manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Rightest
  6. •
  7. Blood Glucose Meter
  8. •
  9. Rightest GT333 User manual

Rightest GT333 User manual

Performing a Test
1) Take one test strip from the vial. Close the vial cap immediately.
2) Insert the strip into the strip port of the meter with the sample window facing up. Push the strip in until it clicks and stops. The meter will
automatically detect the code number.
3) When the blood drop icon flashes on the display window, blood sample is ready to be applied to the test strip port (apply within 2
minutes).
4) Place the lancing device to your fingertip and press the release button.
5) Touch and hold the drop to the edge of sample entry until you hear a " beep " (if volume is turned on) and the View Window is totally
filled with blood. If the View Window is not totally filled with blood or the test does not start, please discard the test strip and repeat the
test with a new test strip.
GT333 Blood Glucose Test Strip INSERT
Intended Use
RIGHTEST GT333 Blood Glucose Test Strip is used by individuals with diabetes. It's for checking on glucose levels of whole
blood from capillary. Capillary blood can be sampled from the fingertip, palm and forearm and it provides results equivalent to
a laboratory instrument (Plasma equivalent). RIGHTEST GT333 Blood Glucose Test Strip aid to diabetes control, could be
used by healthcare professionals in clinical setting, also by people to use at home.
RIGHTEST GT333 Blood Glucose Test Strip is intended for self-testing outside the body (in vitro diagnostic use) by people with
diabetes at home as an aid to monitor the effectiveness of diabetes control. RIGHTEST GT333 Blood Glucose Test Strip
should not be used for the diagnosis of, or screening for diabetes or for neonatal use. Alternative site testing should be done
only during steady - state times (when glucose is not changing rapidly).
RIGHTEST GT333 Blood Glucose Test strip is designed for use with RIGHTEST GT300 or GT333 Blood Glucose Meter to
obtain accurate results.
Please use a minimum blood drop size of 0.75 to perform µL
the test on glucose monitoring system. Blood drop sizes
greater than 3.0 are too large and may contaminate the test µL
strip port of the meter.
Sample Size Example
Alternate site testing: palm or forearm blood sampling
- To perform a test using samples obtained from alternative sites, install the
clear cap on the lancing device (For more information on how to install,
see the Instructions for the lancing device).
- To increase the blood flow, massage the puncture area of palm or
forearm for a few seconds.
- Immediately after massaging the puncture area, press and hold the
lancing device with the clear cap against palm or forearm.
- Then press the release button.
- Continue holding the lancing device against palm or forearm and
gradually increase pressure for a few seconds until the blood sample size
is sufficient ( Refer to Instructions for the lancing device ).
0.75 µL 1.0 µL µL µL 3µL 1.5 2.0 .0
6) Touch and hold the drop to the edge of sample port until you hear a " beep " (if volume is turned on) and the sample
window is completely filled with blood. If the view window is not completely filled with blood the test will not start. If this is
the case, discard the test strip and repeat the test with a new test strip.
7) The countdown mode will begin on the display window. After 5 seconds your test result will appear.
8) Remove the test strip from the meter. Please follow your local regulations to properly discard the used strip.
9) To remove the lancet, pull off the depth adjustable cap of the lancing device. Without touching the used disposable lancet,
insert the lancet tip into the protective cover. Hold the release button of the lancing device in one hand and pull on the
plunger with the other hand to safely eject the used disposable lancet into an appropriate puncture-proof or biohazard
container.
For more information on how to use your meter and understand your test results, see the User’s Manual.
6
7
8
Precautions
- Check the expiration date printed on the test strip vial. Do not use expired test strips.
- Close the test strip vial immediately after taking test strip out from the vial.
