Rinstrum 32 Series Instruction manual

32x Series
Digital Indicator
Quick Start Manual
003R-632-200-M04

Copyright
All Rights Reserved. No part of this document may be
copied, reproduced, republished, uploaded, posted,
transmitted, distributed, stored in or introduced into a
retrieval system in any form, or by any means (electronic,
mechanical, photocopying, recording or otherwise)
whatsoever without prior written permission of
Rinstrum Pty Ltd.
Disclaimer
Rinstrum Pty Ltd reserves the right to make changes
to the products contained in this manual in order
to improve design, performance or reliability.
The information in this manual is believed to be accurate in
all respects at the time of publication, but is subject to
change without notice. Rinstrum Pty Ltd assumes no
responsibility for any errors or omissions and disclaims
responsibility for any consequences resulting from the use
of the information provided herein.
SPECIAL NOTE
Trade Use of the Instrument
This manual may occasionally make reference to
Trade Use settings of the instrument.
Some individual settings may not be legal for trade use.
Please check regulations with the appropriate
Weights and Measures Authority.

Quick Start Manual Rev 2.45 - Software Versions 3.xx
003R-620-245 Page 19
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
BUILD DP⊗Decimal Point Position 000000, 00000.0, 0000.00, 000.000, 00.0000, 0.00000 Accept
CAP⊗Maximum Capacity 000100 to 999999 Default = 003000
<SEL> changes position, <EDT> changes digit.
Accept
RES⊗Resolution (Count-By) 1, 2, 5, 10, 20, 50, 100 Accept
UNITS⊗Units kg, lb, t, none, g (K303: kg, lb, t, none, g) Accept
GRADS⊗
(K305 only)
Graduations (Range 1) 000100 to 030000 Default = 003000
<SEL> changes position, <EDT> changes digit.
Accept
HI.RES⊗Resolution x 10 Mode OFF, ON Accept
CABLE⊗4-Wire or 6-Wire 4, 6 Accept
OPTION USE⊗Scale Use INDUST, NSC, OIML, NTEP Accept
FILTER Reading Average
(time in seconds)
none, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0 Accept
MOTION⊗Motion Detection OFF, 0.5-1.0, 1.0-1.0
Default = 0.5 Graduations per Second
Accept
INIT.Z Initial Zero on Startup OFF, ON Accept
Z.TRAC⊗Zero Tracking Sensitivity OFF, SLOW (0.5 grads/sec), FAST (10 grads/sec) Accept
Z.RANGE⊗Zero Operating Range -2_ 2, -1_ 3, -20_20, FULL
Default = -2_ 2 (–2% to +2%)
Accept
Z.BAND⊗Zero 'Dead' Band <SEL> changes position, <EDT> changes digit.
Default = 0
Accept

Quick Start Manual Rev 2.45 - Software Versions 3.xx
Page 20 003R-620-245
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
OPTION
(continued) R.ENTRY⊗Enable Rear Pushbutton
for Full Setup Entry
OFF, ON Accept
CAL ZERO⊗Zero Calibration Routine
Set between zero and full
scale
(Current weight displays)
<SEL> key to start. Current weight displays.
Remove all weight.
<SEL>, <EDT> or <OK> starts routine (Z.in P displays).
<ITM> key to exit, <SEL>, <EDT> or <OK> to repeat routine.
SPAN⊗Span Calibration Routine
Set between zero and full
scale
(Current weight displays)
<SEL> key to start <EDT> key to select clear linearisation
points ( CLr.L y displays). or not ( CLr.L n displays).
<OK> key to select alternative.
Current weight displays.
Add test weight.
<SEL> or <OK> to show calibration weight value.
Set correct weight. <SEL> changes position, <EDT> chan
g
es
digit.
<ITM> or <OK> starts routine (S.in P displays).
<ITM> key to exit, <SEL>, <EDT> or <OK> to re-edit
calibration weight and repeat routine.

