Rinstrum R300 Series User manual

R300 Series
(R310, R320 and R321)
Digital Indicato
r
Operator Manual
For use with Software Versions 1.1 and above
R300-601-150

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 2
Copyright
All Rights Reserved. No part of this document may be copied,
reproduced, republished, uploaded, posted, transmitted, distributed,
stored in or introduced into a retrieval system in any form, or by any
means (electronic, mechanical, photocopying, recording or otherwise)
whatsoever without prior written permission of Rinstrum Pty Ltd.
Disclaimer
Rinstrum Pty Ltd reserves the right to make changes to the products
contained in this manual in order to improve design,
performance or reliability.
The information in this manual is believed to be accurate in all respects
at the time of publication, but is subject to change without notice.
Rinstrum Pty Ltd assumes no responsibility for any errors or omissions
and disclaims responsibility for any consequences resulting from the
use of the information provided herein.
“Everything should be made as simple as possible, but not simpler.”
- Albert Einstein -

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 3
Table of Contents
1. INTRODUCTION ................................................................. 4
1.1. Approvals................................................................... 4
1.2. Features..................................................................... 4
1.3. Manuals ..................................................................... 4
2. SAFETY .............................................................................. 6
2.1. Operating Environment .............................................. 6
2.2. Electrical Safety ......................................................... 6
2.3. Cleaning..................................................................... 6
3. BASIC OPERATION ........................................................... 7
3.1. User Interface Display and Controls .......................... 7
3.2. Operation Keys .......................................................... 8
3.3. Stability Considerations ........................................... 10
3.4. Annunciators............................................................ 11
4. BASIC WEIGHING ............................................................ 12
4.1. Normal Weighing ..................................................... 12
4.2. Using Tare ............................................................... 12
5. SPECIAL FUNCTIONS (R320 ONLY) .............................. 13
5.1. Testing the Display .................................................. 13
5.2. Counting .................................................................. 13
5.3. Units Switching (kg / lb) ........................................... 14
5.4. Hold ......................................................................... 14
5.5. Peak Hold ................................................................ 14
5.6. Live Weighing .......................................................... 15
5.7. Showing Totals ........................................................ 16
6. ERROR MESSAGES ........................................................ 17
6.1. Weighing Errors ....................................................... 17
6.2. Setup and Calibration Errors.................................... 18
6.3. Diagnostic Errors ..................................................... 19

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 4
1.Introduction
The Rinstrum R300 Series are precision digital indicators using
the latest Sigma-Delta A/D technology to ensure fast and
accurate weight readings.
This manual contains information on the operation of the R310,
R320 and R321.
Note: The R321 is the R320 enclosed in a stainless steel
housing. In this manual any reference to the R320 includes the
R321.
Unless otherwise indicated, all information will apply to the
R310, R320 and R321. For the purposes of readability, where
all instruments are being described, they will be referred to as
the R300 Series.
1.1. Approvals
• NSC approval (4000 divisions at 0.8µV/division).
• NMI approval (4000 divisions at 0.8µV/division).
• C-tick approved and CE approved.
1.2. Features
• R310 has Zero and Tare functionality.
• R320 has Zero and Tare functionality as well as battery
backed clock/calendar, special function key (for counting, live
weight averaging, peak-hold, etc.).
1.3. Manuals
For more information on the R300 Series refer to the R300
Series Reference Manual,R300 Series Quick Start Manual
and R300 Series Communications Manual (available free of
charge from www.rinstrum.com).

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 5
R310
R320
R321

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 6
2.Safety
2.1. Operating Environment
• Operating Temperature: –10 to 50°C
• Humidity: <90% non-condensing
• Operating Voltage: Shown on Rear Label
2.2. Electrical Safety
• For your protection all mains electrical hardware must be
rated to the environmental conditions of use.
• Pluggable equipment must be installed near an easily
accessible power socket outlet.
• To avoid the possibility of electric shock or damage to the
instrument, always switch off or isolate the instrument from
the power supply before maintenance is carried out.
2.3. Cleaning
• To maintain the instrument, never use harsh abrasive
cleaners or solvents. Wipe the instrument with a soft cloth
slightly dampened with warm soapy water.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 7
3.Basic Operation
3.1. User Interface Display and Controls
R310
R320

