seal rings for cable grips (3) Figure 2-6 on page 10
514621 Kep nuts, 6-32NC hex (4–flat enclosure; 6–sloped) Figure 2-8 on page 12
615626 Cable grips, PG9 (3) Figure 2-6 on page 10
715627 Locknuts, PCN9 (3)
815650* Cable tie mounts (8) Figure 2-7 on page 11
10 19538 Cable grip plugs (3) Figure 2-6 on page 10
11 44676 Sealing washer for setup switch access screw (1)
12 42640 Setup switch access screw, 1/4 x 28NF x 1/4 (1)
13 41965 Power cord assembly, 115VAC (1)
45254 Power cord assembly, 230VAC (1)
15 16892 Ground/Earth label (1) Figure 2-7 on page 11
16 45402 Bezel, sloped front (1) Figure 2-9 on page 13
41399 Bezel, flat front (1)
17 41386 Switch panel membrane (1) Figure 2-9 on page 13
18 41400 Backplate gasket (1) Figure 2-6 on page 10
19 45043 Ground wire, 4 in w/ No. 8 eye connector (1) Figure 2-7 on page 11
21 46027 Setup switch mounting bracket (1) Figure 2-8 on page 12
24 44844 Setup switch assembly (1)
25 68403 Wing knobs for tilt stand (2) Figure 2-10 on page 13
26 29635 Tilt stand (1)
27 15144 Nylon washers for tilt stand, 1/4 x 1 x 1/16 (2)
28 45891 Line filter assembly (1) Figure 2-7 on page 11
29 14862* Screws, 8-32NC x 3/8 (4) Figure 2-10 on page 13
30 16903 Model/serial number label (1) —
31 46252 Bezel gasket, sloped front (1) Figure 2-9 on page 13
45076 Bezel gasket, flat front (1)
36 45401 CPU board mounting tab, sloped front models (1) Figure 2-8 on page 12
37 15134 Lock washers, No. 8 (4) Figure 2-7 on page 11
38 48027 Nylon spacers for board mounting (3) Figure 2-8 on page 12
39 45042* Sealing washers (4) Figure 2-10 on page 13
40 15369 Standoffs, fem 6-32NC x 3/4 (3) Figure 2-7 on page 11
41 64956 Display and CPU board assembly, 115 VAC (1) Figure 2-8 on page 12
64960 Display and CPU board assembly, 230 VAC (1)
—40698 VFD display (1)
42 19644 3V cylindrical lithium battery
Table 2-5. Replacement Parts
Discontinued Product

10 IQ plus 710 Installation Manual
Figure 2-6. Backplate Assembly
45 65981 Brass spacer (1) Figure 2-8 on page 12
46 15130 Internal lock washers (2)
49 50959 Metal cord grip, PG9 (1) Figure 2-6 on page 10
50 50962 Metal cord grip nut, PG9 (1)
51 73769 Grounding clips (4)
—70600 7-position connectors for J4, and J12 (2) Figure 2-4 on page 6
—70599 6-position connector for J1 (1)
—70601 10-position connectors for J7 and J8 (2)
—45484 160 mA TR5 subminiature fuses (2), 115 VAC F1 and F2 in Figure 2-4 on page 6
45107 80 mA TR5 subminiature fuses (2), 230 VAC
* Additional parts included in parts kit.
To protect against the risk of fire, replace fuses only with same type and rating fuse.
See Section 10.10 on page 74 for complete fuse specifications.
Ref
Number PN Description (Quantity) Figure
Table 2-5. Replacement Parts (Continued)
Caution
13
10/3X
6/3X
12
11
4/3X
2
7/3X
18
GROUND
STUDS
115V/ 230V PLUG
49
50
51/4X
Discontinued Product

Installation 11
Figure 2-7. Enclosure and Line Filter Assembly
28
D
D
8/6X/D
19
B
37/4X
15
C
T
O BACKPLATE
GROUND STUD
TO CPU BOARD P2
POWER CORD
BLUE
BROWN
CABLE TIE
Discontinued Product

12 IQ plus 710 Installation Manual
Figure 2-8. Enclosure and CPU Board Assembly
40/3X/C
5/6X/A
42
41
38/4X
21
24
A
A
36
TO JUMPER JP4
SLOPED FRONT MODELS ONLY
Discontinued Product

Installation 13
Figure 2-9. Bezel Assembly
Figure 2-10. Tilt Stand Assembly
3/13X/B
44/3X
1
17
31
16
29/4X
39/4X
30
27/2X
25/2X
26
Discontinued Product

