Solinst Levelogger Gold User manual

Levelogger Series
(Levelogger Gold, Barologger Gold, Levelogger Junior, LTC Levelogger Junior and Rainlogger)
User Guide - Software Version 3.4.1
June 21, 2011


Levelogger User Guide - Table of Contents
1 Introduction to Solinst Levelogger Series and Software 1
Levelogger Gold 1
Barologger Gold 2
Levelogger Junior 3
LTC Levelogger Junior 4
Rainlogger 5
Leveloader Gold 6
Software Communication 6
1.1 Level 7
1.1.1 Pressure Calibration 7
1.2 Barologger 7
1.3 Temperature 8
1.4 Conductivity 8
1.4.1 Calibration of the Conductivity Sensor 8
1.5 Total Rainfall 9
1.6 Backwards Compatibility 9
2 System Requirements 10
3 Software Installation 10
3.1 Installing USB Drivers for Levelogger Software 11
USB Installation for Windows 98 11
USB Installation for Windows 2000 12
USB Installation for Windows XP 13
USB Installation for Windows Vista 15
USB Installation for Windows 7 17
Manual USB Installation 19
4 Startup, Configurations and Settings 20
Startup 20
Communication Port Settings 20
4.1 Communicating with a USB Port 20
4.2 Com Port Designation Set Up: 21
4.3 Data Directory 21
5 Levelogger Settings 22
5.1 Levelogger Information Settings 22
5.2 Setting up Channel Information 24
The ‘Level’ Channel 24
The ‘Temperature’ Channel includes the following: 25
The ‘Conductivity’ or Channel 25
5.3 Levelogger Status 26
5.4 Setting up the Levelogger Time 26
5.5 Program Levelogger Settings 26
5.6 Starting and Stopping the Levelogger 26
5.7 Saving and Retrieving Levelogger Settings Files 27
6 LTC Levelogger Junior Calibration 28

Levelogger User Guide - Table of Contents
7 Rainlogger 32
7.1 Rainlogger Communication Interface 32
Rain Gauge Connector 32
7.2 Rainlogger Setup 33
8 Data Control Window 35
8.1 Downloading Options 35
8.2 Graph Manipulation and Zoom Function 37
8.3 Data Compensation 39
9 Real Time View Window 41
10 File Export and Print Function 42
11 Installation and Maintenance of Leveloggers 43
11.1 Installation 43
11.1.1 Free Suspended Installations 44
Suspension Wire/Rope Installation 44
Direct Read Cable Assembly Installation 45
11.1.2 Fixed Installations 46
Artesian Monitoring 46
Vacuum Monitoring 47
Barologger 48
Levelogger 48
11.1.3 Barologger Installation 48
Barometric Efficiency 48
11.1.4 Installation in Extreme Environments 50
Freezing or High Temperature Conditions 50
Marine or Brackish Installations 51
11.2 Levelogger Maintenance 52
12 Manual Barometric Compensation 53
13 Diagnostics Utility 56
13.1 Read Levelogger Information 56
13.2 Run Self-Test 57
13.3 Read Memory Dump 58
13.4 Create Report 59
13.5 Restore Factory LTC Calibration 60
13.6 LTC Calibration History Report 60
14 Firmware Uploader Utility 61
15 Trouble Shooting 62
15.1 Problems During Installation of Levelogger Gold Software 62
15.2 Error During Software Uninstall Process 62
15.3 Problems During Installation of RS232 to USB Converter 62
15.4 Data Has Been Erased Accidentally 63
15.5 Error Messages During Use of Software 63