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 8
3.2. Operation Keys
Key Description
POWER: The <POWER> key is used to turn the
instrument on and off.
• To initially turn the instrument ON: Press and
hold the <POWER> key until the display starts up.
• To turn the instrument OFF: Press and hold the
<POWER> key for three seconds. The instrument
will display OFF followed by the three-second
countdown.
Note: If the <POWER> key has been locked, the
instrument cannot be turned off from the front
keypad.
• Battery Operation: When using batteries the
backlight will automatically turn off to conserve
power after a short period of inactivity. A short
press of the <POWER> key will turn the backlight
on again.
• Automatic Operation: The <POWER> key has a
memory function associated with it. This means
that the state of the power setting is remembered
even if external power is interrupted. It is therefore
possible to turn the instrument on in the safe
knowledge that it will operate whenever external
power is available and will not need to be manually
turned on again if the power is interrupted.
ZERO: The <ZERO> key is used to perform a Zero
adjustment on the scale display when an empty scale
has drifted away from a true zero reading.
• On the R320, the Zero adjustment is stored when
power is removed and is re-used when next
powered up.
• Long Press: When the indicator is set to Industrial
mode a long press of the <ZERO> key will remove
any stored zero adjustment.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 9
Key Description
TARE: The <TARE> key is used to temporarily set
the scale to zero (such as cancelling the weight of a
carton before performing a filling operation). The
display will show the Net weight and the Net
annunciator will be lit.
• The <TARE> key can operate in all modes
(ie. Industrial, OMIL and NTEP).
• The weight tared is deducted from the allowable
range of the scale, reducing the maximum weight
that can be displayed.
• On the R320, the Tare adjustment is stored when
power is removed and is re-used when next
powered up.
GROSS/NET: The <GROSS/NET> key toggles the
weight display between the Gross weight and the Net
weight (provided that a Tare has previously been
acquired using the <TARE> key).
R320
Only
PRINT: The <PRINT> key on the R320 will trigger an
output of the current weight reading if a printer or
computer has been attached to the instrument and
the manual print function has been selected.
• The PRINT prompt is displayed while waiting for
the printer to accept data.
• If the printer is offline the PRINT prompt will remain
for a maximum of 10 seconds before the operation
is cancelled.
• Each weight printed is automatically added to an
internal total weight.
• Long Press: A long press of the <PRINT> key will
print the total. The total weight is then cleared
automatically.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 10
Key Description
R310
Only
TEST: The <TEST> key on the R310 is used to
initiate a display test.
• Once pressed the display is shown with all
segments clear, then all segments lit and then all
segments clear again before returning to normal
operation.
• On the R320 the equivalent function is available
only when TEST has been assigned to the user
defined <FUNCTION> key. Refer to Testing the
Display page 13.
R320
Only
FUNCTION: The <FUNCTION> key on the R320 is
programmable to suit customer requirements. A label
identifying the special function will be attached. Refer
to Special Functions page 13 for details of the
available functions.
• Long Press: A long press of the <FUNCTION>
key may be used for certain functions depending
on the primary function of the key.
3.3. Stability Considerations
Once a <ZERO>, <TARE> or <PRINT> key is pressed the unit
waits for a stable valid reading before performing the associated
operation. If the weight readings remain unstable or invalid due
to some diagnostic error for longer than 10 seconds, the
operation is cancelled and the STABLE ERROR message is
displayed.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 11
3.4. Annunciators
Symbol Name Description
ZERO Visible when the gross reading is
within ±¼ of a division of true zero.
NET Visible when the displayed reading
represents Net weight.
MOTION Visible when the displayed reading is
not stable.
OVER Visible when the setpoint weight is
over the setpoint target.
UNDER Visible when the setpoint weight is
under the setpoint target.
ZERO
BAND
Visible when the displayed weight is
within the zero 'dead' band setting.
(The zero band symbol shows near
the top right corner of the display.)
HOLD Visible when the displayed reading is
held.
LOW
BATTERY
Visible when battery voltage is too low
and batteries need replacing or
recharging. (The low battery symbol
shows in the top right corner of the
display.)