14 IQ plus 710 Installation Manual
3.0 Configuration
To configure the IQ plus 710 indicator, the indicator
must be placed in setup mode. The setup switch is
accessed by removing the large fillister head screw on
the enclosure backplate. Switch position is changed
by inserting a screwdriver into the access hole and
pressing the switch.
When the indicator is placed in setup mode, the word
CONFIG is shown on the display. The CONFIG menu
is the first of ten main menus used to configure the
indicator. Detailed descriptions of these menus are
given in Section 3.2. When configuration is complete,
return to the CONFIG menu and press the (ZERO)
key to exit setup mode, then replace the setup switch
access screw.
3.1 Configuration Methods
The IQ plus 710 indicator can be configured by using
the front panel keys to navigate through a series of
configuration menus or by sending commands or
configuration data to the EDP port. Configuration
using the menus is described in Section 3.1.3.
Configuration using the EDP port can be
accomplished using either the EDP command set
described in Section 5.0 or Version 2.5 or later of the
Revolution™configuration utility.
3.1.1 Revolution Configuration
The Revolution configuration utility provides the
preferred method for configuring the IQ plus 710
indicator. Revolution runs on a personal computer to
set configuration parameters for the indicator. When
Revolution configuration is complete, configuration
data is downloaded to the indicator.
Figure 3-1. Sample Revolution Configuration Display
Revolution supports both uploading and downloading
of indicator configuration data. This capability allows
configuration data to be retrieved from one indicator,
edited, then downloaded to another.
To use Revolution, do the following:
1. Install Revolution (Version 3.0 later) on an
IBM-compatible personal computer running
Windows®98 or later.
Minimum system requirements include a
processor speed of at least 133MHz, 32MB of
memory (64MB recommended), and at least
20MB of available hard disk space for
installation.
2. With both indicator and PC powered off,
connect the PC serial port to the RS-232 pins on
the indicator EDP port.
3. Power up the PC and the indicator. Use the setup
switch to place the indicator in setup mode.
4. Start the Revolution program.
Figure 3-1 shows an example of one of the Revolution
configuration displays.
Revolution provides online help for each of its
configuration displays. Parameter descriptions
provided in this manual for front panel configuration
can also be used when configuring the indicator using
Revolution: the interface is different, but the parameters
set are the same.
3.1.2 EDP Command Configuration
The EDP command set can be used to configure the
IQ plus 710 indicator using either a personal
computer, terminal, or remote keyboard. Like
Revolution, EDP command configuration sends
commands to the indicator EDP port; unlike
Revolution, EDP commands can be sent using any
external device capable of sending ASCII characters
over a serial connection.
EDP commands duplicate the functions available
using the indicator front panel and provide some
functions not otherwise available. EDP commands can
be used to simulate pressing front panel keys, to
configure the indicator, or to dump lists of parameter
settings. See Section 5.0 on page 41 for more
information about using the EDP command set.
Discontinued Product

Configuration 15
3.1.3 Front Panel Configuration
The IQ plus 710 indicator can be configured using a series of menus accessed through the indicator front panel
when the indicator is in setup mode. Table 3-1 summarizes the functions of each of the main menus.
Figure 3-2. Five-Key Keypad Functions in Setup Mode
Four front panel keys are used as directional keys to navigate through the menus in setup mode (see Figure 3-2).
The UNITS ( ) and PRINT () keys scroll left and right (horizontally) on the same menu level; ZERO ( ) and
GROSS/NET ( ) move up and down (vertically) to different menu levels. The TARE key ( ) serves as an Enter key
for selecting parameter values within the menus. A label over each of these keys identifies the direction provided
by the key when navigating through the setup menus.
Menu Menu Function
CONFIG Configuration Configure grads, zero tracking, zero range, motion band, overload, A/D sample rate, tare
function, power-up mode, analog and digital filtering parameters.
FORMAT Format Set format of primary and secondary units, decimal format, and display rate.
CALIBR Calibration Calibrate indicator. See Section 4.0 for calibration procedures.
SERIAL Serial Configure EDP and printer serial ports.
PROGRM Program Set date and time formats, truck mode, passwords, keyboard locks, regulatory mode, and
initial consecutive number value; enable accumulator; define setpoint and macro prompts, and
program macros.
PFORMT Print Format Set print format used for header, gross, net, truck in/out, setpoint, and EDP format tickets. See
Section 6.0 for more information.
SETPTS Setpoints Configure setpoints and batching mode.
DIG IN Digital Input Assign digital input functions.
ALGOUT Analog Output Configure analog output module. Used only if analog output option is installed.
VERSION Version Display installed software version number.
Table 3-1. IQ plus 710 Menu Summary
ZERO GROSS
NET
TARE UNITS PRINT
B/N
Units
Move LEFT /
Previous
Move RIGHT /
Next
Move UP /
Increment Value
Move DOWN /
Decrement Value
ENTER Value
SETUP MODE 5-KEY FUNCTIONS
Discontinued Product

16 IQ plus 710 Installation Manual
Figure 3-3. Setup Mode Menu Navigation
To select a parameter, press or to scroll left or
right until the desired menu group appears on the
display, then press to move down to the submenu or
parameter you want. When moving through the menu
parameters, the default or previously selected value
appears first on the display.
To change a parameter value, scroll left or right to
view the values for that parameter. When the desired
value appears on the display, press to select the
value and move back up one level. To edit numerical
values, use the numeric keypad on the indicator front
panel.
3.2 Menu Structures and Parameter Descriptions
The following sections provide graphic representations of the IQ plus 710 menu structures. In the actual menu
structure, the settings you choose under each parameter are arranged horizontally. To save page space, menu
choices are shown in vertical columns. The factory default setting appears at the top of each column. Parameters
shown surrounded by a dotted-line box only appear under the special circumstances explained under each box.
Most menu diagrams are accompanied by one or more tables that describe all parameters and parameter values
associated with that menu option. Default parameter values are shown in bold type.
1
st
Level
Parameter
Default value
2
nd
Level
Parameter
Value Value Value
1
st
Level
Parameter
2
nd
Level
Parameter
When moving through values below the first menu level, press to return to the level
above. Press or to move to the next parameter on the level above.
Discontinued Product
Other manuals for IQ plus 710
2
Table of contents
Other Rice Lake Monitor manuals

Rice Lake
Rice Lake LaserLight User manual

Rice Lake
Rice Lake 1280 Enterprise Series Installation manual

Rice Lake
Rice Lake LaserLight2 ConnexLink 167091 User manual

Rice Lake
Rice Lake CLS-680 User manual

Rice Lake
Rice Lake LaserLT-60 User manual

Rice Lake
Rice Lake MSI-8000HD User manual

Rice Lake
Rice Lake LaserLight3 User manual

Rice Lake
Rice Lake LaserLight3 User manual

Rice Lake
Rice Lake SURVIVOR IQ F Series User manual

Rice Lake
Rice Lake LaserLight3 User manual