Solinst SRU 4001 User manual

Model 4001
Solinst Readout Unit (SRU)
User Guide
September 15, 2021


SRU User Guide - Contents
1.0 Introduction 5
1.1 Solinst Readout Unit (SRU) 5
1.2 SRU Specifications 5
1.3 Compatibility 6
2.0 SRU Operation 7
2.1 Start Up 7
2.2 LCD Display and Navigation 7
2.3 Main Menu 8
2.4 LCD Contrast 8
2.5 SRU Status 9
3.0 Connecting Dataloggers 10
3.1 Levelogger Connection 10
3.2 LevelVent Connection 10
3.3 AquaVent Connection 10
4.0 Real-Time Readings 11
4.1 Take a Reading 11
4.2 Barometric Compensation 11
4.3 Real-Time Logging 12
5.0 Downloading and Viewing Data Files 13
5.1 Download Logger 13
5.2 View Data Files 13
6.0 Datalogger Status 14
7.0 Levelogger Software and Data Management 15
7.1 Software Requirements and Installation 15
7.2 Transfer Data 16
7.3 View Data 18
7.4 Erase Data 19
7.5 Synchronize SRU Time 19
8.0 Maintenance 20
8.1 Battery Installation 20
8.2 Software and Firmware Updates 20
8.3 Firmware Upgrade Utility 21
Appendix 1: Full SRU Menu Sequence 22


SRU User Guide
Page 5
1.0 Introduction
1.1 Solinst Readout Unit (SRU)
The Solinst Readout Unit (SRU) is an in-field device designed to take and display instant water level readings from a
deployed Solinst datalogger.
When connected, you can also select a real-time sampling rate and start a real-time monitoring session independent of
the datalogger’s internal logging mode. Real-time logs can be saved in the SRU memory.
The SRU has a built-in barometer, which provides the option of showing and saving real-time water level readings
corrected to ambient air pressure.
The SRU also functions as a data download and storage device. It has a robust internal flash memory allowing you
to download and save multiple log files from a number of dataloggers. Saved log files can be viewed on the SRU and
transferred to a PC using Levelogger Software.
The SRU provides access to useful information about the connected datalogger’s status, including battery level, memory,
serial number, firmware version, and logging mode. You can also access information about the SRU status.
1.2 SRU Specifications
Figure 1-1 Solinst Readout Unit (SRU)
SRU Specications
Materials: ABS enclosure, 316 stainless steel connector, silicone rubber cover
Operating Temperature: -20ºC to +50ºC
IP Rating: IP64 (dust and splash resistant)
Battery: Standard 9V replaceable battery
Battery Life: Up to 2 years
Memory: 20 full LTC logs or 18 full LT logs (~1.8 million LTC readings or 2.7 million LT readings)
Size and Weight: 8.9 cm x 3.2 cm x 14.5 cm (3.5” x 1.3” x 5.7”) 351 grams ( 12.4 ounces)
Internal Barometer Accuracy: ±0.2 kPa
Compatible With: 5 Series and Edge Series Solinst dataloggers
USB-C Connection
for PC Communication
Serial Number
LCD Screen
Menu Push-Buttons
Datalogger
Interface Cable
Internal Barometric
Pressure Sensor
Interface
Cable Holder
Rugged Rubber
Outer Cover
9V Battery
Compartment

SRU User Guide
Page 6
1.3 Compatibility
The SRU is compatible with Solinst dataloggers using the following minimum firmware versions, or higher:
Datalogger Firmware Version
Levelogger 5 1.004
Barologger 5 1.004
Levelogger 5 Junior 1.004
Levelogger 5 LTC 1.004
Rainlogger 5 1.004
LevelVent 5 1.004
AquaVent 5 1.004
Levelogger Edge 3.004
Barologger Edge 3.004
Levelogger Junior Edge 3.004
LTC Levelogger Edge 1.003
Rainlogger Edge 3.001
LevelVent 1.000
AquaVent 1.000
It is also important to ensure you are using the most recent SRU firmware version. To check this, go to
www.solinst.com/downloads/
To determine what firmware version your SRU is using see Section 2.5.
To delete and transfer data files from the SRU to a PC, you will need to use Levelogger Software version
4.6.1, or higher.