- Do not perform a control solution test with expired control solution.
- Do not bend or twist the test strip. A damaged test strip may cause incorrect test results.
- Do not reuse test strips and lancets.
- Discard the used disposable lancet and strip into an appropriate puncture-proof or biohazard container.
- If RIGHTEST Meter and Test Strips is exposed to a significant change in temperature, please wait at least 45 minutes before
performing a test.
- If you want to purchase new control solutions, please contact Customer Service.
- Test results may vary if blood samples are taken from different sites or under certain conditions where glucose levels can
change rapidly such as: following a drink, a meal, an insulin dose or exercise. In these cases, only the fingertip should be
used.
- Users should wash their hands thoroughly with soap and water after handling the meter, lancing device, control solution and
test strips.
Test Result
- Blood glucose test results are shown on the meter as mg/dL or mmol/L, depending on the preset of your meter.
- If your blood glucose result is unusually high or low, or if you question your results, repeat the test with a new test strip. You
can also run a Quality Control Test with the RIGHTEST Control Solution GC570 to check your meter and test strip. If the test
result still remains unusually high or low, contact your Physician or the nearest Emergency Healthcare Services immediately.
- If you are experiencing symptoms that are not consistent with your blood glucose test results and you have followed all the
instructions in this insert, contact your Physician or the nearest Emergency Healthcare Services immediately.
- RIGHTEST Meter display results between 10 and 600 mg/dL. If your test result is below 10 mg/dL,
" Lo " will appear on the screen. Please repeat your test with a new strip. If you still get a " Lo " result, you should immediately
contact your Physician or the nearest Emergency Healthcare Services.
- If your test result is above 600 mg/dL, " Hi " will appear on the screen. Please repeat your test with a new strip. If you still get
a " Hi " result, you should immediately contact your Physician or the nearest Emergency Healthcare Services.
(1)
Expected glucose values without diabetes
Releasing
button
Limitations
- The meter readings of the blood glucose may be significantly lower than " true glucose levels " in a hyperglycemic-
hyperosmolar state, with or without ketosis.
- Caution is advised when glucose values in the interpretation of glucose values are below 50 mg/dL or above 250 mg/dL .
Consult a Physician as soon as possible, if values in this range are obtained.
- Healthcare professionals should evaluate their technique and their patients' technique regarding the use of RIGHTEST
Blood Glucose Monitoring System regularly . To accomplish this, it is recommended that blood glucose monitoring results
be compared with a concurrently obtained laboratory measurement on the same blood sample. A proven clinical
laboratory method employing hexokinase or glucose oxidase should be used as the comparative method.
- Hands and fingers contaminated with sugar from foods or beverages may cause falsely elevated results.
- Flouoride should not be used as a preservative when collecting blood glucose samples.
- Storage of strips near bleach as well as bleach containing products will affect the results of RIGHTEST Blood Glucose T est
strip.
- Inaccurate test results may be obtained at altitudes greater than 10,000 feet (3,048 meters) above sea level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- Do not perform the blood glucose test at temperatures below 10°C (50°F) or above 40°C (104°F), nor below 10% or above
90% relative humidity.The suggested temperature range for the control solution test is 15 - 40 °C (59 - 104 °F).
- RIGHTEST Blood Glucose Test strip are designed for use with capillary whole blood samples. Do not use serum or plasma
samples.
Warning
- Keep the test strips and vial cap away from children; they may cause a choking hazard. If a test strip or vial cap is swallowed,
contact your Physician immediately.
- Please refer to the cleaning and disinfecting instructions in the User's Manual.
Preparing the Lancing Device