Levelogger User Guide - Software Version 3.4.0 or Higher
Page 1
Note:
This version
of software is
not compatible
with older models of the
Levelogger (Silver LT and
LTC, Black LTC and old
Rain Loggers). To program
and use the old versions,
Levelogger 3.1.1 Software
and User Guides can still be
accessed at: www.solinst.
com/Downloads/
For Software version 2.0.3,
contact Solinst.
1 Introduction to Solinst Levelogger Series and Software
Levelogger Gold Software Version 3.4.0 is intuitive and easy-to-use. It is designed
to support the Levelogger Gold Series of products.
Levelogger Gold
The Levelogger Gold is an absolute (non-vented) datalogger, which
measures groundwater and surface water levels and temperature. Water
levels are displayed as temperature compensated pressure readings, and
can be barometrically compensated with the aid of a Barologger Gold.
Solinst recommends customers use Version 3.4.0 Software using with V2.006
firmware or higher.
Levelogger Gold Technical Specications
Level Sensor: Piezoresistive Silicon in 316L Stainless Steel
Accuracy (Typical): 0.05% FS
Stability of Readings: Superior, low noise, 6 times better than previous versions
Resolution: 0.002 to 0.0006% FS
Normalization: Automatic Temp Compensation
Temperature Sensor: Platinum Resistance Temperature Detector
Temp. Sensor Accuracy: ± 0.05°C
Temp. Sensor Resolution: 0.003°C
Temp. Comp. Range: -10 to +40°C
Battery Life: 10 Years - based on one reading/min
Clock Accuracy: ± 1 minute/year
Operating Temperature: -20°C to 80°C
Maximum # Readings: 40,000 of level and temperature
Memory: Superior reliability EEPROM Slate, rollover
and redundant backup of last 1200 readings
Communication: Optical Infra-Red Interface, Serial at 9600 Baud, Conversion
to RS232 or USB Computer Connection
Size: 7/8" x 6" (22 mm x 154 mm)
Weight: 6.3 oz (179 grams)
Backwards Compatibility: Full
Corrosion Resistance: Zirconium Nitride (ZrN) Coating
Other Wetted Materials: 316-L Stainless Steel, Delrin, Viton
Sampling Modes: Linear, Event and User-Selectable with
30 separate line items
Measurement Rates: 0.5 sec to 99 hrs
Barometric Compensation:
Software Wizard and one Barologger in local area
(approx. 20 miles/30 km radius and or per 1000 ft/300 m
change in elevation)
LT Models Full Scale
(FS)
Accuracy
(Typical) Resolution
F15, M5 16.4 ft., 5 m ± 0.010 ft., 0.3 cm 0.001% FS
F30, M10 32.8 ft., 10 m ± 0.016 ft., 0.5 cm 0.0006% FS
F65, M20 65.6 ft., 20 m ± 0.032 ft., 1 cm 0.0006% FS
F100, M30 98.4 ft., 30 m ± 0.064 ft., 1.5 cm 0.0006% FS
F300, M100 328.1 ft., 100 m ± 0.164 ft., 5 cm 0.0006% FS
Note:
Let Solinst keep
you up-to-date
with each new
software and firmware
release. Register your
software at:
www.solinst.com/
Registration/ to receive
these updates automatically.

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 2
Barologger Gold
The Barologger Gold uses algorithms based on air pressure only. It measures
and logs changes in atmospheric pressure, which are then used to compensate
water level readings recorded by a Levelogger Gold, or a Levelogger Junior.
LT Models Full Scale
(FS)
Accuracy
(Typical) Resolution
Barologger 0.05% ± 0.003 ft., 0.1 cm 0.002% FS

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 3
Levelogger Junior
The Levelogger Junior provides an inexpensive alternative for measuring
groundwater and surface water levels and temperature. The Levelogger Junior is
compatible with all Levelogger Gold accessories. Solinst recommends customers
use Version 3.4.0 Software with V2.006 firmware or higher.
Levelogger Junior Technical Specications
Level Sensor: Piezoresistive Silicon in 316L SS
Ranges: F15/M5, F30/M10
Accuracy (typical): 0.1% FS
Resolution: 0.028%, 0.021%
Normalization: Automatic Temp Compensation
Temperature Sensor: Platinum RTD
Accuracy: ± 0.1˚C
Resolution: 0.1˚C
Temp Compensation Range: - 10˚C to 40˚C
Battery life: 5 Years
Clock Accuracy: ± 1 minute/year
Operating Temperature: - 20˚C to 80˚C
Memory: Non-volatile EEPROM, FRAM back-up
Maximum # Readings 32,000 sets of readings
Communication: Optical Infra-Red to USB or RS232
Size: 7/8” x 5.5” (22 mm x 140 mm)
Weight: 154 g (5.4 oz)
Wetted Materials: 316L Stainless Steel, Delrin, Viton
Sampling Mode: Linear, SDI-12, Real Time View
(from 0.5 sec to 99 hrs)
Barometric Compensation:
Software Wizard and one Barologger in local area
(approx. 20 miles/30 km radius and or per 1000 ft/300 m
change in elevation)
LT Models Full Scale
(FS)
Accuracy
(Typical) Resolution
F15, M5 16.4 ft., 5 m ± 0.020 ft., 0.6 cm 0.028% FS
F30, M10 32.8 ft., 10 m ± 0.032 ft., 1.0 cm 0.021% FS

