9
Measuring procedure
(1) Press the key. All digits on the LCD will be lit for one second and the unit enters the measurement
mode. (“.” character on the pH value display area will blink.)
(2) Remove the sensor cap and glass electrode cover.
Note: There is phthalate solution inside the glass electrode cover to prevent the electrode from drying
out.
(3) Immerse the glass electrode at least 3 to 5 cm deep into the material to be measured. The measurement
starts when the measured pH value is stable.
(4) After measurement, press the key to turn the power off and wash the glass electrode with water
(distilled, pure or tap water) before storing.
(5) Always put the glass electrode cover back on to avoid any possible damages. Also it is important to have
the glass electrode dipped in water (distilled, pure or tap water) or in pH standard solution.
About the glass electrode
(1) Storing
If the instrument is unused for a long time, the glass electrode becomes too dry to measure the pH values
accurately.
The pH electrode works best and shows accurate readings when used with moderate moisture. When using
the instrument for the first time after purchasing or a long storage period, we recommend placing the glass
electrode in water (distilled, pure or tap water) or the pH standard solution for at least 12 hours and
performing calibration before measuring.
(2) Check and calibration
For an accurate pH measurement, perform calibration using the pH standard solutions.
If there is a deviation from the expected value, adjust it before performing a pH measurement.
Note: In some cases, the pH value of the standard solution changes from its original value. This can happen
if the pH standard solution is dirty or if storing the solution for a long time. It is advisable to use a new
pH standard solution, where applicable.
(3) Cleaning
Be sure to wash the pH electrode after each pH calibration or measurement with water (distilled, pure or tap
water). If left unwashed, the liquid measured tends to form a film on the surface of the electrode, causing
measurement errors. Always wash the pH electrode every time after use.
(4) Usable life
The pH electrode is a consumable. Its useful life varies depending on the liquid to be measured, pH value or
temperature.
(5) Deposit of white potassium chloride crystals
A deposit of white crystals is sometimes found on the tip of the electrode or protection bottle; this does not
pose a problem when using the instrument. The deposit is easily washed off with tap water.
If it does not come off with tap water, soak the electrode in warm water to remove it.
Do not lick or swallow the potassium chloride crystals. If the eyes or skin becomes contaminated with the
crystals, immediately consult a medical attention.
Caution: Spare probe model 610S-II is the exclusive probe for SK-610PH-II. Spare probe model 610S
cannot be used on this.