Quick Start Manual Rev 2.45 - Software Versions 3.xx
003R-620-245 Page 21
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
CAL
(continued) ED.LIN⊗
(K302 only)
Edit Linearisation Points
Set between zero and full
scale
L1. - - - Select Linearisation
point 1 to 10 (L1, L2, L3, L4,
L5, L6, L7, L8, L9, L10).
(Approx. % of full scale)
<SEL> to step through list of points.
<OK> to change selected point or <ITM> to exit.
After <OK> current weight displays.
Add test weight.
<OK> to enter corrected weight. <SEL> changes position,
<EDT> changes digit.
<OK> starts routine (L.in P displays).
<ITM> key to exit, <OK> to repeat routine.
CLR.LIN⊗
(K302 only)
Clear Linearisation Points
L1. - - - Select Linearisation
point 1 to 10 (L1, L2, L3, L4,
L5, L6, L7, L8, L9, L10)
<SEL> to step through list of points.
<OK> to select point to clear or <ITM> to exit.
After <OK> press <ITM> to choose Cont. Y (Yes) or Cont. N
(No).
<ITM> to return to CLR.LIN.
DIR.ZER⊗
Direct Zero Calibration
(Current weight displays)
<OK> key to start. Current weight displays.
<OK> to enter direct zero setting (in mV/V).
<SEL> changes position, <EDT> changes digit.
<OK> key to store new zero calibration.
<ITM> to exit or <OK> to repeat operation.
DIR.SPN⊗
Direct Span Calibration
(Current weight displays)
<OK> key to start. Current weight displays.
<OK> to enter direct span setting (in mV/V).
<SEL> changes position, <EDT> changes digit.
<OK> key to store new span calibration.
<ITM> to exit or <OK> to repeat operation.

Quick Start Manual Rev 2.45 - Software Versions 3.xx
Page 22 003R-620-245
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
CAL
(continued) G.INST⊗
(K305 only)
Installation Gravity
(Current setting displays)
9.750 to 9.860 Default = 9.810
Gravity Compensation Only
Accept
G.FAC⊗
(K305 only)
Factory Gravity
(Current setting displays)
9.750 to 9.860 Default = 9.810
Gravity Compensation Only
Accept
G.FIRST⊗
(K305 only)
First Powerup Prompt
(Current setting displays)
ON, OFF
Prompts user to enter G.INST on first powerup
Accept
FAC.CAL⊗Factory Calibration
Cont. N (No)
Cont. Y (Yes)
Cont. N
Warning: Choosing Cont. Y will restore default factory
calibration in BUILD and CAL menus.
Accept
SPEC SAFE.PC Safe Setup Passcode (000000 no passcode). Set up to 6 digit passcode.
<SEL> changes position, <EDT> changes digit.
Activated only when FULL.PC is also set.
Accept
FULL.PC Full Setup Passcode (000000 no passcode). Set up to 6 digit passcode.
<SEL> changes position, <EDT> changes digit.
Accept
KEY.LOC Front Panel Key Locking
P12345
(P for Power key.
Other keys numbered from
the left, ie. Zero=1.)
P12345
Character indicates key is unlocked.
(–) Dash indicates key is locked.
<SEL> changes position, <EDT> changes digit.
Accept
KEY.FN <FUNCTION> Key Setting NONE, TEST, COUNT, UNITS, HOLD, PEAK.H, LIVE.WT,
SHOW.T, HI.RES, A.TARE, SET.PT
Accept

Quick Start Manual Rev 2.45 - Software Versions 3.xx
003R-620-245 Page 23
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
SPEC
(continued) AUT.OFF Auto Power Off / Battery
Operation
NEVER, 1, 5, 10 (time in minutes)
Default = Never powers off automatically
(Battery Mode: powers down after 30 minutes)
Accept
B.LIGHT Backlight Operation ON, OFF
(Automatically turns backlight off after 10 seconds of
inactivity)
FORCED (Permanently on)
Accept
REM.FN Remote Function NONE, KEY1 to KEY5, BLANK
(Activated only when SERIAL:TYPE is set to AUTO, PRINT or
AUTO.PR)
Accept
REM.CHR Remote Input Transmit Idle
Character
000 to 255 Default = 005
<SEL> changes position, <EDT> changes digit.
Accept
BAT.VLT Battery Voltage 4.8, 7.2, 9.6, 12, 24, PWR (External Power) Accept
SERIAL TYPE Serial Output Type NET, AUTO, SINGLE, PRINT, AUTO.PR Accept
FORMAT Serial Output Format MASTER, CUSTOM, FMT_1, FMT_2, Accept
BAUD Serial Baud Rate 2400, 4800, 9600 Accept