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 4
LTC Levelogger Junior
The LTC Levelogger Junior provides an inexpensive and convenient method
to measure level, temperature and conductivity all in one probe. The LTC
Levelogger Junior is compatible with all Levelogger Gold accessories. Solinst
recommends customers use Version 3.4.0 Software with V2.001 firmware or
higher.
LTC Levelogger Junior Technical Specications
Level Sensor: Piezoresistive Silicon in 316L SS
Ranges: F30/M10, F100/M30
Accuracy (typical): 0.1% FS
Resolution: 0.08" (0.2 cm), 0.24" (0.6 cm)
Normalization: Automatic Temp Compensation
Temperature Sensor: Platinum RTD
Accuracy: ± 0.1˚C
Resolution: 0.1˚C
Temp Compensation Range: -10˚C to 40˚C
Conductivity Sensor: 4-Electrode Platinum
Full Range: 0 - 80,000 µS/cm
Calibrated Range: 500 - 50,000 µS/cm
Accuracy: 2% of reading or 20 µS/cm
Resolution: 1 mS
Temp. Compensation Range: 0˚C to 40˚C
Normalization: Specic Conductance @ 25˚C
General:
Battery Life: 5 Years (5 minute sampling rate)
Clock Accuracy: ± 1 minute/year from 0 to 40˚C
Operating Temperature: - 20˚C to 80˚C
Memory: Non-volatile EEPROM, FRAM back-up
Maximum Readings: 16,000 sets of readings
Communication: Optical Infra-Red to USB or RS232
Size: 7/8" x 7.5" (22 mm x 190 mm)
Wetted Materials: Platinum, 316 Stainless Steel, Delrin®, Viton®
Sampling Mode: Linear, (5 sec to 99 hrs) SDI-12, Real Time View
Barometric Compensation: Software Wizard and Barologger Gold
LT Models Full Scale
(FS)
Accuracy
(Typical) Resolution
F30, M10 32.8 ft., 10 m ± 0.032 ft., 1.0 cm 0.021% FS
F100, M30 98.4 ft., 30 m ± 0.128 ft., 1.5 cm 0.021% FS

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 5
Rainlogger
The Rainlogger is designed for use with most standard tipping-bucket rain gauges
with a reed switch output. It measures total rainfall per sampling period and a
five-minute maximum rainfall (intensity). Now users can add rainfall data to their
Levelogger network. Compatible with Version 3.4.0 Software using V1.003
firmware or higher
Rainlogger Technical Specications
Battery Life: 10 years (logging two parameters/10 minutes)
Clock Accuracy: ± 1 minute/year
Operating Temperature: -20˚C to 80˚C
Resolution: 0.004 to 0.008" (0.1 to 0.2 mm) (dependent on rain gauge type)
Maximum # Readings: 40,000 sets of readings
Memory: Non-volatile EEPROM / FRAM back-up
Communication: Optical infra-red to RS232 or USB
Sampling Mode: Linear
Sampling Rate: 5 minutes to 99 hours (5 minute intervals)
Size: 7/8" x 7" (22 mm x 175 mm)
Weight: 4.8 oz (135 grams)
Materials: Stainless Steel, Delrin, Lexan, Viton

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 6
Leveloader Gold
The Leveloader Gold is a data transfer device for use with all versions of
Leveloggers. It can be used to download, store, and transfer data from Levelogger
to PC, as well as, save settings files for transfer to a Levelogger. Also allows real-
time viewing of Levelogger readings. Compatible with Version 3.4.0 Software
or higher using V1.200 firmware or higher.
Software Communication
Leveloggers connect to a laptop or desktop PC with an optical reader cable.
The optical reader cable uses an infrared data reader/port connected to the
Levelogger and an RS232 or USB Com Port to transfer information between
the Levelogger and computer.
Version 3.4.0 Software provides Levelogger Gold users with many convenient
features. Users can view and program Levelogger settings, begin logging
sessions, monitor real-time readings, download data, manage data files, perform
compensations, and save and export files, with this software version.

