4) Sample Size Example
4
101-3GM555-XXX
EN
Bionime GmbH
Tramstrasse 16
9442 Berneck
Switzerland
4
0.75 µL 1.0 µL 1.5 µL 2.0 µL 3.0 µL
13
2
8
12
9 10 11
5
4) Twist off and set aside the protective cover of the disposable lancet.
5) Replace the depth adjustable cap.
6) Choose a depth of penetration by rotating the top portion of the depth adjustable cap until your
desired setting depth is visible in the window. Settings are based on skin type:
" " for soft or thin skin;
" " for average skin;
" " for thick or calloused skin.
7) Hold the hub in one hand and pull on the plunger with the other hand. The device will be cocked.
Release the plunger and it will automatically move back to its original position near the hub.
4675
1) Wash and dry your hands. Take one strip from the vial. Close the vial cap immediately.
2) Insert the strip into the strip port on the meter with the indication symbol facing up. Push the strip in
until it clicks and stops. The meter will automatically detect the code number.
3) When you see the flashing blood drop icon on the display window, hold the lancing device to the
side of your fingertip and press the release button.
4) The meter only requires a small blood sample. A minimum of 0.75 µL of blood is needed to test.
Blood sample size above 3.0 µL might contaminate the meter. Not enough blood (<0.75 µL) might
cause inaccurate results.
5) Touch and hold the drop to the edge of sample entry until you hear a "beep" and the view window is filled
with blood. If the view window is not completely filled with blood the test will not start. Discard the test
strip and repeat the test with a new test strip.
6) The countdown mode will begin on the display window. After 5 seconds, your test result will appear. For
more information about your test results, please see your User Manual.
For more information on how to use your meter and understand your test results, please see the
User’s Manual.
Alternative site testing: palm or forearm blood sampling
1. Choose the clear cap and follow step 1 through step 3
"Performing a Test".
2. Massage the puncture area of your palm or forearm for a few
seconds.
3. Immediately after massaging the puncture area, press and
hold the lancing device with the clear cap against your palm or
forearm.
4. Press the release button.
5. Continue holding the lancing device against your palm or
forearm and gradually increase pressure for a few seconds
until the blood sample size is sufficient.
6. Follow step 8 to discard the used disposable lancet.
Releasing
button
7) Remove the test strip from the meter. Please follow your local regulations to discard the used strip
properly.
8) Pull off the depth adjustable cap. Without touching the used disposable lancet, insert the lancet tip into
the protective cover.
9) Hold the release button in one hand and pull on the plunger with the other hand to safely eject the used
disposable lancet.
10) Discard the used disposable lancet into an appropriate puncture-proof or biohazard container.
11) Replace the depth adjustable cap after finishing the test.
3a 3b
45
7
BIONIME CORPORATION
No. 100, Sec. 2, Daqing St., South Dist.,
Taichung City 40242, Taiwan
Tel: +886 4 23692388
Fax: +886 4 22617586
http: //www.bionime.com
6
Rotate
Performing a Test
Pressing for
a few seconds
Palm
Forearm
Rev. Date: 2019-10