SRU User Guide
Page 7
Idle
Status: indicates whether the SRU is idle, logging, or downloading/viewing data logs.
Battery icon: represents the battery life remaining for the SRU. Battery life is also numerically displayed
as a percentage to the left of the icon.
Note: It is recommended to replace the battery when the level falls below 20%.
Arrows: are used to navigate through and select the SRU menu items. Pressing and holding any of the arrow
buttons for two seconds, turns the SRU on/off.
Figure 2-2 SRU LCD Display
2.2 LCD Display and Navigation
2.0 SRU Operation
2.1 Start Up
Press and hold any arrow button for two seconds to turn the SRU on. When starting, the SRU will perform
a self-test. The start-up screen is shown first; following a short pause the main menu will be displayed. The
SRU has an auto-off after 2 minutes of inactivity.
Figure 2-1 Start-up Screen
Use the up/down arrows to scroll through menu items.
Use the left/right arrows to select the menu actions shown at the bottom left/right
of the LCD (e.g. Next, Back, etc.)
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD

SRU User Guide
Page 8
2.3 Main Menu
There are three main menu items for the SRU.
Take a Reading: allows you to display an instant reading from the connected datalogger, turn on/off
barometric compensation, and start and save a real-time logging session.
Download/View Files: allows you to download and save the data from the connected datalogger's current
internal logging session, and view all data files saved in the SRU.
Status/Setup: allows you to view information about the SRU and connected datalogger, including battery
level, memory, serial number, firmware version and logging mode. You can also adjust the contrast of the
LCD screen.
Figure 2-3 SRU Main Menu
2.4 LCD Contrast
To adjust the contrast of the LCD display, use the arrows to navigate to the Status/Setup menu, then select
LCD Contrast. Use the up and down arrows to adjust the contrast.
Figure 2-4 LCD Contrast
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD

SRU User Guide
Page 9
2.5 SRU Status
To view information on the SRU status use the arrows to navigate to the Status/Setup menu, then select
SRU Status. The SRU’s serial number, memory status, and firmware version will be displayed.
Figure 2-5 SRU Status
Note: An error message on the SRU screen will indicate when the memory is full.
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD

SRU User Guide
Page 10
3.0 Connecting Dataloggers
3.1 Levelogger Connection
The SRU cable connects to the top of a Levelogger’s Direct Read Cable. Hold the Direct Read Cable connection
and thread the SRU cable connector onto the Direct Read Cable.
Figure 3-1 Levelogger Connection
Note: An L5 Threaded or Slip Fit Adaptor can also be used to connect a Levelogger to the SRU cable.
3.2 LevelVent Connection
The SRU cable connects to a LevelVent’s Wellhead. Hold the Wellhead and thread the SRU cable connector
onto the Wellhead.
Figure 3-2 LevelVent Connection
3.3 AquaVent Connection
The SRU cable connects to an AquaVent’s (App/DataGrabber) Wellhead Connector Cable. Hold the AquaVent
cable and thread the SRU cable connector onto the AquaVent cable.
Figure 3-3 AquaVent Connection

SRU User Guide
Page 11
4.0 Real-Time Readings
4.1 Take a Reading
Select Take a Reading to instantly display the current water level (water column height above the pressure
sensor), temperature, and conductivity being detected by the connected datalogger.
Note: Only LTC Leveloggers will display conductivity readings.
Note: See the Levelogger User Guide (Section 1.2) and Vented Dataloggers User Guide (Section 1.2) for an explanation of the
pressure sensor measurement fundamentals.
Figure 4-1 Take a Reading
The water level reading will indicate "Comp" as it is a compensated level reading. See Section 4.2 for more
details.
4.2 Barometric Compensation
Levelogger 5 dataloggers use an absolute pressure sensor, therefore, detect total pressure above the sensor
(water plus barometric pressure).
To eliminate barometric pressure and display water level only, select Next after taking a reading (Section 4.1).
Select to turn barometric compensation on. The SRU contains an internal barometer that barometrically
compensates the level readings. The barometer reading will be displayed on this screen.
Figure 4-2 Turn Barometric Compensation On/Off
LevelVent 5 and AquaVent 5 vented water level dataloggers use gauged pressure sensors, therefore, barometric
effects are already eliminated from the level reading through the vented cable.
Barometric compensation can be turned off in the same screen.
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD

