
Installation
Wall Mounting. Usingthefourscrewsprovided, attach
the two foot brackets to the sides of the enclosure. Then
attachthe unit toany secure surfaceby inserting screws
through the mounting holes in the foot brackets.
Panel Mounting. Follow the dimensions given for
“Panel Cutout”. Be sure to include the four corner
screw holes. After cutting and drilling, place the front
plate on the front side of the panel with its gasket
against the panel, and the remainder of the square
housing on the back side. Slide the screws through the
four holes drilled in the panel, and into the threaded
holes in the housing. Tighten until the gasket is firmly
compressed against the panel.
Expose Terminals. Removethe four screws which hold
the front plate to its flange. Remove the front plate. The
display board is attached to this front plate. It is also
connected to the power board by a ribbon cable. For
convenience, this cable can be disconnected while
making connections. Connections can be made inside
the enclosure, or the terminals can be unplugged for
easier access, by gently tugging on them.
Sensor Connection. Followthe “Connections”diagram
to connect either two or three wires from the flow meter
or flow sensor.
MonitorAlarm Connection.Connect thealarm devices
to the appropriate relay terminals. Note that the relays
havebothnormally-open (NO)and normally-closed (NC)
contacts.
Analog Connection. This output can be configured 4-
20 mA, 0-10 V or 0-5 V by placing a jumper in the cor-
rect position on the analog header. The swith next to
the header selects active (powered) or passive
(unpowered) output. When using the 12 VDC powered
input you may only select passive output.
Power Connection. Connect AC or DC power as
desired to the appropriate terminals. For safety, if using
AC power, be sure to connect the ground terminal
provided to a good earth ground. If using DC power, be
Settings
Set Units. Press . Use to select volume
units. Use to select the particular unit desired
(gallons, liters, etc.). Then use to switch to time
units. Again, select the unit desired. Press for
next menu item.
Set Decimal. Use to select zero, one, or two
decimalplaceson the flowunits. Press toadvance
to the next menu.
Set K-Factor. The unit will not function properly
until this number is entered. It is simply the number
of pulses which the flow meter or flow sensor puts
out per gallon of liquid. It is marked on the Model/
Serial tag of SeaMetrics flow meters and flow
sensor fittings. On adjustable flow sensors, the K-
factor must be taken from the chart in the flow
sensor instructions, based on pipe size. Press
for next menu item.
Set Pulse Out. An output pulse is activated at the
selected volume intervals if this feature is in use.
Otherwise, it does not need to be set. Press for
next menu item.
Set Time Period. This is the monitoring period at
Caution: When the
control is powered up, relay
or analog outputs may be
present. If this could be a
hazard, wait to make external
connections until program-
ming is complete.
2 of 4
!
SETTINGSOPERATION MODE
SET
SET
SET
SET
SET
SET
SET
TO CLEAR TOTAL
PRESS UP & LEFT
SET 20 mA RATE
123456 GPM
return to
operation mode
Flow
Monitor
Set Pulse Out
1.00 Gal
GAL
MINUTE
0:00:00
0 GAL
ELAPSED
0 GPM
0 GAL
Flow
Rate
SET
SET
Alarm Point:
1234 GAL
Time Period:
12 HOURS
Set Decimal
0.00
Set K-factor
60.00
Displays
SET
SET