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 7
1.1 Level
The Levelogger Gold and Levelogger Junior use a high quality piezoresistive
silicon pressure transducer packaged in 316L stainless steel housing. It gives high
accuracy and high stability. The Levelogger Gold body is coated with Zirconium
Nitride (ZrN) to give corrosion resistance. All Leveloggers measure total or
absolute pressure. When the Levelogger is operating in open air, it is recording
barometric pressure and converting that pressure reading to its water level
equivalent above the datalogger’s pressure zero point of 950 cm (31.17ft). When
submerged, it is recording the combination of barometric pressure and water
pressure. The Levelogger converts the total pressure reading to its corresponding
water level equivalent. Actual water level is obtained by compensating for
barometric pressure. The best method to compensate for barometric pressure
is to employ a Barologger above the water level, somewhere on site, to obtain
records of barometric pressure. The Levelogger Software includes a Barometric
Compensation Wizard, which guides the user through the automated process of
barometric compensation. Manual methods can be employed to determine the
absolute water level using barometric records collected on-site or available from
a local weather station (i.e. Airport). Water level readings from the Levelogger
Gold and Levelogger Junior models are temperature-compensated.
1.1.1 Pressure Calibration
The Levelogger Gold and Levelogger Junior are calibrated against a range of set
reference points to an accuracy of 3 decimal places. The units of pressure are
in pounds per square inch. The conversion factor for pounds per square inch
relates to pressure as follows:
1 pound per square inch = 0.0689476 bar
= 0.703070 m
= 703.070 kg/m2
= 6.895 kPa
= 2.31 ft. H20 @ 4°C.
During the calibration procedure, the Levelogger is fully submerged in a highly
accurate water bath. The bath is set to 15°C and allowed to stabilize. The
pressure is then calibrated to six separate pressure points covering the entire
range of pressure for that particular Levelogger to check for any non-linearity.
The process is repeated again at 35°C to check for temperature effects. Once
done, the Levelogger is approved after all specifications for accuracy, precision,
stability and hysteresis have been met. The Levelogger should be calibrated for
the life-time of the instrument, as long as it is used within its specified range.
1.2 Barologger
The Barologger Gold is designed for use in air only. It has a specific range and
firmware algorithms based on air pressure rather than water pressure. This
makes the Barologger less accurate if used in water, but more accurate if used as
intended, in air. Using a Barologger is the most accurate and convenient method
of obtaining atmospheric pressure and air temperature. When programmed
with the same sampling parameters as the Leveloggers on site, a Barologger
can avoid barometric data time lags and any errors introduced due to moisture
buildup, kinking or damage to vented cable. The Barometric Compensation
Wizard in the Levelogger Software simplifies the adjustment of the level
measurements for barometric pressure changes, by using the synchronized data
from all Leveloggers on site and the site Barologger.

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 8
1.3 Temperature
Levelogger Gold and Levelogger Junior, record temperature compensated
water levels. Groundwater and surface water temperature measurements
are particularly important in situations where temperatures may fluctuate
significantly, and when temperature is used in determining liquid level, viscosity
and hydraulic conductivity. Temperature monitoring is vital in the calculation
of certain temperature dependent contaminant reaction rates. A Platinum
Resistance Temperature Detector is used to accurately compensate for
temperature changes within the range of -10 to +40˚C. The Leveloggers will
record temperature in its thermal range of -20 to +80˚C, but outside the range
of -10 to +40˚C compensation will be less accurate.
1.4 Conductivity
The LTC Levelogger Junior provides the added feature of electrical conductivity
measurement. Conductivity measurement is particularly useful in monitoring
saltwater intrusion, surface water infiltration and mixing, as well as the monitoring
of certain pollutants and contaminant parameters. Conductivity is measured
via a platinum 4-electrode sensor, which produces highly stable and consistent
readings. The sensor is relatively insensitive to dirt and deposits, keeping
maintenance to a minimum. Conductivity calibration is performed by using a
liquid solution with a known conductivity value and the Calibration Wizard in
the Levelogger Windows Software. Raw conductivity measurements can be
converted to conductivity at 25˚C, which is referred to as Specific Conductance.
When programming the LTC the user can select either raw conductivity
measurements or specific conductance.
1.4.1 Calibration of the Conductivity Sensor
The conductivity sensor of the LTC Levelogger Junior is a highly sensitive device
requiring regular calibration by the user. The conductivity calibration frequency
is dependent on the water quality of the Levelogger’s monitoring environment.
When monitoring in freshwater with good water quality, conductivity calibration
may only be required once per year. On the other hand, conductivity calibration
will be required more frequently when monitoring in polluted, salty, brackish,
eutrophic or highly conductive water. To determine whether an LTC Levelogger
Junior should be recalibrated, test the unit in a solution with a known electrical
conductivity value at a reference temperature. If that reference temperature is
25°C, set the LTC to record Specific Conductance in the ‘Conductivity’ Channel,
and observe the current reading by using the Read Now function in the Real Time
View tab. If this reading varies from the known Specific Conductance of the
solution by greater than 2% of the Full Scale of Conductivity measurement, the
unit should be recalibrated using the Conductivity Calibration Wizard procedure
outlined in Section 6. Ensure that conductivity calibration is performed when
the LTC is being set up for its initial use and after long periods of dry storage.