SRU User Guide
Page 12
4.3 Real-Time Logging
The SRU allows you to select a real-time sampling rate and start a real-time logging session independent of the
datalogger’s internal logging mode.
Figure 4-3 Real-Time Logging – Compensation On
Figure 4-4 Real-Time Logging – Compensation Off
To set up real-time logging, after you have turned barometric compensation on/off (Section 4.2), select Next to select
the Sample Rate. The Sample Rate can be set from 2 seconds to 99 hours.
Select Start to begin logging. The level, temperature and conductivity readings will update at the specified sampling
interval.
Note: Only LTC Leveloggers will display conductivity readings.
At any time you can select to Stop and/or Save the real-time logging session. The real-time logs will be saved in the
SRU memory. See Section 5.2 to view the saved log files list and Section 7.2 to transfer files to a PC.
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD

SRU User Guide
Page 13
5.0 Downloading and Viewing Data Files
5.1 Download Logger
To download a datalogger’s current internal logging session, select Download/View Files from the main
menu, scroll to Download Logger and select Next.
The download progress will be displayed at a percentage. All data from the current logging session will be
saved in the SRU memory as a *.xle data file.
Note: An error message on the SRU screen will indicate when the memory is full. You can also view the status of the SRU memory
using the Statu/Status menu (see Section 2.5).
The file will be identified using the datalogger’s serial number and the date and time of the download.
Note: Downloading the data will not stop/interrupt the datalogger from logging. Datalogger memory will not be erased.
Figure 5-1 Download Logger
Note: *.xle files can be opened using Levelogger Software for further analysis and exported for use in other programs.
5.2 View Data Files
After downloading data from the connected datalogger, you can view a list of all the saved files on the SRU,
including real-time logging sessions, by selecting Download/View Files from the main menu then View Files.
The files will be identified as follows:
D = Downloaded Data File
C = Compensated Real-Time Data File
U = Uncompensated Real-Time Data File
Figure 5-2 View Files
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading
Download
99%
99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading
Download
99%
99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD

SRU User Guide
Page 14
6.0 Datalogger Status
To view information about the connected datalogger, scroll to Status/Setup, select Next, then Logger Status.
Figure 6-1 Logger Status
Select Next to continue viewing information including serial number, battery level, memory, logging mode and
status, project information, and firmware version.
Figure 6-2 Logger Information
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD
NextBack
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Idle 99%
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
Next
On
Back
Turn Compensation:
Baro Reading: 98.89kPa
Idle 99%
Back Start
12 mins
Sample Rate:
Compensation: On
Idle 99%
Stop Save
Level (Comp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Home
Data Saved
Compensated
Barometrically
Log 57 99%
Home
Data Saved
Uncompensated
Log 57 99%
Stop Save
Level (Uncomp): 10.045m
Temperature: 28.234ºC
Conductivity: 125.4µS
Log 56 99%
Back Start
12 mins
Sample Rate:
Compensation: Off
Idle 99%
Off
Next
Take a Reading
Download/View Files
Status/Setup
55%
Downloading Download
99% 99%
Cancel
Downloading Logs:
Next
Take a Reading
Download/View Files
Status/Setup
Idle 99%
NextBack
LCD Contrast
SRU Status
SRU Status
Logger Status
Idle 99%
NextBack
LCD Contrast
Logger Status
SRU Status
Idle 99%
Download Logger
Idle 99%
Idle 99%
View Files
NextBack
Download Logger
View Files
Idle
NextBack
99%
NextBack
LCD Contrast
Logger Status
Idle 99%
Home
Logger Data
Saved to file
1081225-01.10.21-12.30.xle
Next 3Home
3/11 Files 99%
1081225-01.10.21-12.30 D
1081225-01.15.21-15.56 C
1081225-01.16.21-22.50 U
NextBack
Idle 99%
Logger Serial: 1081234
Logger Memory: 50%
Logger Battery: 85%
NextBack
Idle 99%
Log Mode: Schedule
Logging Rate: 15 secs
Logger Status: Started
Back Home
Idle 99%
Location: Wellhead 234
Name: Zanzibar 1
Firmware: V1.004
Back
Idle 99%
SRU Serial: 123456
SRU Memory: 50%
Firmware: V1.000
Back Home
Idle 99%
55%
Contrast:
Set LCD

