Sirius Audio Converters www.snellgroup.com Hybrid Audio Cards
Issue 3 Rev 2 Page 12 © 2014 Snell Limited
SW5
Position Description
SW5-1
Select either a free running count in the channel status sample address field or Legacy
mode for older or non compliant AES equipment.
SW5-1 UP
Selects a dynamic free-running count in the Channel Status Sample-Address
field. Each block of Channel Status will have a different value in this field and
calculates a new CRCC-value. This tests downstream equipment for correct
CRCC-detection and could be used to track missing samples.
SW5-1
DOWN
Selects “Legacy-mode”. This sets the Sample-Address field at all zeros. This
results in static Channel Status, as may be required to satisfy older or
non-compliant AES equipment.
SW5-2
On power-up each ADC must run its Calibration routine at least once. This measures
DC-offsets on the input pins and then subtracts this value from all future conversions. This
results in a better, lower noise-floor figure.
If the AES Reference is removed & re-connected, the disruption this causes to the internal
clocks usually invalidates the calibration.
For best noise-floor when the Reference is lost, or never present, use “Once-only
calibration” (SW5-2 UP).
For best noise-floor after the Reference is reconnected, use “Re-calibrate after Ref-lock”
(SW5-2 DOWN).
SW5-2 UP Selects “Once-only Calibration”, this occurs only on power-up.
This is the normal situation for a system where the AES reference is not
expected to change.
SW5-2
DOWN
Selects “Re-calibrate after Reference-lock”. Every time the card detects the
change from No-reference (red NOREF Led On) to a good Reference (NOREF
Led off), the calibration routine is run after a delay set by SW5-3.
This may cause a short period of muted samples during measurement, but
gives a better noise-figure once completed. Ideally any analogue signals
connected to the ADC inputs should be muted or unpowered during
re-calibration for best results.
SW5-3
Select the delay before the calibration routine is run.
SW5-3 UP Selects a 4-second delay before running the calibration routine.
SW5-3
DOWN
Selects a 10-second delay before running the calibration routine.
This setting applies to both power-up calibration and Re-calibrate after Ref-lock
(if used).
SW5-4
Enable or disable gain adjustment buttons.
SW5-4 UP Disables the UP and DOWN buttons for gain adjustment.
SW5-4
Down The UP and DOWN buttons increment and decrement the gain of the mono
analogue channel, see Table 4.
Table 3. SW5 DIL Switch Settings