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 9
1.5 Total Rainfall
The Solinst Rainlogger is designed to count the tips of an external tipping-bucket
rain gauge within a user defined sample interval and output the total rainfall over
that sample interval. The Rainlogger can store 40,000 readings. The Rainlogger
is designed to be compatible with the Solinst Levelogger series of products. The
Rainlogger is programmed and data viewed and exported using the Levelogger
Software from a PC, it can be communicated with using a Leveloader and can
be integrated into an STS Telemetry System.
1.6 Backwards Compatibility
The Levelogger Gold and the Software Versions 3.4.0 or higher are backward
compatible, with limitations. If a Levelogger Gold is to be used with the
Leveloader I, Leveloader II, a Protocol Converter, or a first generation STS
or RRL Telemetry System, the datalogger must be programmed with the old
Levelogger Version 2.0.3 Software. Programming Levelogger Gold loggers
using Version 2.0.3 Software limits the capabilities of the Gold unit, to be the
same as the older stainless steel Levelogger units. If the user has a mix of
Levelogger Gold and older units, they can use the Levelogger Gold Version 3.1.1
Software to program and read all the dataloggers. They will get the accuracy
and features inherent in the older stainless steel Leveloggers, but will obtain the
higher accuracy and enhanced features and functions of the Levelogger Gold
dataloggers.
The LT Levelogger Junior is only compatible with Software Version 3.2.1 and
higher, the LTC Levelogger Junior only is compatible with Version 3.3.0 and
higher, and the Rainlogger is only compatible with Software Version 3.2.2 and
higher.
Note:
This version
of software is
not compatible
with older models of the
Levelogger (Silver LT and
LTC, Black LTC and old
Rain Loggers). To program
and use the old versions,
Levelogger 3.1.1 Software
and User Guides can still be
accessed at: www.solinst.
com/Downloads/
For Software version 2.0.3,
contact Solinst.

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 10
2 System Requirements
The minimal hardware and software requirements for software installation and
operation are:
Hardware Software
Memory: 32MB or more OS: Windows 98, 2000, XP, Vista, or 7
Display: VGA: 800 x 600 pixels, 256 colour
Ports: USB or RS232 Serial Port
Bits per second 9600
Data bits 8
Stop bits 1
Flow control None
Figure 3-1 Software Installation Wizard
Communication Port Setting for Levelogger Communications:
3 Software Installation
• Place the Levelogger CD in the CD ROM drive or download the
software from www.solinst.com/Downloads/ and save to hard drive.
• Open Windows Explorer and double click the setup.exe icon.
• At this point Windows will prompt you through the remainder of the
installation process. Figure 3-1 shows the Levelogger Installation
Wizard.
• Restart the computer after installation is completed. Default
Directory is C:\Program Files\Solinst\Levelogger3_4 (or higher)

