manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Sloan
  6. •
  7. Measuring Instrument
  8. •
  9. Sloan CX Guide

Sloan CX Guide

CX Manual Flushometers
TROUBLESHOOTING GUIDE
1
2
Item No. Part No. Description
1See Table Piston Assembly
2See Table Activation Assembly
Code No. Description Plug Color*
3340048 1.6 gpf/6.0 Lpf closet activation assembly Green
3340049 1.28 gpf/4.8 Lpf closet activation assembly Purple
3340051 0.5 gpf/1.9 Lpf urinal activation assembly Red
3340052 0.25 gpf/1.0 Lpf urinal activation assembly Burgundy
ACTIVATION ASSEMBLY SELECTION GUIDE
* Colors may differ. Consult factory to conrm you have the correct piston assembly.
Item Code No. Description
1 0323182 V-651 Vacuum Breaker
2 0323279 Short Vacuum Breaker Tube
3 0306394 Adjustable Tube
4 0306395 Poly Washer Coupling (set of two)
5 0306392PK Elbow
6 3306002 1-1/2" Coupling with S-21 Gasket
7 0306031PO 1 1/2" (38 mm) Flanged outlet tube RB, 6"
8 0396669PK F-102 1-1/2" (38 mm) Outlet Tube CP 8"*
9 0306238PK F-7 Flange CP*
10 0306146PK F-5-A 1-1/2" Spud Coupling Assembly CP*
11 0306396 Urinal Connection
FLUSH CONNECTION PARTS
Plug
The information contained in this document is subject to change without notice.
Manufactured in the U.S.A by Sloan Valve Company under one
or more of the following patents: U.S. Patents. 5,295,655;
5,542,718; 5,558,120; 5,564,460; 5,730,415; 5,865,420;
5,887,848; 5,967,182. Other Patents Pending.
Bak-Chek®, Para-flo®, PERMEX®, Turbo-Flo®
* Consult factory for alternate nish options
1
2
3
410
5
4
8
7Install with
Flange against spud
For Model 158, Wall Hung For Model 158, Floor Mount
9
4
For Model 198
11
Code No. Description Piston Color*
3340053 1.6 gpf/6.0 Lpf closet activation assembly Green
3340054 1.28 gpf/4.8 Lpf closet activation assembly Purple
3340055 0.5 gpf/1.9 Lpf urinal activation assembly Red
3340056 0.25 gpf/1.0 Lpf urinal activation assembly Burgundy
PISTON ASSEMBLY SELECTION GUIDE
CX Manual Flushometers
TROUBLESHOOTING GUIDE
1. Flushometer does not function (no water/flush).
A. Ensure water supply to valve is turned on.
B. Shut off sleeve is in closed position. Turn counterclockwise to open water ow.
C. Plunger screw is not adjusted properly. Adjust screw using depth guide.
2. Volume of water is not sufficient to evacuate the bowl.
A. Shut off sleeve is not open wide enough. Adjust shut off sleeve for desired delivery of water volume.
B. Urinal piston parts installed in a Closet ushometer. Replace piston with proper Closet piston (1.6 gpf/6.0 Lpf or 1.28 gpf/4.8 Lpf)
C. Water supply volume or pressure is inadequate (low). If the xture does not siphon or evacuate the bowl, steps must be taken to increase the
water supply pressure and/or volume. Contact the xture manufacturer for minimum water supply requirements of the xture.
D. Plunger screw is not adjusted properly. Adjust screw using depth guide.
3. Flushometer closes off immediately.
A. Urinal piston parts installed in a Closet ushometer. Replace piston with proper Closet piston (1.6 gpf/6.0 Lpf or 1.28/4.8 Lpf).
B. Piston lip seal may be damaged or worn. Replace with new correct gpf/Lpf piston.
C. Plunger screw not set properly. Adjust screw using depth guide.
4. Length of flush is too short (Short Flush).
A. Incorrect piston is installed in Flushometer; for instance, Urinal assembly inside a Closet Flushometer,
or High Efciency assembly inside a higher consumption xture. Determine the ush volume required by the xture and replace with proper gpf/
Lpf piston. Use valve label and markings on xture for reference.
B. Piston lip seal may be damaged or worn. Replace with new correct gpf/Lpf piston.
C. Plunger screw is not adjusted properly. Adjust screw using depth guide.
5. Length of flush is too long (Long Flush) or continuous.
A. Metering bypass hole in piston is clogged. Remove the piston from the valve, slide o ring off lter, and wash under clean running water.
Reposition o-ring before reinstalling piston. Replace piston if cleaning does not correct the problem.
B. Supply line water pressure has dropped and is not sufcient to close the valve. Close shut off sleeve until pressure is restored.
D. Piston is damaged. Replace with new proper gpf/Lpf piston
E. Incorrect Piston is installed in Flushometer; for instance, Closet assembly inside a Urinal ushometer. Determine the ush volume required by the
xture and replace piston. Use valve label and markings on xture for reference.
6. Chattering noise is heard during flush.
A. Reduce ow pressure by turning shut off sleeve clockwise.
CARE AND CLEANING
DO NOT use abrasive or chemical cleaners (including chlorine bleach) to clean Flushometers that may dull the luster and attack the chrome or special
decorative nishes. Use ONLY mild soap and water, then wipe dry with clean cloth or towel. While cleaning the bathroom tile, protect the Flushometer
from any splattering of cleaner. Acids and cleaning uids will discolor or remove chrome plating.