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 12
4.Basic Weighing
4.1. Normal Weighing
• Ensure instrument is On and Zero annunciator is lit.
• Place your item on the weigh platform.
• Read the weight display.
4.2. Using Tare
• The indicator displays zero with Zero annunciator lit.
• Place the container on the weigh platform.
• Press the <TARE> key.
• The indicator will show the displayed zero weight and the
Net annunciator will be lit.
• Fill the container to the required weight.
• Press the <GROSS/NET> key to toggle
between the Net weight and the Gross (total)
weight.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 13
5.Special Functions (R320 Only)
5.1. Testing the Display
• Press the <TEST> key to clear the display then
show all segments of the display then clear the
display again before returning to normal operation.
Note: This feature is also available on the R310. For more
information refer to TEST key page 10.
5.2. Counting
• Place the container on the weigh platform and
press <TARE> if required.
• Place the sample pieces to be counted on the weigh
platform.
• Press and hold the <COUNT> key for two
seconds. The default number of pieces in the
sample will be displayed.
• Use the <GROSS/NET> and
<PRINT> keys to alter the
number of pieces. and
• Press <COUNT>. The current sample will be
stored against the entered pieces. The letter p(for
pieces) displays when in counts display.
• Press the <COUNT> key to toggle between the weight
display and the counts display.
• If printing is enabled the sample quantity and weight will be
printed.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 14
5.3. Units Switching (kg / lb)
• Press the <UNITS> key to switch the display
between kilograms and pounds.
• Printing and serial communications will use the units
displayed (either lb or kg).
5.4. Hold
• Press the <HOLD> key to hold the displayed weight
at its current weight.
• The Hold annunciator will be lit.
• Press the <HOLD> key again to release the weight
reading and return the display to normal weighing.
• All printouts that print the displayed weight will use the held
weight reading if it is currently being displayed.
5.5. Peak Hold
• Press the <PEAK> key to show the largest absolute
weight, either positive or negative (eg. -30 is larger
than 25). The R320 compares the current weight
reading with the stored peak and updates the peak
reading whenever a larger weight is detected.
• The Hold annunciator will be lit.
• Press the <PEAK> key to toggle between the
current weight and the peak weight.
• When displaying the peak weight the Hold
annunciator will be lit.
• When displaying the peak weight, press and hold
the <PEAK> key for two seconds to clear the peak
value and reset back to 0 (zero).
• All printouts that print the displayed weight will use the held
weight reading if it is currently being displayed