Wash your hands with warm soapy water and dry thoroughly before you start the test.
1) Hold the depth adjustable cap in one hand and hold the hub in the other hand. Bend the cap towards the down side, until a
gap appears between the cap and hub.
2) Pull the cap and hub off in opposite directions, remove the cap.
3) Insert a new disposable lancet firmly into lancet carrier .
4) Twist off and set aside the protective cover of the disposable lancet.
5) Replace the depth adjustable cap.
6) The adjustable cap with 7 depth levels allows you to select the depth of penetration by rotating the cap until the preferable
depth display in the window. Settings are based on skin type " " for soft or thin skin; " " for average skin;
" " for thick or calloused skin.
7) Hold the hub in one hand and pull on the plunger in the other hand. The device will be cocked. Release the plunger, it will
automatically move back to its original position near the hub.
Test Procedure
GLUCOSE LEVEL
From 70 to 99 mg/dL (3.9 to 5.5 mmol/L)
From 100 to 125 mg/dL (5.6 to 6.9 mmol/L)
126 mg/dL (7.0 mmol/L) and above on more than one testing occasion
INDICATION
Normal fasting glucose
Pre-diabetes (Impaired fasting glucose)
Diabetes
Fasting Blood Glucose
7654
3
1 2
2
4
5
3
9
12
Pressing for
a few seconds
Forearm
Palm
1
TM
Rotate
101-3GS570-XXX
EN
- Not for screening or diagnosis of diabetes mellitus.
- Not for use on critically ill patients, patients in shock, dehydrated patients or hyper-osmolar patients.
- Not for neonatal use
- Alternate site testing should not be used to calibrate continuous glucose monitoring systems (CGMs).
- Alternate site testing should only be done during steady-state times (when glucose is not changing rapidly).
- Results from alternate site testing should not be used in insulin dose calculation.
Storage and Handling
- Store the strips in the original capped vial at temperatures between 4°C to 30°C (39°F to 86°F) and 10% - 90% relative
humidity.
- Replace vial cap immediately and close cap tightly after removing a test strip. If the strip is exposed to the air too long, it will
absorb the moisture and cause inaccurate test results.
- When you open a new vial of test strips please write the opening date on the label. Use test strips within 12 months after first
opening or until the expiration date printed on the label (whichever comes first).
Quality Control
Please refer to the Quality Control section of the User’s Manual.
Troubleshooting and Customer Service
For more information on error messages and troubleshooting, please refer to the Error Messages and Troubleshooting section
of RIGHTEST User’s Manual.
You may also contact your local distributor for customer service or or visit our website at www.bionime.com. (Off-work hour
you could contact your healthcare professional for help)
NOTE
- Keep meter free of dust and liquids including water.
- Suggest not to use this meter close to source of strong electromagnetic radiation, to avoid interference with proper
operation.
(2)
Detection Principle
The glucose oxidase and potassium ferricyanide in the strip react with the glucose in the sample to produce an electrical
current which is proportional to the amount of glucose in the sample. The meter measures the current and converts it to the
corresponding glucose concentration.
Performance Characteristics
Data generated using Rightest GT333 Meter, this meter is the representative model of RIGHTEST GT300 and GT333 Meter.
Precision
The precision was evaluated with (i) venous whole blood sample-the blood samples are collected over a span of time not to
exceed one day per meter and reagent lot combination. (ii) 3 levels glucose control solutions in period of 10 days, by 10
meters and 3 batches of strips.
Accuracy
For the alternate site testing:
The accuracy of Rightest GT333 Blood Glucose Monitoring System was tested by comparing fingertip whole blood (plasma
equivalent) glucose values measured by Rightest GT333 Meter with plasma glucose values obtained from a YSI 2300
(Calibrated with NIST 917c) reference instrument.
The result is shown in the tables below.
(i) venous whole blood sample
(ii) Control solution
Sample
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV (%)
P-01
300
46.4 (2.6)
1.4 (0.08)
3.0 %
P-02