Quick Start Manual Rev 2.45 - Software Versions 3.xx
Page 24 003R-620-245
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
SERIAL
(continued) BITS Serial Format Options
Position 1: Parity
Position 2: Data Bits
Position 3: Stop Bits
Position 4: DTR Handshake
N 8 1 - (Default Serial Format Options)
<SEL> changes position, <EDT> changes digit.
Parity: N None, OOdd, EEven
Data Bits: 7 or 8data bits
Stop Bits: 1 or 2 stop bits
DTR: (-) DTR disabled or dDTR enabled
Accept
ADDRES Instrument Address 01 to 31
<SEL> changes position, <EDT> changes digit.
Accept
RST.CON⊗Reset Printed Consecutive
Numbers to 1
Cont. N (No)
Cont. Y (Yes)
Cont. N
<ITM> to choose Cont. Y (Yes) or Cont. N (No)
Accept
SET.PTS SETPT.1 Control Type
Setpoint 1 (Output 1)
OFF, ON, OVER, UNDER, MOTION, ZERO, NET, ERROR
(Displayed)
Accept
SRC.1 Weight Source GROSS, DISP, NET (Displayed) Accept
TARG.1 Target for Over and Under
Weight
Setpoint 1 (Output 1)
-99999 to 999999
Default = 000000
Accept
SETPT.2 Control Type
Setpoint 2 (Output 2)
OFF, ON, OVER, UNDER, MOTION, ZERO, NET, ERROR
(Displayed)
Accept
SRC.2 Weight Source GROSS, DISP, NET (Displayed) Accept

Quick Start Manual Rev 2.45 - Software Versions 3.xx
003R-620-245 Page 25
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
SET.PTS
(continued) TARG.2 Target for Over and Under
Weight
Setpoint 2 (Output 2)
-99999 to 999999
Default = 000000
Accept
CLOC FORMAT Date Format Setting dd.mm.yy or mm.dd.yy
(K303: mm.dd.yy or dd.mm.yy)
Accept
YEAR Year Setting 2000 to 2099
<SEL> changes position, <EDT> changes digit.
Accept
MONTH Month Setting 01 to 12
<SEL> changes position, <EDT> changes digit.
Accept
DAY Day Setting 01 to 31
<SEL> changes position, <EDT> changes digit.
Accept
HOUR Hour Setting 00 to 23
<SEL> changes position, <EDT> changes digit.
Accept
MINUTE Minute Setting 00 to 59
<SEL> changes position, <EDT> changes digit.
Accept
TEST SCALE Scale Base Test Display
Display reads in millivolts-per-volt
(factory calibrated to 0.1%)
Verify the correct load cell capacity and/or load cell wiring is
used. Remove weight from scale.

Quick Start Manual Rev 2.45 - Software Versions 3.xx
Page 26 003R-620-245
GROUP
(GRP)
ITEM
(ITM)
SELECT
(SEL)
EDIT
(EDT)
Underline = Defaults
OK
TEST
(continued) FRC.OUT Force Outputs <EDT> advances through outputs
(ie. ON.1 and ON.2)
<OK> turns outputs off and exits test.
Default = OFF
O.LOAD Input Overload Count Displays the number of occurrences of an input overload. Return
CLR.OLD Clear Overload Count
Cont. N (No)
Cont. Y (Yes)
Cont. N
Choosing Cont. Y will clear the overload count.
Accept
FACTRY DEFLT Restore Factory Defaults
Cont. N (No)
Cont. Y (Yes)
Cont. N
Warning: Choosing Cont. Y will clear all stored data except
BUILD and CAL menus.
Accept
- END - EXIT SETUP Save settings and return to
normal weighing mode
Accept

Quick Start Manual Rev 2.45 - Software Versions 3.xx
003R-620-245 Page 27
8. Error Messages
8.1. Weighing Errors
•Check Setup = This item can be checked on site by service
personnel
Error Description Resolution
(U - - - - -) The weight is below the
minimum allowable weight
reading.
Increase the weight or
decrease the minimum
allowable weight
reading.
(O - - - - -) The weight is above the
maximum allowable weight
reading.
Warning - overloading
may damage mechanical
scale elements.
Check the condition of
load cell connections.
Check for damaged load
cell.
(ZERO)
(ERROR)
The weight reading is
beyond the limit set for
Zero operation. The
operation of the <ZERO>
key is limited in the setup
during installation. The
indicator cannot be Zeroed
at this weight.
Increase the Zero Range
(Z.RANGE) or use the
<TARE> key instead.
(STABLE)
(ERROR)
Scale motion has
prevented a <ZERO>,
<TARE> or <PRINT>
operation from occurring
on command.
Try the operation again
once the scale is stable.

