Display Description
Conv PCS Piece counting
Conv % Percentage conversion
CAL Calibration
P.C.-on Auto Power Off function activated
P.C.-oF Auto Power Off function disabled
A-1 Fast Response Mode
A-2 Normal Response Mode
P-1 Internal stability detection width: Small
P-2 Internal stability detection width: Large
d-on Zero tracking: ON
d-off Zero tracking: OFF
iF RS-232C communication specification settings
SG Specific gravity measurement
*1Unit kg Kilogram
Unit g Gram
Unit ct Carat
*2Unit oz Ounce
*2 Unit ozt Troy ounce
*2 Unit dwt Penny weight
*2 Unit GN Grain
*2 UnitH tl Hong Kong tael
*2 UnitS tl Singapore tael
*2 Unitt tl Taiwan tale
*2 Unit mo Momme
CLEAr Canceling registered units
*1ELB12K only
*2Not applicable to Leagal Mesuring Instruments in
using area.
Operating Guide
ELB Series
[Menu Selection]
(1) Press the [MODE] key. The currently selected
conditions are displayed in symbols.
(2) The display changes in the order shown in the
table at right each time the [MODE] key is
pressed. Press the [TARE] key when the item
to set is displayed.
(3) “SEt” is displayed and it reverts to the weight
display.
CAUTION! The internal values of the balance
are being revised while “SEt” is
displayed. Do not turn off the
power while “SEt” is displayed.
CAUTION! Menu selection is not possible in
the specific gravity measurement
mode. Display a different unit by
pressing the [UNIT] key to start
menu selection.
[Registering and Deleting Units]
Registration
(1) Press the [TARE] key when the desired unit is displayed. Already registered units are displayed with the
“→” mark.
(2) “SEt” is displayed and then it changes to the weight display in the registered unit.
Deletion
(1) Press the [TARE] key when the unit to delete is displayed. Already registered units are displayed with the
“→” mark.
(2) “SEt” is displayed and then it changes to the weight display.
Delete All
(1) Select “CLEAr” and press the [TARE] key.
(2) “SEt” is displayed and then it changes to the weight display in grams.
[Sensitivity Calibration]
(1) Select “CAL” (sensitivity (span) calibration) in menu selection with nothing
on the pan.
(2) Press the [TARE] key. The value of the weight to be used for calibration flashes.
(3) Place the calibration weight on the pan.
(4) Press the [TARE] key when the stability mark illuminates after placing the
weight.
(5) Zero flashes on the display.