manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. SciCan
  6. •
  7. Laboratory Equipment
  8. •
  9. SciCan STATIM User manual

SciCan STATIM User manual

Other SciCan Laboratory Equipment manuals

SciCan QUANTIM B User manual

SciCan

SciCan QUANTIM B User manual

SciCan Hydrim L11w User manual

SciCan

SciCan Hydrim L11w User manual

SciCan HYDRIM C61w G4 User manual

SciCan

SciCan HYDRIM C61w G4 User manual

SciCan Statim 2000 Installation guide

SciCan

SciCan Statim 2000 Installation guide

SciCan statim 2000 G4 User manual

SciCan

SciCan statim 2000 G4 User manual

SciCan Statim 5000 User manual

SciCan

SciCan Statim 5000 User manual

SciCan STATCLAVE G4 User manual

SciCan

SciCan STATCLAVE G4 User manual

SciCan HYDRIM L110w User manual

SciCan

SciCan HYDRIM L110w User manual

SciCan SALUS User manual

SciCan

SciCan SALUS User manual

SciCan Statim 2000 User manual

SciCan

SciCan Statim 2000 User manual

SciCan BRAVO 17 User manual

SciCan

SciCan BRAVO 17 User manual

SciCan STATIM 900J User manual

SciCan

SciCan STATIM 900J User manual

SciCan Bravo User manual

SciCan

SciCan Bravo User manual

SciCan HYDRIM L110w G4 User manual

SciCan

SciCan HYDRIM L110w G4 User manual

SciCan Statim 2000 Installation guide

SciCan

SciCan Statim 2000 Installation guide

SciCan statim 5000 G4 User manual

SciCan

SciCan statim 5000 G4 User manual

SciCan HYDRIM C61w G4 User manual

SciCan

SciCan HYDRIM C61w G4 User manual

SciCan statim 2000 G4 User manual

SciCan

SciCan statim 2000 G4 User manual

SciCan Statim 2000 User manual

SciCan

SciCan Statim 2000 User manual

SciCan Quantim 16 User manual

SciCan

SciCan Quantim 16 User manual

SciCan statim 5000 G4 Guide

SciCan

SciCan statim 5000 G4 Guide

SciCan STATCLAVE G4 User manual

SciCan

SciCan STATCLAVE G4 User manual

SciCan STATIM 7000 Installation guide

SciCan

SciCan STATIM 7000 Installation guide

SciCan Statim 2000 User manual

SciCan

SciCan Statim 2000 User manual

Popular Laboratory Equipment manuals by other brands

Radleys Carousel 6 Plus instructions

Radleys

Radleys Carousel 6 Plus instructions

Hemotherm 400MR Operation & technical manual

Hemotherm

Hemotherm 400MR Operation & technical manual

mrclab DVO Series Operation manual

mrclab

mrclab DVO Series Operation manual

CoolLED pE-800 quick start guide

CoolLED

CoolLED pE-800 quick start guide

Luminex Guava easyCyte Maintenance Guide

Luminex

Luminex Guava easyCyte Maintenance Guide

IKA A10 basic operating instructions

IKA

IKA A10 basic operating instructions

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Belden

Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler K1223 Series Operation and instruction manual