Quick Start Manual Rev 2.45 - Software Versions 3.xx
Page 28 003R-620-245
8.2. Setup and Calibration Errors
Error Description Resolution
The instrument may be in
Safe Setup and an item
that needs Full Setup has
been selected for editing.
Access Full Setup to edit
the item.
(ENTRY)
(DENIED)
When accessing setup,
more than three attempts
have been made with the
incorrect passcode.
Turn the instrument off.
When the instrument is
turned back on, enter the
correct passcode to
access setup.
(LIN.PT)
(LO)
An attempt has been
made to place a linear-
isation point below zero.
Incorrect linearisation
point entered (must be
between zero and full
scale).
(PT.TOO)
(CLOSE)
An attempt has been
made to place a
calibration point too close
to an existing calibration
point.
Re-enter the calibration
point. Points must be
spaced by at least 2% of
full scale from each other.
(RES)
(LO)
The scale build is
configured for less than
100 graduations.
Check the resolution
(count-by) and capacity
settings.
(RES)
(HIGH)
The scale build is
configured for more
than 30,000 graduations.
(K303 60,000 graduations)
Check the resolution
(count-by) and capacity
settings.
(SPAN)
(LO)
The load cell signal range
(span) is too small for
these settings.
Incorrect span weight
entered (must be between
zero and full scale). Scale
wiring incorrect. Wrong
load cell capacity (too
large). Wrong or no
calibration weight added
to scale.
(SPAN)
(HI)
The load cell signal range
(span) is too large for
these settings.
Incorrect span weight
entered (must be between
zero and full scale). Scale
wiring incorrect. Load cell
capacity too small for
application.

Quick Start Manual Rev 2.45 - Software Versions 3.xx
003R-620-245 Page 29
Error Description Resolution
(ZERO)
(LO)
An attempt has been
made to calibrate zero
below
-2mV/V.
Scale wiring incorrect.
(ZERO)
(HI)
An attempt has been
made to calibrate zero
above +2mV/V.
Remove all weight from
scale. Scale wiring
incorrect.

Quick Start Manual Rev 2.45 - Software Versions 3.xx
Page 30 003R-620-245
8.3. Diagnostic Errors
•Check: This item can be checked on site by service personnel.
•Return for Service: The instrument must be returned to the
manufacturer for factory service.
Error Description Resolution
(E0001) The power supply voltage is too
low.
Check supply
(E0002) The power supply voltage is too
high.
Check scale /
cables
(E0010) The temperature is outside of
allowable limits.
Check location
(E0020) Scale build is incorrect. The
number of graduations has been
set too low or too high.
Fix up scale build
(E0100) The digital setup information has
been lost.
Re-enter setup
(E0200) The calibration information has
been lost.
Re-calibrate
(E0300) All setup information has been lost Enter setup and
calibrate
(E0400) The factory information has been
lost.
Return for Service
(E0800) The EEPROM memory storage
chip has failed
Return for Service
(E2000) ADC Out of Range Error. This
may be caused from a broken
load cell cable.
Check
BUILD:CABLE
setting. Check
load cell cable,
wiring, etc.
(E4000) The battery backed RAM data has
lost data.
Re-enter setup
(E8000) The FLASH program memory is
incorrect
Return for Service
The Etype error messages are additive. For example if instrument
is running off batteries and the temperature drops, the battery
voltage may be too low. The resulting error messages will be
E0011 (0001 + 0010). The numbers add in hexadecimal as follows:
1 - 2 - 3 - 4 - 5 - 6 - 7 - 8 - 9 - A - B - C - D - E - F
(For example, 2 + 4 = 6, or 4 + 8 = C)

Quick Start Manual Rev 2.45 - Software Versions 3.xx
003R-620-245 Page 31
Notes:

Quick Start Manual Rev 2.45 - Software Versions 3.xx
Page 32 003R-620-245
Notes:

Notes:

Table of contents
Other Rinstrum Measuring Instrument manuals

Rinstrum
Rinstrum M4202 User manual

Rinstrum
Rinstrum 400 Series User manual

Rinstrum
Rinstrum 400 Series User manual

Rinstrum
Rinstrum 400 Series User manual

Rinstrum
Rinstrum K405 User manual

Rinstrum
Rinstrum 5200 Instruction manual

Rinstrum
Rinstrum R300 Series User manual

Rinstrum
Rinstrum 2100 Instruction manual

Rinstrum
Rinstrum C300 Series User manual

Rinstrum
Rinstrum R300 Series Instruction manual