
12 QB Series Operation Manual
2.9 Operating the Counting Function
The QB Series balance is capable of functioning as a counting device, enabling it to determine quantity as
opposed to weight. In order to perform this function, the balance must be programmed with the average piece
weight of the sample being evaluated. Once programmed, the balance performs the following calculation to
determine the total number of pieces to display:
Weight on platform - tare weight (if applicable) / average piece weight = displayed piece count
NOTE: When the counting mode is activated, the comparator function is automatically deactivated. Should you desire to
return to the comparator function, all of the most recently entered parameters are automatically recalled when the
counting function is deactivated. If you need to return to the counting mode at a later time, the last average piece weight
utilized is still active.
2.9.1 Counting Function Activation
Prior to using the counting function, the parameters for the operation must be programmed into the balance.
1. Power up the balance (see Section 1.7 on page 5).
2. The balance displays the default weighing mode. If the balance does not show a 0.00 value after the
normal start sequence, press and release the Tkey.
3. Press and hold the Fkey until the word Func is displayed.
4. Release the Fkey. The display changes to show either 1.SET.1 or 1.SET.2. If the display shows 1.SET.2,
continue below with step 5. If the display shows 1.SET.1, press the Tkey once to change the value to
1.SET.2 and then continue with step 5 below.
5. Press and release the Pkey once. The display advances to the weighing mode with a 0value and the P
(counting pieces) indicator illuminated.
2.9.2 Programming the Average Piece Weight
To program the average piece weight, use the steps outlined below.
NOTE: The QB series balance can determine the average piece weight by evaluating initial samples of 10, 30, 50, or 100
pieces. We strongly recommend that you always use a sample of 100 pieces to ensure the most uniform sample piece
calculation resulting in the most accurate counting operation possible based on the samples provided. It is important to
remember that an accurate counting function relies on a uniformity of parts being counted. If there is a large variance
from piece to piece, that error is introduced into the counting function with a resultant error in the value displayed by the
balance.
1. Press and release the Tkey once to ensure that the value on the display is 0.
2. If you desire to use a container to hold the parts to be counted, place the empty container on the weighing
platform.
3. The display may change to indicate a counting value. Press and release the Tkey to return the display to
a 0value.
NOTES:
•The tare value must be ≥+/- 4 display divisions from zero. If the tare value is ≤+/- 3 divisions from
zero, the balance automatically re-establishes a new zero setting before the T button can be
activated. The tare function is not available. To disable this “auto-zero” feature, please refer to
Section 4.0 on page 15.
•Whenever the tare function is activated on a QB balance, the effective weighing range of the device
is reduced by the value of the tare. For example, on a QB-600 with a total capacity of 600g, if a tare
of 100g is entered into the device, the total effective NET weighing range of the balance is only 500g
(total rated capacity [600g] - tare value [100g] = 500g. Any attempt to weigh over the effective
weighing range results in a “o-Err” error message.
4. Press and hold the Fkey until the display reads U Set. Then release the Fkey.
5. The display shows on xxx where “xxx” is either 10, 30, 50, or 100 depending on what value was last used
to determine an average piece weight.
6. Repeatedly press and release the Tkey until the display reads on 100 (recommended) or the quantity you
desire to use as a sample.
7. With the display flashing the number of pieces you have chosen to use as a sample, place those items on
the weighing platform or in the empty container.
8. Press and release the Fkey once. The display momentarily blanks, except for the flashing Mindicator
which confirms that the balance is memorizing the average piece weight.