Other Sloan Measuring Instrument manuals

Sloan Optima User manual

Sloan

Sloan Optima User manual

Sloan CX 158 Use and care manual

Sloan

Sloan CX 158 Use and care manual

Sloan G2 Optima Plus Use and care manual

Sloan

Sloan G2 Optima Plus Use and care manual

Sloan EL-750/Hardwired Use and care manual

Sloan

Sloan EL-750/Hardwired Use and care manual

Sloan TruFlush Use and care manual

Sloan

Sloan TruFlush Use and care manual

Sloan NAVAL Operating manual

Sloan

Sloan NAVAL Operating manual

Sloan TruFlush TRF 156 Use and care manual

Sloan

Sloan TruFlush TRF 156 Use and care manual

Sloan TruFlush TRF 8156 Use and care manual

Sloan

Sloan TruFlush TRF 8156 Use and care manual

Sloan CX 8154-1.6 User manual

Sloan

Sloan CX 8154-1.6 User manual

Sloan Royal Use and care manual

Sloan

Sloan Royal Use and care manual

Sloan TRF 8156-1.28 User manual

Sloan

Sloan TRF 8156-1.28 User manual

Sloan G2 Optima Plus Instruction sheet

Sloan

Sloan G2 Optima Plus Instruction sheet

Sloan CX Series User manual

Sloan

Sloan CX Series User manual

Sloan G2 Optima Plus Instruction sheet

Sloan

Sloan G2 Optima Plus Instruction sheet

Sloan EBV-200-A SMOOTH User manual

Sloan

Sloan EBV-200-A SMOOTH User manual

Sloan CX 8158 Use and care manual

Sloan

Sloan CX 8158 Use and care manual

Popular Measuring Instrument manuals by other brands

YOKOGAWA TM20 instruction manual

YOKOGAWA

YOKOGAWA TM20 instruction manual

Harbor Freight Tools Cen-Tech 5645 Assembly & operating instructions

Harbor Freight Tools

Harbor Freight Tools Cen-Tech 5645 Assembly & operating instructions

Value WT26 Series manual

Value

Value WT26 Series manual

Ebro FOM 310 manual

Ebro

Ebro FOM 310 manual

BEKA BA558C manual

BEKA

BEKA BA558C manual

HomeFree FDU 800 user manual

HomeFree

HomeFree FDU 800 user manual

GE PACE 1000 Safety Instructions and User Guide

GE

GE PACE 1000 Safety Instructions and User Guide

Siemens 7MC3052-1AB68 operating instructions

Siemens

Siemens 7MC3052-1AB68 operating instructions

Analytical Technology A15 owner's manual

Analytical Technology

Analytical Technology A15 owner's manual

Vanguard Instruments Company WRM-10P user manual

Vanguard Instruments Company

Vanguard Instruments Company WRM-10P user manual

Endress+Hauser Proline Promag 50 operating instructions

Endress+Hauser

Endress+Hauser Proline Promag 50 operating instructions

LaserLiner LaserRange-Master i3 manual

LaserLiner

LaserLiner LaserRange-Master i3 manual

Rosemount 2700 instruction manual

Rosemount

Rosemount 2700 instruction manual

Olympus 38DL PLUS user manual

Olympus

Olympus 38DL PLUS user manual

Reely 1719875 operating instructions

Reely

Reely 1719875 operating instructions

HIOS HIT Series owner's manual

HIOS

HIOS HIT Series owner's manual

Extech Instruments MG310 user manual

Extech Instruments

Extech Instruments MG310 user manual

Orbitalum ORBmax Translation of original operating instructions

Orbitalum

Orbitalum ORBmax Translation of original operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.