METER ATEX
Safety, installation, use and maintenance instructions
Warning: this document is an integral part of the 825B105P manual and must be read and understood in conjunction with it.
Ultrasonic level transmitter European directive: 2014/34/UE:
ATEX standard references: EN 60079-0:2012+A11:2013 - EN 60079-1:2014 - EN 60079-11:2012 - EN 60079-26:2015 - EN 60079-31:2014
Nominal charateristics, Marking
Model : METER_
ATEX Group : II 1/2G; II 1D
Protection mode: ia/db IIC T4 Ga/Gb; ta IIIC T135°C Da
Ambient temperature: -20 ÷ +60°C
Protection degree IP: 66
Max process temperature: +60°C
1 - CERTIFIED ATEX VERSION IDENTIFICATION
1. Product Code
2. Notiedbodyidenticationcode
2 - SAFETY INSTRUCTIONS
This ultrasonic level meter is designed to be installed in a potentially explosive atmosphere due to the presence of gas and dust.
Makesurethatthemarkingofthedevicecomplieswiththeclassicationofthepotentiallyexplosivearea.
Verifythattheinstallationoftheprobedoesnotaectthetypeofprotection.
The device can only be opened under voltage in the absence of a potentially explosive atmosphere, to be checked by
means of an explosimeter.
Incorrect use of the device could cause damage to people, to the product and to the equipment connected to it.
Always respect the plate data for power supply and electrical connections.
3 - INSTALLATION AND COMMISSIONING
Theinstallationmustonlybecarriedoutbyqualiedpersonnel,adequatelytrained,incompliancewiththeregulationsinforce
EN 60079-14.
Theequipmentmustbeusedonlyaerhavingcorrectlyunderstoodtheinstructionsinthisdocument,togetherwiththe
825B105P manuals.
The device must be mounted in such a way as to minimize the risk of shocks to the case and to the sensitive part (transducer).
Useonlythethreadedttingtoinserttheprobeintotheprocess;donotforceinsertionusingthecaseasapivotpoint.
Properly tighten the cable glands and the cover before putting the instrument into operation.
Make sure that the metal case is connected to earth through the appropriate connection.
If it is necessary to replace the cable entries with other similar devices, make sure that the marking of the latter is compatible with the
installation area and the type of protection and that the minimum degree of dust tightness is respected (IP6X).
SI18001A_EN
3.Certicatenumber
4. Serial number
5. Marking
6. Supply voltage