300
102.8 (5.7)
2.0 (0.11)
2.0%
P-03
300
123.8 (6.9)
2.0 (0.11)
1.6%
P-04
300
221.1 (12.3)
3.8 (0.21)
1.7%
P-05
300
332.8 (18.5)
5.7 (0.32)
1.7%
Glucose levels
(1) Total test numbers (n)
(2) Mean mg/dL (mmol/L)
(3) SD mg/dL (mmol/L)
(4) CV (%)
CS-L
300
46.0 (2.6)
1.2 (0.07)
2.6%
CS-N
300
118.2 (6.6)
1.8 (0.10)
1.5%
CS-H
300
298.5 (16.6)
3.8 (0.21)
1.3%
Additional Information for Healthcare Professionals
BIONIME CORPORATION
No. 100, Sec. 2, Daqing St., South Dist.,
Taichung City 40242, Taiwan (R.O.C)
Made in Taiwan
Tel: +886 4 23692388
Fax: +886 4 22617586
E-mail: info@bionime.com
http://www.bionime.com
Bionime GmbH
Tramstrasse 16
9442 Berneck
Switzerland
E-mail: [email protected]
Note: * Acceptance criteria in ISO 15197 : 2013 are that 95% of all differences in glucose values should be within ± 15 mg/dL
(0.83 mmol/L) at glucose concentrations < 100 mg/dL (5.55 mmol/L), and within ± 15% at glucose concentrations ≧ 100
mg/dL (5.55 mmol/L).
Note: For glucose concentrations < 100 mg/dL (5.55 mmol/L), difference values are expressed in mg/dL (mmol/L), and for
glucose concentrations ≧ 100 mg/dL (5.55 mmol/L), difference value are compared in percentage.
Lay User Evaluation
A total of 106 users were enrolled. Each user tested their fingertip blood samples with 3 lots of RIGHTEST GT333 Blood
Glucose Test strip and RIGHTEST GT333 Blood Glucose Meter. Then the professional collected blood samples were
centrifuged immediately after collection to obtain plasma. Analyze the plasma by the lab instrument (YSI 2300 analyzer). 100%
of RIGHTEST GT333 BGMS values were within ± 15% of YSI values at glucose concentrations ≧ 100 mg/dL and within ± 15
mg/dL at glucose concentrations < 100 mg/dL.
Reagents
Each Blood Glucose Test Strip contains the following reagents:
Glucose Oxidase (Aspergillus niger) (GOD)
Potassium ferricyanide
Non-reactive ingredients
References
1) Diabetes Information - American Association for Clinical Chemistry (AACC)〔Electronic Version〕Retrieved Mar. 02,2018
from www.labtestsonline.org/understanding/analytes/glucose/test.html
2) Review of Glucose Oxidases and Glucose Dehydrogenases: A Bird's Eye View of Glucose Sensing Enzymes. J Diabetes
Sci Technol 2011 Sep; 5(5): 1068–1076 (2011).
Table 2: represents samples for glucose results lower than 100 mg/dL.
Difference range:
RIGHTEST GT333 Blood
Glucose Meter and YSI
Within ± 5 mg/dL
Within ± 10 mg/dL
Within ± 15 mg/dL
The percent ( and number ) of different sampling sites.
Fingertip
140/198 (70.7%)
190/198 (96.0%)
198/198 (100.0%)
Palm
133/198 (67.2%)
188/198 (94.9%)
198/198 (100.0%)
Forearm
117/198 (59.1%)
181/198 (91.4%)
198/198 (100.0%)
Bias range:
RIGHTEST GT333 Blood
Glucose Meterand YSI
Within ± 5 mg/dL
Within ± 10 mg/dL
Within ± 15 mg/dL
The percent ( and number ) of different sampling sites.
Fingertip
373/462 (80.7%)
459/462 (99.4%)
462/462 (100.0%)
Palm
321/462 (69.5%)
453/462 (98.1%)
458/462 (99.1%)
Forearm
310/462 (67.1%)
442/462 (95.7%)
456/462 (98.7%)
Table 3: represents samples for glucose results greater than 100 mg/dL.
Hematocrit(Hct)
Hematocrit should be between 20% - 60%. If you do not know your hematocrit, ask your healthcare professional.
Interferences
Substance and possible level may interfere the glucose measurement:
Uric acid ≧ 16 mg/dL (0.95mmol/L),
Glutathione reduced ≧ 70 mg/dL (2.28 mmol/L),
Ascorbic acid (Vitamin C) ≧ 6 mg/dL (0.34 mmol/L)
26 toxic amount tested substances (Acetaminophen, Ascorbic acid, Dopamine, EDTA, Gentisic acid, Heparin, Ibuprofen,
L-dopa, Methyldopa, Pralidoxime iodide, Salicylic Acid, Tetracycline, Tolazamide, Tolbutamide, Bilirubin, Cholesterol, Creatinine,
Glutathione, Hemoglobin, Triglycerides, Uric acid, Maltose, Xylose, Galactose, Lactose, Icodextrin) in two blood sample
concentrations.
For in vitro diagnostic use Manufacturer
IVD For single use only EC Representive
LOT
Lot number
CE-mark (with No. of notified body)
Consult the instruction for use
Store between tempe rature 4°C and 30°C ( 39°F and 86 °F ) Biological risks Expiry date
Rev. Date:2019-10
18.8%
37.7%
43.5%
Test range in mg/dL (mmol/L)
Within ± 15mg/dL (0.83 mmol/L) or within ± 15 %
Fingertip
26 - 740
660 (100.0%)
Palm
25 - 475
660 (100.0%)
Forearm
26 - 474
660 (100.0%)
Table 1: Accuracy basic information

