
4.1 Place a 205ltr drum onto the trolley, locate the drum using the drum holder
support(g1.2)andsecurewithknob(g.1.1).
4.2 Installing the pump
4.2.1 Remove the bung from the 2"BSP opening of the drum.
4.2.2 Slacken off the locking nut on the pump probe and insert the pump into
the drum until the probe bottoms out, tighten the barrel nut and then
tightenthelockingnuttoxthepumptothedrum.
4.3 Setting up the pump
4.3.1 Ensure that the air regulator valve is fully wound anti-clockwise and then
connect the air supply onto the quick release connector on the air
regulator.
4.3.2 Lift the knob on the regulator and slowly turn clockwise until the
recommended operating pressure of 70psi is reached, push down on the
regulator knob to lock the position.
Note: Check the system for leaks, should an air leak be discovered, release the
pressure, disconnect the air line and repair the joint or seal before
continuing.
4.4 Operating the gun
❐ WARNING! Ensure that the trigger is not depressed before opening the
nozzle valve.
4.4.1 Press the Reset button for 1 sec to zero the display.
4.4.2 With the nozzle pointing into the receiving container, open the nozzle valve
asshowning.4.
4.4.3 Depress the trigger and dispense the liquid the amount dispensed is
shownonthedisplay(g.5.4).
4.4.4 Release trigger and close the nozzle valve.
4.5 Determining the Correction Factor
4.5.1 Dispense at least 1 litre of liquid into a calibrated container.
4.5.2 Record the quantity of liquid in the container and record the quantity of
liquid dispensed as indicated on the meter display. Press the Move +
Resetbuttons(g.5.1andg.5.2)simultaneouslytodisplaythecurrent
correction factor.
Qty Dispensed
Qty Indicated x Current Correction Factor = New Correction Factor.
4.6 Programming the Correction Factor
4.6.1 Remove the rubber protective cover from the meter and unscrew the four
phillipsscrews(g.6)holdingthedigitalmeter,removefromthegun.
4.6.2 Pressbutton"2"ontherearofthemeter(g.7.2)for1sectoenterthe
Correction Factor mode.
4.6.3 Press the Move button to select the digit to be changed, each press of the
Move button will select the next digit.
4.6.4 Press the Reset button to add 1 on the digit, press as many times as
required until the correct number is reached.
4.6.5 Press "2" for 1 sec to quit the Correction Factor mode.
4.6.6 Press the Move + Reset buttons simultaneously to display the new
correction factor.
4.7 Changing the Units
Note: The meter is pre-set to display litres.
4.7.1 Press button "2" on the rear of the meter for 1 sec to enter the Correction
Factor mode.
4.7.2 PresstheMovebuttonvetimestoselecttheunits(g.5.3).
4.7.3 Press the Reset button to scroll through the Units (L/Gal/PT/QT), select
the Unit required.
4.7.4 Press "2" for 1 sec to quit the Correction Factor mode.
4. OPERATION
AK4561D Issue No: 1 - 14/11/08
g.5
3
4
12
g.7
2
1
g.4
g.6
Meter
Screws
Battery Cover
3.3 Pre-Operation
3.3.1 Liftuptheknob(g.2A)ontheairregulatorandensurethattheknobis
wound fully anti-clockwise.
3.3.2 Removethellerplug(g.2B)fromthelubricatorandllthelubricatorwith
aqualityairtooloil(SealeyATO1000S),retthellerplug.
3.3.3 The lubricator feed is set when the unit is running, adjust the feed knob
(g.2C)thedripratecanbeseenthroughthesightglassofthedome.The
rate of oil feed will automatically increase or decrease inline with the air
ow.