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 15
5.6. Live Weighing
• Press and hold the <LIVE WT> key to switch
between normal weighing and live weight mode.
The display will briefly show NORMAL or
LIVE.WT.
• Note: During normal weighing, this key operates
exactly like a manual <HOLD> key.
• When in Live-Weight mode and while the Net weight is
within the zero 'dead' band, the instrument shows the
current weight.
• Press the <TARE> or <ZERO> key to
clear any residual weight and return the
scale to the zero state. or
• Move the animal on to the weigh platform.
• Once the weight moves outside the zero 'dead' band the
instrument begins to calculate a long-term average that
compensates for any movement in the mass. The
instrument flashes the Hold annunciator and shows the
current average value.
• The Hold annunciator is steady when the final sample
weight is shown on the display.
• Press the <LIVE.WT> key to force the sample to
be recalculated.
• Once the animal is removed, the R320 automatically clears
the previous reading ready for the next animal.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 16
5.7. Showing Totals
• The <PRINT> key is used not only to print the
current weight but also to add that weight to the
current total.
• When the <TOTAL> key is pressed, the indicator
displays COUNT followed by the number of items
in the total.
• After this the indicator displays TOTAL followed by the
current total weight.
• If the total weight is too large to display in six digits, the
weight is shown in two sections labelled with the upper six
digits displayed before the lower six digits.
• Press and hold the <PRINT> key to cause the
total accumulated weight to be printed and then
cleared.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 17
6.Error Messages
6.1. Weighing Errors
Error Description Resolution
(U - - - - -) The weight is below the
minimum allowable weight
reading.
Increase the weight
or decrease the
minimum allowable
weight reading.
(O - - - - -) The weight is above the
maximum allowable weight
reading.
Warning - overloading may
damage mechanical scale
elements.
Check the condition
of load cell
connections. Check
for damaged load
cell.
(ZERO)
(ERROR)
The weight reading is beyond
the limit set for Zero
operation. The operation of
the <ZERO> key is limited in
the setup during installation.
The indicator cannot be
Zeroed at this weight.
Increase the Zero
Range (OPTION:
Z.RANGE) or use
the <TARE> key
instead.
(STABLE)
(ERROR)
Scale motion has prevented a
<ZERO>, <TARE> or
<PRINT> operation from
occurring on command.
Try the operation
again once the
scale is stable.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 18
6.2. Setup and Calibration Errors
Error Description Resolution
The instrument may be in
Safe Setup and an item
that needs Full Setup
has been selected for
editing.
Access Full Setup to edit the
item.
(ENTRY)
(DENIED)
When accessing setup,
more than three attempts
have been made with the
incorrect passcode.
Turn the instrument off. When
the instrument is turned back
on, enter the correct passcode
to access setup.
(LIN.PT)
(LO)
An attempt has been
made to place a linear-
isation point below zero.
Incorrect linearisation point
entered (must be between
zero and full scale).
(PT.TOO)
(CLOSE)
An attempt has been
made to place a
calibration point too close
to an existing calibration
point.
Re-enter the calibration point.
Points must be spaced by at
least 2% of full scale from
each other.
(RES)
(LO)
The scale build is
configured for less than
100 graduations.
Check the resolution (count-
by) and capacity settings.
(RES)
(HIGH)
The scale build is
configured for more
than 30,000 graduations.
Check the resolution (count-
by) and capacity settings.
(SPAN)
(LO)
The load cell signal
range (span) is too small
for these settings.
Incorrect span weight entered
(must be between zero and
full scale). Scale wiring
incorrect. Wrong load cell
capacity (too large). Wrong or
no calibration weight added to
scale.
(SPAN)
(HI)
The load cell signal
range (span) is
too large for these
settings.
Incorrect span weight entered
(must be between zero and
full scale). Scale wiring
incorrect. Load cell capacity
too small for application.
(ZERO)
(LO)
An attempt has been
made to calibrate zero
below -2mV/V.
Scale wiring incorrect.
(ZERO)
(HI)
An attempt has been
made to calibrate zero
above +2mV/V.
Remove all weight from scale.
Scale wiring incorrect.

Rinstrum - R300 Series Digital Indicator Operator Manual Rev 1.5
Page 19
6.3. Diagnostic Errors
• Check: Service personnel can check this item on site.
• Return for Service: The instrument must be returned to the
manufacturer for factory service.
Error Description Resolution
(E0001) The power supply voltage is too low. Check supply
(E0002) The power supply voltage is too high. Check scale /
cables
(E0010) The temperature is outside of allowable
limits.
Check location
(E0020) Scale build is incorrect. The number of
graduations has been set too low or too
high.
Fix up scale
build
(E0100) The digital setup information has been
lost.
Re-enter setup
(E0200) The calibration information has been
lost.
Re-calibrate
(E0300) All setup information has been lost Enter setup
and calibrate
(E0400) The factory information has been lost. Return for
Service
(E0800) The EEPROM memory storage chip
has failed
Return for
Service
(E2000) ADC Out of Range Error. This may be
caused from a broken load cell cable.
Check
BUILD:CABLE
setting. Check
load cell cable,
wiring, etc.
(E4000) The battery backed RAM data has lost
data.
Re-enter setup
(E8000) The FLASH program memory is
incorrect
Return for
Service
The Etype error messages are additive. For example if instrument is
running off batteries and the temperature drops, the battery voltage may be
too low. The resulting error messages will be E0011 (0001 + 0010). The
numbers add in hexadecimal as follows:
1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - A - B - C - D - E - F
(For example, 2 + 4 = 6, or 4 + 8 = C

Other manuals for R300 Series
1
This manual suits for next models
3
Table of contents
Other Rinstrum Measuring Instrument manuals

Rinstrum
Rinstrum 400 Series User manual

Rinstrum
Rinstrum 400 Series User manual

Rinstrum
Rinstrum M4202 User manual

Rinstrum
Rinstrum 5200 Instruction manual

Rinstrum
Rinstrum 400 Series User manual

Rinstrum
Rinstrum K405 User manual

Rinstrum
Rinstrum 400 Series Instruction manual

Rinstrum
Rinstrum 5200 Use and care manual

Rinstrum
Rinstrum C300 Series User manual

Rinstrum
Rinstrum 32 Series Instruction manual