
5
1. Overview
The ToxiLtd is a single sensor gas
detector that is designed to detect either
oxygen (O2) or one of a variety of toxic
gases. The ToxiLtd’s sensor type is
shown on the front of the instrument and
is also shown on the display during the
startup sequence.
ToxiLtd gas detectors are disposable
instruments that are designed to operate
for 2 full years from the date of activation.
ToxiLtd models equipped with a toxic gas
sensor (CO, H2S or SO2) must be
activated within 6 months of the
manufacturing date. Oxygen versions
must be activated within 1 month of the
manufacturing date. If the instrument is
not turned on by the “must be activated by
date”, the instrument will activate itself
and the 2-year countdown will begin
automatically.
Upon delivery, the screen of the ToxiLtd
will be blank except for the negated horn
icon.
If the “must be activated by date” has
passed and the instrument has activated
itself, the instrument will show three lines
on the screen with the negated horn icon.
Contact Sperian or your distributor if
three lines are displayed when you first
receive the detector.
The ToxiLtd is delivered with a pre-
installed lithium thionyl chloride (LiSOCl2)
battery capable of powering the
instrument for the full 2 years of the
instrument’s operating life. After the 2-
year instrument life cycle is up, the
instrument will turn itself off.
Note: Early production runs of the
oxygen version of the ToxiLtd were
referred to as the “OxyLtd”. The
instructions contained in this manual
that refer to the ToxiLtd Oxygen (O2)
are valid for the OxyLtd with similar
instrument firmware.
1.1 Methods of sampling
The ToxiLtd may be used in diffusion
mode, or with the manual sample draw kit
that is available separately. In either
mode, the atmosphere must reach the
sensor for the instrument to register a
reading. In diffusion mode, the
atmosphere reaches the sensor by
diffusing through the sensor port on the
front of the instrument. Normal air
movements are enough to carry the
sample to the sensor. During remote
sampling, the gas sample is drawn into
the sensor compartment through the
probe assembly and a length of tubing.
See section 2.6 for more details on
sampling the atmosphere.
1.2 Sensors
Toxic sensor versions of the ToxiLtd use
an electrochemical toxic gas sensor. The
ToxiLtd O2uses a galvanic oxygen
sensor. Both types of sensor have been
designed to minimize the effects of
common interfering gases. These
sensors provide accurate, dependable
readings for gases commonly
encountered in industrial applications.
1.2.1 Sensor ranges and resolutions.
Sensor Range Resolution
CO 0-1000 PPM 5 PPM
H2S 0 -200 PPM 2 PPM
SO20 – 20 PPM 0.1 PPM
O20-30%/Vol. 0.2%/Vol.
1.2.2 ToxiLtd O2warm-up phase
The ToxiLtd O2requires a one-hour
warm-up phase from the moment of
activation until actual use. See section
2.1.1 for further details.
1.2.3 F1: Missing sensor during
startup
If the ToxiLtd fails to detect the sensor
during startup, it will show “F 1” with the
caution symbol following the software
version screen at startup. This is covered
in greater detail in section 2.1.3.