- 4 -
•To obtain peak readings, press the HOLD Button for 2 to 3 seconds.
The words “Peak HOLD” will appear on the LCD. To clear the peak
reading, press the HOLD Button once (the display will flash). To exit
this function, press the HOLD Button for at least 2 seconds.
•To record the maximum and minimum measurements, press the REC
MAX/MIN Button and “REC” appears on the LCD.
•Press the REC MAX/MIN. Button again and “REC Max” appears on the
LCD, along with the maximum measurement.
•Press the REC MAX/MIN Button again and “REC Min” appears on the
LCD, along with the minimum measurement.
•To exit this function, press the REC MAX/MIN Button for at least 2
seconds until the display reverts to the current reading.
•Press the POWER Button to turn the unit off.
Notes: Keep the lens cap on the PHOTO SENSOR at all times except
when actually taking a reading. Fluctuations in the reading are generally due
to shadows or fluctuations in the line voltage. Ambient temperature, drafts
and ventilation also affect the luminous flux output. Avoid range overload
and do not store in areas of high temperature and/or humidity. Remove the
battery for long-term storage.
ZERO ADJUSTMENT
With the sensor cover securely in place and the meter turned on, and select
a the 40.00 Lux range (digits appear on the LCD). Press the ZERO button
and the LCD should indicate “0.00”.
BATTERY REPLACEMENT
Replace the battery when the low battery icon is displayed in the
left corner of LCD. In-spec measurements may be made for
several hours after the low battery indicator appears. Remove
the battery compartment screw and slide the cover away from
the meter, remove the battery and replace with a 9V battery (alkaline or
heavy duty type). Replace the cover.
AUTOMATIC SHUT OFF
The instrument has an automatic shut off function in order to prolong battery
life. After approximately 10 minutes without activity (no buttons pushed), the
meter will automatically shut off. To disable this feature, press the REC
Button.
CALIBRATION
To maintain accuracy, annual calibration is recommended. For complete
details and current fees contact SPER SCIENTIFIC Ltd, 7720 East Redfield
Rd, Suite 7 ,Scottsdale, Arizona 85260, USA. Phone: (480) 948-4448,
web: www.sperscientific.com, email: service@sperscientific.com.