
EM_aSENSE_all_housings_installation_Feb’04
4 of 8
Mounting of
on to the duct.
11
10
10 Hole with 25 mm diameter
11 Temperature sensor with 110 mm cable mounted in the sampling probe
Mounting Instruction
Since there might be a substantial pressure difference in duct mounting applications, it is
essential to avoid ambient air from suction into the duct mounting box. For correct function it
is indispensible that the sealings of the the box cover, the cable entry bushings, the cable feed
through and the duct entrance are absolutly tight. The duct entrance may need extra sealing
paste in order to prevent leakage. The PCB must be handed carefully and protected from
electrostatic discharge.
1) Electrical cable entry: The box has a factory mounted cable entry bushing in dimension
PG9. Never feed more than one cable through each cable entry bushing, or else gas might
leak through!
2) Mounting the tube: Drill a hole (10) with 25 mm diameter (or 1 inch) for the sampling
probe and two holes with 4 mm diameter for the screws (5) into the air duct and mount
the tube (1) with the gasket (2). The sampling probe should be mounted with the largest
locking nob on top. The unit can be mounted with the air coming from the left or right.
3) Attaching the sensor box is made to the sampling probe by a snap-in bayonet fitting.
First, carefully stick the temperature probe (11) into the sampling probe. (1). Orient the
box onto the sampling probe so that the box upside is on the same side as the largest
locking nob (3). When the probe is fitted into the notches of the box, then turn the box
clockwise until stop (see Figure 1). Position 1 indicates open where the box can be
removed from the sampling probe. In position 2 the box is locked to the probe
.