18 • Chapter 3: Operation GX-3R Condensed Operator’s Manual
3. If CAL RMDR is set to ON (factory setting) and a calibration is due, the screen that appears next
depends on how CAL EXPD is set in User Mode (see page 47). The three possible screens are
described below. If a calibration is not due, the instrument shows how many days are left until a
calibration is due.
CAL EXPD set to
CONFIRM (factory setting)
CAL EXPD set to
CANT USE
CAL EXPD set to
NONE
LCD
Sound Buzzer sounds double pulsing
tone
Buzzer sounds double pulsing
tone
None
Action • Option A, Perform calibration:
Press and release POWER
MODE to perform a calibration.
The instrument takes you
straight to the AUTO CAL CYL
A screen in User Mode’s GAS
CAL\AUTO CAL item. If
Password Protection is set to On
using the SDM-3R PC program,
you must enter a password. See
page 58 for calibration
instructions.
If the calibration is successful,
the screen above will not appear
again until the unit is due for
calibration. If the calibration is
not successful, the screen above
will again appear in the startup
sequence.
• Option B, Bypass message: To
continue without performing a
calibration, press and release
AIR.
The GX-3R cannot be used until a
successful calibration is per-
formed. Press and release
POWER MODE to perform a cal-
ibration. The instrument takes you
straight to the AUTO CAL CYL
A screen in User Mode’s GAS
CAL\AUTO CAL item. If Pass-
word Protection is set to On
using the SDM-3R PC program,
you must enter a password. If you
don’t press POWER MODE, the
instrument automatically goes to
the AUTO CAL CYL A screen
after 6 seconds. If Password Pro-
tection is set to On using the
SDM-3R PC program, you must
enter a password. See page 58 for
calibration instructions.
If the calibration is successful, the
screen above will not appear
again until the unit is due for cali-
bration. If the calibration is not
successful, the screen above will
again appear in the startup
sequence.
• Option A, Perform calibration:
To perform a calibration, press
and release POWER MODE.
The instrument takes you
straight to the AUTO CAL CYL
A screen in User Mode’s GAS
CAL\AUTO CAL item. If
Password Protection is set to
On using the SDM-3R PC
program, you must enter a
password.
• Option B, Bypass message: To
continue without performing a
calibration, wait a few seconds
for the instrument to continue
with its startup sequence.