Other Rightest Blood Glucose Meter manuals

Rightest GS100 User manual

Rightest

Rightest GS100 User manual

Rightest GT100 User manual

Rightest

Rightest GT100 User manual

Rightest GM250S User manual

Rightest

Rightest GM250S User manual

Rightest GM700SB User manual

Rightest

Rightest GM700SB User manual

Rightest GM700 Pro 2 User manual

Rightest

Rightest GM700 Pro 2 User manual

Rightest Max User manual

Rightest

Rightest Max User manual

Rightest Wiz User manual

Rightest

Rightest Wiz User manual

Rightest GS280 User manual

Rightest

Rightest GS280 User manual

Rightest GS100 User manual

Rightest

Rightest GS100 User manual

Rightest HM312 User manual

Rightest

Rightest HM312 User manual

Rightest GT200 User manual

Rightest

Rightest GT200 User manual

Rightest GT333 User manual

Rightest

Rightest GT333 User manual

Rightest Max Tel User manual

Rightest

Rightest Max Tel User manual

Rightest BIONIME GM300 User manual

Rightest

Rightest BIONIME GM300 User manual

Rightest GM700S User manual

Rightest

Rightest GM700S User manual

Popular Blood Glucose Meter manuals by other brands

Infopia Element Instructions for use

Infopia

Infopia Element Instructions for use

GlucoRx HCT owner's manual

GlucoRx

GlucoRx HCT owner's manual

Livongo BG1000 user guide

Livongo

Livongo BG1000 user guide

AIDIAN Nova StatStrip Brief instructions

AIDIAN

AIDIAN Nova StatStrip Brief instructions

Advocate PetTest BMB-EV099X manual

Advocate

Advocate PetTest BMB-EV099X manual

EPS Bio Technology EasyMax LTC Long term care manual

EPS Bio Technology

EPS Bio Technology EasyMax LTC Long term care manual

MICROPOINT qLabs ElectroMeter user manual

MICROPOINT

MICROPOINT qLabs ElectroMeter user manual

ISCtech LASER DOC Quick reference guide

ISCtech

ISCtech LASER DOC Quick reference guide

Roche ACCU-Chek Guide Me quick start guide

Roche

Roche ACCU-Chek Guide Me quick start guide

veri-Q BALANCE PRO user manual

veri-Q

veri-Q BALANCE PRO user manual

DailyChek HL568AA user manual

DailyChek

DailyChek HL568AA user manual

Tykess TD-4116 owner's manual

Tykess

Tykess TD-4116 owner's manual

SD Biosensor SD GlucoNavii NFC User instruction guide

SD Biosensor

SD Biosensor SD GlucoNavii NFC User instruction guide

MediSure DG001MM-MULTI quick start guide

MediSure

MediSure DG001MM-MULTI quick start guide

Abbott FreeStyle Libre 3 quick start guide

Abbott

Abbott FreeStyle Libre 3 quick start guide

Accu-Chek ACTIVE Getting started guide

Accu-Chek

Accu-Chek ACTIVE Getting started guide

Sugarteck TD-4207B owner's manual

Sugarteck

Sugarteck TD-4207B owner's manual

Accu-Chek Mobile user manual

Accu-Chek

Accu-Chek Mobile user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.