Koehler

Koehler K1223 Series Operation and instruction manual

Globe Scientific GCM-12 quick start guide

Globe Scientific

Globe Scientific GCM-12 quick start guide

Getinge 86 SERIES Technical manual

Getinge

Getinge 86 SERIES Technical manual

CORNING Everon 6000 user manual

CORNING

CORNING Everon 6000 user manual

Biocomp GRADIENT MASTER 108 operating manual

Biocomp

Biocomp GRADIENT MASTER 108 operating manual

Cole Parmer Stuart SB3/1 instruction manual

Cole Parmer

Cole Parmer Stuart SB3/1 instruction manual

BEVS 2501/2A user manual

BEVS

BEVS 2501/2A user manual

Qiagen DML 3000 user manual

Qiagen

Qiagen DML 3000 user manual

Queensgate NANOSCAN OP400 Quick start instructions

Queensgate

Queensgate NANOSCAN OP400 Quick start instructions

Parr Instrument 4560 Operating instructions manual

Parr Instrument

Parr Instrument 4560 Operating instructions manual

Integra DOSE IT operating instructions

Integra

Integra DOSE IT operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

STATIM
STATIM
CASSETTE AUTOCLAVE
TROUBLESHOOTING GUIDE
Version 2.1
Original Statim Error Codes
The Statim Sterilizer should never leak steam or have water dripping from under
the machine. If either of these conditions occur, check your cassette and seal. All
leaks should be corrected before proceeding with troubleshooting procedures.
Check Cassette 1
Chamber temperature does not rise more than 2°C in three minutes. In all cases: No power to the
boiler. Check thermal fuse. Replace if necessary.
Check Cassette 2
http://www.sterilizers.com/statim-trouble-shooting-manual.html (1 of 26) [3/27/2004 9:23:44 PM]
STATIM
Cycle Fault 6
The software has detected a steam generator (boiler) temperature 5°C greater than the chamber,
within 7.2seconds after a purge during the sterilizing phase of a cycle. Calibrate boiler. Refer to
Boiler Calibration Section
.
Cycle Fault 7
The cassette temperature has dropped 4°C below set point. If visible steam leaks appear replace
cassette seal, lid or tray as needed. Disassemble solenoid valve and check for debris. Make sure
plunger slides smoothly in plunger tube. Refer to Solenoid Valve Section.
Cycle Fault 8
The software has detected a steam generator (boiler) temperature 5'less than the chamber, within
7.2 seconds after a purge during the sterilizing phase ofa cycle. Calibrate boiler. Refer to Boiler
Calibration Section. Check for kinks in the exhaust tubing. Check to see if condenser bottle is full
of hot water.
Cycle Fault 9 N/A
Cycle Fault 10
The cassette temperature has failed to drop to 115°C during the Unwrapped or Wrapped Cycle or
the temperature has failed to drop to 110°C during the Rubber and Plastics Cycle in the purge
conditioning stage. Check for kinks in the exhaust tubing. Check for a clogged duct in the rear of
the cassette tray. Blow out with compressed air. Check solenoid valve for proper operation.
Cycle Fault 11
The cassette temperature has failed to drop to 102°C within 60 seconds of the end of a cycle
during venting. Check for kinks in the exhaust tubing. Check for a clogged duct in the rear of the
cassette tray. Blow out with compressed air. Check the solenoid valve for proper operation.
Cycle Fault 12
http://www.sterilizers.com/statim-trouble-shooting-manual.html (5 of 26) [3/27/2004 9:23:44 PM]
STATIM
Press Stop To Reset This message is displayed on all error faults. Press STOP to clear message.
Touch pads do not work
Disconnect keypad plug from PCB. Be sure blue plastic piece for keypad plug on PCB is pushed
up on the plug pins as far as possible. Reconnect keypad and check. Replace keypad if necessary.
No display or garbled display on LCD.
Check plug connections from cover to PCB. Check that microprocessor is seated firmly in socket.
Statim makes a clicking noise with cassette removed
This is caused by steam leaking from the cassette. The steam gets into the micro switch causing
the contacts to open and close and the solenoid clicks. Repair cassette leak and clicking should
stop in approximately 24 hours.
Loud buzzing noise
Clean or replace solenoid as needed.
Noise during drying cycle only
Some check valve noise is normal. Check the air filter. Replace if dirty. If filter is wet replace
check valve and compressor if necessary.
Water dripping from drain tube under Statim
Replace seal or repair cassette as needed.
Steam is escaping from Condenser Bottle vent hole
Ensure that condenser bottle is always filled to MIN line.
Steam is leaking from Push-In-Fitting at rear of Statim
Ensure that exhaust tube is fully inserted in fitting. Push past initial resistance until
tube seats.
Wraps remain wet after drying
http://www.sterilizers.com/statim-trouble-shooting-manual.html (8 of 26) [3/27/2004 9:23:44 PM]
STATIM
Ensure that cassette is clean and has been treated with Stat Dri. Drain tube must run
directly to condenser bottle with no dips or loops. Do not stack wraps. Invert mesh
rack to provide air space below wraps. Set bubble level to 4 or 5 o'clock position.
Cover Removal
Unplug Statim and remove cassette, filler cap, 4 side screws and 3 rear screws.
On the Statim 5000 the middle screw in the rear is behind the biological filter holding the filter
bracket.
With facia overhanging edge of counter or table, lift up on rear of cover until it
clears internal components, then slide cover forward about an inch.
Gently rotate the cover up and pivot it around the front left corner of the machine.
When the cover is clear of the machine, rest it on its side. Take care not to strain
the ribbon cables connecting the display and keypad.
Refer to illustration on next page
Solenoid Valve Inspection and Repair
(Honeywell solenoid only)
Ensure that the power is off. Disassembly of the solenoid valve with power on
will cause valve coil damage.
Remove retaining nut, yoke and coil from top of valve.
Unscrew the plunger tube assembly. The tube may be removed by carefully
http://www.sterilizers.com/statim-trouble-shooting-manual.html (9 of 26) [3/27/2004 9:23:44 PM]