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 11
3.1 Installing USB Drivers for Levelogger Software V3.4.0 or Higher
Levelogger Software V3.4.0 or higher, for use with Solinst USB Optical Reader,
PC Interface Cable and Leveloader Gold, comes equipped with USB drivers for:
Windows 98, 2000, XP, Vista, and Windows 7.
USB Installation for Windows 98
1) Plug the USB device into the computer, and Windows will automatically
detect the connected device and start the Hardware Installation Wizard.
2) A message may prompt the user that an ‘FTDI USB Drives Disk’ must be
inserted (Figure 3-2). A disk is not needed. Just click ‘OK’.
Figure 3-2 Insert Disk Window
Figure 3-2 Insert Disk Window
3) On the next screen select ‘Display a list of all the drivers in a specific location,
so you can select the driver you want’ (Figure 3-3).
4) On the next screen use the ‘Browse’ function to select the appropriate
driver. The directory should be:
C:\Program Files\Solinst\Levelogger3_4\USB Drivers\Win_98
Complete and finish the installation. Please note that a system restart may
be required.
Note:
After plugging in
the USB device,
if the Hardware
Installation Wizard does
not automatically appear,
proceed to the Manual USB
Installation instructions.

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 12
USB Installation for Windows 2000
1) Plug the USB device into the PC and Windows will automatically detect the
connected unit.
Figure 3-4 Found New Hardware Pop-up Window
2) This will start the ‘Found New Hardware Wizard’, click next.
3) In the ‘Install Hardware Device Drivers’ window, select ‘Search for a suitable
driver for my device (recommended)’.
4) In the ‘Locate Driver File’ window (Figure 3-5), select ‘Specify a location’
and click ‘Next’.
Figure 3-5 Found New Hardware Wizard
5) In the next window point the browser to the USB driver folder located in the
Levelogger 3_4 folder. Default Destination:
C:\Program Files\Solinst\Levelogger3_4\USB Drivers\
Win_2000_XP_Vista_7
6) Once the installation completes, a system restart may be required.
Note:
After plugging in
the USB device,
if the Found
New Hardware Wizard does
not automatically appear,
proceed to the Manual USB
Installation instructions.

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 13
USB Installation for Windows XP
1) Plug the USB device into the computer, and Windows will automatically
detect the connected device. This will begin the ‘Found New Hardware
Wizard’.
2) The ‘Wizard’ will give the option to connect to Windows Update to search
for software components, select: “No, not at this time”.
3) Click ‘Install from a list or specific location’, then click the ‘Next’ Button.
(Figure 3-6).
Figure 3-6 Found New Hardware Wizard Window
4) Select the installation option, ‘Include this location in the search’ (Figure3-7),
then click the ‘Browse’ Button to search for the appropriate directory:
C:\Program Files\Solinst\Levelogger3_4\USB Drivers\
Win_2000_XP_Vista_7
Figure 3-7 Found New Hardware Window and Figure 3-8 Browse for Folder Window
Note:
After plugging in
the USB device,
if the Found
New Hardware Wizard does
not automatically appear,
proceed to the Manual USB
Installation instructions.

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 14
Figure 3-9 Found New Hardware Search Window
5) A warning message will then prompt that the software has not passed the
Windows Logo Test. Select ‘Continue Anyway’ (Figures 3-10). This will
complete the installation process. A system restart may be required.
Figure 3-10 Hardware Installation Window
Figure 3-11 Found New Hardware Completed Window

Levelogger User Guide - Software Version 3.4.0 or Higher
Page 15
USB Installation for Windows Vista
1) Plug the USB Optical Reader, Leveloader Gold or PC Interface cable into the
desired PC. The system will automatically detect the connected unit and start
the ‘Found New Hardware’ Wizard.
2) From the provided list select “Locate and install driver software
(recommended)”.
Figure 3-12 Found New Hardware Window
3) On the next screen select, “I don’t have the disc. Show me other options.”
Figure 3-13 Found New Hardware - USB Serial Port Window
Note:
These instructions
are for use
with Vista 32
bit Operating
Systems.
Note:
After plugging in
the USB device,
if the Found
New Hardware Wizard does
not automatically appear,
proceed to the Manual USB
Installation instructions.
This manual suits for next models
4
Table of contents
Other Solinst Data Logger manuals

Solinst
Solinst LevelVent User manual

Solinst
Solinst Levelogger Series User manual

Solinst
Solinst Levelogger 3001 Quick reference guide

Solinst
Solinst AquaVent M5 User manual

Solinst
Solinst AquaVent 5 3500 User manual

Solinst
Solinst Barologger 3001 Gold User manual

Solinst
Solinst Levelogger Series User manual

Solinst
Solinst Levelogger 5 Series User manual

Solinst
Solinst Remote Radio Link User manual

Solinst
Solinst LevelVent 5 User manual