SRU User Guide
Page 15
7.0 Levelogger Software and Data Management
7.1 Software Requirements and Installation
The minimal hardware and software requirements for software installation and operation are:
Hardware Software
Processor: 1 GHz or faster processor or SoC OS: Windows 10
RAM: 1 GB for 32-bit or 2 GB for 64-bit
Hard disk space: 128 MB
Display: 800 x 600
Ports: USB or RS-232
Web Download
1) Download the newest version of Levelogger Software by visiting: https://www.solinst.com/downloads/
2) If the installer does not automatically start, to activate the software install click on the ‘setup.exe’ file.
3) The Software Installation Wizard will guide you through the remaining installation process. Figure 7-1 shows
the Levelogger Installation Wizard.
4) Restart the computer after installation is completed. Default Directory is C:\Program Files\Solinst\Levelogger4_6
Figure 7-1 Software Installation Wizard

SRU User Guide
Page 16
7.2 Transfer Data
To transfer data files from the SRU to a PC, Connect the SRU to the PC using a USB cable. The SRU will
automatically turn on and the LCD screen will show a message indicating the SRU is connected to the PC.
Figure 7-2 SRU to PC Connection
To start the Levelogger Software, click , or click the Start button and select:
Programs > Solinst > Levelogger 4 > Levelogger 4.6.1
Go to the SRU tab and select the appropriate Com Port for the connected SRU from the centre drop-down
menu and Click the ‘Retrieve SRU Settings’ icon .
A destination folder window will open where you can choose the location your log files will be saved.
Figure 7-3 Levelogger Software SRU Tab – Select Data File Location

SRU User Guide
Page 17
Select the file(s) from the list that you want to download, and click the ‘Download Data’ icon
.
Figure 7-4 Levelogger Software SRU Tab

SRU User Guide
Page 18
7.3 View Data
Data files can be viewed in the Data Control tab by selecting Finish and Open all, or select Finish if you do
not need to view the data right away.
Figure 7-5 Finish Data Download

SRU User Guide
Page 19
Data can be viewed later by going to the Data Control tab and selecting the ‘Open Files’ icon. From here, you
can also print data or export data to a .csv or .xml file.
Note: For more information on the Data Control tab, see the Levelogger Series or Vented Dataloggers User Guides.
Figure 7-6 Data Control Tab
7.4 Erase Data
All data files can be erased from the SRU by selecting the ‘Erase Files’ icon in the SRU tab.
Note: An error message on the SRU screen will indicate when its memory is full. You can also view the status of the SRU memory
using the Statu/Status menu (see Section 2.5).
7.5 Synchronize SRU Time
The attached SRU system time is displayed in the bottom left corner of the SRU tab. To synchronize the SRU
to the PC clock, simply click Synchronize.
Figure 7-7 Clock Synchronization
Note: To synchronize to a different time/ time zone you first must adjust the computer time using the Windows Operating System
Date and Time Settings menu. Then re-start the Solinst Levelogger Software and synchronize your SRU to the new time.

SRU User Guide
Page 20
8.0 Maintenance
8.1 Battery Installation
The SRU uses one 9V replaceable alkaline or lithium battery. To install or replace the battery, remove the
SRU from the protective outer case. Push up to remove the battery enclosure door from the back of the SRU.
Connect the battery lead to the 9V battery and install the battery in the housing. Replace the battery enclosure
door and the protective outer case of the SRU.
Note: It is recommended to replace the battery when the level falls below 20%.
Figure 8-1 SRU Battery Installation
8.2 Software and Firmware Updates
Each time the software is opened, it automatically checks on-line for software updates. If there is an update,
“Software Update Available” will appear in the top right of the menu bar. When you click the message, it will
open a web page where you can download the software update.
Note: Your PC must have an Internet connection to check for software and firmware updates.
The Software also checks for firmware updates once you have retrieved settings from a connected SRU.
“Firmware Update Available” will appear in the top right menu bar. Clicking the message will open a web page
where you can download the firmware update. See Section 8.3 for firmware update instructions.
This manual suits for next models
1
Table of contents
Other Solinst Data Logger manuals

Solinst
Solinst AquaVent 5 3500 User manual

Solinst
Solinst Barologger 3001 Gold User manual

Solinst
Solinst AquaVent M5 User manual

Solinst
Solinst LevelVent 5 User manual

Solinst
Solinst Levelogger Series User manual

Solinst
Solinst Levelogger Gold User manual

Solinst
Solinst Levelogger 3001 User manual

Solinst
Solinst Levelogger Series User manual

Solinst
Solinst Remote Radio Link User manual

Solinst
Solinst Levelogger 3001 Quick reference guide