Section 2: General Information
2.1 GENERAL DESCRIPTION
Operated by a 16-bit PLC (Programmable Logic Con-
troller), the SurgiGraphic™ 6000 Image Guided Surgical
Table features lateral tilt, trendelenburg/reverse
trendelenburg, and adjustable height functions.
The SurgiGraphic Table features a manually controlled
four-way float tabletop. The "float" range is 16" (406 mm)
longitudinal and 8.0" (203 mm) lateral (when tabletop is
paralleltotheearthplane). However, this table enables the
operator to have 16" (406 mm) longitudinal float while
tabletop is in a tilt position and 8.0" (203 mm) lateral float
when table is in trendelenburg/reverse trendelenburg
positions.
The table accepts positioning commands from four
sources:
1. A customizable Hand Control (which includes ON,
LOCK, UNLOCK, HOME or LEVEL, TREND, RE-
VERSE TREND, HEIGHT UP, HEIGHT DN,
TILT L, TILT R, EARTH PLANE, and four Pro-
grammable Memory buttons). Refer to Figure 4-5 of
this document for control feature locations.
2. A surgeon's Palm Handle Control (which includes
Four-Way Manual Float control with LOCKOUT
switch, TREND, REVERSE TREND, HEIGHT UP,
HEIGHT DOWN, TILT R, TILT L, and HOME or
LEVEL functions). Refer to Figure 4-6 of this docu-
ment for control feature locations.
3. A limited Emergency Backup Buttons (which in-
cludes LEVEL, Battery Backup [6000CE/6000FB] and
Floor Lock functions). Refer to Figure 2-10 or 2-11 of
this section of the document for control feature loca-
tions.
4. Anoptional physician-controlled Foot Control (which
includes Trendelenburg Tilts, Lateral Tilts, and Height
functions).
2.2 TECHNICAL SPECIFICATIONS
2.2.1 Overall Size
Original / 6000CE: 31" wide x92" long x 32" to 42" high
(787 mm wide x 2337 mm long x 813 to 1067 mm high).
6000FB: 27" wide x92" long x 30.5" to 39.5" high
(686 mm wide x 2337 mm long x 775 to 1003 mm high).
2.2.2 Ranges of Motion
Height:
Original / 6000CE: 32" to 42" (813 to 1067 mm).
6000FB: 30.5" to 39.5" (775 to 1003 mm).
Trendelenburg Range: 16° ±1°.
Lateral Tilt Range: 16° ±1°.
Tabletop Motion: Four-Way Float 16" longitudinal
(410 mm), 8.0" lateral (205 mm).
2.2.3 Weight
The table weighs 990 lb (449 kg). Patient weight capacity
is 500 lb (227 kg).
2.2.4 Power Requirements
Line Power Input:
100 VAC, 60 Hz, One-Phase, 5.0 Amps.
120 VAC, 60 Hz, One-Phase, 4.2 Amps.
220 VAC, 60 Hz, One-Phase, 2.4 Amps.
230/240 VAC, 60 Hz, One-Phase, 2.0 Amps.
NOTE: Each table is shipped from the factory configured to the
electrical requirement specified on the factory order. If this
electrical configuration needs to be changed in the field, consult
STERIS for the needed procedure and required materials.
2.2.5 Anatomical Imaging Area
Top Extended: 71" Long, 21" Wide (1803, 533 mm).
Top Retracted: 55" Long, 21" Wide (1397, 533 mm).
2.3 PINCH POINT WARNINGS
WARNING - PINCHING HAZARD:
• Pinch points are created during tabletop ar-
ticulation. Carefully review the photographs
inFigures2-1through2-8.
• To avoid serious injury, keep limbs, fingers,
and other body areas clear of all pinch points
when positioning the table.
CAUTION - POSSIBLE EQUIPMENT DAM-
AGE: Failure to keep all personnel and equipment
clear of the table before actuating any inertia-driven
or power-driven movement could result in table
damage.
When inertia-driven or power-driven table parts close,
pinch-point hazards exist (see Figs. 2-1 through 2-8). All
personnel involved in tabletop positioning should exam-
ineandbeaware of these points before operating the table.